TUTTNAUER PREVACUUM STEAM HEATED AUTOCLAVE, MODELS 6990 SP-1A, 66120 SP-1A, 69150 SP-2A, 69120 L USA SP-2A, 69180 SP-2HA
K063711 · Tuttnauer Co. , Ltd. · FLE · Feb 2, 2007 · General Hospital
Device Facts
Record ID
K063711
Device Name
TUTTNAUER PREVACUUM STEAM HEATED AUTOCLAVE, MODELS 6990 SP-1A, 66120 SP-1A, 69150 SP-2A, 69120 L USA SP-2A, 69180 SP-2HA
Applicant
Tuttnauer Co. , Ltd.
Product Code
FLE · General Hospital
Decision Date
Feb 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
This Tuttnauer autoclave is designed for use in healthcare settings for the sterilization of heat stable medical devices; wrapped solids, hollow and porous products. Sterilization of liquid loads should only be used in non-clinical applications in open or closed (but not sealed) containers. Because of the difficulty to standardize the liquid cycle load. This cycle in not intended for sterilization of materials that come into contact with patients and should only be used for non-clinical applications only.
Device Story
Prevacuum steam-heated autoclaves; utilize saturated steam for sterilization of heat-stable medical devices, wrapped solids, hollow/porous products, and non-clinical liquids. Device comprises pressure vessel, steam jacket, heating elements, water reservoir, water/vacuum pumps, and electronic control panel. Operator selects pre-programmed cycles via control panel; microcomputer manages cycle parameters (time, temperature, pressure) automatically. Includes Bowie & Dick and Air Leakage test programs. Used in healthcare settings; operated by clinical staff. Provides automated sterilization cycle; ensures consistent sterilization parameters; benefits patient by ensuring sterile medical devices.
Clinical Evidence
Bench testing only. Validation studies conducted in accordance with ANSI/AAMI ST8 (2001) confirmed compliance with physical and microbiological performance requirements. Successful sterilization achieved in all tests.
Technological Characteristics
Steam sterilizer; pressure vessel with steam jacket, heating elements, water reservoir, water pump, vacuum pump. Electronic control panel with microcomputer. Twelve pre-programmed cycles plus two test programs (Bowie & Dick, Air Leakage). Automatic shutdown capability. Pneumatic or hydraulic door operation depending on model.
Indications for Use
Indicated for sterilization of heat-stable medical devices, wrapped solids, hollow, and porous products in healthcare settings. Liquid loads are indicated for non-clinical applications only in open or closed (non-sealed) containers. Not indicated for sterilization of patient-contact materials in liquid cycles.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
K063711
## 510(k) Summary
FEB - 2 2007
Tuttnauer USA Co. Ltd. 25 Power Drive Hauppauge, NY 11788 Phone: 631-737-4850 Fax: 631-737-0720
Robert R. Basile, V.P.
November 21, 2006
Electronic pre-vacuum autoclave, steam sterilizer
Tuttnauer Prevacuum Steam Heated Autoclave with Automatic Hinged Door (Model 6990 SP-1A)
Tuttnauer Prevacuum Steam Heated Autoclave with Automatic Hinged Door (Model 66120 SP-1A)
Tuttnauer Prevacuum Steam Heated Autoclave with Two Automatic Hinged Doors (Model 69150 SP-2A)
Tuttnauer Prevacuum Steam Heated Autoclave Equipped with Two Automatic Hinged Doors (Model 69120 L USA SP-2A)
Tuttnauer Steam-Heated Prevacuum Autoclave with Two Horizontal Sliding Doors (Model 69180 SP-2HA)
Tuttnauer Prevacuum Steam Heated Autoclave Equipped with Vertical Sliding Door (Model 6690 L USA SP-1V)
Steam Sterilizer Class II Device - 21 C.F.R. § 880.6880
Submitter:
Contact Name:
Date Prepared:
Common Name (for all):
Trade Names:
Classification Name (for all):
{1}------------------------------------------------
### Substantial Equivalence:
All six Tuttnauer autoclave models referenced above (Models 6990 SP-1A; 66120 SP-1 A: 69150 SP-2A: 69120 L USA SP-2A: 69180 SP-2HA; and 6690 L USA SP-1V) are claimed to be substantially equivalent to the autoclaves cleared under the following 510(k) numbers:
K850264 (non-computer controlled autoclaves) K850264A (non-computer controlled autoclaves) K032192 (computer controlled autoclaves)
The six Tuttnauer autoclave models covered by this 510(k) notification differ in the size of their chambers (dimensions and volume) and in their door configurations. All six models have been previously cleared by FDA in their manual form; the six models covered herein only differ from these previously cleared devices in that they contain electronic control panels to allow for automatic usage. As there is no change to factors affecting sterilization, such as cycle time, temperature, or pressure, these six devices should be equivalent to the predicate devices cleared under the three 510(k) numbers listed above.
### General Description:
Tuttnauer autoclave models 6990 SP-1A: 66120 SP-1A: 69150 SP-2A: 69120 L USA SP-2A; 69180 SP-2HA; and 6690 L USA SP-1V are all prevacuum autoclaves that utilize saturated steam as the sterilizing agent. Each of these autoclaves is designed for sterilization of heat stable medical devices; wrapped solids, hollow and porous products, as well as liquids for non-clinical applications in open or closed (but not sealed) containers.
The key differences between the six models covered by this 510(k) notification are (1) chamber dimensions/chamber volume, and door configuration. These differences are summarized here:
| Model | Chamber Dim. (w x h x d) | Chamber Volume | Door Configuration |
|-------------------|--------------------------|----------------|--------------------------------------------------------------------|
| 6990 SP-1A | 24" x 36" x 36" | 21 ft. 3 | automatic hinged door<br>driven by pneumatic<br>drive |
| 66120 SP-1A | 26" x 26" x 50" | 21.2 ft. 3 | automatic hinged door<br>driven by pneumatic<br>drive |
| 69150 SP-2A | 24" x 36" x 60" | 33 ft. 3 | two automatic hinged<br>doors driven by<br>pneumatic drive |
| 69120 L USA SP-2A | 24" x 36" x 48" | 750 l. 3 | two automatic hinged<br>doors driven by<br>pneumatic drive |
| 69180 SP-2HA | 24" x 36" x 83" | 1165 l. 3 | two horizontal sliding<br>doors operated by<br>hydraulic operators |
{2}------------------------------------------------
through an oil cylinder
## 6690 L USA SP-1 V 26" x 26" x 39"
450 1. 3
one vertical sliding door operated by hydraulic system through two oil cylinders mounted laterally on both sides of door
The six Tuttnauer models covered herein are substantially the same as models cleared under 510(k) numbers K850264 and K850264A. Both of these 510(k) notifications and clearances cover the manual versions of the automatic autoclaves subject to this 510(k) application. Tuttnauer has modified the control panel of the manual autoclaves, replacing the current manual control panel with an electronic control panel. CDRH has already cleared a similar manual-toelectronic change in 510(k) number K032192. Therefore, all three of these 510(k) notifications and subsequent clearances are claimed as predicate devices for all six of the autoclaves in this 510(k) application.
### Design and Materials:
Each of the six Tuttnauer autoclave models is a steam sterilizer composed of a pressure vessel with steam jacket, heating elements, a water reservoir, a water pump, and a vacuum pump, Each autoclave is pre-programmed with twelve sterilization cycles, and each has two test programs: (1) a "Bowie & Dick Test," and (2) an "Air Leakage Test." The six autoclaves are equipped with the capability to shut down automatically. An emergency stop push button that is mounted on the front panel is an available option for customers with specific needs.
The control software, which is identical to that used in the autoclayes cleared under K032192, allows for automatic operation of all six autoclaves. The microcomputer technology employed in the autoclaves ensures that the entire cycle will operate on a fully-automatic basis. Therefore, no further intervention is necessary after setting the pre-selected parameters and starting operation of the autoclaye.
### Comparison to Cleared Devices:
The only material component that is being changed on these six devices is the substitution of a manual control panel with an electronic control panel, including operational software. This electronic control panel/operational software package was previously cleared as a modification to two existing Tuttnauer autoclaves under 510(k) number K032192. This control panel/software substitution does not affect the intended use of the six devices included in this notification, and the fundamental scientific technology of the six devices is not altered.
### Intended Use:
All six Tuttnauer autoclave models referenced above (Models 6990 SP-1A; 66120 SP-1A; 69150 SP-2A; Model 69120 L USA SP-2A; 69180 SP-2HA; and Model 6690 L USA SPl V) are designed for sterilization of heat stable medical devices; wrapped solids, hollow and
{3}------------------------------------------------
porous products, as well as liquids for non-clinical applications in open or closed (but not sealed) containers. Indication for use statement is attached
Because of the difficulty to standardize the liquid cycle load. This cycle in not intended for sterilization of materials that come into contact with patients and should only be used for non-clinical applications only.
### Technology Considerations:
Each of the six Tuttnauer autoclave models is a steam sterilizer composed of a pressure vessel with steam jacket, heating elements, a water reservoir, a water pump, and a vacuum pump.
### Non-Clinical Testing:
Tuttnauer conducted validation studies in accordance with ANSI/AAMI ST8 (2001). Testing showed that the sterilizers meet all aspects of the standard, including physical and microbiological performance requirements. Successful sterilization was accomplished in all tests performed.
### Conclusion:
For all of the foregoing reasons, Tuttnauer believes that the six Tuttnauer autoclave models described in this 510(k) notification may be safely marketed in the United States.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized eagle with its wings spread. The eagle is facing to the left.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tuttnauer Company Limited C/O Mr. Mark Yacura Buchanan Ingersoll, P.C. 1776 K Street N.W., Suite 300 Washington, DC 20006-2365
FEB - 2 2007
Re: K063711
Trade/Device Name: Tuttnauer Prevacuum Steam Heated Autoclaves Models 6990 SP-1A, 66120 SP-1A, 69150 SP-2A, 69120 L USA SP-2A, 69180 SP-2HA, 6690 L USA SP-1V Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: December 13, 2006 Received: December 14, 2006
Dear Mr. Yacura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Yackura
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
### 510(k) Number: K063711
Device Name: Tuttnauer Prevacuum Steam Heated Autoclaves models 6990 SP-1A, 66120 SP-1A, 69150 SP-2A, 69120 L USA SP-2A, 69180 SP-2HA, 6690 L USA SP-1V
Indications for Use: This Tuttnauer autoclave is designed for use in healthcare settings for the sterilization of heat stable medical devices; wrapped solids, hollow and porous products. Sterilization of liquid loads should only be used in non-clinical applications in open or closed (but not sealed) containly si
Maximum loads depend upon the chamber size, and are listed below.
## MODEL NUMBER 6990 SP-1A
#### Cycle Program Temperature Maximum Maximum Number ofi Load (Ibs) Trays 01.02 . Unwrapped 250 72 2 03, 04 Unwrapped 270/274 72 2 05, 06, 08 Wrapped 270 72 2 07 Wrapped 274 72 2 09, 10, 11, 12 Liquids 250 72 2 13 B & D test 274 72 2 14 Air leak test . .
## CHAMBER (inches) 24x36x36
Maximum load for one tray may not exceed the maximum load for the autoclave
AND/OR
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shule M. Maugham K.D.
Manager, General Manager
and all Associates
Page 1 of 4
{7}------------------------------------------------
# MODEL NUMBER 66120 SP-1A
## CHAMBER (inches) 26x26x50
| Cycle<br>Number | Program | Temperature<br>°F | Maximum<br>Load (lbs) | Maximum<br>Trays |
|-----------------|---------------|-------------------|-----------------------|------------------|
| 01, 02 | Unwrapped | 250 | 96 | 2 |
| 03, 04 | Unwrapped | 270/274 | 96 | 2 |
| 05, 06, 08 | Wrapped | 270 | 96 | 2 |
| 07 | Wrapped | 274 | 96 | 2 |
| 09, 10, 11, 12 | Liquids | 250 | 96 | 2 |
| 13 | B & D test | 274 | 96 | 2 |
| 14 | Air leak test | - | - | - |
Maximum load for one tray may not exceed the maximum load for the autoclave
## MODEL NUMBER 69150 SP-2A
## CHAMBER (inches) 24x36x60
| Cycle<br>Number | Program | Temperature<br>°F | Maximum<br>Load (lbs) | Maximum<br>Trays |
|-----------------|---------------|-------------------|-----------------------|------------------|
| 01, 02 | Unwrapped | 250 | 192 | 2 |
| 03, 04 | Unwrapped | 270/274 | 192 | 2 |
| 05, 06, 08 | Wrapped | 270 | 192 | 2 |
| 07 | Wrapped | 274 | 192 | 2 |
| 09, 10, 11, 12 | Liquids | 250 | 192 | 2 |
| 13 | B & D test | 274 | 192 | 2 |
| 14 | Air leak test | - | - | - |
Maximum load for one tray may not exceed the maximum load for the autoclave
{8}------------------------------------------------
# MODEL NUMBER 69120 SP-2A
## CHAMBER (inches) 24x36x48
| Cycle<br>Number | Program | Temperature<br>°F | Maximum<br>Load (lbs) | Maximum<br>Trays |
|-----------------|---------------|-------------------|-----------------------|------------------|
| 01, 02 | Unwrapped | 250 | 192 | 2 |
| 03, 04 | Unwrapped | 270/274 | 192 | 2 |
| 05, 06 | Wrapped | 270 | 192 | 2 |
| 07 | Wrapped | 274 | 192 | 2 |
| 08 | Wrapped | 250 | 192 | 2 |
| 09, 10, 11, 12 | Liquids | 250 | 192 | 2 |
| 13 | B & D test | 274 | 192 | 2 |
| 14 | Air leak test | - | - | - |
Maximum load for one tray may not exceed the maximum load for the autoclave
## MODEL NUMBER 69180 SP-2HA
## CHAMBER (inches) 24x36x83
:
:
| Cycle<br>Number | Program | Temperature<br>°F | Maximum<br>Load (lbs) | Maximum<br>Trays |
|-----------------|---------------|-------------------|-----------------------|------------------|
| 01, 02 | Unwrapped | 250 | 256 | 2 |
| 03, 04 | Unwrapped | 270/274 | 256 | 2 |
| 05, 06 | Wrapped | 270 | 256 | 2 |
| 07 | Wrapped | 274 | 256 | 2 |
| 08 | Wrapped | 250 | 256 | 2 |
| 09, 10, 11, 12 | Liquids | 250 | 256 | 2 |
| 13 | B & D test | 274 | 256 | 2 |
| 14 | Air leak test | - | - | - |
Maximum load for one tray may not exceed the maximum load for the autoclave
{9}------------------------------------------------
## MODEL NUMBER 6690 SP-1V
# CHAMBER (inches) 26x26x39
| Cycle<br>Number | Program | Temperature<br>°F | Maximum<br>Load (lbs) | Maximum<br>Trays |
|-----------------|---------------|-------------------|-----------------------|------------------|
| 01, 02 | Unwrapped | 250 | 144 | 2 |
| 03, 04 | Unwrapped | 270/274 | 144 | 2 |
| 05, 06, 08 | Wrapped | 270 | 144 | 2 |
| 07 | Wrapped | 274 | 144 | 2 |
| 09, 10, 11, 12 | Liquids | 250 | 144 | 2 |
| 13 | B & D test | 274 | 144 | 2 |
| 14 | Air leak test | - | - | - |
Maximum load for one tray may not exceed the maximum load for the autoclave
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.