CENTRICITY RADIOLOGY RA600, CARDIOLOGY CA1000 AND DIGITAL HARDCOPY
K063628 · Ge Healthcare Integrated IT Solutions · LLZ · Dec 26, 2006 · Radiology
Device Facts
Record ID
K063628
Device Name
CENTRICITY RADIOLOGY RA600, CARDIOLOGY CA1000 AND DIGITAL HARDCOPY
Applicant
Ge Healthcare Integrated IT Solutions
Product Code
LLZ · Radiology
Decision Date
Dec 26, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
RA600/CA1000/Digital Hardcopy is intended for viewing and diagnostic interpretation of images acquired from CT, MR, CR, DR, US, XA and other DICOM-compliant medical imaging systems when installed on suitable commercial-standard PC hardware. RA600 / CA1000 is intended for use as a primary diagnostic and analysis workstation in Radiology/ Cardiology or other departments. It is also intended for use as a clinical review workstation throughout the healthcare facility and may be part of a larger PACS configuration. Digital Hardcopy is intended for use primarily as a workstation for the high volume burning of CDs or DVDs containing DICOM medical images and associated diagnostic report or analysis information. CD /DVD burning and disk labeling are done via a commercially available external robotics device. RA600/CA1000/Digital Hardcopy receives imaging studies and data over LAN, WAN, intranet or internet from a PACS server or directly from a DICOM -compliant modality or archive utilizing both lossless and lossy compression. It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application. The RA600/CA1000/Digital Hardcopy workstation may interface with various information systems within the healthcare environment, such as the HIS, RIS, and CVIS. It may be sold as software only, or as a turnkey system. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
PC-based DICOM workstation platform for medical imaging management; accepts, transmits, prints, displays, stores, edits, and processes medical images/data. Operates as standalone software on commercial PCs or turnkey hardware systems. Used by radiologists, cardiologists, physicians, technologists, and nurses in clinical settings (radiology, cardiology, general healthcare facilities). Integrates with PACS, HIS, RIS, and CVIS via DICOM networking. Digital Hardcopy component manages high-volume CD/DVD burning via external robotics. Supports third-party software (RIS, voice recognition, advanced imaging analysis). Facilitates diagnostic interpretation and clinical review; aids clinical decision-making by providing centralized access to imaging studies.
Clinical Evidence
Bench testing only. Quality assurance measures included risk analysis, requirements and design reviews, unit-level module verification, system integration testing, and final acceptance testing/validation.
Technological Characteristics
PC-based DICOM workstation; software-based or turnkey hardware configuration. Supports DICOM networking (LAN/WAN/intranet/internet). Compatible with commercial-standard PC hardware. Supports lossless and lossy compression. Mammographic interpretation requires 5 Mpixel resolution monitors.
Indications for Use
Indicated for radiologists, cardiologists, physicians, technologists, and nurses for viewing and diagnostic interpretation of medical images (CT, MR, CR, DR, US, XA) and clinical review. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Accusketch Cardiac Quantitative System W/ Advanced Analysis Components (K023100)
Submission Summary (Full Text)
{0}------------------------------------------------
Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital Hardcopy 510k Premarket Notification Submission
# K663628
### Section 2 510(k) Summary of Safety and Effectiveness
| Date: | December 19, 2006 | DEC 26 2006 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Submitter: | GE Healthcare Integrated IT Solutions<br>540 W Northwest Highway<br>Barrington, IL 60010 | |
| Contact Person: | Karen M. Lunde<br>Sr. Regulatory Affairs Leader<br>GE Healthcare Integrated IT Solutions<br>Phone: (847 277-6092<br>Fax: (414) 755-0655 | |
| Device: | | |
| Trade Name: | Centricity Radiology RA600 / Centricity Cardiology CA1000 /<br>Centricity Digital Hardcopy | |
| Common/Usual Name: | Picture Archiving and Communications Systems and Workstation | |
| Classification Names: | 21 CFR 892.2050 System, Image Processing, Radiological | |
| Predicate Device: | K042525 Centricity Radiology RA600 / Centricity Cardiology CA1000 /<br>Centricity Digital Hardcopy Workstation<br>K023178 Innova 4100 (Specific features)<br>K023100 Accusketch Cardiac Quantitative System W/ Advanced<br>Analysis Components (Specific features) | |
| Device Description: | Centricity Radiology RA600 / Centricity Cardiology CA1000 /<br>Centricity Digital Hardcopy is a PC-based DICOM workstation platform<br>which provides scaleable image and data management solutions for<br>medical imaging. This software-based product provides capabilities<br>for the acceptance, transmission, printing, display, storage, editing<br>and digital processing of medical images and associated data. | |
| | RA600/CA1000 / Digital Hardcopy may be combined with a PACS<br>network or connected directly to a modality through the use of DICOM<br>networking. The RA600/CA1000 / Digital Hardcopy software<br>application may be sold as a standalone product for use with 'off the<br>shelf' PC hardware that meets minimum specifications or as a turnkey<br>solution integrated with hardware components to be configured to<br>meet the users specific needs. | |
| | RA600 / CA1000 / Digital Hardcopy can also provide the hardware and<br>OS platform for a user to operate 3rd party software and/or other GE<br>software applications such as RIS, voice recognition, or advanced<br>imaging analysis, and view any data presented through those<br>applications. | |
| | RA600 / CA1000 can act as an image repository for the Centricity Web<br>Viewer application. | |
| | Targeted users of this system are trained professionals, including | |
Targeted users of this system are trained professionals, including radiologists, cardiologists, physicians, technologists and nurses.
{1}------------------------------------------------
Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital Hardcopy 510k Premarket Notification Submission
KUL 3628
RA600/CA1000/Digital Hardcopy is intended for viewing and Intended Use: diagnostic interpretation of images acquired from CT, MR, CR, DR, US, XA and other DICOM-compliant medical imaging systems when installed on suitable commercial-standard PC hardware. RA600 / CA1000 is intended for use as a primary diagnostic and analysis workstation in Radiology/ Cardiology or other departments. It is also intended for use as a clinical review workstation throughout the healthcare facility and may be part of a larger PACS configuration.
> Digital Hardcopy is intended for use primarily as a workstation for the high volume burning of CDs or DVDs containing DICOM medical images and associated diagnostic report or analysis information. CD /DVD burning and disk labeling are done via a commercially available external robotics device.
> RA600/CA1000/Digital Hardcopy receives imaging studies and data over LAN, WAN, intranet or internet from a PACS server or directly from a DICOM -compliant modality or archive utilizing both lossless and lossy compression. It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application. The RA600/CA1000/Digital Hardcopy workstation may interface with various information systems within the healthcare environment, such as the HIS, RIS, and CVIS. It may be sold as software only, or as a turnkey system.
> Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
- The Centricity Radiology RA600 / Centricity Cardiology CA1000 / Technology: Centricity Digital Hardcopy employs the same functional scientific technology as its predicate devices.
- Test Summary: The Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital Hardcopy complies with the voluntary standards as detailed in Section 12 Specific Standards and Guidance. The following quality assurance measures were applied to the development:
- Risk Analysis .
- . Requirements Reviews
- Design Reviews .
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final acceptance testing (Validation)
- . Performance testing
- Safety testing .
- Conclusion: GE considers features of the Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital Hardcopy are equivalent to those of the predicate devices.
{2}------------------------------------------------
30-Dcc-2004 (Vcr. 2) Page I of I
1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1
11.00 11
# Third Party Review Quality Assessment
Section 1 - Submission Information
| 510(k) No.: | K063628 |
|------------------------------------|--------------|
| Third Party Organization: | Underwriters |
| Third Party's Primary Reviewer(s): | Silvia Ankwa |
| ODE/OIVD Division: | DRARD |
| Branch/Team: | RDB |
Section 2 - 510(k) Decision
| Third party recommendation: | SE <span style="text-decoration:underline;">SE</span> NSE ____ Other (specify): ________ |
|-----------------------------|------------------------------------------------------------------------------------------|
| ODE/OIVD final decision: | SE <span style="text-decoration:underline;">SE</span> NSE ____ Other (specify): ________ |
Section 3 - Assessment of Third Party Review
| Review Element | Rating (check one) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------|----------------|
| | Adequate | Minor Issue(s) | Major Issue(s) |
| a. Determination of device eligibility for third party review | ✓ | | |
| b. Extent of pre-submission consultation with ODE/OIVD division | ✓ | | |
| c. Organization and format of review documentation | ✓ | | |
| d. Determination of 510(k) administrative completeness (screening review) | ✓ | | |
| e. Summary of device characteristics, intended use, and performance (including accessories, if applicable) and reason for 510(k) submission | ✓ | | |
| f. Comparison to legally marketed devices-identification and analysis of key similarities and differences | ✓ | | |
| g. Rationale for conclusions and recommendation | ✓ | | |
| h. Use of guidance documents and standards | ✓ | | |
| i. Resolution of 510(k) deficiencies and FDA requests for additional information | ✓ | | |
| j. Scope of reviewer expertise and use of consulting reviewers | ✓ | | |
| k. Other (specify): | | | |
Comments (explanation of ratings/tssues):
Section 4 ODE/OIVD Assessar Information
Assessed by:
Date: 12-13-06
Tel. No.: 24
3-06 Tel. No.: 240-271-3664
Routing: Division Christompleted assessment (this page only) to inside front cover of 510(k). DMC Forward this page only to Fire Rechen, POS/ODE, Rm. 120J, Corp. Rival (HF/ 30) 1
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized graphic in the center. The text around the perimeter reads "DEPARTMENT OF HEALTH & HUMAN SERVICES". The graphic in the center appears to be a stylized representation of an eagle or bird, with three curved lines forming the wings and body.
## DEC 2 6 2006
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
GE Healthcare Integrated IT Solution % Ms. Silvia Ankova Senior Project Engineer Underwriters Laboratories, Inc. 333 Pfingsten Rd. NORTHBROOK IL 60062
Re: K063628
Trade/Device Name: Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital
Hardcopy Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 5, 2006 Received: December 6, 2006
Dear Ms. Ankova:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/11 description: The image is a black and white circular logo or seal. The text "1906-2006" is at the top of the seal. The letters "DA" are in the center of the seal in a bold, blocky font. The word "Centennial" is below the letters "DA" in a cursive font.
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I tods be actised has i Brinination that your device complies with other requirements of the Act that I Dri has made a actor regulations administered by other Federal agencies. You must or any I odolar bakated and strenging, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K I at 0077, idocinig (21 CFR Part 820); and if applicable, the electronic forul in the quality by tions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ough finding of substantial equivalence of your device to a legally prematicated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 cr from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
510(k) filed on October 16, 2006
Centricity Radiology RA600 / Centricity Cardiology CA1000 / Centricity Digital Device Name: Hardcopy
K063628;
#### Indications for Use:
RA600/CA1000/Digital Hardcopy is intended for viewing and diagnostic interpretation of images acquired from CT, MR, CR, DR, US, XA and other DICOM-compliant medical imaging systems when installed on suitable commercial-standard PC hardware. RA600 / CA1000 is intended for use as a primary diagnostic and analysis workstation in Radiology/ Cardiology or other departments. It is also intended for use as a clinical review workstation throughout the healthcare facility and may be part of a larger PACS configuration.
Digital Hardcopy is intended for use primarily as a workstation for the high volume burning of CDs or DVDs containing DICOM medical images and associated diagnostic report or analysis information. CD /DVD burning and disk labeling are done via a commercially available external robotics device.
RA600/CA1000/Digital Hardcopy receives imaging studies and data over LAN, WAN, intranet or internet from a PACS server or directly from a DICOM -compliant modality or archive utilizing both lossless and lossy compression. It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application. The R4600/CA1000/Digital Hardcopy workstation may interface with various information systems within the healthcare environment, such as the HIS, RIS, and CVIS. It may be sold as software only, or as a turnkey system
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Per 21 CFR 801.109) OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Barbara Bucher for NCB
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K063628
-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.