← Product Code [LON](/productcode/LON) · K063492

# VITEK 2 GRAM POSITIVE ERYTHROMYCIN FOR STREPTOCOCCUS PNEUMONIAE (K063492)

_bioMerieux, Inc. · LON · Dec 21, 2006 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K063492

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Dec 21, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** AI/ML

## Indications for Use

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae is a qualitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents, The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilii, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

## Device Story

VITEK®2 AST card inoculated with standardized organism suspension; incubated in VITEK®2 instrument; optically monitored via transmittance optics. System measures light transmittance through wells to track organism growth; algorithm calculates MIC results once predetermined growth thresholds are reached. Used in clinical laboratories by technicians; results reported as MIC and category interpretation (SIR). Facilitates clinical decision-making regarding antimicrobial therapy for S. pneumoniae infections.

## Clinical Evidence

Clinical study at three sites compared VITEK®2 Erythromycin card to broth microdilution (MH broth with lysed horse blood). Combined clinical and challenge data (n=437) showed 98.9% category agreement. Performance metrics: 2 minor, 2 major, and 1 very major discrepancy. Reproducibility demonstrated across three sites using manual and automated inoculation methods.

## Technological Characteristics

Miniaturized broth microdilution card; automated incubation and optical growth monitoring. System platforms: VITEK 2 (automated) and VITEK 2 Compact (manual fill/seal). Connectivity: Integrated system. Software-based analysis of growth kinetics.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK®2 Gram Positive Telithromycin for Streptococcus pneumoniae (k053186)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k063492

B. Purpose of Submission:
Addition of Erythromycin to the VITEK®2 and VITEK®2 Compact Systems
Antimicrobial Susceptibility Test (AST) System

C. Measurand
Erythromycin ≤ 0.25 - ≥ 1 μg/mL

D. Type of Test:
Qualitative growth based detection algorithm using predetermined growth thresholds

E. Applicant:
bioMerieux, Inc.

F. Proprietary and Established Names:
VITEK®2 Gram Positive Erythromycin

G. Regulatory Information:
1. Regulation section:
866.1645 Short-Term Antimicrobial Susceptibility Test System
2. Classification:
II
3. Product Code:
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
4. Panel:
83 Microbiology

H. Intended Use:
1. Intended use(s):
The VITEK®2 Antimicrobial Susceptibility Test is intended to be used
with the VITEK®2 System for the automated quantitative or qualitative
susceptibility testing of isolated colonies for most clinically significant aerobic
gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus
agalactiae, and S. pneumoniae.

The VITEK®2 Gram Positive Susceptibility Card is intended for use with the
VITEK®2 system in clinical laboratories as an in vitro test to determine the
susceptibility of Streptococcus pneumoniae to antimicrobial agents when used as
instructed in the Online Product Information.

{1}

Page 2 of 6

2. **Indication(s) for use:**
This submission is for the addition of the VITEK® 2 Gram Positive Erythromycin for *Streptococcus pneumoniae* at a concentration of 0.06, 0.125, and 0.25 µg/mL and a calling range of range of ≤0.25, and ≥ 1 µg/mL.

3. **Special condition for use statement(s):**
Prescription use only

VITEK® 2 does not report intermediate results for *Streptococcus pneumoniae*/Erythromycin. At this time, intermediate strains for *Streptococcus pneumoniae*/Erythromycin are very rare.

4. **Special instrument Requirements:**
Not applicable

I. **Device Description:**
The VITEK® 2 AST card containing the test is inoculated with a standardized organism suspension. The card is incubated within the instrument and optically monitored throughout the incubation cycle. Results are automatically calculated once a predetermined growth threshold is reached and a report is generated that contains the final result.

J. **Substantial Equivalence Information:**

1. **Predicate device name(s):**
VITEK® 2 Gram Positive Telithromycin for *Streptococcus pneumoniae*

2. **Predicate K number(s):**
k053186

3. **Comparison with predicate**
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Determine antimicrobial susceptibility to antimicrobial agents | Same  |
|  Instrument | VITEK®2 System | Same  |
|  Test Card | VITEK®2 card, including the base broth | Same  |
|  Test organism | Colonies of Gram-Positive cocci | Same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Antibiotic | Erythromycin at specific concentrations | Telithromycin at specific concentrations  |
|  Reading algorithm | Unique for erythromycin | Unique for telithromycin  |

K. **Standard/Guidance Document Referenced (if applicable):**

{2}

Page 3 of 6

Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; CLSI M7 (M100-S16) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”.

## L. Test Principle:

Optics systems use visible light to directly measure organism growth. These transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. An interpretive call is made between 4 and 16 hours for a “rapid” read but may be extended to 18 hours in some instances. The VITEK®2 Susceptibility Card test is based on the microdilution minimum inhibitory concentration technique with concentrations equivalent to standard method concentrations. Several parameters based on the growth characteristics observed are used to provide appropriate input for the MIC calculations. Discriminate analysis is used to develop the algorithm that determines the susceptibility result for all antimicrobials on the VITEK®2 system. The MIC result must be linked to organism identification in order to determine a category interpretation. A category interpretation (SIR) will be reported along with a MIC.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility:

Ten strains of *Streptococcus pneumoniae* were tested at three sites. This testing was performed using both the manual dilution of the inoculum and also the automatic dilution method. Acceptable reproducibility was demonstrated with only category agreement (S, R) since that is all that is detected.

b. Linearity/assay reportable range:

Not applicable

c. Traceability (controls, calibrators, or method):

The recommended QC isolate was tested on every test occasion with the reference method and the VITEK®2. The reference method QC results were in range for every day tested. The VITEK®2 was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended range.

Quality Control was performed during the studies using both the auto-dilution and the manual method of diluting the organisms. Results demonstrated that methods were comparable with the same mode.

{3}

Page 4 of 6

|  ORGANISM | Test Results | VITEK®2 AUTO-DIL | VITEK®2 MAN-DIL | Reference Conc. (ug/mL) | Reference  |
| --- | --- | --- | --- | --- | --- |
|  |   |   |   |   |   |
|  S. pneumoniae | ≤0.25 | 79 | 79 | ≤0.03125 | 41  |
|  ATCC 49619 |  |  |  | 0.0625 | 19  |
|  Expected Range: |  |  |  | 0.125 | 19  |
|  0.03125 – 0.125 μg/mL |  |  |  |  |   |
|  |   |   |   |   |   |

Inoculum density control was monitored using the DensiChek instrument. This was standardized weekly with all results recorded and in the expected range. Verification was performed during internal testing.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

A clinical study was conducted at three sites using the VITEK®2 gram positive cards with erythromycin and the broth microdilution method using Mueller Hinton (MH) broth with lysed horse blood prepared as recommended by CLSI. Inoculum was prepared with direct colony suspension. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. Two methods of inoculation (manual and automated) were evaluated. Clinical testing was performed using the automated method of inoculation and the challenge set was tested using both the manual and the automated method. The test device had a growth rate of &gt;90%. All isolates grew in the VITEK®2 cards in less than 16 hours. Essential agreement was not calculated because the VITEK®2 card contained &lt;5 dilutions of erythromycin. A comparison was provided to the reference method with the following agreement.

Summary Table for S. pneumoniae

|   | CA Tot | CA N | CA % | #R | Min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 379 | 375 | 98.9 | 99 | 1 | 2 | 1  |
|  Challenge | 58 | 57 | 98.3 | 16 | 1 | 0 | 0  |
|  Combined | 437 | 432 | 98.9 | 115 | 2 | 2 | 1  |

{4}

Page 5 of 6

CA-Category Agreement
R-resistant isolates
min-min discrepancies
maj-major discrepancies
vmj-very major discrepancies

Category Agreement is when the interpretation (SIR) of the reference method agrees exactly with the interpretation (SIR) of the VITEK®2 results.

Manual Dilution:
The challenge set of organisms was also tested at one site using the manual method of inoculation with the following performance that demonstrated that there no difference between the two inoculation methods.

Summary Table for S. pneumoniae
|   | CA Tot | CA N | CA % | #R | Min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Challenge | 58 | 57 | 98.3 | 16 | 1 | 0 | 0  |

b. Matrix comparison: Not applicable

3. Clinical studies:
a. Clinical sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a and b are not applicable): Not Applicable

4. Clinical cut-off: Not applicable

5. Expected values: S. pneumoniae ≤0.25 (S), 1 (I), ≥2 (R)

N. Labeling
The expected value range, interpretive criteria and QC for gram negative panels are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

{5}

Page 6 of 6

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K063492](https://fda-staging.innolitics.com/device/K063492)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
