← Product Code [LPH](/productcode/LPH) · K063822

# MODIFICATION TO HIPSTAR FEMORAL STEM (K063822)

_Stryker Orthopaedics · LPH · Jan 25, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063822

## Device Facts

- **Applicant:** Stryker Orthopaedics
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Jan 25, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The HIPSTAR hip stem is a sterile, single-use device intended for total hip arthroplasty.

## Device Story

The HIPSTAR Femoral Stem is a straight, titanium alloy hip prosthesis designed for total hip arthroplasty. It features a flared geometry for rotational stability and a grit-blasted surface to enhance bone-to-implant interface. The device has a 127° neck angle and is intended for cementless fixation. It is a sterile, single-use implant used by orthopedic surgeons in a clinical/surgical setting. The device replaces the femoral head and neck to restore joint function, correct deformities, or treat fractures. It provides structural support for the hip joint, potentially reducing pain and improving mobility for patients with degenerative joint disease or trauma.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design and technological similarities to the predicate device.

## Technological Characteristics

Material: Titanium alloy. Design: Straight stem with flare for rotational stability, 127° neck angle, grit-blasted surface for bone-to-implant interface. Fixation: Cementless. Sterilization: Sterile, single-use.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- HIPSTAR femoral hip stem ([K051223](/device/K051223.md))

## Reference Devices

- Osteolock™ HA Acetabular Cup

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063822

# 510(k) Summary of Safety and Effectiveness for the HIPSTAR Femoral Stem

JAN 2 5 2007

Proprietary Name:

Common Name:

Classification Name and Reference

Regulatory Class:

Device Product Code:

For Information contact:

Date Summary Prepared:

HIPSTAR Femoral Stem

Total Hip Joint Replacement Prosthesis

Hip joint, metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prostheses, 21 CFR §888.3353

Hip joint metal/polymer/metal semiconstrained porous-coated uncemented prosthesis 21 CFR §888.3358

Class II

87 MEH - prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calcium-phosphate

87 LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or nonporous, uncemented

87 LPH - prosthesis, hip, semi-constrained, metal/polymer, porous uncemented

Tiffani Rogers Regulatory Affairs Specialist Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, New Jersey 07432 Phone: (201) 831-5612 Fax: (201) 831-6038 E-Mail: Tiffani.Rogers@stryker.com

January 25, 2007

{1}------------------------------------------------

## Device Description

The Hipstar Femoral Stem is a straight hip stem manufactured from titanium alloy, The Hipstar Temoral Stein is a Statight any varies a flare for rotational stability, with a TMZF - - The Fripstal Inp realersbility. The body of the stem, with the exception of the narrow unstal stem for implain cases arrit blasted for increased bone to implant interface. The subject hip stem has a 127° neck angle.

## Intended Use:

The HIPSTAR hip stem is a sterile, single-use device intended for total hip arthroplasty.

### Indications

- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- necrosis;
rheumatoid arthritis (excepting the Osteolock™ HA Acetabular Cup and Peri-● Apatite coated prostheses);
- correction of functional deformity; .
- correction or adures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal . treatment of head involvement that are unmanageable using other techniques
- intended for cementless fixation. .

### Substantial Equivalence:

The subject Hipstar femoral stem is a modification to the Hipstar stem cleared in The Subject Hipstar relination of the substantial equivalence of the Hipstar hip stem is based on its similarities in intended use, design and sterilization to the previously cleared Hipstar femoral hip stem (K051223 cleared March 06, 2006).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three figures representing people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circle around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Stryker Howmedica Osteonics Corporation % Ms. Tiffani Rogers Regulatory Affairs Specialist Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey 07430

JAN 2 5 2007

Re: K063822

Trade/Device Name: HIPSTAR Femoral Stem Regulation Number: 21 CFR 888.3358, 21 CFR 888.3353 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH Dated: December 21, 2006 Received: December 22, 2007

Dear Ms. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Tiffani Rogers

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sincerely yours,

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

510(k) Number (if known): ______________________

Device Name: _________________________________________________________________________________________________________________________________________________________________

#### Indications

- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- rheumatoid arthritis (excepting the Osteolock™ HA Acetabular Cup and Peri-Apatite . coated prostheses);
- correction of functional deformity; .
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ● with head involvement that are unmanageable using other techniques.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X

OR (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Pla
(Division

**Division of Gene**
**and Neurological**

---

**Source:** [https://fda-staging.innolitics.com/device/K063822](https://fda-staging.innolitics.com/device/K063822)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).