← Product Code [OBZ](/productcode/OBZ) · K063498

# HYDRAGEL 18 A1AT ISOFOCUSING KIT (PN 4356) (K063498)

_Sebia · OBZ · Jun 11, 2007 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063498

## Device Facts

- **Applicant:** Sebia
- **Product Code:** [OBZ](/productcode/OBZ.md)
- **Decision Date:** Jun 11, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5130
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The HYDRAGEL 18 A1AT ISOFOCUSING kit is designed for the qualitative detection and identification of the different phenotypes of Alpha-1 antitrypsin (A1AT). Phenotyping results in conjunction with clinical findings and other laboratory assays aid in the diagnosis of Alpha-1 antitrypsin deficiency. The analysis is performed on human sera separated into electrophoretic patterns ready for qualitative analysis. The procedure includes isoelectrofocusing on agarose gel, performed on the semi-automatic HYDRASYS system, followed by immunofixation with anti-Alpha-1 antitrypsin antiserum. The use of enzyme labeled anti-Alpha-1 antitrypsin antiserum enhanced the detection and identification of the different phenotypes. For in vitro diagnostic use only. The A1AT Controls are designed for the migration control of the human Alpha-1 antitrypsin isoforms pattern obtained with HYDRAGEL 18 A1AT ISOFOCUSING* isoelectric-focusing procedure. The A1AT Controls should be used as human sera. For in vitro diagnostic use only.

## Device Story

The HYDRAGEL 18 A1AT ISOFOCUSING kit is an in vitro diagnostic test for identifying A1AT phenotypes. Input consists of human serum samples. The device utilizes isoelectric focusing on agarose gel, performed on the semi-automatic HYDRASYS system, followed by immunofixation using enzyme-labeled anti-A1AT antiserum. The process separates A1AT isoforms based on their isoelectric points. The resulting electrophoretic patterns are evaluated visually by laboratory personnel. Results are used in conjunction with clinical findings and other laboratory assays to assist in diagnosing A1AT deficiency. The system provides a standardized method for phenotype characterization compared to traditional polyacrylamide gel methods.

## Clinical Evidence

Clinical study evaluated 68 serum samples (16 congenital deficiency, 15 pulmonary, 8 hepatic, 1 infertility, 1 panniculitis, 23 normal, 4 indeterminate). Compared to predicate method, demonstrated 100% agreement. Clinical sensitivity 100% (39/39); specificity 92% (23/25); overall agreement 97%.

## Technological Characteristics

Agarose gel-based isoelectric focusing; semi-automated electrophoresis (HYDRASYS system); immunofixation with enzyme-labeled anti-A1AT antiserum; visual evaluation of electrophoretic patterns; in vitro diagnostic use.

## Regulatory Identification

An alpha-1-antitrypsin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the alpha-1-antitrypsin (a plasma protein) in serum, other body fluids, and tissues. The measurements aid in the diagnosis of several conditions including juvenile and adult cirrhosis of the liver. In addition, alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema.

## Predicate Devices

- Polyacrylamide gel isoelectric focusing technique (not specifically named, cited as predicate technique used in clinical diagnostic laboratories)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k063498

B. Purpose for Submission:
New Device

C. Measurand:
Alpha-1-Antitrypsin Phenotype

D. Type of Test:
Qualitative, isofocusing

E. Applicant:
SEBIA, Inc.

F. Proprietary and Established Names:
HYDRAGEL 18 A1AT ISOFOCUSING Kit

G. Regulatory Information:

1. Regulation section:
21 CFR § 866.5130 Alpha-1-antitrypsin immunological test system
21 CFR § 862.1660, Quality Control Material (Assayed and Unassayed)

2. Classification:
Class II, Device
Class I, Quality Control Material

3. Product codes:
OBZ, Alpha-1-antitrypsin, qualitative, phenotype
JJX, Single (Specified) Analyte Controls (Assayed and Unassayed)

4. Panel:
Immunology 82
Chemistry 75

H. Intended Use:

1. Intended use(s):
The HYDRAGEL 18 A1AT ISOFOCUSING kit is designed for the qualitative detection and identification of the different phenotypes of Alpha-1 antitrypsin (A1AT).
Phenotyping results in conjunction with clinical findings and other laboratory assays aid in the diagnosis of Alpha-1 antitrypsin deficiency. The analysis is performed on human sera separated into electrophoretic patterns ready for qualitative analysis. The procedure includes isoelectrofocusing on agarose gel, performed on the semi-automatic HYDRASYS system, followed by immunofixation with anti-Alpha-1 antitrypsin antiserum. The use of enzyme labeled anti-Alpha-1 antitrypsin antiserum enhanced the detection and identification of the different phenotypes. For in vitro diagnostic use only.

The A1AT Controls are designed for the migration control of the human Alpha-1 antitrypsin isoforms pattern obtained with HYDRAGEL 18 A1AT ISOFOCUSING* isoelectric-focusing procedure. The A1AT Controls should be used as human sera. For in vitro diagnostic use only.

{1}

2. Indication(s) for use:
Same as Intended Use.

3. Special conditions for use statement(s):
The device is for prescription use only.

4. Special instrument requirements:
HYDRASYS System PN 1211, with option HYDRASYS FOCUSING PN 1235, or PN 1212

I. Device Description:
The HYDRAGEL 18 A1AT ISOFOCUSING kit includes: 10 agarose gels (ready-to-use); 3 mL vial ethylene glycol solution (ready-to-use); 50 mL vial anodic solution (ready-to-use); 50 mL vial cathodic solution (ready-to-use); 10 packs of 2 strips each; 85 mL vial sample diluent CSF/A1AT (ready-to-use); 0.4 mL vial anti-A1AT-PER antiserum (stock solution); 3 mL vial antiserum diluent (ready-to-use); two 70 mL vials Rehydrating solution (ready-to-use); two 20 mL vials TTF3 solvent (ready-to-use); two 0.5 mL vials TTF3 (stock solution); 1 pack of 10 (18 teeth) applicators (ready-to-use); 1 pack of 2 buffer containers (ready-to-use); 1 pack of 10 single trough antiserum segments (ready-to-use); 1 pack of 10 thin filter papers; 3 packs of 10 each thick filter papers.

The A1AT Controls pack contains three lyophilized vials of three phenotypes: one Normal A1AT Control vial (MM) and two Pathological A1AT Control vials (MZ and SZ). The A1AT Control pack is sold separately.

J. Substantial Equivalence Information:
1. Predicate device name(s):
Polyacrylamide gel isoelectric focusing
2. Predicate 510(k) number(s):
Pre-amendment method
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Device  |
|  Intended use/ Indication for use | For the detection and identification of A1AT phenotyping in human serum by isoelectric focusing as an aid in A1AT deficiency | Same  |
|  Sample type | Serum | Same  |
|  Results | Qualitative interpretation of patterns of A1AT phenotypic stained fractions after electrophoretic migration | Same  |
|  Technology | Gel electrophoretic migration | Same  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Methodology | Isoelectric focusing and Immunofixation | Isoelectric focusing  |
|  Gel | Agarose | Polyacrylamide  |
|  Equipment | Semi-automated HYDRASYS electrophoresis apparatus | Standard isoelectric focusing apparatus  |
|  A1AT Antisera | Antibody specificity to A1AT | No antisera  |
|  A1AT Antisera enhancement | Peroxidase enzyme | None  |
|  Migration | 20°C for about 1 hour (total 510 Vh) | Premigration 30 minutes 1200 V for 3 hours  |
|  Antiserum application/incubation | 10 minutes at 20°C | None  |
|  Gel treatment | A. At 20°C:
Blotting #.1: 3 min
Gel rehydration #1: 5 min.
Blotting #2: 3 min
Gel rehydration #2: 5 min.
B. At 30°C:
TTF3 incubation: 10 min.
Blotting #3: 3 min.
C. At 50°C:
Drying: 3 min.
D. Washing & final drying: 18 min. | Fixation with 12.5% trichloracetic acid; washing with ethanol & acetic acid solution; Coomassie Blue staining; Destaining  |
|  Sample volume | 10 μL | 2.5 μL  |
|  Sample dilution | 1:10 with Sebia diluent | None  |
|  Lowest detectible Limit | 5 mg/dL | None specified  |

K. Standard/Guidance Document Referenced (if applicable):
None provided.

L. Test Principle:
A1AT is a polymorphic glycoprotein found in normal human serum. It belongs to a group of proteins which inhibit a variety of human proteases including trypsin, cathepsin G, neutrophil elastase, etc. A1AT is mainly produced in hepatic cells and by immune system. The gene that encodes A1AT is located at a single locus on chromosome 14 and has three alleles designated by letters with M for normal, Z and S for deficiency, and null. Individuals with ZZ, SZ, MZ or null phenotypes show increased risks of disease which may vary in patients exhibiting the same phenotype. A1AT deficiency is a genetic disorder first described in patients with pulmonary emphysema.

{3}

The assay is carried out in two stages: isoelectrofocusing on agarose gel to fractionate the proteins in the serum samples and immunofixation with enzyme (peroxidase)-labelled anti-AlAT antiserum to detect the different phenotypes of AlAT. The semi-automated HYDRASYS system performs all the steps needed to obtain gels ready for interpretation.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

The within-run gel reproducibility was determined by testing eight samples fifteen or eighteen times each on a single gel of 15 or 18 tracks. Two gels of the same lot number were used in this study. Two normal serum samples with M1 and M1M2 phenotypes and six pathological serum samples with MS, SS, MZ, ZZ, MX phenotypes reproduced the corresponding phenotype correctly.

The between-run gel reproducibility was determined by testing fifteen samples and three controls twelve times on two lots of gels. Seven normal serum samples (2 M1M2, 4 MM, 1M1M3); eight pathological samples (2 MS, 1 SS, 1 MZ, 1 ZZ, 1 SZ, 1 MX, 1 MI [I allele is rare and associated with decreased A1AT levels]); and three controls (1 normal MM and 2 pathological MZ and SZ) reproduced the corresponding phenotype correctly.

The within-run and between-run antiserum reproducibility studies were determined by testing fifteen samples and three controls on four different lots of antiserum. The 18 samples were applied in 18 tracks of 4 gels. Nine normal samples with MM phenotype; six pathological samples with 2 SS, 2 M2Z, 1 MP, 1 M undetermined variant; and three controls with 1 normal MM and 2 pathological MZ and SZ reproduced the corresponding phenotype correctly.

The within-run and between-run control reproducibility studies were determined by testing each of the three controls (MM, MZ, SZ phenotypes) six times in four gels for two days. All three controls reproduced the corresponding phenotype correctly.

#### b. Linearity/assay reportable range:

Not applicable.

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

No reference standards and method available.

Stability: The expiration date claims for the products are as follows:

|  Product | Temperature | Expiration Date Claim  |
| --- | --- | --- |
|  A1AT Antiserum | 2-8 °C | 2 years  |
|  Control: Lyophilized | 2-8 °C | 3 years  |
|  Control: Reconstituted | 2-8 °C | 1 week  |
|  Control: Reconstituted | -20 °C | 6 months  |

{4}

Expected values: Qualitative Interpretation of normal pattern for A1AT 'MM' phenotype fraction.

d. Detection limit:
The detection limit was determined by testing diluted A1AT (concentration of 27 mg/dL with pathological ZZ phenotype) down to 5 mg/dL. The ZZ phenotype was detectable at 5 mg/dL.

e. Analytical specificity:
Interference study: Two studies were performed. For each interfering substance, four concentrations were tested. Study 1 consisted of aliquots from one normal serum pool sample (normal MM phenotype) and three controls (1 normal MM and 2 pathological MZ and SZ). One set of these aliquots was spiked with hemoglobin up to 0.35 g/dL and another set with bilirubin up to 0.20 mg/mL. Study 2 consisted of aliquots from six normal MM serum and three controls (1 normal MM and 2 pathological MZ and SZ). One set of these aliquots were spiked with total cholesterol up to 6.9 mmol/L and another with triglycerides up to 8.0 mmol/L. No interference was observed.

f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:
i. Study design: A total of 68 serum samples were tested on HYDRAGEL 18 A1AT ISOFOCUSING kit using HYDRASIS apparatus and on polyacrylamide gels using a standard isoelectric focusing apparatus. Forty-one (41) were pathological serum samples, 4 unclear diagnoses, and 23 were normal serum samples. There was 100% agreement between the two methods. Results of the phenotype interpretation are summarized below:

|  Phenotype Qualitative Results | N | Agreement  |
| --- | --- | --- |
|  Normal Samples  |   |   |
|  MM | 22 | 22  |
|  M1M2 | 1 | 1  |
|  Pathologic Samples  |   |   |
|  MZ | 14 | 14  |
|  MS | 9 | 9  |
|  M2S | 1 | 1  |
|  ZZ | 9 | 9  |
|  SS | 4 | 4  |
|  SZ | 4 | 4  |
|  Others (unclear diagnosis)  |   |   |
|  IM or FM | 1 | 1  |
|  IM | 2 | 2  |
|  Heterogeneous (Indeterminate) | 1 | 1  |
|  Total | 68 | 68  |

{5}

b. Matrix comparison:

Not applicable.

# 3. Clinical studies:

a. Clinical Sensitivity and specificity:

The clinical sensitivity and specificity studies were evaluated on 68 clinically characterized sera from patients with the following diagnosis: 16 congenital deficiency, 15 pulmonary disorders, 8 hepatic disorders, 1 infertility and 1 panniculitis, 23 normals and 4 indeterminate diagnoses. Sensitivity was  $100\%$  and specificity was  $92\%$ . The overall agreement was  $97\%$ .

|   | Diagnosis  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal* | Negative | Total  |
|  HYDRAGEL 18 A1AT ISOFOCUSING kit | Positive | 39 |  | 2** | 41  |
|   |  Equivocal* |  | (4) |  |   |
|   |  Negative |  |  | 23 | 23  |
|   |  Total | 39 | (4) | 25 | 64  |

*Equivocals were excluded from the calculation. The four equivocal samples were from patients with normal A1AT quantitative measurements being ruled out for infertility, pulmonary and hepatic disorder and congenital deficiency. The A1AT phenotypes were IM or FM, IM, IM, and heterogeneous pattern respectively.
**The two false positive samples were from asymptomatic patients, one with a diagnosis of infertility and the other, panniculitis and the A1AT phenotypes were MZ and MS respectively. These two patients (without underlying pulmonary or hepatic disorders) are assumed to be carriers of S and Z deficiency alleles.

Sensitivity:  $100\%$  (39/39)

Specificity:  $92\%$  (23/25)

Overall percent Agreement:  $97\%$  (62/64)

b. Other clinical supportive data (when a. is not applicable):

Not applicable.

# 4. Clinical cut-off:

Same as Expected values.

# 5. Expected values/Reference range:

Expected values in the normal population should be the presence of MM phenotype fraction. The sponsor has also provided literature on the A1AT concentrations for the different A1AT phenotypes (see table below).

|  A1AT phenotype | A1AT concentration (mg/dL)  |
| --- | --- |
|  MM | 103 - 200  |
|  MS | 100 - 180  |
|  SS | 70 - 105  |
|  MZ | 66 - 120  |
|  SZ | 45 – 80  |
|  ZZ | 10 - 40  |
|  Null/Null | 0  |

Vidal et al. (Guidelines for the diagnosis and management of  $\alpha$ -1-antitrypsin deficiency Arch Bronconeumol 2006

{6}

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K063498](https://fda-staging.innolitics.com/device/K063498)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
