INTEGRITY LOW SPEED DENTAL AIR HANDPIECE AND ATTACHMENT, MODELS: MOTOR INT-E, MOTOR INT-D AND ATTACHMENT INT-SNC
Applicant
Osseo Scientific, LLC
Product Code
EFB · Dental
Decision Date
Jan 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The INTEGRITY Air Powered Low Speed Handpieces are intended for removing curious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and polishing teeth for use by a trained professional in general dentistry.
Device Story
Integrity Low Speed Dental Air Handpiece is a 5-vane air-powered motor system; operates at speeds up to 20,000 RPM. Compatible with standard 2, 3, or 4-hole air connections; accepts ISO E-Type or U-Type attachments. Used in clinical dental settings by trained professionals for restorative procedures, cavity preparation, and tooth polishing. Device functions via pneumatic power; provides mechanical rotation to dental burs or polishing tools. Healthcare providers use the handpiece to manually manipulate tooth structure; output is the physical removal or finishing of dental material. Benefits include efficient tooth preparation and restoration maintenance.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through comparison of design specifications, materials, and performance characteristics against legally marketed predicate devices.
Technological Characteristics
5-vane air motor; max speed 20,000 RPM. Compatible with ISO E-Type/U-Type attachments and standard 2, 3, or 4-hole air connections. Complies with ISO 7785-2 (low speed handpieces), ISO 1797 (shank dimensions), ISO 9168 (hose connectors), and ISO 3964 (coupling dimensions). Sterilizable.
Indications for Use
Indicated for use by trained dental professionals for removal of carious material, reduction of hard tooth structure, cavity preparation, finishing of preparations, restorations, and tooth polishing in general dentistry patients.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
K070155
### 510(k) SUMMARY
## OSSEO SCIENTIFIC, LLC INTEGRITY LOW SPEED HANDPIECE & ATTACHMENT
1. Submitter Name, Address, Phone/Fax
Registration No. 3005405362
Osseo Scientific, LLC 3138 Veeder Ave. Toms River, NJ 08753 USA Telephone # 201-490-5421 Facsimile # 201-490-5423
2. Contact Person-Owner
Michael Feldman Telephone # 732-597-5971
3. Date summary prepared: October-28-2006
4. Device Name:
Trade Name: INTEGRITY Low Speed Dental Air Handpiece and Accessories. Model No's INT-E, INT-D, INT-SNC.
Common Name: Handpiece Low Speed Dental.
Classification Name: Dental Handpiece and Accessories. ( 21CFR 872.4200 )
Product Code: 76 EFB ( "Handpiece, Air-Powered, Dental" ).
Device Class: Class I
5. Indications for Use:
The INTEGRITY Air Powered Low Speed Handpieces are intended for removing curious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and polishing teeth for use by a trained professional in general dentistry.
'JAN 2 5 2007
{1}------------------------------------------------
- 6. Predicate Device Name:
For Model INT-E: A-dec/W&H Synea & TreND Low Speed (K944713 & K993526) and Dabi Atlante Low Speed (K926154).
For Model INT-SNC: A-dec/W&H Synea & TreND ( K993526), Dabi Atlante (K926155) and MK-dent Straight Handpiece (K051872)
For Model INT-D: Promident Low Speed (K781545)
7. Description of Device:
The Integrity Low Speed Handpiece is a 5-vane motor, ranging in speed up to 20,000 RPM, and is compatible to all standard 2, 3 or 4 hole fixed air connection. It accepts all brands of ISO E-Type or U-Type attachments. The Motors and Straight Nose Cone are used for general dentistry procedures.
- 8. Substantially Equivalence- Safety and Effectiveness:
The INTEGRITY Low Speed handpieces share virtually all specifications and design characteristics of the predicate devices and are Identical or similar based on comparisons of formulations and intended uses as shown below. The handpieces are constructed of materials of identical specification as the predicate device. Air Connection, and interface with other dental attachments were Identical. Air Pressure and, speed Results were similar. It is therefore Substantially Equivalent to one or more dental handpieces currently marketed in the USA.
The Handpieces conform to applicable ISO standards. The ability to repeatedly adequately sterilize the device is being confirmed by validation protocol, and will be available for inspection before the device is marketed.
| > Intended use; | Identical |
|------------------------------------------------|-----------|
| > Indications for use; | Identical |
| > Target population; | Identical |
| > Anatomical sites; | Identical |
| > Where used (hospital, home, ambulance, etc); | Identical |
| > Energy used and/or delivered; | Identical |
| > Human factors; | Identical |
| > Design; | Similar |
| > Performance; | Identical |
| > Standards met | Identical |
| > Materials | Identical |
{2}------------------------------------------------
| > Biocompatibility | |
|------------------------------------------------------------------|--|
| > Compatibility with the environment and other devices Identical | |
| > Sterility | |
| > Mechanical safety | |
| > Electrical safety | |
| > Chemical safety | |
| > Thermal safety | |
| > Radiation safety | |
8. Conclusion: Based on information in the notification Osseo Scientific, LLC believes that INTEGRITY Air Powered Low Speed Handpieces are substantially equivalent to the claimed predicate, i.e.,
For Model INT-E: A-dec/W&H Synea & TreND Low Speed (K944713 & K993526) and Dabi Atlante Low Speed (K926154).
For Model INT-SNC: A-dec/W&H Synea & TreND ( K993526), Dabi Atlante (K926155) and MK-dent Straight Handpiece (K051872)
For Model INT-D: Promident Low Speed (K781545)
#### Voluntary Standard Compliance:
ISO Standard 7785-2: Low Speed Dental Handpiece.
ISO Standard 1797: Shank Dimensions.
ISO Standard 9168: Hose Connectors.
ISO Standard 3964: Coupling Dimensions
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 5 2007
Osseo Scientific, LLC C/O Mr. Neil E. Devine Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
Re: K070155
Trade/Device Name: Integrity Low Speed Air Dental Handpiece and Attachment Regulation Number: 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: January 16, 2007 Received: January 17, 2007
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sujata Y. Michael D.M.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Integrity Low Speed Air Dental Handpiece and Attachment.
Models: Motor INT-E, Motor INT-D and Attachment INT-SNC
Indications for Use:
The INTEGRITY Air Powered Low Speed Handpieces are intended for removing carious material, reducing hard tooth structure, savity preparation, finishing tooth preparations, restorations and polishing teeth for use by a trained professional in general dentistry.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runns
10) 54
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.