MODULAR DISINFECTION SYSTEM (MDS) FOR ENDOSCOPE REPROCESSING
K063876 · Medivators Reprocessing Systems · FEB · Jun 14, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K063876
Device Name
MODULAR DISINFECTION SYSTEM (MDS) FOR ENDOSCOPE REPROCESSING
Applicant
Medivators Reprocessing Systems
Product Code
FEB · Gastroenterology, Urology
Decision Date
Jun 14, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Medivators Reprocessing Systems Modular Disinfection System (MDS) for Endoscope Reprocessing tests, disinfects and rinses flexible endoscopes, such as fiberoptic and video endoscopes between patient uses. The MDS is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes. It is indicated for use with FDA cleared, liquid high level disinfectants. The MDS does not replace manual cleaning of endoscopes, they must be precleaned and cleaned according to the endoscope manufacturer's instructions and professional guidelines prior to reprocessing them in the MDS.
Device Story
Modular Disinfection System (MDS) is an electro-mechanical device for automated high-level disinfection of flexible endoscopes. Input: pre-cleaned flexible endoscopes. Operation: software-controlled cycle performs leak testing, blockage testing, connectivity verification, detergent flush, high-level disinfection, and rinsing. Asynchronously processes two scopes. Output: disinfected endoscopes and printed records of cycle results (serial number, test results). Used in healthcare facilities; operated by trained personnel. Safety features include cycle interruption and alarms for compromised disinfection conditions (e.g., filter blockage, connectivity failure). Benefits: ensures standardized, documented high-level disinfection, reducing risk of cross-contamination between patient uses.
Clinical Evidence
Bench testing only. Efficacy validated via simulated-use and in-use testing per FDA 1993 guidance. Simulated-use demonstrated >10^6 log reduction of Mycobacterium terrae. In-use testing confirmed no viable organisms post-cycle. Material compatibility and biocompatibility studies confirmed no significant deterioration and safe disinfectant residue levels. Performance testing verified leak, blockage, connectivity, disinfection, rinse, and drying cycles, including alarm functionality for failure modes.
Technological Characteristics
Electro-mechanical automated endoscope reprocessor. Features water filtration system, connector blocks for scope interface, and integrated printer. Compatible with FDA-cleared liquid high-level disinfectants. Software-based control system manages cycle parameters (temperature, time, flow).
Indications for Use
Indicated for high-level disinfection of heat-sensitive, semi-critical flexible endoscopes (fiberoptic and video) between patient uses. Requires prior manual cleaning per manufacturer instructions. Not for rigid endoscopes.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
K063876
p. 1 of 4
# Medivators Reprocessing Systems Modular Disinfection System for Endoscope Reprocessing
# 510(k) Summary of Safety and Effectiveness
JUN 1 4 2007
| Manufacturer: | Medivators Reprocessing Systems, a Division of Minntech Corporation |
|-------------------|---------------------------------------------------------------------|
| Address: | 14605 28th Avenue North<br>Mpls., MN 55447<br>USA |
| Official Contact: | Lynn Lueders |
Director, Regulatory Affairs
Medivators has supplied the following information to the U.S. Food and Drug Administration to support substantial equivalence of the Medivators Reprocessing Systems Modular Disinfection System (MDS) for Endoscope Reprocessing to other endoscope reprocessors currently marketed in the U.S.
#### 1. Device Description
The Modular Disinfection System (MDS) is an electro-mechanical system intended to test and high level disinfect heat sensitive semi-critical endoscopes. It is not intended for reprocessing rigid endoscopes. The machine can use any FDA cleared, reusable liquid germicide which is on the market that is labeled for high level endoscope disinfection. The testing provided to the FDA used both Medivators Rapicide® High Level Disinfectant and Cidex® OPA High Level Disinfectant.
The MDS does not replace manual cleaning of endoscopes, they must be precleaned and cleaned according to the endoscope manufacturer's instructions and professional guidelines prior to reprocessing them in the MDS.
During its software based reprocessing cycle, the machine tests the endoscopes for leaks and blockages. Before and during the machine eveles, connectivity with the machine through the connector blocks is evaluated. If the scopes pass these tests, they are flushed with water and detergent, disinfected and rinsed. The disinfection cycle takes approximately 30 minutes. The MDS is capable of asynchronously reprocessing two scopes at a time.
{1}------------------------------------------------
K063876
p. 2 of 4
The machine has many built in safety features which stop the cycle and alarm when certain conditions exist which could indicate that disinfection might be compromised. These alarms and causes are defined in the directions for use for the product.
The machine also prints records by endoscope serial number indicating the results of testing, disinfection, number of disinfections, etc. which are required for permanent records.
#### 2. Intended Use
Medivators Reprocessing Systems Modular Disinfection System (MDS) for Endoscope Reprocessing tests, disinfects and rinses flexible endoscopes, such as fiberoptic and video endoscopes between patient uses. The MDS is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes. It is indicated for use with FDA cleared, liquid high level disinfectants.
The MDS does not replace manual cleaning of endoscopes, they must be precleaned and cleaned according to the endoscope manufacturer's instructions and professional guidelines prior to reprocessing them in the MDS.
### 3. Comparison to Another Device in Commercial Distribution Within the United States
The MDS is equivalent in function and indications to the Medivators DSD-91ETM Endoscope Disinfector. This machine was found to be substantially equivalent in 1994 (K914145). Both machines have the same indications for use, the same methods of providing disinfection, can use the same reuseable disinfectants at the same temperatures and times. Both machines can reprocess two scopes asynchronously, one in each bay.
#### Summary of Testing 4.
Medivators has provided testing to show that the MDS is safe and effective for its intended use following the requirements listed in the FDA's Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washers, Washer/Disinfectors, and Disinfectors Intended for Use in Health Care Facilities (dated August 1993). This testing included:
### Efficacy Testing
The MDS was tested using standard methods defined by the FDA guidance document. This testing included simulated use and in use testing.
{2}------------------------------------------------
The simulated use testing showed a >10° log reduction in the organism Mvcobacterium terrae after the endoscopes were reprocessed in the MDS.
663876
In use testing showed no viable organisms remained on the endoscopes following the disinfection cycle.
Additional in-use testing was performed to show that the MDS provided disinfected endoscopes when exposed to the entire endoscope cycle of disinfecting and rinsing.
### Material Compatibility
The effect of disinfectants (Rapicide and Cidex OPA) on the materials used in the MDS was evaluated and showed that the materials had no significant deterioration over their use life. Studies were also presented to show that the filters used in the water filtration system were compatible with common disinfectants.
### Biocompatibility
The amount of disinfectant residue left on endoscopes and in the MDS after the disinfection and rinsing cycles was evaluated and compared to determined safe levels. The results of the testing showed that any remaining residues would not have an effect on patients or users of the machine.
### Performance Data
Data was provided to the FDA to show that the machine performs as required. This evaluation included testing to show that the leak check. blockage check, connectivity checks, disinfection cvcle, rinse cycles, and drying cycles performed correctly. Any error messages were tested to ensure they function properly to notify users of any possible failure modes. Testing was performed to show that all critical parameters of the machine function correctly.
Testing was presented that showed that disinfectants which require heating remained at their required temperature for the length of time required in the labeling for the disinfectant.
Testing was completed that showed that the MDS self disinfection cvcle works properly by disinfecting all areas of the machine, including the water filtration system.
{3}------------------------------------------------
K063876
p. 48/4
Studies were performed to show that the water filtration system will function appropriately over time and that the machine will alarm if water filters are plugged to a point which will lower the water pressure below required input. This study also showed that the filters remained bacterial retentive even when plugged to a point that causes the machine to alarm.
#### 5. Summary of Substantial Equivalence
Minntech Corporation has provided the above information in the form of a 510(k) to support the claim that the MDS is safe and effective when used in accordance with the device labeling.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medivators Reprocessing Systems C/O Ms. Lynn Lueders Director, Regulatory Affairs A Minntech Corporation Business Group 14605 28th Avenue, North Minneapolis, Minnesota 55447-4822
JUN 1 4 2007
Re: K063876
Trade/Device Name: Medivators Reprocessing Systems Modular Disinfection System for Endoscope Reprocessing Regulation Number: 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FEB Dated: May 6, 2007 Received: June 7, 2007
Dear Ms. Lueders:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Lueders
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chris Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if Known): K063876
Device Name:
Medivators Reprocessing Systems Modular Disinfection System for Endoscope Reprocessing
Indications for Use:
Medivators Reprocessing Systems Modular Disinfection System (MDS) for Endoscope Reprocessing tests, disinfects and rinses flexible endoscopes. such as fiberoptic and video endoscopes between patient uses. The MDS is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes. It is indicated for use with FDA cleared, liquid high level disinfectants.
The MDS does not replace manual cleaning of endoscopes, they must be precleaned and cleaned according to the endoscope manufacturer's instructions and professional guidelines prior to reprocessing them in the MDS.
AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-the C
(21 CFR 801
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(
Children of Anesthesiology, General Hospital, infection of Anesthestology, Gene
C10(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.