← Product Code [KNW](/productcode/KNW) · K063697

# SPECIAL SPROTTE, CHIBA, DR STEINHOFF KIT (K063697)

_PAJUNK GmbH Medizintechnologie · KNW · Feb 14, 2007 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063697

## Device Facts

- **Applicant:** PAJUNK GmbH Medizintechnologie
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Feb 14, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are not appropriate for bone biopsies. Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy. Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue. Pajunks Breast localization cannulas/ needles are intended for diagnostic sampling of breast tissue during breast biopsy procedures. They are to be used for diagnostic purposes only and are not intended for therapeutic uses. The Breast localization cannulas/ needles are indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. This device is not indicated for use under MR based imaging technologies, such as MRI.

## Device Story

Manual surgical instruments for soft tissue biopsy, puncture, and aspiration; includes various needle/cannula configurations (Chiba, Sprotte, Steinhoff kit, breast localization). Devices consist of medical-grade stainless steel, plastic luer hubs, stylets, and optional depth-stops. Used by clinicians in clinical settings for percutaneous procedures. Operates via manual insertion to obtain tissue samples or perform aspiration. Breast localization needles used under mammographic guidance to sample breast lesions. Provides tissue for histologic examination. Benefits include minimally invasive diagnostic tissue acquisition. Single-use, EO-sterilized, latex-free.

## Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993. Sterilization validated per AAMI/ISO 10993-7.

## Technological Characteristics

Medical-grade stainless steel needles/cannulas; plastic luer hubs; stylets; depth-stops. EO-sterilized (SAL < 10^-6). Single-use. Manual operation. No electronic components or software.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Manan MRI Chiba ([K980211](/device/K980211.md))

## Reference Devices

- Pajunk cannulas ([K040965](/device/K040965.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063697

s 1 of 4

្រុងប្រទេស MEDIZINTECHNOLOGIE

# 510(k) Premarket Notification Submission:

# Summary of Safety and Effectiveness

Date of Preparation: November 29th, 2006

FEB 1 4 2007

| Submitter Information/ production site:                         | USA Contact:                                                                                                                                            |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pajunk GmbH                                                     | Pajunk Medical Systems                                                                                                                                  |
| Karl-Hall-Strasse 01                                            | German American Trade Center                                                                                                                            |
| 78187 Geisingen                                                 | 5126 South Royal Atlanta Drive                                                                                                                          |
| Germany                                                         | 30084 Tucker, Georgia                                                                                                                                   |
| Fon: +49(0)7704-9291-586                                        | USA                                                                                                                                                     |
| Fax: +49(0)7704-9291-605                                        |                                                                                                                                                         |
| Establishment Registration Number: 9611612                      |                                                                                                                                                         |
| Contact:                                                        | Contact                                                                                                                                                 |
| Christian Quass, Director Regulatory Affairs                    | Stefan Dayagi                                                                                                                                           |
| Fon: +49(0)7704-9291-586                                        | Fon: +01(0)770-493-9305                                                                                                                                 |
| Fax: +49(0)7704-9291-605                                        |                                                                                                                                                         |
| E-Mail: christian.quass@pajunk.com                              | E-Mail: stefan.dayagi@pajunk-usa.com                                                                                                                    |
|                                                                 | Contract Sterilizer:                                                                                                                                    |
|                                                                 | Sterigenics SteriPro Lab & EO Facility                                                                                                                  |
|                                                                 | Dreieichstrasse 7                                                                                                                                       |
|                                                                 | 64546 Moerfelden                                                                                                                                        |
|                                                                 | RN: 3002807090                                                                                                                                          |
| Device Information:                                             |                                                                                                                                                         |
|                                                                 | Trade Names: Chiba, Chiba Special, Special Sprotte, Kit acc.<br>Dr. Steinhoff, Initial Puncture needle/ cannula,<br>Breast Localization cannula/ needle |
|                                                                 | Common Name: Pajunks soft tissue biopsy, puncture and<br>aspiration cannulas and needles                                                                |
|                                                                 | Classification Name: Gastroenterology-urology biopsy instrument<br>Handheld biopsy instrument                                                           |
|                                                                 | Classification Reference: 21 CFR § 876.1075, April 1st, 2006<br>21 CFR § 878.4800, April 1st 2006                                                       |
| Additional Classification: Kit, Needle, Biopsy                  |                                                                                                                                                         |
|                                                                 | Additional Classification Reference: Gastroenterology-urology biopsy instrument.                                                                        |
| Poposed Classification: Regulatory Class II                     |                                                                                                                                                         |
| Product Classification Code: KNW                                |                                                                                                                                                         |
| Additional Product Classification Code: FCG                     |                                                                                                                                                         |
|                                                                 | Classification Panel: Gastroenterology/ Urology                                                                                                         |
| Additional Review Advisory Committee: General & Plastic Surgery |                                                                                                                                                         |
| Predicate Devices:                                              | 1. K980211 Manan MRI Chiba, spinal,<br>breast localization etc.                                                                                         |

{1}------------------------------------------------

K063697

pg 2 of 4

DIZINTECHNOLOGIE

#### Indications for use

#### Breast localization cannulas/ needles

Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue.

#### Soft tissue puncture and aspiration kit acc. Steinhoff

Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy.

#### Soft tissue biopsy needle/ cannula, Soft tissue puncture cannula/ needle, Initial Puncture Cannula/ needle, Soft tissue aspiration cannula/ needle, Chiba needle for soft tissue biopsy, puncture and aspiration

Chiba Special tip grinding cannulal needle for soft tissue biopsy, puncture and aspiration

#### Special Sprotte tip cannula/ needle for aspiration, puncture and biopsy

Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are EO-sterilized, latex-free devices for single use.

It is not appropriate for bone biopsies.

#### Device Description

### Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles

The cannulas and needles consist of medical grade steel, plastic luer hub and stylet (stabilizing mandrin inside cannula). For safety reasons, the graduated puncture cannula also has an additional depthstop installed at the shaft of the cannula.

#### Aspiration puncture cannula type Chiba Special

The cannula Type Chiba Special has been developed by PAJUNK® with an approx. 1 cm matted finish of the cannula tip and three additional ring-markings.

#### Aspiration puncture cannula type Chiba with beveled tip

For safety reasons, the graduated puncture cannula/ needle has an additional depth-stop installed at the shaft. As a standard, the cannulas/ needles with beveled tip are normally equipped with a handle plate and a depth stop.

#### Initial puncture cannula

This universally usable puncture cannula/ needle made of high-grade stainless steel is suitable for all percutaneous punctures with guidance wires. It is available in a 2-parted and in a 3-parted version, with and without graduation. The outer cannula has a rounded, blunt tip, and it is designed to match the inside stylet-cannula.

#### Special puncture cannula with atraumatic Sprotte tip

This cannula with a Special Sprotte tip is intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. It is an EO-sterilized, latex-free device for single use.

#### Interventinal micro cannula Chiba type

For better depth control, the interventional micro-cannula features graduated collars at intervals of 1 cm.

{2}------------------------------------------------

K063697

pg 3 of 4

MEDIZINTECHNOLOGIE

#### Aspiration-puncture cannula with injecting tube acc. Steinhoff

The puncture kit according Dr. Steinhoff contains a Chiba-type puncture cannula/ needle with a cannula rider and a flexible tube. The millimeter-precise introduction of the puncture needle by means of the cannula rider is facilitated with the aid of the centimeter-graduation and the corresponding markings at intervals of 5 mm. Thereby, the needle can be quided safely, without lacking the necessary flexibility.

#### Breast Localization cannulas

Pajunks breast localization cannulas can be used in Mammographic procedures to obtain breast lesion tissue.

#### Predicate Devices:

Predicate Device for Pajunks soft tissue biopsy and aspiration cannulas and needles are the cannulas and needles cleared for market by Manan in K980211. These devices have identical indications for use and the same technical specification in materials and grinding as Pajunks cannulas.

Pajunks Chiba cannulas/ needles are already cleared for a different intended use (anesthesia conduction, spinal and epidural) in K040965.

The detailed discussion of substantial eqivalence can be found in Section 12 of this submission.

#### Sterilization

Pajunks soft tissue biopsy and aspiration cannulas and needles are supplied packed in a blister as single-use, sterile, pyrogenfree and latex free disposable cannulas.

The contract sterilizer and the sterilizing process at Sterigenics is the same as that one used for all further Pajunk products already cleared for market in the USA. All cannulas are very similar in dimensions and materials.

Sterilization of Pajunks soft tissue biopsy and aspiration cannulas and needles at Sterigenics is being done according to a documented process. This Ethylene Oxide sterilization process is recurrently evaluated for suitability and effectiveness and the results are acceptable.

Sterilization of the Pajunks soft tissue biopsy and aspiration cannulas and needles is accomplished using Ethylene Oxide (EO) sterilization to a Sterility Assurance Level (SAL) of < 10°. After manufacture the cannulas are arranged in a foil bag which is sealed. EO sterilization is accomplished with exposure to 100% EO in accordance with AAMI/ISO 10993-7.

#### Packaging and Labeling

The cannulas and needles subject to this premarket notification are packed and labeled employing the same validations, in-process-controls, procedures and materials as Pajunks disposable anesthesia conducting cannulas/ needles already cleared for market.

A process descriptions for labeling and packaging can be found in section 13.0 of this submission.

#### Biocompatibility status

All materials employed in the manufacturing process that may come in contact with blood, tissue or fluids to be injected have been cleared in Pajunks former 510(k) applications. Therefore they are deemed to be biocompatible. Testing according to ISO 10993 has been conducted successfully employing cannulas/ needles that consists of identical materials.

Furthermore these materials are long term proven materials for the use with medical devices.

PMN

{3}------------------------------------------------

K063697

pg 4 of 4

MEDIZINTECHNOLOGIE

#### Standards

There are no special standards applicable for the cannulas and needles. Applicable sections of FDA´s GUIDANCE FOR THE CONTENT OF PREMARKET NOTIFICATIONS FOR BIOPSY DEVICES USED IN GASTROENTEROLOGY AND UROLOGY have been taken into regard.

A list of all standsrds applicable as well as a declaration of conformity with this standards can be found in section 9.0 of this submission.

#### Conclusion:

The comparison between the predicate devices and the proposed device in section 12 of this submission demonstrates that the proposed device is safe and effective, as well as substantially equivalent to the predicate devices.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines forming the snake and staff. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Pajunk GmbH % Christian Quass Director Regulatory Affairs Karl-Hall-Strasse 01 78187 Geisingen, Germany

FEB 1 4 2007

Re: K063697

Trade/Device Name: Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles Initial puncture cannula/needle Special puncture cannula with atraumatic Sprotte tip Aspiration puncture cannula type Chiba Special Aspiration puncture cannula type Chiba with beveled tip Interventional micro cannula Chiba type Soft tissue puncture and aspiration kit acc. Steinhoff Breast localization cannulas/needles Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: II Product Code: KNW. FCG Dated: December 9, 2006 Received: December 13, 2006

Dear Christian Quass:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

#### Page 2 - Christian Quass

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legal/y marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may ohtain other general information on your responsibilities under the Act from the Division of Swall Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K063697

g l ots

K063697

Image /page/6/Picture/3 description: The image shows the word "PAJUNK" in large, bold, white letters. Below this word is the word "MEDIZINTECHNOLOGIE" in smaller, black letters. The background of the image is a textured pattern.

#### Indications for use

510(k) Number:

Device Name:

Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles Initial puncture cannula/ needle Special puncture cannula with atraumatic Sprotte tip Aspiration puncture cannula type Chiba Special Aspiration puncture cannula type Chiba with beveled tip Interventional micro cannula Chiba type

Indications for Use:

Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration.

They are not appropriate for bone biopsies.

Prescription Use (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRA, Office of Device Evaluation (ODE)

P. Mo

(Division Sign-Of Division of General, Rest prative, and Neurological Devices

**510(k) Number** /K06362

Page 1 of 1

AK063697.doc

{7}------------------------------------------------

K063697

pg 2 of 3

Image /page/7/Picture/3 description: The image contains the word "PAJUNK" in large, bold, white letters. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, thinner letters. The background of the image is a grainy, dark texture, which contrasts with the white text.

Indications for use

| 510(k) Number:       | K063697                                                |
|----------------------|--------------------------------------------------------|
| Device Name:         | Soft tissue puncture and aspiration kit acc. Steinhoff |
| Indications for Use: |                                                        |

Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy.

Prescription Use****X****
(Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{8}------------------------------------------------

**Indications for use**

| 510(k) Number: |
|----------------|
|----------------|

K063697

Device Name:

Breast localization cannulas/ needles

Indications for Use:

Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue.

Pajunks Breast localization cannulas/ needles are intended for diagnostic sampling of breast tissue during breast biopsy procedures. They are to be used for diagnostic purposes only and are not intended for therapeutic uses.

The Breast localization cannulas/ needles are indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

This device is not indicated for use under MR based imaging technologies, such as MRI.

Prescription Use ***X***
(Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

## Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K063697](https://fda-staging.innolitics.com/device/K063697)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).