K070050 · Encore Medical L.P. · LPH · May 23, 2007 · Orthopedic
Device Facts
Record ID
K070050
Device Name
VITALITY HIP SYSTEM MODEL#420
Applicant
Encore Medical L.P.
Product Code
LPH · Orthopedic
Decision Date
May 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; - rheumatoid arthritis; - correction of functional deformity; - femoral fracture . This device may also be indicated in the salvage of previously failed surgical attempts. The Encore hip systems are for total hip replacement except for the bipolar or unipolar hip systems which are for hemi-arthroplasty applications. The CLP Offset, CLP-R, and Stamina hip systems are for either total or hemi-applications.
Device Story
Encore Medical Hip System consists of various modular hip prostheses (stems, cups, liners, heads) for total hip replacement or hemi-arthroplasty. Modification involves updating Instructions for Use (IFU) to consolidate documentation and reflect current clinical practice. Devices are implanted by orthopedic surgeons in clinical/hospital settings. System includes cementless, cemented, and dual-application options. Output is a restored hip joint function. Benefits include relief of disability from degenerative disease, arthritis, or fracture. No changes to device design, materials, or manufacturing processes.
Clinical Evidence
No clinical data provided; this is a labeling-only modification.
Technological Characteristics
Prosthetic hip system components including femoral stems, acetabular cups, liners, and heads. Materials include metal, polymer, and ceramic combinations. Fixation methods include cemented and cementless (porous-coated) options. Design features are identical to previously cleared predicate devices.
Indications for Use
Indicated for patients with disability due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or femoral fracture. Applicable for total hip replacement or hemi-arthroplasty depending on specific system configuration. Indicated for salvage of failed surgical attempts.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
# Summary of Safety and Effectiveness
Date: April 18, 2007
Manufacturer: Encore Medical, L.P. 9800 Metric Blvd Austin, TX 78758
Contact Person: Teffany Hutto Regulatory Affairs Specialist Phone: (512) 834-6255 Fax: (512) 834-6313 Email: Teffany Hutto@encoremed.com
| Product | 510(k) Number, Clearance Date, Classification | Product Code |
|-----------------------------------------------------------------------|-------------------------------------------------------------------|---------------|
| CLP I (Q System 28 Hip Stem) & CLP-R | K910010 – March 12, 1991<br>Class II | LWJ |
| All Poly Acetabular Cup | K934169 – May 3, 1994<br>K942611 – September 28, 1994<br>Class II | JDI |
| Foundation Cemented Hip Stem | K935449 – March 30, 1995<br>Class II | JDI |
| Bipolar | K953510 – August 9, 1995<br>Class II | KWY |
| Vitality Hip Stem | K962560 – September 19, 1996<br>Class II | JDI |
| Foundation Press-Fit Hip Stem & CoCr Femoral Heads | K973302 – December 2, 1997<br>Class II | LPH, LZO |
| Unipolar with Modular Neck Length Sleeves | K973614 – December 18, 1997<br>Class II | KWI |
| Revelation Hip Stem | K973685 – December 19, 1997<br>Class II | LPH |
| Foundation Fracture Hip Stem | K973809 – January 2, 1998<br>Class II | LWJ |
| FMP Acetabular Shells (Spiked, Hemispherical, and Flared) | K974093, K974095, and K973119 –<br>January 28, 1998<br>Class II | LPH |
| Lincar Hip Stem | K974294 – January 12, 1998<br>Class II | LPH, LZO |
| Stamina Hip Stem | K980473 – April 16, 1998<br>Class II | LZO, LWJ, KWY |
| Keystone Modular Hip Stem | K000521 – May 10, 2000<br>Class II | LPH |
| ALFA II (Omega II Modular Total Hip System) & Modular<br>Femoral Neck | K000817 – June 8, 2000<br>Class II | JDI, LPH |
| R120 (R120 Modular Total Hip System) | K011774 – September 5, 2001<br>Class II | JDI, LPH |
| R120 PC (R120 Modular Total Hip System) | K021822 – July 23, 2002<br>Class II | JDI, LPH |
| FMP Constrained Acetabular Liners | K023794 – April 1, 2003<br>Class II | KWZ |
| Product Code | Regulation and Classification Name |
|--------------|-------------------------------------------------------------------------------------------------------------------|
| LWJ | Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis per 21 CFR 888.3360 |
| JDI | Hip joint metal/polymer semi-constrained cemented prosthesis per 21 CFR 888.3350 |
| KWY | Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis per 21 CFR 888.3390 |
| LPH | Hip joint metal/polymer/metal semi-constrained porous coated uncemented prosthesis per 21 CFR 888.3358 |
| LZO | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis, per 21 CFR 888.3353 |
| KWL | Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis per 21 CFR 888.3360 |
| KWZ | Hip joint metal/polymer constrained cemented or uncemented prosthesis per 21 CFR 888.3310 |
{1}------------------------------------------------
### Summary of Safety and Effectiveness - cont. Encore Medical - Hip System IFU
The modification consists of a change to the Instructions for Use for the devices Description: listed above to minimize the necessity for multiple IFU's and to update the contents to reflect current practice.
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis . of the natural femoral head;
- rheumatoid arthritis; .
- correction of functional deformity; .
- femoral fracture .
This device may also be indicated in the salvage of previously failed surgical attempts.
The Encore hip systems are for total hip replacement except for the bipolar or unipolar hip systems which are for hemi-arthroplasty applications. The CLP Offset, CLP-R, and Stamina hip systems are for either total or hemi-applications.
The following Hip Systems are for Cementless application:
| Linear | Stamina |
|----------------------|---------------------|
| Foundation Press-Fit | CLP Offset |
| Keystone Modular | CLP I |
| Revelation | Foundation Fracture |
The following hip systems are for cemented application: Foundation Cemented Vitality
The following hip systems are for either cementless applications: CLP R Alpha II R120 PC R120
Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same materials, design, indications, packaging, and sterilization.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is presented in black and white, giving it a formal and official appearance.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 2 3 2007
Encore Medical, L.P. % Ms. Teffany Hutto Regulatory Affairs Specialist 9800 Metric Boulevard Austin, Texas 78758
Re: K070050
Trade/Device Name: Encore Medical Hip System IFU Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LWJ, LZO, JDI, KWY, KWL, KWZ Dated: April 18, 2007 Received: May 9, 2007
Dear Ms. Hutto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Teffany Hutto
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
barbare Rouchup
Mark N. Melk Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): ___K070050
Device Name: Encore Medical Hip System IFU
Indications for Use:
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral . head;
- rheumatoid arthritis; .
- correction of functional deformity;
(Part 21 CFR 801 Subpart D)
- femoral fracture .
This device may also be indicated in the salvage of previously failed surgical attempts.
The Encore hip systems are for total hip replacement except for the bipolar or unipolar hip systems which are for hemiarthroplasty applications. The CLP Offset, CLP-R, and Stamina hip systems are for either total or hemi-applications.
| The following Hip Systems are for Cementless application: | | | |
|-------------------------------------------------------------------------------|---------------------|------------------|----------------------|
| Linear | Stamina | Keystone Modular | Revelation |
| Foundation Press-Fit | CLP Offset | CLP I | Foundation Fracture |
| The following hip systems are for cemented application: | | | |
| Vitality | Foundation Cemented | | |
| The following hip systems are for either cemented or cementless applications: | | | |
| Alpha II | CLP R | R120 | R120PC |
| Prescription Use | X | AND/OR | Over-The-Counter Use |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jubacebneupforma
Division Sign-Off
Restorative, and Neurological Devices
Page 1 of 1
(21 CFR 801 Subpart C)
(Posted November 13, 2003)
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.