ROCHE TOTAL MYCOPHENOLIC ACID ASSAY, CALIBRATORS AND CONTROLS
K063520 · Roche Diagnostics · OAV · Feb 16, 2007 · Clinical Toxicology
Device Facts
Record ID
K063520
Device Name
ROCHE TOTAL MYCOPHENOLIC ACID ASSAY, CALIBRATORS AND CONTROLS
Applicant
Roche Diagnostics
Product Code
OAV · Clinical Toxicology
Decision Date
Feb 16, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3840
Device Class
Class 2
Indications for Use
The COBAS INTEGRA Mycophenolic Acid cassette (MPA) contains an in vitro diagnostic reagent system intended for use of COBAS INTEGRA systems for the quantitative determination of total mycophenolic acid in human serum or plasma as an aid in the management of mycophenolic acid therapy in renal and cardiac transplant patients. The Roche Total MPA Calibrators are designed for the calibration of the Roche Total MPA assay for the quantitative determination of mycophenolic acid in human serum and plasma on Roche automated clinical chemistry analyzers. The diluent is negative human serum and may be used for dilution of high samples or as a blank sample. The Roche Total MPA controls are for use in quality controls by monitoring accuracy and precision for the quantitative methods as specified in the enclosed value sheet.
Device Story
Device is an in vitro diagnostic reagent system for quantitative determination of total mycophenolic acid (MPA) in human serum or plasma. Operates on COBAS INTEGRA clinical chemistry analyzers. Principle: enzyme-mimicking assay using mutant IMPDH II; MPA concentration inversely proportional to NADH formation rate measured at 340 nm. Used in clinical laboratories by technicians to monitor MPA therapy in transplant patients. Output is a quantitative concentration value (µg/mL) used by physicians to manage immunosuppressant dosing and minimize rejection risk. Benefits include therapeutic drug monitoring (TDM) to guide personalized therapy.
Clinical Evidence
Method comparison study across four sites (US, Germany, France) compared Roche Total MPA assay to HPLC. Tested 265 renal/cardiac transplant patient samples (Site 1), 147 renal samples (Site 2), and 159 renal/cardiac samples (Site 3). Passing/Bablok regression showed high correlation (r=0.992–0.997) and slopes ranging from 1.01 to 1.10. Precision studies (n=21 days) showed total CVs ranging from 0.7% to 4.5% across various concentrations. Linearity validated from 0.4 to 15.0 µg/mL.
Technological Characteristics
Dual-reagent system. Reagent 1: mutant IMPDH II, IMP stabilizers, buffers. Reagent 2: NAD, stabilizers, buffers. Sensing principle: spectrophotometric measurement of NADH at 340 nm. Form factor: liquid reagents for automated clinical chemistry analyzers. Measuring range: 0.4–15.0 µg/mL (extendable to 50 µg/mL via dilution).
Indications for Use
Indicated for renal and cardiac transplant patients undergoing mycophenolic acid therapy. For prescription use only.
Regulatory Classification
Identification
A sirolimus test system is a device intended to quantitatively determine sirolimus concentrations in whole blood. Measurements are used as an aid in management of transplant patients receiving therapy with sirolimus.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Sirolimus Test Systems.” See § 862.1(d) for the availability of this guidance document.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k063520
B. Purpose for Submission:
New Device
C. Measurand:
Mycophenolic Acid (MPA)
D. Type of Test:
Quantitative
E. Applicant:
Roche Diagnostics Corp
F. Proprietary and Established Names:
Total Mycophenolic Acid
Total MPA Calibrators
Total MPA Controls
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OAV | II | 21 CFR 862.3840 | Clinical Toxicology |
| DLJ | II | 21 CFR 862.3200 | Clinical Toxicology |
| LAS | I | 21 CFR 862.3280 | Clinical Toxicology |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The COBAS INTEGRA Mycophenolic Acid cassette (MPA) contains an in vitro diagnostic reagent system intended for use of COBAS INTEGRA systems for the quantitative determination of total mycophenolic acid in human serum or plasma as an aid in the management of mycophenolic acid therapy in renal and cardiac transplant patients.
The Roche Total MPA Calibrators are designed for the calibration of the Roche Total MPA assay for the quantitative determination of mycophenolic acid in human serum and plasma on Roche automated clinical chemistry analyzers. The diluent is negative human serum and may be used for dilution of high samples or as a blank sample.
The Roche Total MPA controls are for use in quality controls by monitoring accuracy and precision for the quantitative methods as specified in the enclosed value sheet.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
For the use on the COBAS INTEGRA systems
I. Device Description:
The Total MPA assay is a dual reagent assay. Reagent 1 is a mutant inosine monophosphate dehydrogenase II enzyme (IMPDH II) and IMP stabilizers and buffers. Reagent 2 is comprised of NAD, stabilizers and buffers.
The Total MPA calibrators are six levels of ready to use calibrators. The kit contains 6 vials of levels A to F (5 mL each) and one vial of diluent (10 mL). The diluent is negative human serum and may be used for dilution of high samples or as a blank sample. Levels A to F correspond to 0, 1, 3, 5, 10 and 15 µg/mL respectively. The product has been prepared exclusively from the blood of donors tested individually and shown by FDA-approved methods to be free from HBsAg and antibodies to HCV and HIV.
{2}
The Total MPA control set is a tri-level ready to use serum liquid based control set. The kit contains 2 vials of levels I to III (5 mL each). Levels I to III corresponds to approximately 0.86, 3.40 and 11.96 µg/mL, respectively. The product has been prepared exclusively from the blood of donors tested individually and shown by FDA-approved methods to be free from HBsAg and antibodies to HCV and HIV.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Microgenics CEDIA Sirolimus Assay
Microgenics CEDIA Sirolimus Calibrators Kit
Microgenics CEDIA Sirolimus Controls Kit
2. Predicate 510(k) number(s):
k034069
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Target Population | Renal and Cardiac Transplant Patients | Renal Transplant Patients |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Quantitative determination for total mycophenolic acid in human serum or plasma as an aid in the management of mycophenolic acid therapy in renal and cardiac transplant patients | Quantitative determination for sirolimus in human whole blood using automated clinical chemistry analyzers as an aid in the management of sirolimus therapy in renal transplant patients taking sirolimus. |
| Methodology | Enzyme-mimicking assay with MPA concentration inversely proportional to the formation of NADH. | Enzyme immunoassay with Sirolimus concentration directly proportional to a generated color change. |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reagent Format | Liquid | Lyophilized |
| Matrix | Nonhemolyzed human serum or EDTA Plasma (K_{2} or K_{3} EDTA) | Whole blood treated with EDTA |
| Range | 0.4 – 15 ug/mL, extendable with post-dilution to 50 ug/mL. | 5.0 – 30 ng/mL |
## K. Standard/Guidance Document Referenced (if applicable):
None referenced by sponsor.
## L. Test Principle:
The Roche Total MPA assay is a two-reagent system that utilizes a fixed amount of mutant IMPDH II (in reagent I). IMPDH II combines with a fixed amount of IMP and NAD to form NADH which is measured at 340 nm. The formation of NADH is inhibited when MPA is present in the serum or plasma samples. MPA concentration is inversely proportional to the rate of NADH formation. The reagents used to measure MPA concentration in serum or plasma are designed to mimic the *in vivo* mechanism of the enzyme.
### In vivo mechanism
IMPDH II + NAD $\xrightarrow{\text{IMP}}$ NADH + XMP $\xrightarrow{\text{If MPA is present}}$ NADH is released from the enzyme but XMP is not
### Roche Total MPA assay mimicking in vivo
IMPDH II + NAD $\xrightarrow{\text{IMP}}$ NADH If MPA is present, the formation of NADH is inhibited.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-run precision was assayed using 21 replicates of three MPA controls and two human plasma pools. MPA-free patient plasma pools were prepared and spiked to low (sub-therapeutic) and high (therapeutic) levels with MPA. The results are summarized in the table below.
{4}
| Material | TDM I | TDM II | TDM III | HSP 1 | HSP 2 |
| --- | --- | --- | --- | --- | --- |
| Mean (μg/mL) | 0.84 | 3.43 | 12.12 | 1.46 | 9.08 |
| SD (ug/mL) | 0.01 | 0.02 | 0.11 | 0.01 | 0.05 |
| CV % | 1.4 | 0.5 | 0.9 | 0.8 | 0.5 |
| N | 21 | 21 | 21 | 21 | 21 |
Within-run, between-day and total imprecision were assessed for seven samples; 3 levels of controls and 4 human plasma pools. The samples were analyzed in triplicate, once a day for 21 days $(n = 63)$ on the COBAS Integra 700. Between-day results were obtained from the second replicate from the triplicate run. Two of the human plasma pools (HP1 and HP2) were MPA-free and spiked to low and high levels. The remaining two human plasma pools were prepared from samples taken from patients on MPA therapy (clinical low and high). Results are summarized below.
| Specimen | MPA I | MPA II | MPA III | HP 1 | HP 2 | Clinical Low | Clinical High |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Total Mean μg/mL | 0.93 | 3.54 | 12.37 | 1.57 | 9.35 | 1.61 | 6.36 |
| Within Run SD μg/mL | 0.011 | 0.015 | 0.056 | 0.013 | 0.041 | 0.014 | 0.044 |
| Within Run CV% | 1.1% | 0.4% | 0.5% | 0.9% | 0.4% | 0.8% | 0.7% |
| Total SD μg/mL | 0.042 | 0.065 | 0.163 | 0.038 | 0.141 | 0.035 | 0.113 |
| Total CV% | 4.5% | 1.8% | 1.3% | 2.4% | 1.5% | 2.2% | 1.8% |
| Between-day μg/mL SD | 0.041 | 0.063 | 0.153 | 0.035 | 0.135 | 0.032 | 0.104 |
| Between-day CV% | 4.4% | 1.8% | 1.2% | 2.2% | 1.4% | 2.0% | 1.6% |
The sponsor also conducted a CLSI precision study on the COBAS 400. Within-run, between-day and total imprecision was conducted on 3 levels of Roche total MPA controls and two human plasma pools. Each sample was assayed in triplicate daily for 10 days. Between day results were obtained from the second replicate from the triplicate run. The results are summarized in the table below.
| Specimen | MPA I | MPA II | MPA III | HP I | HP II |
| --- | --- | --- | --- | --- | --- |
| Total Mean μg/mL | 0.91 | 3.48 | 12.22 | 1.55 | 9.11 |
| Within Run SD μg/mL | 0.019 | 0.037 | 0.034 | 0.027 | 0.074 |
| Within Run CV% | 2.0% | 1.1% | 0.3% | 1.7% | 0.8% |
| Total SD μg/mL | 0.041 | 0.061 | 0.090 | 0.047 | 0.081 |
| Total CV% | 4.5% | 1.8% | 0.7% | 3.0% | 0.9% |
| Between-day SD μg/mL | 0.037 | 0.049 | 0.083 | 0.039 | 0.034 |
| Between-day CV% | 4.1% | 1.4% | 0.7% | 2.5% | 0.4% |
{5}
The sponsors' two clinical sites also conducted within-run, between-day and total imprecision studies on 3 levels of Roche Total MPA controls, one level of MPA calibrator and 3 levels of human plasma pools. The material was assayed in triplicate measurement in 1 run per day on each of 21 days. Between-day results were calculated using the second replicate from each 3-fold determination. Plasma pools were prepared using samples from subjects who were not on MPA therapy. These pools were spiked to sub-therapeutic or low-end reportable range. The reports for the 2 sites are summarized in the tables below.
Site 1- Integra 400
| Specimen | MPA Control I | MPA Control II | MPA Control III | Calibrator BI | HP 1 | HP 2 | HP 3 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Total Mean μg/mL | 0.873 | 3.443 | 12.011 | 1.039 | 0.781 | 2.471 | 9.924 |
| Within Run SD μg/mL | 0.023 | 0.045 | 0.161 | 0.028 | 0.021 | 0.027 | 0.229 |
| Within Run CV% | 2.6 | 1.3 | 1.3 | 2.7 | 2.7 | 1.1 | 2.3 |
| Total SD μg/mL | 0.024 | 0.051 | 0.209 | 0.033 | 0.026 | 0.041 | 0.229 |
| Total CV% | 2.8 | 1.5 | 1.7 | 3.1 | 3.4 | 1.6 | 2.3 |
| Between-day SD μg/mL | 0.009 | 0.024 | 0.133 | 0.016 | 0.015 | 0.030 | 0.0 |
| Between-day CV% | 1.1 | 0.7 | 1.1 | 1.5 | 1.9 | 1.2 | 0.0 |
Site 2- Integra 800
| Specimen | MPA Control I | MPA Control II | MPA Control III | Calibrator BI | HP 1 | HP 2 | HP 3 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Total Mean μg/mL | 0.895 | 3.519 | 12.298 | 1.065 | 0.840 | 2.585 | 10.339 |
| Within Run SD μg/mL | 0.018 | 0.026 | 0.094 | 0.019 | 0.023 | 0.022 | 0.093 |
| Within Run CV% | 2.0 | 0.7 | 0.8 | 1.8 | 2.7 | 0.9 | 0.9 |
| Total SD μg/mL | 0.027 | 0.038 | 0.106 | 0.026 | 0.028 | 0.041 | 0.131 |
| Total CV% | 3.0 | 1.1 | 0.9 | 2.4 | 3.3 | 1.6 | 1.3 |
| Between-day SD μg/mL | 0.020 | 0.028 | 0.049 | 0.017 | 0.016 | 0.034 | 0.093 |
| Between-day CV% | 2.3 | 0.8 | 0.4 | 1.6 | 1.9 | 1.3 | 0.9 |
b. Linearity/assay reportable range:
The sponsor conducted six linearity recovery studies to validate the claimed measuring range of 0.4 to 15.0 μg/mL for the Total MPA assay.
Full-range linearity (up to 15 μg/mL) was assessed for the following:
- non clinical spiked plasma pools (donors not on MPA therapy),
- clinical plasma pools from renal transplant patients taking MPA and spiked with additional MPA
{6}
- clinical plasma pools from cardiac transplant patients taking MPA and spiked with additional MPA.
The samples were spiked with MPA to >20% higher than the upper limit of the assay range (15 ug/mL) and combined with a 0 or low clinical plasma pool in varying portions to create systematic dilutions from low levels to >15 ug/mL.
Low-end linearity (0 to approximately 5 ug/mL) was assessed for the following:
- non clinical plasma pools (donors not on MPA therapy),
- clinical plasma pools from renal transplant patients taking MPA and spiked with additional MPA
- clinical plasma pools from cardiac transplant patients taking MPA and spiked with additional MPA.
The results of the linearity recovery studies in the form of recovery ranges (%) for the claimed assay range is shown in the table below.
| | Measured Value (ug/mL) | Theoretical Value (ug/mL) | Recovery % |
| --- | --- | --- | --- |
| Study 1- Non-clinical plasma pools spiked with MPA (full-range) | 15.73 | 17.00 | 92.53 |
| | 14.27 | 14.88 | 95.53 |
| | 12.70 | 12.75 | 99.61 |
| | 10.89 | 10.63 | 102.49 |
| | 8.90 | 8.5 | 104.71 |
| | 6.53 | 6.38 | 102.43 |
| | 4.25 | 4.25 | 100.00 |
| | 1.93 | 2.13 | 90.82 |
| Study 2-clinical plasma pools from renal transplant patients taking MPA (full-range) | 14.71 | 15.01 | 97.98 |
| | 12.67 | 12.67 | 100 |
| | 10.84 | 10.33 | 104.98 |
| | 8.19 | 7.98 | 102.61 |
| | 5.66 | 5.64 | 100.4 |
| | 2.99 | 3.29 | 90.77 |
| | 0.95 | 0.95 | 100 |
| Study 3- Clinical plasma pools from cardiac transplant patients taking MPA (full-range) | 14.81 | 15.07 | 98.27 |
| | 13.02 | 12.73 | 102.28 |
| | 10.8 | 10.4 | 103.95 |
| | 8.46 | 8.05 | 105.09 |
| | 5.7 | 5.71 | 99.82 |
| | 3.13 | 3.37 | 92.88 |
| | 1.03 | 1.03 | 100 |
{7}
| | Measured Value (ug/mL) | Theoretical Value (ug/mL) | Recovery % |
| --- | --- | --- | --- |
| Study 4- Non-clinical plasma pools spiked with MPA (low-end) | 5.12 | 4.91 | 104.1 |
| | 4.54 | 4.43 | 102.54 |
| | 3.94 | 3.94 | 100 |
| | 3.44 | 3.44 | 100 |
| | 2.9 | 2.95 | 98.25 |
| | 2.38 | 2.46 | 96.8 |
| | 1.88 | 1.97 | 95.6 |
| | 1.45 | 1.48 | 98.25 |
| | 1.02 | 0.98 | 103.7 |
| | 0.56 | 0.492 | 113.83* |
| Study 5- Clinical plasma pools from renal transplant patients taking MPA (low-end) | 5.51 | 5.59 | 98.6 |
| | 4.57 | 4.69 | 97.35 |
| | 3.45 | 3.69 | 93.54 |
| | 2.31 | 2.57 | 89.86 |
| | 1.53 | 1.68 | 91.26 |
| | 0.96 | 1.01 | 95.44 |
| | 0.88 | 0.89 | 98.43 |
| | 0.68 | 0.671 | 101.4 |
| | 0.58 | 0.56 | 103.79 |
| | 0.5 | 0.45 | 111.84* |
| | 0.4 | 0.335 | 119.3* |
| Study 6- Clinical plasma pools from cardiac transplant patients taking MPA (low-end) | 5.63 | 6 | 93.83 |
| | 5.48 | 5.88 | 93.2 |
| | 4.47 | 4.92 | 90.85 |
| | 3.34 | 3.72 | 89.78** |
| | 2.38 | 2.64 | 90.15 |
| | 1.52 | 1.68 | 90.48 |
| | 0.96 | 0.96 | 100 |
| | 0.87 | 0.84 | 103.57 |
| | 0.75 | 0.72 | 104.17 |
| | 0.67 | 0.6 | 111.67* |
| | 0.56 | 0.48 | 116.67* |
| | 0.48 | 0.36 | 133.33* |
| | 0.38 | 0.24 | 158.33* |
*Recovery exceeds 10% but is ≤0.3 ug/mL
** Recovery exceeds 10% and 0.3 ug/mL (by 0.08 ug/mL)
The sponsor conducted an extended measuring range study for the MPA assay. A manual dilution of samples with a post dilution factor of 5 is recommended in the product labeling. A 2000 ug/mL stock solution of MPA in methanol was spiked into normal human serum to produce serum samples with MPA concentrations of 15, 20, 25, 30, 33, 40, 50 and 55 ug/mL. Each sample was diluted 1:5 and 1:1 with the diluent from the
{8}
Total MPA calibrator kit (level A). Samples were measured in triplicate on both methods and the median results were compared. The 1:5 diluted samples had a recovery range of 97 to 105% and the 1:1 diluted samples has a recovery range of 92 to 108%. The results demonstrate acceptable recovery for dilution of samples with concentrations between 15 and 50 ug/mL with the diluent provided.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Total MPA assay calibrators and controls are prepared gravimetrically to target levels and are verified against a validated HPLC method. The MPA used in the reagent, calibrators, and controls are traceable to an internally prepared mycophenolic acid. No USP material is available for MPA.
The Roche Total MPA Reagent was evaluated for closed shelf-life (accelerated and real-time) and open on-board stability for the MPO reagent. The results support a 15 month shelf life claim at 4 to 8° C for the MPO reagent. Real time stability studies are ongoing. The open on-board stability claim is 84 days.
Real time stability studies are ongoing for the Total MPA Controls (3 levels- 0.80, 3.5 and 12 µg/mL) Calibrators (2 levels- 5.0 and 15.0 µg/mL) and have been tested to date for 22 months. The accelerated study results support an 18 month closed stability claim. Open bottle stability support the open bottle package insert claims for calibrators and controls stability (6 months when stored at 2- 8 °C).
d. Detection limit:
The sponsor conducted a lower detection limit study with zero calibrator. The zero calibrator was assayed 21 times and the mean and SD of the 21 measured absorbance values was calculated. The mean mAbs value- 2SD of the replicates was calculated as 0.3 µg/mL.
The sponsor conducted a functional sensitivity study to determine the concentration at which acceptable assay precision is observed. The testing was performed using 8 samples from patients on MPA therapy with concentrations between 0.3 and 1.06 ug/mL. Samples were assayed in triplicate in one run per day for 10 days (n=30). Samples at 0.4 ug/mL showed 20% CV. The sponsor designated 0.4 g/mL as the low end of the assay's measuring range.
e. Analytical specificity:
9
{9}
The sponsor conducted specificity studies for the major MPA metabolites, mycophenolic acid glucuronide (MPAG), and the acyl glucuronide of MPA (AcMPAG). The metabolites were added to normal human plasma containing two concentrations of MPA (1.7 and 8.3 ug/mL). Samples and controls were run in triplicate. Samples containing a metabolite were run on the HPLC simultaneously in order to measure and subtract out the amount of MPA hydrolyzed from the metabolites. The percent cross reactivity equation is:
[Control Recovery- (sample recovery-HPLC)]/ Metabolite concentration *100 or
[Column F- (Column D- Column E)]/ Column C *100
The concentrations and results are shown in the table below.
| Column A | Column B | Column C | Column D | Column E | Column F | Column G |
| --- | --- | --- | --- | --- | --- | --- |
| Compound | MPA Conc in Sample (ug/mL) | Cross Reactant Conc (ug/mL) | Observed MPA test Conc (ug/mL) in sample | HPLC MPA Conc (ug/mL) | MPA Control Conc (ug/mL) | % Cross Reactivity |
| MPAG | 1.7 | 1000 | 2.63 | 1.03 | 1.56 | 0.004 % |
| MPAG | 8 | 1000 | 8.96 | 1.03 | 7.76 | 0.017 % |
| AcMPAG | 1.7 | 10 | 2.37 | 0.27 | 1.54 | 5.60 % |
| AcMPAG | 8 | 10 | 8.39 | 0.27 | 7.64 | 4.80 % |
Endogenous substances and co-administered drugs were tested in an interference study. Eight endogenous substances were tested to the highest concentration expected and one hundred and three co-administered drugs were tested to a 3-fold level higher than the highest concentration reported following therapeutic dosage. Samples were measured in triplicate with two levels of MPO (high and low) and the median was used to determine the deviation from the control sample. The following charts are a summary of the reported results.
| | Interferent conc Range (mg/dL) | MPO analyte result range (ug/mL) | MPO recovery range (%) |
| --- | --- | --- | --- |
| Hemoglobin (μg/mL) | | | |
| [MPA] High- 8.53 | 1 -1108 | 8.60 – 8.24 | 100.2- 96.6 |
| [MPA] Low- 1.322 | 0- 1226 | 1.35 – 1.29 | 102.3 – 98.5 |
| Labeling states “No Significant interference up to 1000 mg/dL”. | | | |
| Conjugated Bilirubin (mg/dL) | | | |
| [MPA] High- 9.49 | 0 – 78 | 9.49 – 10.08 | 106.5- 102.0 |
| [MPA] Low- 1.57 | 0 – 79 | 1.57 – 1.68 | 100.0 – 107.0 |
{10}
| Unconjugated Bilirubin | | | |
| --- | --- | --- | --- |
| [MPA] High- 9.15 | 0 – 28 | 9.15 – 10.02 | 102.3- 109.5 |
| [MPA] Low- 1.49 | 0 – 64 | 1.49 – 1.63 | 101.3 – 109.4 |
| Labeling states “No significant interference up to an I index of 66 for conjugated and 17 mg/dL for unconjugated bilirubin”. | | | |
| Lipemia (with intralipid) | | | |
| [MPA] High- 9.18 | 43 – 154 | 9.18 – 10.08 | 101.2 – 109.8 |
| [MPA] Low 1.45 | 58 – 166 | 1.45- 1.58 | 100.0 – 109.0 |
| Labeling states “No interference up to 93 mg/dL for intralipid” and the limitations states to “Avoid the use of lipemic specimens”. | | | |
| Clinical Triglycerides | | | |
| [MPA] High- 12.19 | 336- 926 | 12.19- 13.34 | 99.3- 109.4 |
| [MPA] Low- 4.70 | 74- 885 | 4.70 – 5.01 | 97.9- 106.6 |
| Labeling states “No significant interference up to a triglycerides level of 500 mg/dL with a recovery specification of +/- 10 % or 600 mg/dL wit ha recovery specification of +/- 15%. No significant interference up to a intralipid level of 93 mg/dL. There is poor correlation between the L index (turbidity) and triglycerides concentration. Avoid the use of lipemic specimens”. | | | |
| Total Proteins | | | |
| [MPA] High- 9.45 | 2 – 11 | 9.14- 10.37 | 96.7 – 109.7 |
| [MPA] Low- 2.23 | 4-11 | 2.12 – 2.37 | 95.1 – 106.3 |
| Labeling states “No significant interference from total protein concentrations of 4 -11 g/dL”. | | | |
| Albumin | | | |
| [MPA] High- 11.0 | 4.51-5.43 | 10.28- 11.23 | 91.5 (only one level) |
| [MPA] Low- 2.2 | 4.51-6.32 | 2.1-2.3 | 91.3-96.5 |
| The sponsor states no significant interference up to 5.4 g/dL albumin. | | | |
| Gamma Globulin | | | |
| [MPA] High- 11.0 | 0 -9.92 | 9.24-10.25 | 90.1- 99.8 |
| [MPA] Low- 2.2 | 0-6.20 | 1.96-2.16 | 90.7-98.6 |
| The sponsor states no significant interference up to 6.2 g/dL Gamma Globulin. | | | |
| Cholesterol | | | |
| [MPA] High- 8.2 | 3- 503 | 8.2- 8.42 | 100.2 – 102.7 |
| [MPA] Low- 1.79 | 3- 525.9 | 1.79- 1.83 | 100.0 – 102.8 |
| Labeling states “No significant interference up to 500 mg/dL cholesterol”. | | | |
| Creatinine | | | |
| [MPA] High- 10.59 | 0.7 – 10.8 | 10.28 – 10.59 | 98.2 – 99.7 |
| [MPA] Low – 1.93 | 0.7- 10.9 | 1.93 – 1.96 | 100.0 – 102.6 |
| Labeling states “No significant interference up to 10 mg/dL”. | | | |
| Uric Acid | | | |
| [MPA] High- 9.96 | 1.84- 21.18 | 9.92- 10.03 | 99.4- 100.7 |
| [MPA] Low – 1.70 | 6.04 – 25.61 | 1.68 – 1.71 | 98.8 – 100.6 |
| Labeling states “No significant interference up to 20 mg/dL uric acid”. | | | |
{11}
Additionally, the sponsor evaluated the effects of 103 potential drug interferences on the MPA assay on two levels of MPA (approximately 1.5 and 9.0 µg/mL). The samples were measured in triplicate and the median was used to determine the deviation from the control sample. The possible drug interferents were tested at 3X the normal plasma concentrations and there were no significant interferences were caused by the drugs (except for intralipid interference) listed in the package insert.
The sponsor conducted additional studies to establish time, temperature and freeze/thaw cycles for the MPA samples used with the Total MPA Assay. Samples were aliquotted and stored at 15-25°C, 2-8°C and -20°C. Frozen samples were thawed on the day tested. Samples were measured in triplicate in several time intervals over the course of at least four weeks. The sponsor states that results for plasma sample are stable for 3 months at -20°C and 2-8°C, and for 4 weeks at 25°C. The sponsor’s serum results support stability for 4 weeks at -20°C, 2-8°C and 25°C and 5 freeze thaw cycles for serum samples.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor conducted a three part, four site clinical study to compare the performance characteristics of the Roche Total Mycophenolic Acid assay on the Integra platform to HPLC MPA measurements. Inclusion/exclusion criteria are described below.
Inclusion criteria:
Patient:
Renal or cardiac transplant recipient who is taking MPA as an immunosuppressant and who consents to participate in the study, where consent required by law.
Persons for whom less than 2 samples have already been used in this trial.
Sample:
Serum collected without a gel separation tube or EDTA K₃ or K₂ plasma in sufficient volume to complete the required method comparison analysis.
Exclusion Criteria
Patient:
All other transplant types besides renal and cardiac
Transplant recipients who are not taking MPA
Any non-transplant patient who is taking MPA
Persons not consenting to the trial, where consent is required by law.
12
{12}
Persons for whom 2 samples have already been used in this trial
Site 1: US
Site 2: Germany
Site 3: France
Site 4: Roche Diagnostics
The first part of the study tested EDTA plasma specimens from renal and cardiac (148 and 117 samples from 209 patients respectively) transplant recipients (n=265) at site 1 whose immunosuppressent drug regime includes mycophenolate mofetil or mycophenolate sodium (MPA). Samples were assayed fresh at the immunochemistry labs then stored frozen at -20 °C. The residual plasma samples were tested in parallel on the HPLC and Integra 400 Plus.
The second part of the study tested banked residual renal EDTA plasma samples (147 samples from 86 patients) from a Roche Pharmacoutical MPA study that was performed at approx. 70 sites in Europe and included renal samples from patients in France (that were stored in Germany). All samples were from patients whose immunosuppressant drug regimen includes mycophenolate mofetil or mycophenolate sodium (MPA). The parallel testing was performed on HPLC and Integra 800 at site 2.
The third part of the study tested 159 residual EDTA plasma specimens from renal (89) and cardiac (70) transplant recipients from Site 1 whose immunosuppressant drug regimen included mycophenolate mofetil or mycophenolate sodium (MPA). The samples obtained from Site 1 were tested on the Integra 700 and were different from the samples that were tested at site 1. All patients had received either renal or cardiac allografts and were on bid (twice daily) MPA therapy.
The sponsor calculated regression slopes, intercepts, and correlation coefficients according to Passing/Bablok methods. The method comparison results for all three sites are shown in the chart below.
| (Site 1) Integra 400 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Methodology vs HPLC | Slope (95% CI) μg/mL | Intercept (95% CI) μg/mL | Correlation Coefficient | Sample Size | Sample Range μg/mL | Neat Sample Range μg/mL | Spiked Sample Range μg/mL |
| Renal | 1.010 (0.988-1.035) | 0.068 (0.034-0.105) | 0.995 | 148 | 0.40-14.80 | 0.40 - 13.70 | 7.9-14.80 |
{13}
| Cardiac | 1.014
(0.996-1.031) | 0.057 (-0.004-0.103) | 0.992 | 114 | 0.40-13.60 | 0.40-11.20 | 6.40-13.60 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Combined | 1.011
(1.000-1.025) | 0.064
(0.038-0.090) | 0.993 | 265 | 0.40-14.80 | | |
| Site 2-Integra 800 | | | | | | | |
| Renal | 1.100
(1.073-1.120) | -0.1200
(-.1920-0.0656) | 0.994 | 147 | 0.50-14.70 | 0.500 - 14.70 | None |
| Site 3-Integra 700 | | | | | | | |
| Renal | 1.062
(1.044 - 1.079) | 0.019
(-0.021-0.073) | 0.997 | 89 | 0.46-13.57 | 0.460-9.077 | 5.34-13.57 |
| Cardiac | 1.088
(1.050-1.125) | -0.023
(-0.110-0.045) | 0.993 | 70 | 0.57-14.22 | 0.573-9.914 | 5.96-14.22 |
| Combined | 1.068
(1.052-1.085) | 0.005
(-0.003-0.050) | 0.995 | 159 | 0.46-14.22 | | |
The sponsor also conducted imprecision testing at the external sites, as discussed in the precision section above.
# b. Matrix comparison:
A serum/plasma comparison study was conducted for all sample types that the sponsor recommends for use with the MPA total assay. The serum/EDTA plasma paired samples from 50 donors were spiked and tested in triplicate with the MPA assay on the Integra 700. The samples were compared in the following combinations:
Serum vs. EDTA,
$\mathrm{K}_3$ EDTA vs. $\mathrm{K}_2$ EDTA,
$\mathrm{K}_3$ EDTA vs. $\mathrm{K}_2$ EDTA (half filled tubes) and
$\mathrm{K}_3$ EDTA vs. $\mathrm{K}_3$ EDTA (half filled tubes)
Each sample was measured in triplicate. The median was calculated and serum and plasma results were compared by a method comparison. The slope, intercept, and correlation coefficient (Passing/Bablok) were determined. The results for each comparison are shown in the chart below.
{14}
| Comparison | Method | N | Range | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | X: K3EDTA | 50 | 1.230-15.310 | 0.988 | -0.003 | 0.9977 |
| | Y: Serum | | | | | |
| 2 | X: K3EDTA | 50 | 1.230-15.310 | 0.992 | -0.054 | 0.9984 |
| | Y: K2EDTA | | | | | |
| 3 | X: K3EDTA | 11 | 1.750-13.550 | 0.951 | 0.195 | 0.9977 |
| | Y: K2EDTA (half) | | | | | |
| 4 | X: K3EDTA | 8 | 1.230-14.300 | 1.016 | 0.018 | 0.9992 |
| | Y: K3EDTA (half) | | | | | |
$\mathrm{K}_2$ EDTA, $\mathrm{K}_3$ EDTA, and serum samples all met the sponsors predetermined acceptance criteria.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable - This information is not typically provided for immunosuppressant TDM assays.
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
The sponsor cites multiple studies from the literature concerning the expected values (therapeutic range) in the package insert and articles to support their claim expected values. The package insert states:
{15}
"The therapeutic range of mycophenolic acid is not fully established and is dependent on transplant type and coadministered drugs. Optimal mycophenolic acid assay values to prevent rejection may vary based on the test system and therefore should be established for each test system".
"Decreased incidences of rejection in the early months post-transplantation have been reported in renal transplant patients with predose MPA concentrations (measured by HPLC) of ≥ 1.3 μg/mL with coadministration of cyclosporine and ≥1.9 μg/mL with coadministration of tacrolimus.¹,² An upper therapeutic range based on development of toxicity has not been established. The clinical ramifications of MPA concentrations beyond the early post transplantation period are not yet known.¹
Cardiac transplant patients, predose MPA concentrations (measured by HPLC) of 1.2 – 3.5 μg/mL have been recommended to minimize incidences of rejection.¹,³ Higher predose concentrations (≥2.5 μg/mL) in the early post-transplantation period (<6 months) have also been suggested.³,⁴,⁵ Pediatric cardiac transplant patients have been shown to require higher doses of MPA in comparison to adults due to differences in MPA metabolism.¹,³,⁴,⁶
1) 3. Van Gelder T, Meur YL, Shaw LM, Oellerich M, DeNofrio D, Holt C, Holt DW, Kaplan B, Kuypers D, Meiser B, Toenshoff B, Mamelock RD. Therapeutic drug monitoring of mycophenolate mofetil in transplantation. Ther Drug Monit 2006; 28(2): 145-154.
2) Cox VC, Ensom MHH. Mycophenolate mofetil for solid organ transplantation: Does the evidence support the need for clinical pharmacokinetic monitoring? Ther Drug Monit(3) 2003; 25(2): 137-157.
3) Meiser BM. Therapeutic drug monitoring of mycophenolic acid in cardiac transplant recipients: does it make sense? Curr Opin Organ Transplant 2005; 10: 350-354.
4) Gajarski RJ, Crowley DC, Zamberlan MC, Lake KD. Lack of correlation between MMF dose and MPA level in pediatric and young adult cardiac transplant patients: Does the MPA level matter? Am J Transplant 2004; 4:1495-1500.
5) Meiser BM, Reichart B. New agents and new strategies in immunosuppression after heart transplantation. Curr Opin Organ Transplant 2002; 7: 226-232.
6) Dipchand AI, Pietra B, McCrindle BW, Rosebrook-Bicknell HL, Boucek MM. Mycophenolic acid levels in pediatric heart transplant recipients receiving mycophenolate mofetil. J Heart Lung Transplant. 2001; 20(10): 1035-1043.
The sponsor notifies users to investigate the transferability of the expected values to its own patient population and if necessary determine its own reference range. For diagnostic purposes the test findings should always be assessed in conjunction with the patient's medical history, clinical examinations, and other findings.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{16}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.