# V.A.C. FREEDOM CANISTER, MODEL M6275421; V.A.C. ATS CANISTER, MODELS K6275422, M6275423 (K063590)

_Kci USA, Inc. · OMP · Jan 25, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063590

## Device Facts

- **Applicant:** Kci USA, Inc.
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Jan 25, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The V.A.C. Therapy System is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. It is indicated for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts. The V.A.C.® GranuFoam® Silver™ Dressing is an effective barrier to bacterial penetration and may help reduce infection in the above wound types.

## Device Story

V.A.C. Freedom and V.A.C. ATS therapy device canisters serve as reservoirs for collecting wound exudate and fluids during negative pressure wound therapy. Canisters are molded plastic, designed for insertion into V.A.C. therapy units. Fluid path is sterile for single use. Features include graduation markings for fluid estimation, hydrophobic membranes to isolate collected fluid from the vacuum pump, and optional gel packets to solidify wound fluid. Device maintains sterility via sterile barriers at fluid path openings. Used in clinical settings to support wound healing by removing infectious material and exudate, reducing edema, and promoting granulation tissue formation.

## Clinical Evidence

Bench testing only. Verification of fluid path sterile barrier integrity was performed using accelerated aging, distribution simulation, and microbial aerosol challenge testing. Sterility assurance level of 10^-6 was achieved.

## Technological Characteristics

Molded plastic canisters with graduation markings. Includes hydrophobic membranes for pump isolation and optional gel packets for fluid solidification. Sterile fluid path maintained by sterile barriers at open ends. Sterilization assurance level 10^-6.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- V.A.C. Family of Devices ([K032310](/device/K032310.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063590

Image /page/0/Picture/1 description: The image shows the logo for KCI. The logo consists of a stylized graphic to the left of the letters "KCI". The graphic is made up of several curved lines that are stacked on top of each other. The letters "KCI" are in a bold, sans-serif font, and there is a registered trademark symbol to the right of the "I".

## JAN 2 5 7007 14 - 1 - 510(k) SUMMARY - 510(k) SUMMARY
Product Name: V.A.C. Freedom® and V.A.C. ATS® Therapy Device Canisters

| Date prepared                                                                                                                                                                                  | November 30, 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Name                                                                                                                                                                                           | KCI USA, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Address                                                                                                                                                                                        | 8023 Vantage Dr., San Antonio, TX 78230                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Fax number                                                                                                                                                                                     | (210) 255-6727                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Name of contact person                                                                                                                                                                         | Christy Hubbard Oviatt, Sr. Regulatory Affairs Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Name of the device                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Trade or proprietary<br>name                                                                                                                                                                   | • V.A.C. 300 ml Canister (with Gel) for V.A.C. Freedom®<br>• 500 ml Canister (with Gel) for V.A.C. ATS®<br>• 500 ml Canister for V.A.C. ATS®                                                                                                                                                                                                                                                                                                                                                                     |
| Common or usual<br>name                                                                                                                                                                        | Accessory to Negative Pressure Wound Therapy Device                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification name                                                                                                                                                                            | Accessory to Powered Suction Pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Legally marketed device<br>to which equivalence is<br>claimed                                                                                                                                  | The canisters are proposed to be substantially equivalent<br>to canisters currently marketed in sterile barrier<br>packaging. These canisters were included in the V.A.C.<br>Family of Devices 510(k) K032310.<br>• V.A.C. 300 ml Canister (with Gel) for V.A.C. Freedom®<br>• 500 ml Canister (with Gel) for V.A.C. ATS®<br>• 500 ml Canister for V.A.C. ATS®                                                                                                                                                   |
| Device description                                                                                                                                                                             | The V.A.C. ATS® and V.A.C. Freedom® canisters are<br>provided with a sterile fluid path for single use with the<br>V.A.C. ATS® and V.A.C. Freedom® therapy units.                                                                                                                                                                                                                                                                                                                                                |
| Device function                                                                                                                                                                                | The V.A.C. ATS® and V.A.C. Freedom® canisters function<br>as reservoirs to collect fluids and wound exudates<br>removed from the wounds during treatment with the<br>negative pressure therapy devices.                                                                                                                                                                                                                                                                                                          |
| Device design                                                                                                                                                                                  | The canisters are molded plastic devices designed to<br>insert into the V.A.C. Freedom® or V.A.C. ATS® Therapy<br>units.                                                                                                                                                                                                                                                                                                                                                                                         |
| Differences in<br>device design<br>from predicate<br>device                                                                                                                                    | The only difference between the proposed and the predicate<br>devices is the means by which sterility of the fluid path of the<br>device is maintained. The predicate products are packaged with<br>sterile barrier packaging while the modified canisters are<br>designed with sterile barriers at the open ends of the fluid path of<br>the product to maintain the sterility prior to use.                                                                                                                    |
| Intended use of the<br>device                                                                                                                                                                  | The V.A.C. Therapy System is intended to create an environment<br>that promotes wound healing by secondary or tertiary (delayed<br>primary) intention by preparing the wound bed for closure,<br>reducing edema, promoting granulation tissue formation and<br>perfusion, and by removing exudate and infectious material. It is<br>indicated for patients with chronic, acute, traumatic, subacute<br>and dehisced wounds, partial-thickness burns, ulcers (such as<br>diabetic or pressure), flaps and grafts. |
|                                                                                                                                                                                                | The V.A.C.® GranuFoam® Silver™ Dressing is an effective<br>barrier to bacterial penetration and may help reduce infection in<br>the above wound types.                                                                                                                                                                                                                                                                                                                                                           |
| Differences in<br>intended use<br>from the<br>predicate                                                                                                                                        | The intended use of the devices has not changed from the<br>intended use of the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                               |
| Summary of the<br>technological<br>characteristics of<br>the device                                                                                                                            | The modified and unmodified V.A.C. ATS® and V.A.C. Freedom®<br>canisters have molded-in graduation markings on the side of the<br>canister to aid in estimation of fluid removal. Hydrophobic<br>membranes secured to the side of the canisters during the<br>manufacturing process isolate the collected fluid from the<br>vacuum pump in each case. A gel packet is included in the<br>V.A.C. Freedom® canister and in one model of the V.A.C. ATS®<br>canisters to solidify the wound fluid in the canister.  |
| Summary of<br>nonclinical tests                                                                                                                                                                | Verification of the integrity of the fluid path sterile barriers of the<br>modified design was through accelerated aging, distribution<br>simulation, and microbial aerosol challenge testing.                                                                                                                                                                                                                                                                                                                   |
| Conclusions drawn<br>from the<br>nonclinical tests<br>that demonstrate<br>that the device is<br>as safe, as<br>effective, and<br>performs as well as<br>or better than the<br>predicate device | The modified V.A.C. ATS® and V.A.C. Freedom® canisters with<br>the sterile fluid path are safe and effective for their intended use.<br>The sterility assurance level of the sterile fluid path of the<br>canisters can be assured to a level of 10-6.                                                                                                                                                                                                                                                           |

{1}------------------------------------------------

K063590 page 2/2

Image /page/1/Picture/1 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left, followed by the letters "KCI" in bold, sans-serif font. There is a small circle with an R inside of it to the right of the I.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings forming three curved lines. The eagle is positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the circumference of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 7 2009

KCI USA, Inc. % Ms. Christy Oviatt 6203 Farinon Drive San Antonio, Texas 78230

Re: K063590

Trade/Device Name: V.A.C.® Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: January 4, 2007 Received: January 5, 2007

Dear Ms. Oviatt:

This letter corrects our substantially equivalent letter of January 25, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not

{3}------------------------------------------------

Page 2 - Ms. Christy Oviatt

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number (if known): 1063590 Device Name: V.A.C. ® Therapy System Indications for Use:

The V.A.C. Therapy System is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. It is indicated for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

The V.A.C.® GranuFoam® Silver™ Dressing is an effective barrier to bacterial penetration and may help reduce infection in the above wound types.

Prescription Use X AND/OR Over-The-Counter Use _______

(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, I

(Posted November 16, 2005) and Neurological Devices

| 510(k) Number: | 12063590 |
|----------------|----------|
|----------------|----------|

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K063590](https://fda-staging.innolitics.com/device/K063590)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).