K063821 · Nova Biomedical Corp. · CGA · Jan 24, 2007 · Clinical Chemistry
Device Facts
Record ID
K063821
Device Name
NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Applicant
Nova Biomedical Corp.
Product Code
CGA · Clinical Chemistry
Decision Date
Jan 24, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nova StatStrip Glucose Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, arterial, and neonate whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control. Nova StatStrip Glucose Test Strips are intended for use only with the StatStrip Glucose Hospital Meter for quantitative tests. The glucose meter is intended to quantitatively measure glucose (sugar) in whole blood. The Glucose Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatStrip Glucose Test Strips are for testing outside the body (in vitro diagnostic use only). Nova StatStrip Control Solutions is intended for use with the Nova StatStrip Glucose Hospital Meter and Nova StatStrip Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Level 1, Level 2, and Level 3). These solutions will be offered for sale separately from the meter. Nova StatStrip Glucose Linearity Kit solutions are used to check the linearity of the Nova StatStrip Glucose Hospital Meter System. There are five levels of linearity solutions, (Level 1, Level 2, Level 3, Level 4, and Level 5).
Device Story
Nova StatStrip Glucose Hospital Meter System is a point-of-care diagnostic device for quantitative glucose measurement in whole blood (capillary, venous, arterial, neonate). Operated by healthcare professionals in clinical settings; provides plasma-equivalent glucose results to aid diabetes management. System includes meter, test strips, control solutions, and linearity kits. Meter performs electrochemical analysis of blood samples applied to test strips; results displayed to clinician for immediate clinical decision-making. Benefits include rapid, accurate bedside glucose monitoring, reducing need for central laboratory testing.
Clinical Evidence
No clinical data provided in the document; substantial equivalence is based on bench testing and performance verification of the glucose measurement system.
Technological Characteristics
Electrochemical glucose measurement system; handheld meter form factor; utilizes disposable test strips; includes liquid control solutions and linearity kits for quality assurance; calibrated for plasma-equivalent results.
Indications for Use
Indicated for quantitative glucose measurement in capillary, venous, arterial, and neonate whole blood by healthcare professionals in clinical settings to monitor diabetes control effectiveness.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k063821
B. Purpose for Submission:
Modification of current glucose monitoring device to expand the reportable range and to add neonate blood as a sample type
C. Measurand:
Whole Blood Glucose
D. Type of Test:
Quantitative, utilizing glucose oxidase technology
E. Applicant:
Nova Biomedical Corp.
F. Proprietary and Established Names:
Nova StatStrip Glucose Hospital Meter System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345 Glucose test system
21 CFR 862.1660 Quality control materials
2. Classification:
Class II
3. Product code(s):
CGA, JJX
{1}
4. Panel:
75 (Chemistry)
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The Nova StatStrip Glucose Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, arterial, and neonate whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control.
Nova StatStrip Glucose Test Strips are intended for use only with the StatStrip Glucose Hospital Meter for quantitative tests. The glucose meter is intended to quantitatively measure glucose (sugar) in whole blood. The Glucose Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatStrip Glucose Test Strips are for testing outside the body (in vitro diagnostic use only).
Nova StatStrip Control Solutions is intended for use with the Nova StatStrip Glucose Hospital Meter and Nova StatStrip Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Level 1, Level 2, and Level 3). These solutions will be offered for sale separately from the meter.
Nova StatStrip Glucose Linearity Kit solutions are used to check the linearity of the Nova StatStrip Glucose Hospital Meter System. There are five levels of linearity solutions, (Level 1, Level 2, Level 3, Level 4, and Level 5).
3. Special conditions for use statement(s):
Provides plasma equivalent results
For professional use only
4. Special instrument requirements:
Nova StatStrip Glucose Hospital Meter system
{2}
3
I. Device Description:
The Nova StatStrip Glucose Hospital Meter System contains the following:
1. StatStrip Meter
2. Charging Station
3. Vial of StatStrip Test Strips (25)
4. Battery (3.7 V Lithium)
5. Quick Reference Guide
6. Instructions for Use Manual
7. StatStrip Log book
J. Substantial Equivalence Information:
1. Predicate device name(s):
Nova StatStrip Glucose Hospital Meter
Roche Accu-Chek Inform Meter
2. Predicate 510(k) number(s):
k060345, k003846
3. Comparison with predicate:
Similarities and Differences
| Characteristic | K063821 Modified Nova StatStrip Glucose Meter System | Predicate Nova StatStripGlucose Meter K060345 | Predicate Roche Accu-Chek Inform Meter K003846 |
| --- | --- | --- | --- |
| Measuring Range | 10-600 mg/dL | 20-600 mg/dL | 10-600 mg/dL |
| Hematocrit Range | 20-65 % | 30-60 % | 20-65 % |
| Operating Principle | No change from StatStrip predicate device | Electrochemical biosensor | Electrochemical biosensor |
| Intended Use (System) | The Nova StatStrip Glucose Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, and arterial and neonate whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control | The Nova StatStrip Glucose Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, and arterial whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control | The Accu-Check Inform Meter is designed to quantitatively measure the concentration of glucose in whole blood samples. The device is indicated for use health care professionals. |
{3}
| Sample type | No change from StatStrip predicate device | Whole blood | Whole blood |
| --- | --- | --- | --- |
| Sample size | No change from StatStrip predicate device | 1.3 uL | 4 uL |
| Sample application | No change from StatStrip predicate device | Test strip capillary draw | Test strip capillary draw |
| Handheld meter? | No change from StatStrip predicate device | YES | YES |
| Meter Calibration | No change from StatStrip predicate device | Automatic, no Calibration Code | Calibration code key needed |
| Data storage | No change from StatStrip predicate device | 1200 QC and patient test results | 4000 results |
| Test Time | No change from StatStrip predicate device | 6 sec | 26 sec |
| Weight | No change from StatStrip predicate device | 8.8 oz. | 12 oz |
| Bar code scanner | No change from StatStrip predicate device | YES | YES |
| Power source | No change from StatStrip predicate device | Rechargeable 3.7 V Lithium battery | Rechargeable 3.7 V Lithium battery |
| Accessories to the Meter: | | | |
| Charging station: | No change from StatStrip predicate device | Charging Station Base Unit | Charging Station Base Unit |
| Controls: | No change from StatStrip predicate device | Liquid, 3 levels | Liquid, 2 levels |
| Linearity Solutions | No change from StatStrip predicate device | 5 levels | Liquid, 6 levels |
| Test Strips – Active reagent: | No change from StatStrip predicate device | Glucose Dehydrogenase and Glucose Oxidase | Glucose dehydrogenase |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition 2004
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline Evaluation of Matrix Effects; Approved Guideline, Second Edition 2003
CLSI EP9-A: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. Vol. 15, No. 17, 1995
CLSI EP17-A: Protocols for determination of limit detection and limit of quantitation; Approved Guideline.
EN ISO 15197: 2003, In Vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
{4}
5
L. Test Principle:
The Test Principle used by this device is electrochemical biosensor technology using Glucose Oxidase. The strip uses the enzyme Glucose Oxidase to produce an electrical current that will stimulate a chemical reaction. This reaction is measured by the Nova StatStrip Glucose Hospital Meter and displayed as a blood glucose result.
Glucose measurement is based on the following equations:
$$
\text{Glucose} + \text{Enzymes (oxidized form)} \rightarrow \text{Gluconic acid} + \text{Enzymes (reduced form)}
$$
$$
\text{Enzymes (reduced form)} + \text{Ferricyanide} \rightarrow \text{Enzymes (oxidized form)} + \text{Ferricyanide}
$$
$$
\text{Ferrocyanide} \xrightarrow{\text{-e-}} \text{Ferricyanide}
$$
The current generated at the electrode is proportional to the glucose concentration of the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed and established in k060345.
Additional precision testing was performed on venous blood samples spiked with 4 target glucose concentrations between 5 and 60 mg/dL (6, 25, 40, 58 mg/dL) and 3 hematocrit levels (19%, 45%, 67%). Three lots of strips and 5 meters were used. Twenty measurements were taken for each lot of strips. The acceptance criteria are: for glucose level ≤ 50 mg/dL, ≤ 8% CV or <4.5 SD; and for glucose level from 51-150, ≤ 6% CV. All measurements met the sponsor's acceptance criteria.
b. Linearity/assay reportable range:
Linearity studies were performed and established in k060345.
Additional linearity studies were performed to verify the linearity range of 10 to 600 mg/dL. Samples were run in duplicate using 5 meters and one lot of test strips. Twelve glucose concentrations were prepared (spiked blood). Target glucose concentrations were at 5-10, 15-25, 30-40, 50-60, 65-75, 75-85, 130-150, 210-230, 300-340, 400-440, 500-550, and 560-600 mg/dL. The concentration range of 7 to 602 mg/dL was tested. The acceptance criteria are:
{5}
a correlation coefficient $r^2 \geq 0.95$ , and the test strip performance will be acceptable if the measurement bias related to YSI (at all glucose levels) is within $15\%$ or absolute error is $\leq 10.5 \, \mathrm{mg/dL}$ , whichever one is greater.
The results of the Modified Nova glucose meter (candidate device) verses the YSI (reference method) results are plotted in the graph below.

All the measurements met the sponsor's acceptance criteria. The sponsor claims the linearity range is 10 to $600\mathrm{mg / dL}$ .
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Please refer to k060345 for traceability and stability.
Controls were cleared in k060345.
d. Detection limit:
Detection limit studies were performed and established in k060345.
Limit of Blank (LoB) and Limit of Detection (LoD) studies were performed according to the CSLI EP17-A document. Blood bank samples were used and glucose concentrations tested using the YSI reference method. The LoD was determined by measuring 40 times of a low glucose sample on the modified Nova glucose meter and the YSI reference method. The sponsor chose the range of $10 - 600\mathrm{mg / dL}$ as the reportable range for the modified Nova glucose meter as determined in the linearity studies above.
e. Analytical specificity:
Analytical specificity studies were performed and established in k060345.
Hematocrit study: The sponsor performed the hematocrit study to verify that
{6}
varying hematocrit levels do not affect the glucose results reported by the candidate device at low glucose concentration levels.
Blood samples were prepared (spiked) with 8 target ranges of glucose at 5-15, 20-35, 40-55, 60-80, 100-150, 200-300, 350-450 and 550-650 mg/dL and at 4 hematocrit levels (18-25%, 40-46%, 48-55% and 63-68%). Samples were run in duplicate on 5 different meters and three different strip lots. The sponsor's acceptance criteria are:
1. Bias measurements related to YSI at all glucose levels must be within 15% or absolute error is < 10.5 mg/dL, whichever is greater.
2. In addition, the test strip performance will be acceptable if the results are within the following specifications:
| Glucose level (mg/dL) | CV % |
| --- | --- |
| ≤ 50 | 8% or SD = 4 mg/dL |
| 51-150 | ≤ 6 |
| 151-400 | ≤ 4 |
| 401-600 | ≤ 4 |
All the measurements met the sponsor's criteria and the sponsor claims that the reportable hematocrit range is 20% to 65% for modified Nova glucose meter (candidate device).
f. Assay cut-off:
None applicable
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison was performed and established in k060345.
In addition, method comparison of arterial samples from 100 neonates and capillary samples from heel sticks of 100 neonates were performed from two different clinical sites.
i.) Arterial samples from discarded whole blood gas syringe samples from 100 neonates from a clinical site were used to evaluate the candidate device verses the predicate device (Accu-Chek Inform Meter, which was cleared for neonate use) and a reference method (YSI). In order to obtain sufficient patient specimens with glucose levels of less than 30 mg/dL, 26 neonatal arterial blood samples were allowed to glycolize. The glucose concentration range tested was 7 to 279 mg/dL and the hematocrit range was 26% to 59%.
7
{7}
The linear regression results of the candidate device and the predicate device as they compare with the YSI reference method are summarized in the table below:
Arterial blood comparison of methods
| | Modified Nova glucose meter vs. YSI | Roche Accu-Chek Inform meter vs. YSI | Roche Accu-Chek Inform meter vs. Nova glucose meter |
| --- | --- | --- | --- |
| Correlation (r2) | 0.992 | 0.984 | 0.984 |
| Slope | 1.109 | 1.080 | 1.015 |
| Intercept | -3.471 | -6.945 | 4.842 |
| N | 200 (100 samples in duplicate) | 200 (100 samples in duplicate) | 200 (100 samples in duplicate) |
The system accuracy results for glucose concentration based on the ISO 15197 document are as follows:
- For glucose values $< 75\mathrm{mg / dL}$ , $100\%$ of the candidate's glucose results are within $\pm 15\mathrm{mg / dL}$ compare to the YSI results.
- For glucose values $>75\mathrm{mg/dL}$ , $98\%$ of the candidate's glucose results are within $\pm 20\%$ of the YSI results.
ii.) Capillary blood samples from the heel sticks of 100 neonates from the NICU department of a children's hospital were used to evaluate the candidate device verses the J&J Vitros Chemistry analyzer (comparator method) and the predicate device (LifeScan SureStep Pro Meter, which was cleared for neonate use). 3 patient samples out of the 100 target samples were pooled together and spiked with a glucose solution in order to obtain glucose levels above $120\mathrm{mg / dL}$ . The age range of the patient tested was 1 to 25 days old, the hematocrit range was $29.7\%$ to $66.9\%$ , and the glucose concentration range tested was 25 to $411\mathrm{mg / dL}$ .
The linear regression results of the candidate device and the predicate device as they compare with the J&J comparator method are summarized in the table below:
{8}
9
Capillary blood comparison of methods
| | Modified Nova glucose meter vs. J&J Vitros | SureStep Pro meter vs. J&J Vitros | Modified Nova glucose meter vs. SureStep Pro meter |
| --- | --- | --- | --- |
| Correlation (r²) | 0.986 | 0.996 | 0.981 |
| Slope | 0.986 | 1.041 | 0.942 |
| Intercept | -3.425 | -6.176 | 2.842 |
| N | 198 (99 samples in duplicate) | 100 | 198 (99 samples in duplicate) |
The system accuracy results for glucose concentration based on the ISO 15197 document are as follows:
- For glucose values < 75 mg/dL, 98% of the candidate’s glucose results are within ± 15 mg/dL compare to the J&J Vitros results.
- For glucose values > 75 mg/dL, 100% of the candidate’s glucose results are within ± 20% of the J&J Vitros results.
Since the neonatal clinical samples did not cover the claimed meter range, results from a study of spiked samples with high glucose concentrations and varying hematocrits were provided by the sponsor to demonstrate acceptable performance (≤ 15% bias or absolute error is < 10.5 mg/dL from the YSI method). See the hematocrit study above.
b. Matrix comparison:
Established in k060345
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{9}
5. Expected values/Reference range:
The reference value for serum and plasma glucose for adult is 70-105 mg/dL and 60-100 mg/dL for a child
The Nova StatStrip Glucose Meter expected values were referenced from the literature:
* Burtis, Carl A. and Ashwood, Edward R., ed. 1999. Tietz Textbook of Clinical Chemistry. Philadelphia, PA: W.B. Saunders Co.
The sponsor states in the labeling that each laboratory should establish and maintain its own reference values. The values given above should be used only as a guide.
N. Instrument Name:
Nova StatStrip Glucose Hospital Meter
O. System Descriptions:
1. Modes of Operation:
The Nova StatStrip Glucose Hospital Meter used with the Nova StatStrip Glucose Test Strips is a single use test system used to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary (fingertip) whole blood, venous and arterial and neonate. The Nova StatStrip Glucose Test Strips are for in vitro diagnostic use only.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
The Nova StatStrip Glucose Hospital Meter memory will store 1000 Patient Tests, 500 QC Tests, and 4000 Operator ID’s.
4. Specimen Sampling and Handling:
Refer to k060345
5. Calibration:
Refer to k060345
{10}
6. Quality Control:
Refer to k060345
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Refer to k060345
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.