← Product Code [NQG](/productcode/NQG) · K070036

# REPROCESSED MEDTRONIC CARDIAC STABILIZATION AND POSITIONING DEVICES (K070036)

_Ascent Healthcare Solutions · NQG · Aug 17, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070036

## Device Facts

- **Applicant:** Ascent Healthcare Solutions
- **Product Code:** [NQG](/productcode/NQG.md)
- **Decision Date:** Aug 17, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4500
- **Device Class:** Class 1
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Reprocessed Cardiac Stabilization and Positioning Devices are indicated for use during performance of minimally invasive cardiovascular surgery through a sternotomy incision approach on the non-arrested heart. Sternum retractor blades are used to provide access to the thoracic cavity and to provide a mount for the tissue stabilizer and heart positioner. It also facilitates the positioning of pericardial sutures. The tissue stabilizer is used to stabilize and minimize the motion of selected sites on the beating heart. The heart positioner aids in positioning the heart by the application of vacuum suction when positioned on the heart.

## Device Story

Reprocessed cardiac stabilization and positioning devices used in minimally invasive coronary artery bypass grafting. Components include tissue stabilizers, heart positioners, and sternum retractor blades. Device provides retraction and stabilization of beating heart; heart positioner utilizes vacuum suction. Used in OR by surgeons. Reprocessing involves cleaning, decontamination, and functional testing of previously used devices to restore to original performance specifications.

## Clinical Evidence

No clinical data. Bench and laboratory testing conducted, including biocompatibility validation, sterilization validation, functional testing, and packaging validation.

## Technological Characteristics

Mechanical cardiovascular surgical instruments. Components: tissue stabilizer, heart positioner, sternum retractor blades. Materials and design identical to original predicate devices. No software or energy source.

## Regulatory Identification

Cardiovascular surgical instruments are surgical instruments that have special features for use in cardiovascular surgery. These devices include, e.g., forceps, retractors, and scissors.

## Predicate Devices

- Medtronic Octopus TE Stabilizer & Accessories ([K041338](/device/K041338.md))
- Medtronic Octopus Tissue Stabilizer ([K964445](/device/K964445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070036

## SECTION 5: 510(k) SUMMARY

AUG | 7 2007

| Submitter:           | Ascent Healthcare Solutions<br>10232 South 51st Street<br>Phoenix, Arizona 85044                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Katie Bray<br>RA Specialist/Biomedical Engineer<br>(480) 763-5300 (o)<br>(480) 763-6089 (f)<br>kbray@ascenths.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date of preparation: | December 29, 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Name of device:      | Trade/Proprietary Name: Reprocessed Cardiac Stabilization and<br>Positioning Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                      | Classification Name: Cardiovascular surgical instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| K041338              | 510(k) Title<br>Medtronic Octopus TE Stabilizer &<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                      | Manufacturer<br>Medtronic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| K964445              | Medtronic Octopus Tissue Stabilizer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                      | Medtronic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                      | Note: These are class I devices that do not require a 510(k) for original (brand new) devices. Therefore, the predicate device may not have a filed 510(k) from the original manufacturer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device description:  | Cardiac stabilization and positioning devices are specially<br>designed devices used in minimally invasive cardiac surgery for<br>coronary artery bypass grafting. The devices offer a retraction<br>as well as a stabilizing function to control the movement of the<br>beating heart. The cardiac stabilization and positioning devices<br>consist of a combination of one or more of the following<br>components: tissue stabilizer, heart positioner, and sternum<br>retractor blades.                                                                                                                                                                                                     |
| Indications for Use: | Reprocessed Cardiac Stabilization and Positioning Devices are<br>indicated for use during performance of minimally invasive<br>cardiovascular surgery through a sternotomy incision approach<br>on the non-arrested heart. Sternum retractor blades are used to<br>provide access to the thoracic cavity and to provide a mount for<br>the tissue stabilizer and heart positioner. It also facilitates the<br>positioning of pericardial sutures. The tissue stabilizer is used to<br>stabilize and minimize the motion of selected sites on the<br>beating heart. The heart positioner aids in positioning the heart<br>by the application of vacuum suction when positioned on the<br>heart. |

CONFIDENTIAL

, … "

:

{1}------------------------------------------------

| Technological<br>characteristics: | The design, materials, and intended use of Reprocessed<br>Cardiac Stabilization and Positioning Devices are identical to the<br>predicate devices. The mechanism of action of Reprocessed<br>Cardiac Stabilization and Positioning Devices is identical to the<br>predicate devices in that the same standard mechanical design,<br>materials, and size are utilized. There are no changes to the<br>claims, intended use, clinical applications, patient population,<br>performance specifications, or method of operation. In addition,<br>Ascent Healthcare Solutions's reprocessing of Cardiac<br>Stabilization and Positioning Devices includes removal of<br>adherent visible soil and decontamination. Each individual<br>Cardiac Stabilization and Positioning Device is tested for<br>appropriate function of its components prior to packaging and<br>labeling operations. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance data:                 | Bench and laboratory testing was conducted to demonstrate<br>performance (safety and effectiveness) of Reprocessed Cardiac<br>Stabilization and Positioning Devices. This included the<br>following tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                   | Biocompatibility Validation of reprocessing Sterilization Validation Function test(s) Packaging Validation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                   | Performance testing demonstrates that Reprocessed Cardiac<br>Stabilization and Positioning Devices perform as originally<br>intended.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Conclusion:                       | Ascent Healthcare Solutions concludes that the modified<br>devices (Reprocessed Cardiac Stabilization and Positioning<br>Devices) are safe, effective, and substantially equivalent to the<br>predicate devices as described herein.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

CONFIDENTIAL

:

:

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 7 2007

Ascent Healthcare Solutions c/o Ms. Katie Bray Regulatory Affairs Engineer 10232 South 51st Street Phoenix, AZ 85044

Re: K070036

Reprocessed Medtronic Cardiac Stabilization and Positioning Devices Regulation Number: 21 CFR 870.4500 Regulation Name: Cardiovascular Surgical Instruments Regulatory Class: Class I non-exempt Product Code: NQG Dated: July 20, 2007 Received: July 23, 2007

Dear Ms. Bray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

B.Jemimon for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| ProCode | Family                                            | Model                     |
|---------|---------------------------------------------------|---------------------------|
| NQG     | Medtronic Heart<br>Stabilizers and<br>Positioners | Octopus 3.0 Stabilizer    |
|         |                                                   | Octopus 4.0 Stabilizer    |
|         |                                                   | Octopus 4.3 Stabilizer    |
|         |                                                   | Starfish Heart Positioner |
|         |                                                   | Starfish 2 Positioner     |
|         |                                                   | Evolution                 |
|         |                                                   | Urchin Positioner         |

{5}------------------------------------------------

## SECTION 4: INDICATIONS FOR USE STATEMENT

## 510(k) Number (if known): 《(070036

Device Name: Reprocessed Medtronic Cardiac Stabilization and Positioning Devices

## Indications For Use:

Reprocessed Cardiac Stabilization and Positioning Devices are indicated for use during performance of minimally invasive cardiovascular surgery through a sternotomy incision approach on the non-arrested heart. Sternum retractor blades are used to provide access to the thoracic cavity and to provide a mount for the tissue stabilizer and heart positioner. It also facilitates the positioning of pericardial sutures. The tissue stabilizer is used to stabilize and minimize the motion of selected sites on the beating heart. The heart positioner aids in positioning the heart by the application of vacuum suction when positioned on the heart.

B.H.mmanama

Division Sign-Division of Cardiove 510(k) Number

AND/OR

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

CONFIDENTIAL

Ascent Healthcare Solutions Reprocessed Cardiac Stabilization and Positioning Devices Traditional 510(k)

Page 10

---

**Source:** [https://fda-staging.innolitics.com/device/K070036](https://fda-staging.innolitics.com/device/K070036)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).