# MODIFICATIONS TO CD HORIZON SPINAL SYSTEM (K063670)

_Medtronic Sofamor Danek · NKB · Jun 28, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063670

## Device Facts

- **Applicant:** Medtronic Sofamor Danek
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jun 28, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications. With the exception of degenerative disc disease, the CD HORIZON LEGACY 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients. The CD HORIZON® SPIRE Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor. When used as a pedicle screw fixation system in skeletally mature patients, the CD HORIZON® AGILE™ Dynamic Stabilization device is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the degenerative spondylolisthesis with objective evidence of neurologic impairment and/or failed previous fusion in the thoracic, lumbar and/or sacral spine. Additionally, when used as a pedicle screw fixation system, the CD HORIZON® AGILE™ Dynamic Stabilization device is indicated for use in patients who are receiving fusions with autogenous graft only; who are having the device fixed or attached to the lumbar or sacral spine; and/or are having the device removed after the development of a solid fusion mass. In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

## Device Story

CD HORIZON® Spinal System is a modular orthopedic construct for spinal immobilization and stabilization. System comprises rods, hooks, screws, plates, and connectors. Subject of this submission is the addition of 5.5mm diameter PEEK Optima 1 rods. Device is implanted by surgeons in clinical/OR settings to facilitate spinal fusion. Components are rigidly locked in tailor-made configurations. System can connect to VERTEX™ Reconstruction System. Benefits include stabilization of spinal segments to support fusion and treat degenerative conditions, trauma, or deformities. AGILE™ Dynamic Stabilization component uses titanium and polycarbonate-urethane for motion-sparing stabilization.

## Clinical Evidence

No clinical data provided. Substantial equivalence based on bench testing and risk analysis.

## Technological Characteristics

System components include rods, hooks, screws, and plates. Materials include PEEK Optima 1, commercially pure titanium, titanium alloy, and polycarbonate-urethane. Modular design allows for rigid locking in various configurations. Non-sterile, intended for surgical implantation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- CD HORIZON® components ([K042962](/device/K042962.md))
- CD HORIZON® components ([K050809](/device/K050809.md))
- CD HORIZON® components ([K032033](/device/K032033.md))
- CD HORIZON® components ([K042025](/device/K042025.md))
- CD HORIZON® components ([K060615](/device/K060615.md))

## Reference Devices

- VERTEX™ Reconstruction System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K063670" in a handwritten style. The text appears to be a code or identifier, with a combination of letters and numbers. The characters are bold and slightly slanted, giving the impression of quick handwriting.

# CD HORIZON® Spinal System Summary of Safety and Effectiveness December 2006

Medtronic Sofamor Danek, Inc. USA I. Company: 1800 Pyramid Place JUN 2 8 2007 Memphis. TN 38132 (901) 396-3133

Contact: Christine Scifert Group Director, Regulatory Affairs

#### II. Proposed Proprietary Trade Name: CD HORIZON® Spinal System

- III. Classification Name(s)/Product Code(s): Spinal Interlaminal Fixation and Spinal Intervertebral Fixation Orthosis and/or Pedicle Screw Spinal System (per 21 CFR Section 888.3050, 888.3060 and/or 888.3070) Product Codes: MNI, MNH, KWP, KWQ, NQP and NKB
#### IV. Product Description

The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD HORIZON® Spinal System can be connected to the VERTEX™ Reconstruction System through a rod connector.

Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws.

CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

The CD HORIZON® AGILE™ Dynamic Stabilization device is comprised of four components and is comprised by a combination of materials. The proximal and distal rod components are manufactured from commercially pure titanium. The cable is fabricated from a 7x7 filament yarn made from titanium alloy, with a cylinder rotary-swaged to the end, made from the same material. The spacer portion of the device is manufactured from Polycarbonate-Urethane.

The purpose of this 510(k) submission is to include additional 5.5mm diameter rods manufactured from PEEK Optima 1 to the CD HORIZON® Spinal System.

{1}------------------------------------------------

#### V. Indications

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications.

With the exception of degenerative disc disease, the CD HORIZON LEGACY 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients.

The CD HORIZON® SPIRE Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.

When used as a pedicle screw fixation system in skeletally mature patients, the CD HORIZON® AGILE™ Dynamic Stabilization device is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the degenerative spondylolisthesis with objective evidence of neurologic impairment and/or failed previous fusion in the thoracic, lumbar and/or sacral spine. Additionally, when used as a pedicle screw fixation system, the CD HORIZON® AGILE™ Dynamic Stabilization device is indicated for use in patients who are receiving fusions with autogenous graft only; who are having the device fixed or attached to the lumbar or sacral spine; and/or are having the device removed after the development of a solid fusion mass.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

#### VI. Substantial Equivalence

Documentation, including a risk analysis, was provided which demonstrated the subject rods to be substantially equivalent to predicate CD HORIZON® components previously cleared in K042962. K050809, K032033, K042025 and K060615.

page 2 of 2

C.30043

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with outstretched wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 8 2007

Medtronic Sofamor Danek, Inc. % Ms. Christine Scifert Group Director, Regulatory Affairs 2600 Sofamor Danek Drive Memphis, Tennessee 38132

Re: K063670

Trade Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, NQP, MNI, MNH, KWP, KWQ Dated: May 30, 2007 Received: May 31, 2007

Dear Ms. Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Ms. Christine Scifert

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or at the Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1 December 2006

510(k) Number (if known):__k06367 ס

CD HORIZON® Spinal System Device Name: ____________

### Indications for Use:

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORLZON® Spinal System may also be used for the same indications.

With the exception of degenerative disc disease, the CD HORIZON LEGACY 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients.

The CD HORIZON® SPIRE Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.

When used as a pedicle screw fixation system in skeletally mature patients, the CD HORIZON® AGILE™ Dynamic Stabilization device is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the degenerative spondylolisthesis with objective evidence of neurologic impairment and/or failed previous fusion in the thoracic, lumbar and/or sacral spine. Additionally, when used as a pedicle screw fixation system, the CD HORIZON® AGILE™ Dynamic Stabilization device is indicated for use in patients who are receiving fusions with autogenous graft only; who are having the device fixed or attached to the lumbar or sacral spine; and/or are having the device removed after the development of a solid fusion mass.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

| Prescription Use<br>(Part 21 CFR 801 Subpart D)                          | X                        | AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) |        |
|--------------------------------------------------------------------------|--------------------------|--------|------------------------------------------------|--------|
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) |                          |        |                                                |        |
| (Division Sign-Off)                                                      |                          |        |                                                |        |
| Division of General, Restorative Device Evaluation (ODE)                 |                          |        |                                                |        |
|                                                                          | and Neurological Devices |        |                                                |        |
| 510(k) Number                                                            | 1663670                  |        |                                                | 000111 |

---

**Source:** [https://fda-staging.innolitics.com/device/K063670](https://fda-staging.innolitics.com/device/K063670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).