ITALITE , MODEL 500-3.1, 550-3.1, 600-3.1, 650-3.1, 450-2.0,400-2.0
K070021 · Jg Medical Products, LLC · HRX · Dec 7, 2007 · Orthopedic
Device Facts
Record ID
K070021
Device Name
ITALITE , MODEL 500-3.1, 550-3.1, 600-3.1, 650-3.1, 450-2.0,400-2.0
Applicant
Jg Medical Products, LLC
Product Code
HRX · Orthopedic
Decision Date
Dec 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Indications for Use
The ITALITE tm Light Wand is indicated for illumination during joint examinations, arthroscopies, biopsies, in minimally invasive procedures of the knee, shoulder, temporal-mandibular joint, ankle, and elbow. The ITALITE tm Light Wand is disposable, single use and is intended to be discarded after the procedure. It is supplied clean but not sterile and must be sterilized before use. The ITALITE tm is for illumination only and is not indicated for viewing. It is to be used by physicians only.
Device Story
ITALITE Light Wand is a semi-rigid, single-use fiber optic device providing illumination for minimally invasive joint procedures. Input consists of light from external, off-the-shelf light sources (e.g., ACMI, Wolff, Stortz, Olympus) connected to the proximal end. The device features a stainless steel tube (2.0 mm or 3.1 mm diameter) containing 50-micron glass fibers. Light is transmitted through the fibers and emitted from a distal window oriented perpendicular to the shaft, providing side-directed illumination. The device is operated by physicians in clinical settings. It does not provide viewing capabilities. By providing localized illumination, it assists the physician in visualizing the surgical field during joint examinations or biopsies, potentially improving procedural accuracy and patient outcomes. The device is supplied clean but non-sterile and requires sterilization by the user prior to each procedure.
Clinical Evidence
Bench testing only. Performance tests confirmed the device met functional specifications, and safety tests ensured compliance with applicable industry standards. Sterilization validation protocol confirmed the device's ability to be sterilized.
Technological Characteristics
Semi-rigid fiber optic light wand; stainless steel tube (2.0 mm or 3.1 mm diameter); 50-micron glass fibers; epoxy-sealed/polished proximal and distal ends; perpendicular distal light emission; compatible with standard light sources. Designed to meet IEC60601-2-18. Manufactured under an approved Quality System. Non-sterile, single-use, disposable.
Indications for Use
Indicated for illumination during joint examinations, arthroscopies, and biopsies in minimally invasive procedures of the knee, shoulder, temporal-mandibular joint, ankle, and elbow. For physician use only. Not for viewing.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
Medscope 4mm Arthroscope (k894303)
ASAP Arthroscope (k031972)
Submission Summary (Full Text)
{0}------------------------------------------------
K07-0021
Pg 1 of 5
ITALITE tm 510k JG Medical Products LLC
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS
0EC 17 2007
This summary of 510k safety and effectiveness information is being submitted in accordance with 21CFR part 807.92
1. Submitters name, address, phone number, contact person and preparation date:
| | Name: JG Medical Products LLC. | | | |
|------------------------------------|-----------------------------------|--|--|--|
| | 110 E. Main Street, Suite 7 | | | |
| Huntington, N.Y. 11743 | | | | |
| | Phone: 631 673 5255 | | | |
| | Fax: 631 673 4949 | | | |
| | Email: wstern@multigon.com | | | |
| | Responsible person: William Stern | | | |
| Official correspondent: | | | | |
| | William Stern | | | |
| | Multigon Industries, Inc. | | | |
| 1 Odell Plaza | | | | |
| | Yonkers, N.Y. 10701 | | | |
| | Phone: 914 376 5200 ext. 27 | | | |
| | Fax: 914 376 6111 | | | |
| | Email: wstern@multigon.com | | | |
| Date of Preparation: | | | | |
| | 12/5/07 | | | |
| 2. Proprietary Name: | | | | |
| | ITALITE tm Models:<br>500-3.1 | | | |
| | 550-3.1 | | | |
| | 600-3.1 | | | |
| | 650-3.1 | | | |
| | 450-2.0 | | | |
| | 400-2.0 | | | |
| Common /Usual Name:<br>Arthroscope | | | | |
| | | | | |
{1}------------------------------------------------
K070021
Pg 2 of 5
ITALITE tm 510k JG Medical Products LLC
Classification Name: 21 CFR888.1100 Arthroscopes
Classification Number: 90HRX
3. Substantially Equivalent Devices
JG Medical Products LLC believes that the ITALITE tm light wand Models 500-3.1; 550-3.1; 600-3.1; 650-3.1; 450-2.0; 400-2.0 is substantially equivalent to the following cleared devices:
| Trade or Proprietary Name | Manufacturer | 510(k) Number |
|---------------------------|--------------------------|---------------|
| Medscope 4mm Arthroscope | Medscope Corp. | k894303 |
| ASAP Arthroscope | ASAP Endoscopic Products | k031972 |
All of the above predicate Arthroscopes have two fiber optic channels, one for viewing and one for illuminating the subject field. The ITALITE tm light wand models above are made of similar materials, fibers, and epoxies.to the predicate devices above. In fact the ITALITE tm light wand uses exactly the same materials and construction methods as the Medscope 4mm Arthroscope k894303, However the ITALITE tm light wand Models have only an illumination channel and are not used for viewing. Illumination is perpendicular to the shaft. Substantial equivalence is claimed for the indications for use for this application.
{2}------------------------------------------------
K070021
P) 3 21-5
ITALITE tm 510k JG Medical Products LLC
# 4. DEVICE DESCRIPTION
The ITALITE tm models of Light Wands range in size as follows:
| MODEL | LENGTH | DIAMETER |
|---------|--------|----------|
| 650-3.1 | 650 mm | 3.1 mm |
| 600-3.1 | 600 mm | 3.1 mm |
| 550-3.1 | 550 mm | 3.1 mm |
| 500-3.1 | 500 mm | 3.1 mm |
| 450-2.0 | 450 mm | 2.0 mm |
| 400-2.0 | 400 mm | 2.0 mm |
The ITALITE tm series of Light Wands are semi rigid type devices which are intended for illumination during joint examinations, biopsies, in minimally invasive procedures of the knee, shoulder, temporal mandibular joint, ankle, and elbow.
The device consists of a thin walled stainless steel tube (either 3.1 mm or 2.0 mm) varying in length from 400 to 650 mm depending on the model number as above. 50 micron glass fibers fill the inner diameter of the tube. At the proximal end the fibers are sealed with epoxy and polished. They are terminated in a light pipe connector compatible with off the shelf light sources such as those manufactured by ACMI, Wolff, Stortz, and Olympus. A stainless steel knob at the proximal end facilitates the direction and placement of the Light Wand.
The distal end of the Light Wand has a window perpendicular to the shaft where the fibers terminate. They are epoxied and polished to form a window perpendicular to the shaft. Hence the light is directed perpendicular to the shaft and provides illumination in that direction instead of head on from the end of the Light Wand.
{3}------------------------------------------------
K070021 pg 4 of 5
ITALITE tm 510k JG Medical Products LLC
#### 5. PERFORMANCE STANDARDS
No performance standards have been established for the ITALITE tm Light Wands under section 514 of the Federal Food and Drug Act. However the ITALITE tm Light Wands have been designed to meet the following standards:
Applicable portions of IEC60601-2-18 Manufacturing under an approved Quality System
#### 6. INDICATIONS FOR USE
The ITALITE tm Light Wand is indicated for illumination during joint examinations, arthroscopies, biopsies, in minimally invasive procedures of the knee, shoulder, temporalmandibular joint, ankle, and elbow .
The ITALITE tm Light Wand is disposable, single use and is intended to be discarded after the procedure. It is supplied clean but not sterile and must be sterilized before use.
The ITALITE tm is for illumination only and is not indicated for
viewing.
It is to be used by physicians only.
#### 7. CONTRA-INDICATIONS
None known at this time.
#### 8. COMPARISON TO PREDICATE DEVICES
The ITALITE tm series of Light Wands have the same device characteristics as the approved predicate devices listed in item 3 above with the commonality of the illumination modality . principles of operation, and materials.
{4}------------------------------------------------
KOFOD'S
105 of
ITALITE tm 510k JG Medical Products LLC
The ITALITE tm series of Light Wands are constructed of materials of the same specifications as the predicate devices to ensure biocompatibility. The ITALITE tm Light Wands conforms to applicable ISO standards. The device will be sold non-sterilized before use by the user. The ability to sterilize the device has been confirmed by the recommended sterilization validation protocol.
Performance tests designed to ensure that the device met all of its functional specifications. Safety tests have been performed to ensure the device complies with applicable industry and safety standards.
The ITALITE tm Light Wand series device labeling includes instructions for safe and effective use, warnings, cautions and guidance for use. It has therefore shown to be safe and effective.
#### 9. SOFTWARE
The ITALITE tm series of Light Wands have no software or firmware associated with them.
## 10. LITERATURE REVIEW
A review of the literature pertaining to the safety and effectiveness of the ITALITE tm Arthroscopes has been conducted and appropriate safeguards have been incorporated in the design of the ITALITE tm Light Wands.
#### 11. CONCLUSIONS
The conclusion drawn from these tests is that the ITALITE tm Light Wands are substantially equivalent in safety and efficacy to the predicate devices listed in item 3 above.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JG Medical Products, LLC % Multigon Industries, Inc. William Stern Official Correspondent l Odell Plaza Yonkers, New York 10701
DEC 7 2007
Re: K070021
Trade/Device Name: ITALITE™ Model 500-3.1, 550.3.1, 600-3.1, 650-3.1, 450-2.0 Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: November 28, 2007 Received: November 30, 2007
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. William Stern
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N Millham
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K070021 pg 1 of 1
ITALITE tm 510k JG Medical Products LLC
32
# Indications for Use
510(k) Number (if known): k070021 TM Device Name: ITALITE LIGHT WAND
Indications For Use:
The ITALITE tm Light Wand is indicated for illumination during joint examinations. arthroscopies, biopsies, in minimally invasive procedures of the knee, shoulder, temporal-mandibular joint, ankle, and elbow.
The ITALITE tm Light Wand is disposable, single use and is intended to be discarded after the procedure. It is supplied clean but not sterile and must be sterilized before use.
The ITALITE tm is for illumination only and is not indicated for viewing. It is
to be used by physicians only.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CD®1vigfiga aff and Neurological Device
**510(k) Number** |
$$k$$
070021|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.