K063469 · Eastern Medikit , Ltd. · FOZ · Dec 1, 2006 · General Hospital
Device Facts
Record ID
K063469
Device Name
MEDICAN ALPHA
Applicant
Eastern Medikit , Ltd.
Product Code
FOZ · General Hospital
Decision Date
Dec 1, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Intravascular catheter is a medical device intended to be used to introduce fluids or medicament through the peripheral veins and / or withdrawal of blood samples. Intravascular catheter comprises of a short flexible plastic tube (the Catheter) which is inserted into a vein over a hollow introducer needle, after which the needle is withdrawn and discarded. The intravascular catheter is used in aseptic environment. Federal law (USA) restricts this device to sale by or on the order of a physician.
Device Story
Over-the-needle peripheral intravascular catheter; consists of flexible plastic tube and hollow introducer needle. Used for accessing peripheral vascular system to introduce fluids/medications or withdraw blood samples. Physician or healthcare provider inserts catheter into vein; needle is withdrawn and discarded. Device used in aseptic clinical environments. Luer cap included as additional feature. Provides sterile access to cardiovascular system for short-term use (up to 30 days).
Clinical Evidence
Bench testing only. Performance verified per ISO 10555-1 and ISO 10555-5 (design qualification, visual/physical analysis). Physicochemical testing (non-volatile residue, heavy metals, residue on ignition, buffering capacity, bacterial endotoxin) and sterility testing performed per USP-29. Biocompatibility verified per ISO 10993-1:1997 for external communicating, circulating blood contact, prolonged exposure (>24 hrs to 30 days).
Technological Characteristics
Over-the-needle peripheral catheter; flexible plastic tube. Complies with ISO 10555-1 and ISO 10555-5. Biocompatibility per ISO 10993-1. Sterilization method not specified but device supplied sterile. Single-use. Includes luer cap.
Indications for Use
Indicated for infusion of I.V. solutions and intermittent intravenous drug administration in patients requiring peripheral vascular access.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
B. Braun Medical Inc. - 1-Introcan W [K 982805]
Becton Dickinson - BD Insyte ™ [K 013073]
Submission Summary (Full Text)
{0}------------------------------------------------
E063469
Page No.: 059
1 of 2
510 (K) Summary of Safety and Effectiveness for 'Medican Alpha'
:
| DATE: | October 27, 2006 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | EASTERN MEDIKIT LIMITED, |
| | 196, Udyog Vihar, Phase-1, |
| | Gurgaon-122 016 (Haryana) |
| CONTACT PERSON: | RAJEEV ASRI |
| PHONE NUMBER: | + 91-124-4001831 (extn. 259) |
| FAX NUMBER: | + 91-124-4001835 |
| E-MAIL: | rajeevasri@medikit.com |
| TRADE NAME: | Medican Alpha |
| COMMON OR USUAL NAME: | CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT TERM (LESS THAN 30 DAYS). |
| | [Various Sizes of IV Catheters / Cannulae] |
| CLASSIFICATION NAME: | Catheter, intravascular, therapeutic, short term less than 30 days |
| DEVICE CLASSIFICATION: | Class II, 21 CFR 880,5200: Intravascular Catheter |
| PREDICATE DEVICE(S): | B. Braun Medical Inc. - 1-Introcan W [K 982805] |
| | Becton Dickinson - BD Insyte ™ [K 013073] |
| DESCRIPTION: | As per ISO 10555-1, the generic definition of an intravascular catheter is - Intravascular catheter is a tubular device, single or multi lumen, designed to be partially or totally inserted into the cardiovascular system for diagnostic and/or therapeutic purpose |
| | The device under this submission is an over-needle-peripheral catheter. |
| | Over - the - needle peripheral intravascular catheters are intended for accessing the peripheral vascular system, supplied in the sterile condition and intended for single use. These catheters are designed for the introduction or withdrawal of liquids into or from the peripheral vascular system. |
| INTENDED USE: | Intravascular catheter is a medical device intended to be used to introduce fluids or medicament through the peripheral veins and / or withdrawal of blood samples. |
| | Intravascular catheter comprises of a short flexible plastic tube (the Catheter) which is inserted into a vein over a hollow introducer needle, after which the needle is withdrawn and discarded. |
| | The intravascular catheter is used in aseptic environment. |
| | Federal law (USA) restricts this device to sale by or on the order of a physician. |
| INDICATIONS FOR USE: | - Infusion of I.V. Solutions. |
| | - Intermittent intravenous Drug administration. |
DEC - 1 2006に、その他、その他、その他、その他、その他、その他、その他、一般的な人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の
ないと、その他の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の
{1}------------------------------------------------
Page No.: 060
K0634168
2 of 2
SUMMARY OF NON-CLINICAL PERFORMANCES AND CONCLUSIONS:
In-vitro performance testing of the Medican Alpha included Design Qualification, visual analysis and Physical Analysis as per ISO 10555-1 & ISO 10555-5 [Sterile, Single-use intravascular catheter - (1) General requirements and (5) Over Needle Peripheral Catheter]. The Medican Alpha has also been tested for Physicochemical testing like Non volatile residue, Heavy metals, Residue on ignition, Buffering capacity, Bacterial Endotoxin tests and Sterility test as per USP-29.
The results have confirmed that Medican Alpha meets the requirement for all general and specific requirements of ISO 10555-1 & ISO 10555-5; and also passes the Physiochemical, Bacterial Endotoxin and Sterility tests as per USP-29
The Biocompatibility of the Medican Alpha has been verified in accordance with ISO 10993-1:1997 (Biological Evaluation of Medical Devices). Test results confirmed Biocompatibility of the catheter when tested as an External Communicating, Circulating Blood contact, and prolonged Exposure (>24 hrs to 30 days) device.
In comparative study of Medican Alpha IV Catheter with Becton Dickinson catheter premarket notification, BD Insyte, it is found that The Medican Alpha IV Catheter is similar in respect of material, design and intended use with Becton Dickinson catheter. "BD Insyte ™". The only two major differences are ") presence of luer cap in Medikit Brand catheter and 2) that of packaging (blister) material. A luer cap is an additional feature in Medican Alpha and does not alter any functional aspect of the device. The packaging material does not affect the functional use of product as both provide a sterile package till the shelf life period. Therefore, Medican Alpha may be considered as substantially equivalent to "BD Insyte TM".
> In comparative study of Medican Alpha IV Catheter with B Braun catheter premarket notification, 1-Introcan W, it is observed that The Medican Alpha IV Catheter is similar in respect of material, design and intended use with B Braun catheter, 1-Introcan W. The two major differences are 1) the safety feature of the predicate device that is not present in the Medikit brand Medical Alpha and 2) presence of luer cap in Medikit Brand catheter; this is an additional feature of Medican Alpha. Therefore, Medican Alpha may be considered as substantially equivalent to B Braun catheter, 1-Introcan W, except for the Safety features provided in B Braun 1-Introcan W.
SUBSTANTIAL EQUIVALENCE:
4
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/12 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC = 1 2006
Eastern Medikit Limited C/O Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
Re: K063469
Trade/Device Name: Medican Alpha Regulation Number: 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: November 15, 2006 Received: November 16, 2006
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fice number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Kunnes
A Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page No.: 061
K063469
## Indications for Use
510(k) Number (if known): Not yet assigned
Device Name: Medican Alpha
Indications for Use:
- Infusion of I.V. Solutions.
- Intermittent intravenous Drug administration.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
and the country of the states
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
lee R.ADW 12/1/20
Con of Anesthesiolo
K0634169
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.