K070118 · Tympany, Inc. · EWO · Mar 7, 2007 · Ear, Nose, Throat
Device Facts
Record ID
K070118
Device Name
OTOGRAM, CA3350
Applicant
Tympany, Inc.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Mar 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Otogram is an auditory diagnostic instrument intended to administer, under supervision by a trained healthcare professional, a battery of diagnostic and screening procedures that include the following: 1) Pure tone air and bone conduction audiometry with automated masking 2) Speech recognition threshold with automated masking 3) Speech discrimination with automated masking 4) Tympanometry 5) Acoustic reflex threshold, both ipsilateral and contralateral 6) Distortion product otoacoustic emissions 7) Pure tone Stenger 8) Patient survey The Otogram is indicated for use by trained healthcare professionals on both adult and pediatric subjects for measurement of audiometric parameters to identify and supply data to help diagnose hearing loss and ear disorders.
Device Story
Otogram CA3350 is a computer-controlled diagnostic instrument combining audiometry, distortion product otoacoustic emissions (DPOAE) analysis, and tympanometry. The device connects to a PC via RS-232 port; control software on the PC manages the instrument. It performs pure tone air/bone conduction audiometry, speech recognition/discrimination, tympanometry, acoustic reflex thresholds, DPOAE, and Stenger tests. Operated by trained healthcare professionals in clinical settings, the device provides diagnostic data to assist in identifying hearing loss and ear disorders. The clinician reviews the output to inform clinical decision-making. Benefits include a comprehensive, integrated diagnostic battery for efficient patient assessment.
Clinical Evidence
Bench and user testing performed. Comparison testing conducted by a Hearing Instrument Specialist Intern to validate performance against predicate devices. No clinical trial data provided.
Technological Characteristics
Computer-controlled audiometric instrument. Interfaces via RS-232. Electrical safety per IEC 60601-1, 60601-1-1, and 60601-1-2 (EMC). Performance/calibration per IEC 1027, ANSI S3.39, and IEC 60645-5. Class II device.
Indications for Use
Indicated for adult and pediatric patients to measure audiometric parameters for the identification and diagnosis of hearing loss and ear disorders. Used by trained healthcare professionals.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
### 510(k) Supplementary Information K070118
1. Submitters Name and Address: Tympany, Inc.
| | 2795 East Cottonwood Parkway, Suite 660<br>Salt Lake City, UT 84121<br>Phone: (801) 365-2800<br>FAX: (801) 365-3000 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Contact Person & Phone: | Mr. Lex J. Pearce<br>Phone: (801) 365-2868<br>FAX: (801) 365-3005<br>lex@sonici.com |
| 3. Date Prepared: | 26 January 2007 |
| 4. Device Name / Classification: | Audiometer, Otoacoustic Emissions Analyzer (21<br>CFR 874.1050, Product Code EWO)<br>Auditory Impedance Tester, Tympanometer (21<br>CFR 874.1090, Product Codes ETY, NAS) |
| 5. Proprietary / Trade Name: | Otogram CA3350 |
#### 6. Device Description:
The Otogram CA3350 is a computer-controlled, audiometric instrument combining the functions of an Audiometer, Distortion Product Otoacoustic Emissions Analyzer and Auditory Acoustic Impedance/Admittance Tester (i.e. Tympanometer). The device is controlled through the use of ASCII commands transmitted over a standard RS-232 communication port. Control software is installed on the supplied PC.
#### 7. Intended Use:
The Otogram is an auditory diagnostic instrument intended to administer, under supervision by a trained healthcare professional, a battery of diagnostic and screening procedures that include the following:
1) Pure tone air and bone conduction audiometry with automated masking
- 2) Speech recognition threshold with automated masking
- 3) Speech discrimination with automated masking
4) Tympanometry
- 5) Acoustic reflex threshold, both ipsilateral and contralateral
- 6) Distortion product otoacoustic emissions
- 7) Pure tone Stenger
- 8) Patient survey
The Otogram is indicated for use by trained healthcare professionals on both adult and pediatric subjects for measurement of audiometric parameters to identify and supply data to help diagnose hearing loss and ear disorders.
{1}------------------------------------------------
| | The previous version of this chart appears in Exhibits 5, 10, and 12. | | | | |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Characteristic | Predicate<br>Impedance Audiometer<br>Madsen Zodiac 901<br>K910247 | Predicate<br>Otoacoustic Emissions<br>Bio-Logic AuDX<br>K974076 | Predicate<br>Tympany Otogram<br>Combined Device<br>Impedance Audiometer<br>Otoacoustic Emissions<br>K041853 | Predicate<br>Tympany Otogram<br>CA3350<br>Combined Device<br>Impedance Audiometer<br>Otoacoustic Emissions<br>K070118 |
| | Intended Use: | To diagnose a variety of<br>middle ear disorders using<br>impedance and static<br>pressure measurements<br>(i.e., tympanometry and<br>acoustic reflex) in the ear<br>canal. | To test cochlear function<br>by measuring otoacoustic<br>emissions. | To diagnose hearing and<br>otologic disorders in the<br>middle-ear and total ear<br>system, using audiometry,<br>Tympanometry, and<br>acoustic reflex. To test<br>cochlear function and<br>presence of otoacoustic<br>emissions. | To administer, under<br>supervision by a trained<br>healthcare professional, a<br>battery of diagnostic and<br>screening procedures that<br>include pure tone air and<br>bone conduction<br>audiometry with automated<br>masking, speech<br>recognition threshold with<br>automated masking, speech<br>discrimination with<br>automated masking,<br>tympanometry, acoustic<br>reflex threshold (both<br>ipsilateral and<br>contralateral), distortion<br>product emissions, pure<br>tone Stenger and patient<br>survey. |
| Characteristic | Predicate<br>Impedance Audiometer<br>Madsen Zodiac 901<br>K910247 | Predicate<br>Otoacoustic Emissions<br>Bio-Logic AuDX<br>K974076 | Predicate<br>Tympany Otogram<br>Combined Device<br>Impedance Audiometer<br>Otoacoustic Emissions<br>K041853 | Tympany Otogram<br>CA3350<br>Combined Device<br>Impedance Audiometer<br>Otoacoustic Emissions<br>K070118 | |
| Physical<br>characteristics: | | | | | |
| Computer<br>interface: | RS 232C | RS 232C | Computer included, RS<br>232C | Computer included, RS<br>232C | |
| Internal printer | Thermal 112 mm, 4.5" | Thermal 59 mm,<br>2 1/4" | Thermal 58 mm,<br>2 1/4" | NA | |
| Printer interface | Parallel port | RS 232C thermal printer | Parallel port or USB<br>Computer LCD 1024 x 768 | Parallel port or USB<br>Computer LCD 1024 x 768 | |
| Display | Graphic supertwist LCD<br>backlight, 256 lines x 128<br>dots<br>White text/graphics on<br>blue background | LCD 1" x 3"<br>Black and white | Color, resistive<br>touchscreen | Color, resistive<br>touchscreen | |
| Control interface | Keyboard | Touchpad | Keyboard, touchscreen | Keyboard, touchscreen | |
| Size/weight-<br>metric | 370 x 385 x 120 mm<br>(W x D x H), 7.6 kg | 98 x 41 x 197 mm<br>(W x D x H), 0.45 kg | 410 x 300 x 480 mm<br>(W x D x H), 20.4 kg | 356 x 348 x 493 mm<br>(W x D x H), 21.1 kg | |
| Size/weight-<br>American | 14.8" x 15.4" x 4.8", 16.8<br>lbs | 3 7/8" x 1 5/8" x 7 3/4", 1<br>lb | 16" x 19", x 12", 45 lbs | 14.0" x 13.7" x 19.4", 46.5<br>lbs | |
| Energy source | AC 50/60 Hz, 100-240 V | AC 50/60 Hz, 100-240 V | AC 50/60 Hz, 100-120 V | AC 47-63 Hz. 90-250<br>VAC | |
| Hardcopy output | Thermal paper or external<br>printer paper | External thermal printer | Thermal paper or external<br>printer paper | External printer paper | |
| Standard and<br>safety<br>characteristics: | | | | | |
| Performance and<br>calibration | IEC 1027, ANSI S3.39 | NA | IEC 1027, ANSI S3.39 | IEC 1027, ANSI S3.39,<br>IEC 60645-5 | |
| Electrical safety | EN 60601-1, class I,<br>type B | EN 55011: 1991 group 1<br>Class B | EN 60601-1, class I,<br>type B | IEC 60601-1 and 60601-1-<br>1, class I, type B and BF<br>and IEC 60601-1-2 (EMC) | |
rs in Ev 8. Updated Substantial Equivalence Chart:
The previous version of this chart appea
{2}------------------------------------------------
.
-
'
Page 5 of 6
:
and the comments of the comments of
{3}------------------------------------------------
# 9. Significant changes from predicate Tympany Otogram combination device:
- Expansion of intended use to include use by "trained healthcare professional" vs. "qualified/trained audiologist"
- The indications for use stated in the 501(k) submission for the Bio-Logic AuDX device (K974076) indicates use by a "trained health care professional (for example an Audiologist)"
- Training for the Otogram CA3350 includes a thorough review and training using the User Instruction Guide included in Exhibit 13 of the 510(k) submission.
- Comparison testing with predicate devices was performed by a Hearing Instrument Specialist Intern - someone who has been trained on the use of the equipment, but is not an audiologist. Test reports are contained in Exhibit 12 of the 510(k) submission.
- Different computer and improved design on various components and assemblies.
- Computer specifications and design specifications are contained in Exhibit 11 of the 510(k) submission.
# 10. Safety and Effectiveness comparison to predicate devices:
The results of bench, user, and laboratory testing indicate that the new device is as safe and effective as the predicate devices.
# 11. Conclusion:
After analyzing both bench and user testing data, it is the conclusion of Tympany Inc. that the Tympany Otogram CA3350 is safe and effective as the predicate devices, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.
MAR 0 7 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tympany, Inc. c/o Lex J. Pearce 2795 East Cottonwood Parkway, Suite 660 Salt Lake City, UT 84121
Re: K070118
Trade/Device Name: Otogram CA3350 Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO, NAS, ETY Dated: January 11, 2007 Received: January 12, 2007
Dear Mr. Pearce:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
M.B. Egelman Si MD
g
Malvina B. Evdelman. M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K070118
Device Name: Otogram CA3350
Indications for Use:
The Otogram is a computer-controlled, audiometric instrument combining the functions of an Audiometer, Distortion Product Otoacoustic Emissions Analyzer and Auditory Acoustic Impedance/Admittance Tester (i.e. Tympanometer) The Otogram is indicated for use by trained healthcare professionals on both adult and pediatric subjects for measurement of audiometric parameters to identify and supply data to help diagnose hearing loss and ear disorders.
The Otogram is an auditory diagnostic instrument intended to administer, under supervision by a trained healthcare professional, a battery of diagnostic and screening procedures that include the following:
1) Pure tone air and bone conduction audiometry with automated masking
2) Speech recognition threshold with automated masking
3) Speech discrimination with automated masking
4) Tympanometry
5) Acoustic reflex threshold, both ipsilateral and contralateral
6) Distortion product otoacoustic emissions
7) Pure tone Stenger
8) Patient survey
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use ___________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------|--|
| Division of Ophthalmic Ear, Nose and Throat Devises | |
| 510(k) Number | K070118 |
|---------------|---------|
|---------------|---------|
| | Prescription Use<br>(Per 21 CFR 801.109) |
|--|------------------------------------------|
|--|------------------------------------------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.