SFB AND CFB IMPLANTS - 5.0 PLATFORM

K063718 · Nobel Biocare AB · DZE · Jan 5, 2007 · Dental

Device Facts

Record IDK063718
Device NameSFB AND CFB IMPLANTS - 5.0 PLATFORM
ApplicantNobel Biocare AB
Product CodeDZE · Dental
Decision DateJan 5, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

Device Story

Endosseous dental implants surgically placed in upper or lower jaw bone; provide support for prosthetic devices (artificial teeth). Restores patient esthetics and chewing function. Used in clinical dental settings by trained professionals. Supports single or multiple unit restorations; splinted or non-splinted applications. Allows for immediate placement and function if specific stability criteria are met. Device serves as mechanical anchor for dental prosthetics.

Clinical Evidence

No clinical data provided; substantial equivalence based on design similarity and intended use.

Technological Characteristics

Endosseous dental implant; 5.0 mm platform diameter. Designed for surgical placement in jaw bone to support prosthetic restorations. Class II device (21 CFR 872.3640).

Indications for Use

Indicated for patients requiring single or multiple unit dental restorations in the upper or lower jaw arches. Supports prosthetic devices to restore esthetics and chewing function. Suitable for immediate placement and function if initial stability requirements are met.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K>63718 ## 1.4 510(k) Summary of Safety and Effectiveness | Submitted by: | Herbert Crane<br>Director, Regulatory Affairs | JAN - 5 2007 | |----------------------------------------|----------------------------------------------------------------------------|--------------| | Address: | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 | | | Telephone: | (714) 282-5074 | | | Facsimile: | (714) 282-9023 | | | Date of Submission: | December 13, 2006 | | | Classification Name: | Endosseous Dental Implant (21 CFR 872.3640) | | | Trade or Proprietary<br>or Model Name: | SFB and CFB – 5.0 Platform | | | Legally Marketed Devices: | SFB and CFB (K061003) | | Device Description: The SFB and CFB - 5.0 Platform implants are similar to the predicate SFB and CFB implants. They differ only in platform diameter. Indications for Use: Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. Nobel Biocare Special 510(k) Notification SFB and CFB -- 5.0 Platform December 2006 000009 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the emblem. The logo is presented in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Nobel Biocare AB JAN 0 5 2007 C/O Mr. Herbert Crane Director, Regulatory Affairs Nobel Biocare USA, Incorporated 22715 Savi Ranch Parkway Yorba Linda, California 92887 Re: K063718 Trade/Device Name: SFB & CFB-5.0 Platform Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: December 13, 2006 Received: December 14, 2006 Dear Mr. Crane: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Crane Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Dunlap Dr. Chia-Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): Device Name: SFB & CFB - 5.0 Platform Indications For Use: Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![Signature](signature.png) (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices | 510(k) Number: | K063218 | |----------------|---------| |----------------|---------| Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...