BASAL DORSAL PLATES

K063831 · Newdeal Sas · HRS · Feb 7, 2007 · Orthopedic

Device Facts

Record IDK063831
Device NameBASAL DORSAL PLATES
ApplicantNewdeal Sas
Product CodeHRS · Orthopedic
Decision DateFeb 7, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Basal Dorsal Plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include: - Moderate to severe Hallux Valgus - Hallux varus

Device Story

Basal Dorsal Plate is a metallic bone fixation device for first metatarsal basal osteotomy; used for forefoot deformity correction (Hallux Valgus, Hallux Varus). Device features anatomical shape matching dorsal curve of first metatarsal base; available in two lengths. Fixation achieved via four CALCANEA 3.5mm variable-angle, locking titanium alloy (Ti-6AI-4V ELI) screws. Device provided sterile. Used by orthopedic surgeons in clinical settings to stabilize bone segments post-osteotomy; provides mechanical support to facilitate healing and correction of deformities.

Clinical Evidence

Bench testing only. Mechanical calculations and bending resistance tests were performed and compared against predicate device specifications. Results confirmed mechanical properties are compatible with intended use.

Technological Characteristics

Material: Titanium alloy (Ti-6AI-4V ELI). Design: Anatomical dorsal curve, two lengths. Fixation: 3.5mm variable-angle locking screws. Sterilization: Provided sterile. Connectivity: None. Energy source: None.

Indications for Use

Indicated for fixation of first metatarsal basal osteotomy in patients with moderate to severe Hallux Valgus or Hallux varus.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### CONFIDENTIAL Newdeal SAS 510(k): PreMarket Notification Basal Dorsal Plate #### 510(K) SUMMARY V. #### BASAL DORSAL Plates #### Submitter's name and address: Newdeal SAS 10, place d'Helvétie 69006 Lyon, France Tel: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52 #### Contact person and telephone number Morgane Grenier Director of Regulatory and Clinical Affairs, EMEA Newdeal SAS 10, place d'Helvétie 69006 Lyon, France Tel: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52 #### Alternate Contacts #### Authorized Agent in the United States Judith E. O'Grady, RN, MSN Sr. Vice President, Regulatory Affairs, Quality Assurance and Clinical Affairs Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536, USA Tel: (609) 936-2311 Fax: (609) 275-9445 E-mail: jogrady@integra-ls.com #### Date Summary was prepared: December 18, 2006 #### Name of the device: | Proprietary Name: | Basal Dorsal Plate | |-----------------------|-------------------------------------------------------------------------------------------------| | Common Name: | Plate, Fixation, Bone | | Classification Name: | Single/multiple component metallic bone fixation appliances and<br>accessories (21CFR 888.3030) | | Device Product Code: | HRS | | Classification Panel: | Orthopedic | #### Substantial Equivalence: The Basal Dorsal Plate is substantially equivalent to the Newdeal B-BOP Plate, K052152 and to the Acumed Lower Extremity Congruent Bone Plate System, K033639. FFB 7 2007 {1}------------------------------------------------ ## CONFIDENTIAL Newdeal SAS Newucar reMarket Notification Basal Dorsal Plate # Device Description: The Basal Dorsal Plate is designed for fixation of basal osteotomy of the 15 metatarsal. In the case of significant deformities of the forefoot (Hallux Valgus, Hallux Varus), the basal In the case of sigments and a generally the more prevalent procedure than the distal osteotomy, of proxing greater corrections of the 1st metatarsal. Characteristics of the plate have taken into account all the requirements associated with basal osteotomy: - Anatomical shape adapted to the dorsal curve of the basis of the 1* metatarsal, in . two different lengths for better adaptation to the specific anatomy of the patient. - Fixation with CALCANEA™ 3.5mm screws in variable angle design for easier . positioning and grip into the bone, and a locking design for better stability of the system. The Basal Dorsal Plate and CALCANEA™ screws are made from titanium alloy (Ti-6AI-4V ELI), which are color-coded for ease of identification. Fixation of the Basal Dorsal Plate is provided by four CALCANEA™ screws, already I inditor it with the CALCANEA™ Plate and Screws system (510(k) K041786). The plates and screws are provided sterile with the Basal Dorsal Plate. #### Intended Use: The Basal Dorsal Plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include: - Moderate to severe Hallux Valgus - Hallux varus #### Testing and Test Results: An evaluation of the bending resistance based upon mechanical calculations has demonstrated that the bending behavior of the Basal Dorsal Plates will be equivalent or greater than the predicate devices (Synthes Modular Foot System). Mechanical tests have been carried out and results were then compared with the expected in vivo specifications performance. All the results show us that the Basal Dorsal Plates have mechanical properties compatible with their intended uses. #### Conclusion The Newdeal Basal Dorsal Plates are substantially equivalent to commercially marketed devices, the Newdeal B-BOP Plate, K052152 and the Acumed Lower Extremity Congruent Bone Plate System, K033639. The Newdeal Basal Dorsal Plates do not raise any new issues of scientific technology, safety or effectiveness. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Newdeal SAS % Judith O'Grady, R.N., M.S.N. Senior Vice President, Regulatory Affairs Integra Lifesciences Corporation 311 Enterprise Drive Plainsboro, New Jersey 08536 2 2007 FR Re: K063831 Trade/Device Name: Basal Dorsal Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: December 21, 2006 Received: December 27, 2006 Dear Ms. O'Grady: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Judith O'Grady, R.N., M.S.N. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Kaubara Buchim Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### IV. Indications for Use 510(k) Number (if known); KO6 3831 Device Name: Basal Dorsal Plate Indications For Use: The Basal Dorsal Plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include: Moderate to severe Hallux Valgus । Hallux varus Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Choubare buelund for mcu 1 (Division Sign-Off Division of General, Restorative, and Neurological Devices **510(k) Number** K063831 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...