The intended use of PerioGlas Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include: - Periodontal/Infrabony defects - Ridge Augmentation (sinusotomy, osteotomy, cystectomy) - Extraction sites maintenance/augmentation, implant (ridge preparation/ placement) - Sinus lifts - Cystic cavities - Cranio-facial augmentation. For larger defects, a mixture of PerioGlas Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Device Story
PerioGlas Putty is a synthetic, osteoconductive bone void filler; composed of bioactive glass (45S5 Bioglass) particulate and a gelatin binder. Supplied sterile; intended for dental/craniofacial surgical use by clinicians. Gelatin binder provides malleability for placement; hydrates when wetted; rapidly absorbed post-implantation to allow cell infiltration. Bioglass particles remain as a scaffold for host bone ingrowth; gradually resorbed and replaced by new bone tissue. Used alone or as an extender with autograft/allograft. Not for load-bearing applications; requires internal/external stabilization for fracture fixation. Benefits include biocompatible scaffold for natural bone regeneration.
Clinical Evidence
In vivo performance data and supporting in vitro data were presented to demonstrate substantial equivalence. No specific clinical trial metrics (e.g., p-values, sensitivity) were detailed in the summary; the device relies on the established safety and efficacy profile of the predicate bioactive glass materials.
Technological Characteristics
Osteoconductive bone void filler; two-component system: 45S5 Bioglass (calcium-phosphorus-sodium-silicate) particulate and gelatin binder. Malleable putty form factor when wetted. Non-structural scaffold; resorbable. Sterilized. No software or electronic components.
Indications for Use
Indicated for patients requiring bone void filling or augmentation in oral, dental intraosseous, and craniofacial defects, including periodontal/infrabony defects, ridge augmentation, extraction sites, sinus lifts, and cystic cavities. Contraindicated in individuals with known allergies to gelatin.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
NovaBone Putty - Resorbable Bone Void Filler (K051617)
Submission Summary (Full Text)
{0}------------------------------------------------
173
2007
Image /page/0/Picture/1 description: The image shows the words "510(k) Summary" on the left side of the image. On the right side of the image, the words "NOVABONE SYNTHETIC BONE GRAFT PRODUCTS" are stacked on top of each other. Below that, the number "K063596" is written in cursive with a line underneath it.
11/30/06
#### Submitter Information: 1.
PerioGlas Putty - Bioactive Bone Graft Gel
| Name: | NovaBone Products, LLC |
|------------|----------------------------------------------------|
| Address: | 13709 Progress Boulevard, #33<br>Alachua, FL 32615 |
| Telephone: | (386) 462-7660 |
| Facsimile: | (386) 418-1636 |
| Contact: | David M. Gaisser |
#### 2. Name of Device:
| Trade Name: | PerioGlas Putty -- Bioactive Bone Graft Gel |
|----------------------|------------------------------------------------------------------------------|
| Common Name: | Osteoconductive Bone Void Filler<br>Synthetic Resorbable Bone Graft Material |
| Classification Name: | Endosseous Implant for Bone Filling and/or Augmentation |
#### 3. Legally Marketed Predicate Device:
| Predicate #1: | PerioGlas -- Synthetic Bone Graft Particulate<br>[K053387, K040278, K992416, K962492, K930115]<br>(Also named as NovaBone per Special 510(k) K000149) |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate #2: | NovaBone Putty - Resorbable Bone Void Filler<br>[K051617] |
#### 4. Device Description
PerioGlas Putty is an osteoconductive, bioactive, bone void filler device. The device is intended for dental intraosseous, oral, and cranio-/maxillofacial bony defects and is supplied sterile. PerioGlas Putty is a two-component device composed of a synthetic bioactive glass particulate mixed with a binder. The a melt-derived calcium-phosphorus-sodium-silicate major component is designed specifically for for its absorbability (Bioglass) particulate and osteoconductive nature. The second component is gelatin powder, selected for its biocompatibility and physical gelation properties to act as a temporary binding agent for the particulate. The gelatin is rapidly absorbed from the graft site to permit tissue infiltration between the Bioglass particles and replacement of the particles by host bone during the healing process.
{1}------------------------------------------------
510(k) Premarket Notification NovaBone Products, LLC NovaBone Putty - Resorbable Bone Void Filler Confidential
K063596
2 of 3
#### 5. Intended Use
PerioGlas Putty is indicated to be packed into bony voids or gaps to fill and/or augment oral, dental intraosseous, and craniofacial defects. These defects may include: periodontal/infrabony defects; alveolar ridge augmentation (sinusotomy. osteotomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); sinus lifts; cystic defects; craniofacial augmentation. PerioGlas Putty may be used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (demineralized freeze dried bone), or it may be mixed with either as a bone graft extender.
#### 6. Technological Characteristics
The technological characteristics of the PerioGlas Putty device are similar to those of the predicates. The device and the predicates are designed as osteoconductive space-filling devices to be gently packed into defect sites and used as nonstructural scaffolds for the body's natural healing and bone regeneration process. The device indications are the same as for the PerioGlas predicate.
The PerioGlas Putty device is identical in composition and formulation to the NovaBone Putty predicate device. The primary component of PerioGlas Putty is identical to the bioactive glass (45S5 Bioglass) particulate found in the PerioGlas and NovaBone Putty predicates. This synthetic material is both biocompatible and osteoconductive. The main technological difference between PerioGlas Putty and the PerioGlas predicate device is that while PerioGlas contains only the particulate bioactive glass component, the PerioGlas Putty contains a second component (gelatin) that acts as a temporary binding agent between the particles when wetted with water. When wetted, this gelatin binder hydrates and fills the space between the Bioglass particles, resulting in a soft malleable putty to aid product handling and placement. The gelatin binder then is rapidly removed from the implantation site via normal physiologic processes, opening space between the bioactive glass particles for cell infiltration and bone formation.
After binder absorption, the residual Bioglass particles of the PerioGlas Putty device remain in the graft site. This particulate acts as a scaffold for bone ingrowth, with gradual absorption and replacement by new bone tissue. The PerioGlas Putty and predicate devices are substantially absorbed within the sixmonth timeframe normally associated with bone remodeling, the devices being replaced by new bone tissue.
#### 7. Warnings and Precautions
PerioGlas Putty does not possess sufficient mechanical strength to support loadbearing defects prior to hard tissue ingrowth. In cases of fracture fixation or
ﺐ ﺍ
{2}------------------------------------------------
Confidential
K063596
373
where load support is required, standard internal or external stabilization techniques must be followed to obtain rigid stabilization in all planes.
PerioGlas Putty is intended for use by clinician familiar with bone grafting and internal/external fixation techniques. PerioGlas Putty must not be used to gain screw purchase or to stabilize screw placement.
PerioGlas Putty contains gelatin. This device should not be used by individuals having known allergies to gelatin products.
#### 8. Complications
Possible complications are the same as to be expected of autogenous bone grafting procedures. These may include: superficial wound infection, deep wound infection, deep wound infection with osteomyelitis, delayed union, loss of reduction, failure of fusion, loss of bone graft, graft protrusion and / or dislodgement, and general complications that may arise from anesthesia and / or surgery. Complications specific to oral/dental use are those as may be typically observed for similar bone grafting procedures and may include: tooth sensitivity, gingival recession, flap sloughing, resorption or ankylosis of the treated root, abscess formation.
The potential for allergic reaction to the gelatin component exists.
#### 9. Conclusion
PerioGlas Putty is claimed to be substantially equivalent to the PerioGlas and NovaBone Putty predicate devices as a non-structural osteoconductive bone void filler for osseous defects. In vivo performance data were presented. Additional supporting in vitro data were supplied.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### FEB 2 2 2007
Mr. David M. Gaisser Vice President NovaBone Products, LLC 13709 Progress Boulevard, #33 Alachua, Florida 32615
Re: K063596
Trade/Device Name: PerioGlas Putty -Bioactive Bone Graft Gel Regulation Number: 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: November 30, 2006 Received: December 6, 2006
Dear Mr. Gaisser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Gaisser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Oves
Chiu-Lin Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Confidential
510(k) Premarket Notification NovaBone Products, LLC PerioGlas Putty - Bioactive Bone Graft Gel
Graft Gel
K063596
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known):
Device Name: __PerioGlas Putty - Bioactive Bone Graft Gel
Indications For Use:
The intended use of PerioGlas Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include:
- Periodontal/Infrabony defects
- · Ridge Augmentation (sinusotomy, osteotomy, cystectomy)
- · Extraction sites maintenance/augmentation, implant (ridge preparation/ placement)
- Sinus lifts
- · Cystic cavities
- · Cranio-facial augmentation
For larger defects, a mixture of PerioGlas Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Prescription Use __XX
OR (Per 21 CFR 801.109) Over-The-Counter Use
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qur
Division of Anesthesiology, Memorial Hospital,
Infection Control, Dental Devices
K.0.63576
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.