# ACUSON ANTARES DIAGNOSTIC ULTRASOUND SYSTEM (K063803)

_Siemens Medical Solutions USA, Inc., Ultrasound DI · IYN · Jan 5, 2007 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063803

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc., Ultrasound DI
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jan 5, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The Antares ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal, Great Vessel, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes.

## Device Story

ACUSON ANTARES is a multi-purpose diagnostic ultrasound system; utilizes various transducers to transmit ultrasonic energy into patients; receives and processes echoes to generate real-time images of anatomical structures and fluid flow. System supports multiple imaging modes: B-mode, M-mode, PWD, CWD, Color Doppler, Amplitude Doppler, and 3D imaging. Used in clinical settings (e.g., radiology, cardiology, surgery) by trained healthcare professionals. Provides anatomical measurements to aid clinical diagnosis. Benefits include non-invasive visualization of internal structures and hemodynamics. Features include tissue harmonic imaging, spatial compounding, tissue equalization, and vascular enhancement technologies.

## Clinical Evidence

Bench testing only. The system complies with recognized safety and performance standards, including UL 60601-1, CSA C22.2 No. 601-1, NEMA UD-2 (acoustic output measurement), and NEMA UD-3 (thermal/mechanical indices). Biocompatibility is addressed per ISO 10993.

## Technological Characteristics

Multi-purpose diagnostic ultrasound system. Transducers include phased, curved, linear, and mechanical 3D arrays. Energy source: ultrasonic. Connectivity: standard diagnostic ultrasound interfaces. Software: proprietary imaging and measurement algorithms. Compliance: UL 60601-1, CSA C22.2 No. 601-1, NEMA UD-2, NEMA UD-3, ISO 10993.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- ACUSON Antares Diagnostic Ultrasound System ([K063138](/device/K063138.md))
- SONOLINE Antares Diagnostic Ultrasound System ([K050034](/device/K050034.md))
- SONOLINE Antares with CLARIFY VE Diagnostic Ultrasound System ([K033196](/device/K033196.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(K) SUMMARY

063805
JAN - 5 2007

## ACUSON ANTARES™ Diagnostic Ultrasound System

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary.

#### 1. Submitted By:

Siemens Medical Solutions USA, Inc., Ultrasound Division 1230 Shorebird way Mountain View, CA 94043

### Contact Person:

Michaela Mahl Regulatory Affairs Specialist Phone: (650) 694 5653 (650) 943 7053 FAX:

## Date Prepared:

12/01/2006

#### 2. Proprietary Name: ACUSON ANTARES™ Ultrasound System

### Common/ Usual Name:

Diagnostic Ultrasound System with Accessories

## Classification Name:

Requiatory Class: II Review Category: Tier II Classification Panel: Radiology

| Ultrasonic Pulsed Doppler Imaging System: | FR # 892.1550 | Product Code 90-IYN |
|-------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System:    | FR # 892.1560 | Product Code 90-IYC |
| Diagnostic Ultrasound Transducer:         | FR # 892.1570 | Product Code 90-ITX |

#### 3. Predicate Device:

- K063138, 11/22/2006, ACUSON Antares™ Diagnostic Ultrasound System 트
- 트 K050034, 01/13/2005, SONOLINE Antares™ Diagnostic Ultrasound System
- K033196, 10/16/2003, SONOLINE Antares™ with CLARIFY VE Diagnostic Ultrasound 포 System

#### Device Description: 4.

The ANTARES system is a multi-purpose diagnostic ultrasound system with accessories and proprietary software.

The ACUSON ANTARES has been designed to meet the following product safety standards:

- UL 60601-1, Medical Electrical Equipment, Part 1: General Requirements for Safety .
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment .

{1}------------------------------------------------

- NEMA UD-2. Acoustic Output Measurement Standard for Diagnostic Ultrasound .
- NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output . Indices on Diagnostic Ultrasound Equipment
- 93/42/EEC Medical Devices Directive .
- Safety and EMC Requirements for Medical Equipment .
- EN 60601-1 .
- . EN 60601-1-1
- EN 60601-1-2 �
- . EN 60601-1-2-37
- EN 60601-1-4 �
- IEC 61157 Declaration of Acoustic Power �
- ISO 10993 Biocompatibility .
- The system's acoustic output is in accordance with ALARA principle (as low as reasonably . achievable)

#### 5. Intended Uses:

The Antares ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal, Great Vessel, and Peripheral Vascular applications.

The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes.

#### દ. Technological Comparison to Predicate Device:

The ACUSON ANTARES™ is substantially equivalent to the predicate devices listed in paragraph 3 above. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures to aid in diagnosis.

### End of 510(k) Summary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions USA, Inc. Ultrasound Group % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

JAN - 5 2007

Re: K063803

Trade Name: ACUSON Antares Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: December 21, 2006 Received: December 22, 2006

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent(for the use indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devises, or unc the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the ACUSON Antares Diagnostic Ultrasound System as described in your premarket 100
notificatio notification:

Image /page/2/Picture/11 description: The image is a circular logo or seal. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo in a bold font. The word "Centennial" is below the letters "FDA". There are three stars below the word "Centennial".

noting Public I

{3}------------------------------------------------

## Transducer Model Number

| CW2                   | PX4-1 Phased Array              |
|-----------------------|---------------------------------|
| CW5                   | CH6-2 Curved Array              |
| C5-2 Curved Array     | CH4-1 Curved Array              |
| CX5-2 Curved Array    | PH4-1 Phased Array              |
| VF7-3 Linear Array    | P10-4 Phased Array              |
| EC9-4 Curved Array    | VF13-5SP Linear Array           |
| VFX9-4 Linear Array   | C5F1 Curved Array Mechanical 3D |
| VF10-5 Linear Array   | C7F2 Curved Array Mechanical 3D |
| VF13-5 Linear Array   | EV9F4 Curved Array              |
| VFX13-5 Multi-D Array | V5Ms Multiplane TEE             |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Andrew Kang (240) 276-3666.

Sincerely yours.

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

510 (k) Number (if known):

K063803

Device Name:

# ACUSON Antares Ultrasound System

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  |   | Mode of Operation |   |     |     |                  |                      |                              |                       |                       |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------|
| Clinical Application             | A | B                 | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)    |
| Ophthalmic                       |   |                   |   |     |     |                  |                      |                              |                       |                       |
| Fetal                            |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Abdominal                        |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Intraoperative<br>(Note 9)       |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Intraoperative<br>Neurological   |   | P                 | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Pediatric                        |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Small Organ<br>(Note 1)          |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Neonatal Cephalic                |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Adult Cephalic                   |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Cardiac                          |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,6,7,8,10 |
| Trans-esophageal                 |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,6,10     |
| Transrectal                      |   | P                 | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Transvaginal                     |   | P                 | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Transurethral                    |   |                   |   |     |     |                  |                      |                              |                       |                       |
| Intravascular                    |   |                   |   |     |     |                  |                      |                              |                       |                       |
| Peripheral vessel                |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Laparoscopic                     |   |                   |   |     |     |                  |                      |                              |                       |                       |
| Musculo-skeletal<br>Conventional |   | P                 | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Musculo-skeletal<br>Superficial  |   | P                 | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Other (specify)                  |   |                   |   |     |     |                  |                      |                              |                       |                       |

N = new indication; P = previously cleared by FDA - K063138, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyrold, penis, prostate, etc.

Note 2 Ensemble tissue harmonic imaging

Note 3 SleClear multi-view spatial compounding

- Tissue Equalization Technology Note 4
- 3-Scape real-time 3D imaging Note 5

Note 6 Cadence contrast agent imaging (cardiac only)

B&W SieScape panoramic Imaging Note 7

Note 8 Power SieScape panoramic imaging

- Note 9 For example: vascular, abdominal
Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Hogdon
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K063803

{6}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use:

CW2 Probe for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Abdominal                        |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Intraoperative<br>(Note 9)       |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Small Organ<br>(Note 1)          |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Cardiac                          |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral vessel                |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                    |

N = new Indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc. For example: vascular, abdominal Note 9

1. Example: Pseudata/abdominal

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy Roagdon

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devic 510(k) Number

Section 4.3, Pg. 2 of 21

{7}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use:

# CW5 Probe for use with ACUSON Antares

Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  |   | Mode of Operation |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A | B                 | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic                       |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Abdominal                        |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Intraoperative<br>(Note 9)       |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Intraoperative<br>Neurological   |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Small Organ<br>(Note 1)          |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Adult Cephalic                   |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Cardiac                          |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Trans-esophageal                 |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral vessel                |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Laparoscopic                     |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |   |                   |   |     | P   |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |   |                   |   |     |     |                  |                      |                              |                       |                    |
| Other (specify)                  |   |                   |   |     |     |                  |                      |                              |                       |                    |

N = new indication; P = previously cleared by FDA - K050034; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc. Note 9 For example: vascular, abdominal

> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

> > Prescription Use (Per 21 CFR 801.109)

Nancy C. Brogdon

(Division Sign-Off Division of Reproductive, and Radiological Devices 510(k) Number _

December 13, 2006

{8}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use:

C5-2 Curved Array Transducer for use with ACUSON Antares

Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Neonatal Cephalic                |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 2 Ensemble tissue harmonic imaging

SieClear multi-view spatial compounding Note 3

Note 4 Tissue Equalization Technology

- 3-Scape real-time 3D Imaging Note 5
- B&W SieScape panoramic imaging Note 7

Note 8 Power SleScape panoramic Imaging

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy Shoagdon
(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number .

Section 4.3, Pg. 4 of 21

{9}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use: CX5-2 Curved Array Transducer for use with ACUSON Antares

Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Neonatal Cephalic                |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Ensemble tissue harmonic imaging Note 2

Note 3 SleClear multi-view spatial compounding

Tissue Equalization Technology Note 4

- 3-Scape real-time 3D imaging Note 5
- Note 7 B&W SieScape panoramic Imaging
- Power SleScape panoramic imaging Note 8

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Harry C. Brogdon

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number_

{10}------------------------------------------------

#### 510 (k) Number (if known): K063803

Device Name: Intended Use: VF7-3 Linear Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously deared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

- Note 2 Ensemble tissue harmonic Imaging
- SteClear multi-view spatial compounding Note 3
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D Imaging
- Note 7 B&W SieScape panoramic imaging
- Note 8 Power SieScape panoramic imaging

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancyc Brogdon
(Division Sign-Off)

Division of Reproductive, A and Radiological Devices 510(k) Number __

{11}------------------------------------------------

#### 510 (k) Number (if known): K063803

Device Name: Intended Use:

# EC9-4 Curved Array Transducer for use with ACUSON Antares

Ultrasound Imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Transvaginal                     |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Superficial  |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

For example: breast, testes, thyrold, penis, prostate, etc. Note 1

Ensemble tissue harmonic imaging Note 2

Note 3 SieClear multi-view spatial compounding

- Tissue Equalization Technology Note 4
- 3-Scape real-time 3D imaging Note 5
- Note 7 B&W SieScape panoramic imaging

Power SieScape panoramic imaging Note 8

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

T

{12}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use:

VFX9-4 Linear Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

- Ensemble tissue harmonic imaging Note 2
- Note 3 SieClear multi-view spatial compounding
- Tissue Equalization Technology Note 4
- 3-Scape real-time 3D imaging Note 5

Note 7 B&W SieScape panoramic imaging

- Note 8 Power SieScape panoramic imaging
Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy C. Brogdon
(Division Sign-Off)

Division of Reproductive, Abdomina and Radiological Devices 510(k) Number

December 13, 2006

{13}------------------------------------------------

510 (k) Number (if known): K063803

Device Name:

Intended Use:

VF10-5 Linear Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- Ensemble tissue harmonic imaging Note 2
- Note 3 SieClear multi-view spatial compounding
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging
- Note 7 B&W SieScape panoramic imaging
- Power SieScape panoramic imaging Note 8

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy C

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number ***_***__ j2063803

{14}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use: ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

VF13-5 Linear Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

- Note 2 Ensemble tissue harmonic imaging
- Note 3 SieClear multi-view spatial compounding
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging

Note 7 B&W SieScape panoramic imaging

- Power SieScape panoramic imaging Note 8
Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy Brogdon
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

{15}------------------------------------------------

510 (k) Number (if known); K063803

Device Name: Intended Use: VFX13-5 Multi-D Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ<br>(Note 1)          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Neonatal Cephalic                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Musculo-skeletal<br>Conventional |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Musculo-skeletal<br>Superficial  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                     |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

- Note 2 Ensemble tissue harmonic imaging
- SieClear multi-view spatial compounding Note 3
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging
- B&W SieScape panoramic imaging Note 7
- Note 8 Power SieScape panoramic imaging

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy C Beagdon
(Division Sign-Off)

Division of Reproductive, Al and Radiological Devices 510(k) Number

{16}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use: PX4-1 Phased Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                       |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)    |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Fetal                            |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Abdominal                        |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Pediatric                        |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Small Organ                      |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Neonatal Cephalic                |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Adult Cephalic                   |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Cardiac                          |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,6,7,8,10 |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Transurethrai                    |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Peripheral vessel                |                   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10   |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Musculo-skeletal<br>Conventional |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Musculo-skeletal<br>Superficial  |                   |   |   |     |     |                  |                      |                              |                       |                       |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                       |

N = new indication; P = previously cleared by FDA - K050034, K033196; E = added under Appendix E

Additional Comments:

Note 2 Ensemble tissue harmonic imaging

Note 3 SieClear multi-view spatial compounding

Note 4 Tissue Equalization Technology

3-Scape real-time 3D imaging Note 5

Note 6 Cadence contrast agent imaging (cardiac only)

- Note 7 B&W SieScape panoramic imaging
Power SieScape panoramic imaging Note 8

Note 10 Clarify VE vascular enhancement technology

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Nancy Crozlor
(Division Sign-Off)
Prescriptio

Division of Reproductive and Radiological Devices 510(k) Number

{17}------------------------------------------------

510 (k) Number (if known): K063803

Device Name: Intended Use:

CH6-2 Curved Array Transducer for use with ACUSON Antares Ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                     |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)  |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Fetal                            |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Abdominal                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intraoperative<br>Neurological   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Pediatric                        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,10 |
| Small Organ                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Neonatal Cephalic                |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Trans-esophageal                 |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                     |
| Peripheral vessel                |                   | P | P | P   |     | P                | P                    |…

---

**Source:** [https://fda-staging.innolitics.com/device/K063803](https://fda-staging.innolitics.com/device/K063803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).