← Product Code [DZE](/productcode/DZE) · K063861

# GS SYSTEM (K063861)

_Osstem Implant Co., Ltd. · DZE · Sep 21, 2007 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063861

## Device Facts

- **Applicant:** Osstem Implant Co., Ltd.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Sep 21, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

GS System is intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. GS System is for one stage surgical procedures. It is not intended for immediate load.

## Device Story

GS System is a titanium dental implant system surgically placed into the upper or lower jaw bone. It serves as a foundation for single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture prosthetics, and provides abutment support for fixed bridgework. The device is used by dental clinicians in a clinical setting. It features surface treatments (R.B.M. or CellNest) to facilitate integration. The system is designed for one-stage surgical procedures and is not indicated for immediate loading. By providing a stable anchor in the jaw, the device restores dental function and aesthetics for edentulous patients.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and performance were verified through product validations and biocompatibility testing to ensure compliance with applicable international and US regulations.

## Technological Characteristics

Titanium metal endosseous dental implant. Surface treatments: Resorbable Blasting Media (R.B.M.) or CellNest. Designed for one-stage surgical placement in maxillae or mandibles.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- SS System, Osstem Co., Ltd, ([K062051](/device/K062051.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K06 3861

# 510(k) Summary

SEP 2 1 2007

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: December 4, 2006

1. Company and Correspondent making the submission:

| - | Submitter's Name : | Osstem Implant Co., Ltd.          |
|---|--------------------|-----------------------------------|
| - | Address :          | #507-8 Geoje3-Dong Yeonje-Gu      |
|   |                    | Busan, 611-804, Republic of Korea |
| - | Contact :          | Mr. Seung Woo Song                |

- 2. Device:

| Proprietary Name:    | GS System                   |
|----------------------|-----------------------------|
| Common Name -:       | Dental Implant System       |
| Classification Name: | Endosseous dental implant   |
|                      | 21CFR 872.3640 and 872.3630 |
|                      | Class II                    |
|                      | DZE, NHA                    |

- 3. Predicate Device:
SS System, Osstem Co., Ltd, K062051

- 4. Description:
The GS System is a dental implant system made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.

The GS System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The differences between them are shape. The surface treatment of the SS System is R.B.M. (Resorbable Blasting Media) and that of GS system is R.B.M. or CellNest.

The GS System is substantially equivalent in design, function and intended use to the SS System of Osstem Implant Co., Ltd.

Osstem Co., Ltd

510(k) Summary Page 1 of 2

{1}------------------------------------------------

K06 3861

## 5. Indication for use:

GS System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or internediate abutment support for fixed bridgework. GS System is for one stage surgical procedures. It is not intended for immediate load.

#### 6. Review:

The GS System has similar material, indication for use, design and technological characteristics as the predicate device.

The GS System has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.

# 7. Conclusions:

Based on the information provided in this premarket notification Osstem concludes that the GS System is safe and effective and substantially equivalent to the predicate device as described herein.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" written around the perimeter of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Osstem Implant Company, Limited C/O Ms. Cathryn N. Cambria Consultant Cambria Regulatory Consulting, Incorporated 5536 Trowbridge Drive Dunwoody, Georgia 30338

Re: K063861

Trade/Device Name: GS System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: August 29, 2007 Received: September 4, 2007

Dear Ms. Cambria:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

SEP 2 1 2007

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2 - Ms. Cambria

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Syptie Y. Michaud MD.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number K 06 3861

Device Name: GS System

GS System is intended for use in partially or fully edentulous Indication for use: mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. GS System is for one stage surgical procedures. It is not for immediate load.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR (Per 21CFR801 Subpart D)

Over-The-Counter Use (Per 21CFR807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K063861

Osstem Implant Co., Ltd

Indication for Use Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K063861](https://fda-staging.innolitics.com/device/K063861)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
