← Product Code [NYO](/productcode/NYO) · K070066

# QUANTA LITE RNA POL III ELISA (K070066)

_Inova Diagnostics, Inc. · NYO · Apr 11, 2007 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070066

## Device Facts

- **Applicant:** Inova Diagnostics, Inc.
- **Product Code:** [NYO](/productcode/NYO.md)
- **Decision Date:** Apr 11, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The QUANTA Lite™ RNA Pol III ELISA is a semiquantitative enzyme-linked immunosorbent assay for the detection of IgG anti-RNA Polymerase III antibodies in patient sera. The presence of these antibodies, when considered in conjunction with other laboratory and clinical findings, is an aid in the diagnosis of systemic sclerosis (scleroderma) with increased incidence of skin involvement and renal crisis.

## Device Story

QUANTA Lite RNA Pol III ELISA is an in vitro diagnostic test for patient serum samples. Device utilizes enzyme-linked immunosorbent assay (ELISA) technology to detect IgG antibodies against RNA Polymerase III. Performed in clinical laboratories by trained personnel. Output is semiquantitative measurement of antibody levels. Results interpreted by clinicians alongside other laboratory and clinical findings to support diagnosis of systemic sclerosis, particularly in patients presenting with skin involvement or renal crisis. Assists in clinical decision-making by identifying specific autoantibody profiles associated with disease severity.

## Clinical Evidence

No clinical data provided in the summary document; substantial equivalence established via bench testing and performance validation of the ELISA assay.

## Technological Characteristics

Semiquantitative enzyme-linked immunosorbent assay (ELISA). Utilizes purified RNA Polymerase III antigen coated on microplate wells. Detection via enzyme-conjugated anti-human IgG. Readout via spectrophotometric measurement of colorimetric reaction. Standalone laboratory test.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- QUANTA Lite Centromere ([K962536](/device/K962536.md))
- QUANTA Lite Scl-70 ([K962536](/device/K962536.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k070066

B. Purpose for Submission:
New device

C. Measurand:
Anti-RNA polymerase III antibodies

D. Type of Test:
Semi-quantitative Enzyme-linked immunosorbent assay (ELISA)

E. Applicant:
INOVA Diagnostics, Inc.

F. Proprietary and Established Names:
QUANTA Lite™ RNA Polymerase III ELISA

G. Regulatory Information:
1. Regulation section:
CFR 866.5100 Antinuclear Antibody Immunological Test System
2. Classification:
Class II
3. Product code:
NYO Autoantibody, anti-ribonucleic acid polymerase (RNAP) III
4. Panel:
IM 82

H. Intended Use:
1. Intended use:
QUANTA Lite™ RNA Polymerase III ELISA is a semi-quantitative enzyme-linked immunosorbent assay (ELISA) for the detection of IgG anti-RNA polymerase III antibodies in patient sera. The presence of these antibodies when considered in conjunction with other laboratory and clinical findings is an aid in the diagnosis of systemic sclerosis, (scleroderma), with increased incidence of skin involvement and renal crisis.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Microwell plate reader capable of measuring OD at 450nm (and 620nm for dual wavelength readings)

I. Device Description:
The QUANTA Lite RNA Polymerase III ELISA consists of a polystyrene microwell ELISA plate coated with purified recombinant immunodominant fragment of RNA Pol III antigen; ELISA negative, low positive and high positive controls; sample diluent; wash concentrate; goat anti-human IgG horseradish peroxidase conjugate; TMB chromogen; and stop solution.

{1}

J. Substantial Equivalence Information:

1. Predicate device name(s):
MBL International Corp. RNA Polymerase III ELISA

2. Predicate 510(k) number(s):
k060431

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | QUANTA Lite RNA Polymerase III ELISA | MBL RNA Polymerase III ELISA  |
|  Intended Use | The detection of anti-RNA Polymerase III antibodies | Same  |
|  Capture antigen | Purified recombinant fragment of RNA Pol III | Same  |
|  Sample matrix | Serum | Same  |
|  Units | Arbitrary ELISA Units | Same  |
|  Method | ELISA | Same  |
|  Solid phase | Polystyrene microwell strips | Same  |
|  Substrate | TMB Chromogen | Same  |
|  Stop solution | 0.344M sulfuric acid | 1N sulfuric acid  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | QUANTA Lite RNA Polymerase III ELISA | MBL RNA Polymerase III ELISA  |
|  Indications for Use | To aid in the diagnosis of systemic sclerosis (scleroderma) with increased incidence of skin involvement and renal crisis | To aid in the diagnosis of systemic sclerosis  |
|  Calculation of results | Compared to a cut-off control | Compared to Calibrator 2  |
|  Result interpretation | ≤20 U/mL = negative; 20-39 = weak positive; 40-80 = mod. positive; >80 strong positive | <28 U/mL = negative ≥28 U/mL = positive  |
|  Controls | Negative, low positive, and high positive | Not provided  |
|  Conjugate | Horseradish peroxidase labeled goat anti-human IgG | Horseradish peroxidase labeled goat anti-human IgG, IgA and IgM  |

{2}

3

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Sample diluent | Tris-buffered saline, Tween 20, protein stabilizers and preservative | PBS buffer with Tween 20  |
|  Wash concentrate | Tris-buffered saline and Tween 20 at 40X | PBS buffer with Tween 20 at 10X  |
|  Incubation times | 30-30-30 minutes | 60-60-30 minutes  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced

## L. Test Principle:

Purified recombinant immunodominant fragment of RNA Pol III antigen is bound to the wells of a polystyrene microwell plate. Pre-diluted controls and diluted patient sera are added to separate wells, allowing any anti-RNA Pol III antibodies present to bind to the immobilized antigen. Unbound sample is washed away and an enzyme labeled goat anti-human IgG antibody is added to each well. A second incubation allows the enzyme labeled anti-human IgG to bind to any patient antibodies which have become attached to the microwells. After washing away any unbound enzyme labeled anti-human IgG, the remaining enzyme activity is measured by adding a chromogenic substrate and measuring the intensity of the color that develops. The assay is evaluated by spectrophotometrically measuring and comparing the color intensity that develops in the patient wells with the color in the control wells. Results determined with the assay are interpreted as negative or weak, moderate, or strong positive and are reported in arbitrary units.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Intra-assay performance for the assay was evaluated by testing 9 specimens a total of 9 times each on 2 lots of kits. The samples tested ranged from 3 to 98 Units with %CVs ranging from 2% to 5%.

|   | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean units | 22 | 31 | 26 | 90 | 39 | 34 | 58 | 2 | 8  |
|  SD | 0.7 | 1.3 | 0.6 | 1.5 | 0.6 | 0.9 | 1.5 | 0.1 | 0.1  |
|  CV% | 3% | 4% | 2% | 2% | 2% | 3% | 3% | 5% | 2%  |

Inter-assay variation was assessed by testing 12 specimens 6 separate times over 4 days, for a total of 5 different runs. This was done for 2 lots of kits. Percent CVs ranged from 2% to 5%.

|   | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean Units | 29 | 23 | 32 | 28 | 91 | 40 | 32 | 35 | 62 | 3 | 9 | 4  |
|  SD | 0.5 | 0.5 | 1.6 | 1.2 | 3.5 | 1.8 | 0.9 | 1.7 | 1.4 | 0.2 | 0.4 | 0.2  |
|  CV% | 2% | 2% | 5% | 4% | 4% | 4% | 3% | 5% | 2% | 5% | 4% | 5%  |

{3}

b. Linearity/assay reportable range:

No claims were made regarding linearity for the assay. Reactivity is related to the quantity of antibody present in a non-linear fashion. While increases and decreases in patient antibody concentrations will be reflected in a corresponding rise or fall in reactivity, the change is not proportional (i.e. a doubling of the antibody concentration will not double the reactivity).

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

There is no recognized standard or reference material for anti-RNA Pol III antibodies.

Accelerated stability studies were conducted on 2 lots of antigen coated plates and 3 lots of low and high positive kit controls. Data supported a shelf life of one year.

d. Detection limit:

To determine the detection limit, a high positive sample was serially diluted until further dilutions did not cause a reduction in the O.D. reading. The lower limit was determined to be 2 Units.

e. Analytical specificity:

To demonstrate whether high levels of endogenous substances could cause a false positive result, sera with known quantities of potential interferant were tested. To determine if the presence of these substances could cause a false negative, the substances were added to three sera with known quantities of anti-RNA polymerase III. The results were compared to mixing the positive sera with normal sera.

|  Substance added: | Norm. sera Units | Pos. sera #1 Units | Recovery % | Pos. sera #2 Units | Recovery % | Pos. sera #3 Units | Recovery %  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Mixed with normal sera | — | 55 |  | 42 |  | 66 |   |
|  Hemoglobin 1000 mg/dL | 4 | 54 | 98% | 39 | 93% | 72 | 109%  |
|  Bilirubin 29.7 mg/dL | 4 | 54 | 98% | 40 | 94% | 65 | 98%  |
|  Cholesterol 354 mg/dL | 4 | 55 | 100% | 39 | 93% | 69 | 104%  |
|  Triglycerides 2173 mg/dL | 5 | 58 | 105% | 37 | 88% | 67 | 100%  |
|  High level RF (OD 2.78) | 4 | 57 | 104% | 42 | 100% | 67 | 100%  |

{4}

The package insert contains the following recommendations: “Microbially contaminated, heat-treated, samples with visible particulate should not be used. Grossly hemolyzed or lipemic specimens should be avoided.”

The testing of 102 sera with other autoimmune disease antibodies and infectious disease antibodies included the following: 60 samples with other autoimmune antibodies (RNP, Sm, Sm/RNP, Scl-70, SS-A, SS-B, Jo-1, Ribo P, and chromatin); 42 various infectious disease patient sera (HCV, HSV, CMV, toxoplasmosis, rubella, and parvovirus). Two (from the infectious disease group) of the 102 (2%) were positive in the new assay.

f. Assay cut-off:

The cut-off was established by testing 629 samples. Five-hundred and twenty-seven random blood donors were tested. Two samples were above the 20 Unit cut-off with values of 27 and 39 Units (99.6% were negative). The mean value of the blood donor samples was 3.4 Units and the standard deviation was 2.6 Units. The cut-off was validated by running disease controls which included other autoimmune disease (n=60, 30 RA and 30 SLE) and 42 sera from patients with antibodies to infectious disease organisms including herpes simplex, CMV, toxoplasmosis, rubella, parvovirus, hepatitis C. Two of the infectious disease sera showed low positive results (both at 22 Units). The specificity for all groups tested was 99.5%.

The cut-off was also validated by ROC analysis comparing the new assay to the laboratory radio-immunoprecipitation method with 1356 samples. At the established cut-off of &lt;20 Units = negative, the specificity was 99.4%.

2. Comparison studies:

a. Method comparison with predicate device:

A total of 51 samples were tested in two clinical laboratories on both the INOVA assay and the predicate assay.

|   | MBL Anti-RNA Polymerase III  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  QUANTA Lite RNA Pol III ELISA | Positive | 32 | 0 | 32  |
|   |  Negative | 2 | 17 | 19  |
|   |  Total | 34 | 17 | 51  |
|  Positive percent agreement: |   | 32/34 = 94.1% |   | (95% CI: 80-99%)  |
|  Negative percent agreement: |   | 17/17 = 100% |   | (95% CI: 80-100%)  |
|  Overall agreement: |   | 49/51 = 96.1% |   |   |

b. Additional comparison study:

In this study 1319 patients with systemic sclerosis as well as 37 normal blood donors were tested by a radio-immunoprecipitation (RIP) assay and the new device.

{5}

6

|   | Radio-immunoprecipitation (RIP)  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  QUANTA Lite RNA Pol III ELISA | Positive | 320 | 20 | 340  |
|   |  Negative | 14 | 1002 | 1016  |
|   |  Total | 334 | 1022 | 1356  |

Positive percent agreement: 320/354 = 95% (95% CI: 93-98%)
Negative percent agreement: 1002/1022 = 98% (95% CI: 80-100%)
Overall agreement: 1350/1356 = 97%

c. Matrix comparison:
Both assays use serum as the test matrix.

3. Clinical studies:

a. Clinical Sensitivity and Specificity:
A total of 1737 patients with systemic sclerosis and 666 blood donors or subjects with infectious or autoimmune diseases were tested on the new assay to calculate the clinical sensitivity and specificity.

|   | Disease  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  positive | negative | Total  |
|  RNA pol III | positive | 394 | 4 | 398  |
|   |  negative | 1343 | 662 | 2005  |
|   |  Total | 1737 | 666 | 2403  |

Clinical sensitivity: 394/1737 = 22.7%  95% CI = 20.7-24.7%
Clinical specificity: 662/666 = 99.4%  95% CI = 98.5-99.8%

b. Other clinical supportive data:

Prevalence of anti-RNA Pol III antibodies in various diseases

|   | Literature |   |   | INOVA RNA Pol III ELISA  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Disease | # Total | # anti-RNA Pol III Pos | % Pos | # Total | # anti-RNA Pol III Pos | % Pos  |
|  All SSc | 2323 | 356 | 15% | 1737 | 394 | 23%  |
|  SLE | 177 | 3 | 2% | 30 | 0 | 0%  |
|  MCTD | 49 | 0 | 0% | ND |  |   |
|  PM/DM | 50 | 0 | 0% | ND |  |   |
|  RA | 84 | 1 | 1% | 30 | 0 | 0%  |
|  Primary | 32 | 1 | 3% | ND |  |   |
|  UCTD | 243 | 0 | 0% | ND |  |   |
|  NBD | 360 | 4 | 1% | 564 | 2 | 0.4%  |
|  Nucleolar | 200 | 3 | 2% | ND |  |   |
|  ID | ND |  |  | 42 | 2 | 5%  |
|  All controls | 968 | 12 | 1% | 666 | 4 | 1%  |

{6}

Abbreviations: Pos=Positive; SSc=systemic sclerosis; SLE=systemic lupus erythematosus; MCTD=mixed connective tissue disease; PM/DM=polymyositis/dermatomyositis; RA=rheumatoid arthritis; UCTD=unspecified connective tissue disorder; NBD=normal blood donor; Nucleolar=nucleolar immunofluorescent pattern; ID=infectious disease. References 1 and 5 used ELISA, 2 used Western blot, and 6, 7 and 8 used radio-immuno precipitation.

Anti-RNA Pol III antibodies in Diffuse versus Limited SSc*

|  # anti-RNA Pol III Pos | # Diffuse Cutaneous | % | # Limited or overlap | % | # Renal Crisis | %  |
| --- | --- | --- | --- | --- | --- | --- |
|  319 | 264 | 83% | 55 | 17% | 86 | 27%  |

* based on literature

4. Clinical cut-off:
See assay cut-off and expected values.

5. Expected values/Reference range:
The expected result in the normal population is negative (≤ 20 Units). At the established cut-off of 20 Units, 99.4% of the non-target populations were negative. Refer to section M.3.b. for the incidence of anti-RNA Pol III antibodies in other disease groups and the normal population as found in the literature.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K070066](https://fda-staging.innolitics.com/device/K070066)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
