CLASSIC/HYPOTHERMIA, FLEXIBLE, 10-SECOND, LEFT-RIGHT HANDED, BASAL, LIGHTWEIGHT-PROBE DIGITAL CLINICAL THERMOMETER
K070028 · Actherm, Inc. · FLL · Feb 8, 2007 · General Hospital
Device Facts
Record ID
K070028
Device Name
CLASSIC/HYPOTHERMIA, FLEXIBLE, 10-SECOND, LEFT-RIGHT HANDED, BASAL, LIGHTWEIGHT-PROBE DIGITAL CLINICAL THERMOMETER
Applicant
Actherm, Inc.
Product Code
FLL · General Hospital
Decision Date
Feb 8, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
For all variant Thermometers the intended use is as medical device supplied by internal power and intended to precisely measure human body temperature. It can be used in the measurement of oral, axillary and rectal temperature. They are used alone for human beings at all ages, especially professional medical practitioners, to diagnosis or observe human states of health. All kinds of Actherm Digital Clinical Thermometer help to detect and to make accurate judgements of states of health, protect user safety and public health.
Device Story
Digital clinical thermometers consisting of a temperature sensor in a metal probe (flexible or inelastic) connected to an IC unit and LCD within a pen-like plastic housing. Devices measure human body temperature (oral, axillary, rectal). Variants include hypothermia monitoring, 10-second rapid measurement, left/right-handed display orientation, and high-accuracy basal temperature monitoring. Operated by professional medical practitioners or patients for self-use. Input is thermal energy; device processes signal via internal IC to display maximum temperature value on LCD. Non-predictive operation. Provides objective temperature data to assist in health state assessment and clinical decision-making.
Clinical Evidence
Bench testing only. Compliance with ASTM E1112-00, EN 12470-3, IEC 60601-1, IEC 60601-1-2, and ISO 10993-1 standards demonstrated through verification and validation testing.
Technological Characteristics
Electronic clinical thermometers with metal probe sensors (flexible or inelastic), IC processing unit, and LCD. Powered by internal battery. Form factor: pen-like plastic housing (except round housing for Lightweight-Probe model). Standards: ASTM E1112-00, EN 12470-3, IEC 60601-1, IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for human beings of all ages for the measurement of oral, axillary, and rectal body temperature to diagnose or observe health states.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Actherm Digital Clinical Thermometer, Model ACT2000, ACT2020 et al. (K010238)
Digital Clinical Thermometer, Model ACT2030, ACT2038 (K021612)
Digital Clinical Thermometer, Model ACT3020 (K021614)
Digital Clinical Thermometer, Model ACT2130, ACT3136 et al. (K031905)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K070028" in a handwritten style. The characters are bold and slightly slanted, giving the impression of quick handwriting. The text appears to be a code or identifier, possibly a serial number or reference code.
# Attachment 4
### 510 (k) Summary
#### Submitter's Identification
8 2007
Actherm Inc. 6th F, # 85, Kuan-Min 6 Road, Jubei, 302 HsinChu Taiwan Tel: +886 3 553 7315 +886 3 553 7319 Fax:
Contact Person: Richard Hsieh
Date prepared: December 25, 2006
#### Name of the devices
Classic/Hypothermia Digital Clinical Thermometer, Model ACT2000/ACT2000+, ACT2010/ACT2010+, ACT2020/ACT2020+, ACT2030/ACT2030+, ACT2130/ACT2130+, ACT2038/ACT2038+;
Flexible Digital Clinical Thermometer, Model ACT3020/ACT3020+, ACT3030/ACT3030+, ACT3136/ACT3136+, ACT3020 Express, ACT3020 Express, ACT3030 Express, ACT3136 Express, ACT3230 Express;
10-Second Digital Clinical Thermometer, Model ACT2010 Express, ACT2020 Express, ACT2030 Express, ACT2130 Express, ACT2230 Express, ACT2330 Express;
Left-Right Handed Digital Clinical Thermometer, Model ACT2030 LR/ACT2030+ LR, ACT2130 LR/ACT2130+ LR, ACT2230 Express LR, ACT3230 Express LR, ACT3136 Express LR;
Basal Digital Clinical Thermometer, Model ACT2010 Basal, ACT2010 Express Basal, ACT3020 Basal, ACT3020 Express Basal, ACT3230 Basal, ACT3136 Basal, ACT2130 Basal, ACT3136 Express Basal, ACT6070 Basal;
Lightweight-Probe Digital Clinical Thermometer, Model ACT6070 Express;
#### Classification Name
Clinical Electronic Thermometer (21 CFR 880.2910 Product Code FLL)
#### Predicate Device Information
| Actherm Digital Clinical Thermometer, Model ACT2000, ACT2020 et al. | (K010238) |
|---------------------------------------------------------------------|-----------|
| Digital Clinical Thermometer, Model ACT2030, ACT2038 | (K021612) |
| Digital Clinical Thermometer, Model ACT3020 | (K021614) |
| Digital Clinical Thermometer, Model ACT2130, ACT3136 et al. | (K031905) |
| Lumiscope Model 2018 Thermometer | (K954792) |
{1}------------------------------------------------
A4: 2 of 3
KO 70028
#### evice Description
The Classic/Hypothermia Digital Clinical Thermometer consists of an electronically sensor which is located in a metal probe and is connected with a PCB and a Liquid Crystal Display all situated in a pen-like plastic housing. This structure is used for all kind of Thermometer mentioned in this description.
Hypothermia Thermometer have a greater measuring range so that they can be used as device for under cooled person temperature monitoring.
A flexible probe tip is provided by the Flexible Digital Clinical Thermometer that makes the measuring of the temperature more comfortable and safer.
With the patented 10-Second Digital Clinical Thermometer every user is able to receive the result of the measurement in a very quick time. The Express Models also operates with the vame efficient reliability and accuracy like the other Thermometer.
Due to a special integrated sensor in the Left-Right Handed Digital Clinical Thermometer display can easily be read whether the Thermometer is held in the right or the left hand. Especially for left-handed people it makes the measuring result read off more pleasant.
The Basal Digital Clinical Thermometer have a fourth digit on the display, because the measurement of this Thermometer is even more accurate compared with normal Thermometer. This precisely measuring method can be used to monitor and interpreting basal temperature changes.
The Lightweight-Probe Digital Clinical Thermometer consists of a flexible probe, connection cord and a display unit embedded in a round housing.
Variants Descriptions: "+" or Hypothermia, Express, Dual scale, Backlight, LR, following with the model numbers indicates different measuring range, different measuring time, scale switchable, with backlight with Left-Right Handed function.
For all Thermometers is essential, that they consist of a temperature sensor embedded in a special designed metal probe, which could be flexible or inelastic. This sensor is connected with the IC unit, which is connected with a Liquid Crystal Display (LCD) to display the measured temperature. All parts are situated in a pen-like plastic housing. During measuring, only the maximum measured value is displayed on LCD window and all devices are non-predictive.
All variant thermometers comply with referenced product standards of ATSM E1112-00, EN12470-3, IEC 60601-1, IEC 60601-1-2 and ISO 10993-1.
#### Intended Use
For all variant Thermometers the intended use is as medical device supplied by internal power and intended to precisely measure human body temperature. It can be used in the measurement of oral, axillary and rectal temperature. They are used alone for human beings at all ages, especially professional medical practitioners, to diagnosis or observe human states of health. All kinds of Actherm Digital Clinical Thermometer help to detect and to make accurate judgements of states of health, protect user safety and public health.
{2}------------------------------------------------
44: 3 UT 3
K070028
#### `omparison to Predicate Devices
### The Classic/Hypothermia, Flexible, 10-Second, Left-Right Handed and the Basal
Digital Clinical Thermometer are substantial equivalent to the Predicate Devices Actherm Digital Clinical Thermometer. All new devices use the same kind of technology and have the same design as the predicate devices. (Exception is Model ACT3230 Express which has a new, slightly changed design.) All devices are only supported by a new Model feature in comparison to the redicate devices to increase the device performances. I.e. the Basal Digital Thermometer has poligher accuracy. The material, chemical composition and the energy source remains the used Predicate Devices.
The Lightweight-Probe Digital Clinical Thermometer is substantial equivalent to the Predicate Device Lumiscope Model 2018 Thermometer (K954792). Only the design and the used technology are adapted to the Actherm Inc. corporate policy. The material and the chemical commonition remains the same as the used Predicate Devices.
#### Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence is as follows:
Compliance to applicable voluntary standards includes ASTM E1112-00, EN 12470-3 as well as IEC 60601-1, IEC 606001-1-2 and ISO 10993-1 requirements.
Guidance Documents included the "FDA Guidance on the Content of the Premarket Notification 510 (k) Submissions for Clinical Electronic Thermometers".
#### Jonclusion
Classic/Hypothermia Digital Clinical Thermometer, The Flexible Digital Clinical Thermometer, 10-Second Digital Clinical Thermometer, Left-Right Handed Digital Clinical Thermometer, Basal Digital Clinical Thermometer and the Lightweight-Probe Digital Clinical Thermometer with all corresponding Models have the same intended use and similar technological characteristics as the cleared devices of Actherm Inc. Digital Clinical Thermometer. Moreover, verification and validation tests contained in this submission clearly demonstrate that any differences in their characteristics do not raise any new questions of safety or effectiveness. Furthermore, those engineering differences do not affect the intended use or alter the fundamental scientific technology of the cleared devices of Actherm Inc. Digital Clinical Thermometer.
Thus, the Classic/Hypothermia Digital Clinical Thermometer, Flexible Digital Clinical Thermometer, 10-Second Digital Clinical Thermometer, Left-Right Handed Digital Clinical Thermometer, Basal Digital Clinical Thermometer, and the Lightweight-Probe Digital Clinical Thermometer with all corresponding Models are substantially equivalent to the Predicate Devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Richard Hsieh Vice President Actherm, Incorporated 6" F, #85 Kuan-Min 6 Road Jubei, HsinChu 302 TAIWAN
FEB 8 2007
Re: K070028
Trade/Device Name: Classic/Hypothermia Digital Clinical Thermometer; Flexible Digital Clinical Thermometer; 10-Second Digital Clinical Thermometer; Left Right Handed Digital Clinical Thermometer; Basal Digital Clinical Thermometer; Lightweight-Probe Digital Clinical Thermometer Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 1, 2007 Received: February 5, 2007
Dear Mr. Hsieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Hsieh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Kunoe
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Attachment 2
Indications for Use Statement
510 (k) Number (if known)
Device Name
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br>1<br>1<br>Children and | - 4 | :<br>-----<br>CAR<br>A B A BREAD - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All - All | Sun and States of States of Status | | --------------------------- |
|---|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 1 | '<br>- | - M | 1 | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | AND AND A CHANAL AND AND AND AND AND A BROAD A BROAD A BROAD A BREAD A BREAD A BREACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH A BEACH | |
Classic/Hypothermia Digital Clinical Thermometer,
Flexible Digital Clinical Thermometer,
10-Second Digital Clinical Thermometer,
Left-Right Handed Digital Clinical Thermometer,
Basal Digital Clinical Thermometer,
Lightweight-Probe Digital Clinical Thermometer;
Indications for Use
All Digital Clinical Thermometers are intended to precisely measure human body temperature. The Thermometers can be used in the measurement of oral, axillary and rectal temperature.
Prescription Use (Per 21 CFR 801 Subpart D) OR
Over-The-Counter Use × (Per 21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cim W.
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.