K063462 · Neat Stitch , Ltd. · GAM · Jan 12, 2007 · General, Plastic Surgery
Device Facts
Record ID
K063462
Device Name
NEATSTITCH NEATCLOSE SUTURING DEVICE
Applicant
Neat Stitch , Ltd.
Product Code
GAM · General, Plastic Surgery
Decision Date
Jan 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeatStitch NeatClose suturing device is intended for the approximation and/or ligation of soft tissues in laparoscopic procedures.
Device Story
NeatClose Suturing Device is a manual surgical instrument for laparoscopic soft tissue approximation/ligation. System consists of a reusable handle, single-use disposable cartridge containing needle guides and needles, and a pre-loaded suture. During use, the device is inserted through a port or access site; needle guides are deployed to contact the inner abdominal wall; needles are deployed to penetrate tissue and are retrieved into the cartridge; needle guides retract automatically. The device is then withdrawn, leaving the suture for the surgeon to tie manually. Used in clinical settings by surgeons. Benefits include efficient, reliable, and user-friendly port access closure compared to traditional hand suturing.
Clinical Evidence
Controlled animal study using two pigs (16 total access sites) compared the NeatClose device to hand suturing and three other marketed suturing devices. Evaluated feasibility, safety, efficacy, ease of use, and duration. Results concluded the device is effective, safe, reliable, and user-friendly for port access closure. Bench testing included USP-standardized validation for sterility (USP 71), tensile strength (USP 881), diameter (USP 861), and needle attachment (USP 871).
Technological Characteristics
Manual surgical instrument; reusable handle with single-use disposable cartridge. Pre-loaded with absorbable poly(glycolide/l-lactide) surgical suture. Operates via mechanical needle deployment and retrieval mechanism. Sterilization validated per international standards.
Indications for Use
Indicated for the approximation and/or ligation of soft tissues in laparoscopic procedures.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
ATRAMAT and SuperSorb Polyglycolic Acid Surgical Sutures (K040282)
Submission Summary (Full Text)
{0}------------------------------------------------
K063462 Pay
NeatClose Suturing Device 11 Nov 2006
# 510(k) summary of Safety and Effectiveness NeatStitch Laparoscopic Closure device
### 1. Sponsor name
NeatStitch Ltd. Hahadas 1 St. POB 132 Or Akiva Israel 30600 Tel: + 972 4 6262876 Fax: + 972 4 6262877
JAN 1 2 2007
#### 2. Device name
Proprietary name: NeatClose Suturing Device Common/Usual name: Manual Surgical Instrument for general use. Classification name: Needle driver Class 1 (exempt) 878.4800 (Product code HCF- Instrument, Ligature
Passing and Knot tying), Manual surgical instrument for general use Class II (510(k)) 878.4493 (Product code GAM- Suture, Absorbable, Synthetic polyclycolic acid) Absorbable poly(glycolide/l-lactide) surgical suture.
# 3. Identification of Predicate or Legally marketed Device
FASTCLOSE Suturing Device (K011105) LSI Solutions Suture Placement Device (K981531) ATRAMAT and SuperSorb Polyglycolic Acid Surgical Sutures (K040282).
### 4. Device description
The NeatClose Suturing Device has three major component: 1) a reusable Handle, 2) a single-use disposable cartridge (containing a natented needle guides and needles), and 3) a suture. The cartridge (including the pre-loded suture) is assembled onto the Handle. The device is then inserted through the access site through the port working channel (or without it), the needle guides are deployed and the device is withdrawn till the needle guides contact the inner abdominal wall. Then the needles are deployed, penetrating the tissue and retrieved in the cartridge. Right after, automatically, the needle guides are retracted to the cartridge as well. Then the device is completely withdraws, leaving the suture to be tied by the surgeon as he/she routinely with
{1}------------------------------------------------
K063462
NeatClose Suturing Device 11 Nov 2006
#### 5. Intended use
The NeatClose Suturing Device is intended for the approximation and/or ligation of soft tissues in laparoscopic procedures.
## 6. Comparison of Technological Characteristics
The NeatClose Suturing Device is substantially equivalent in intended use and/or function to the following predicate devices: FASTCLOSE Suturing Device (K011105), LSI Solutions Suture Placement Device (K981531), and the ATRAMAT and SuperSorb Polyglycolic Acid Surgical Sutures (K040282).
The NeatClose Suturing Device is pre-loaded with a commercial cleared suture.
The intended use of each of the predicate devices is soft tissue approximation and/or ligation of soft tissue. The NeatClose Suturing Device is used for the same intended use of soft tissue approximation and/or ligation as the predicate devices.
The operating principle of the NeatClose Suturing Device is the same as that of the predicate devices: a manual instrument is used to pass needles through tissue for suturing.
#### 7. Performance Testing
The NeatClose Suturing Device performs the same function as a standard needle driver or guide. Since the NeatClose Suturing Device is pre-loaded with a non-sterile thread the suture was tested and showed that the suture characteristics as determined by the suture manufacturer (including degradation) and according to USP, are preserved after sterilization and packaging.
In addition, prototype NeatClose Suturing Device was tested in a controlled animal study with two pigs (another pig was used later as an acute experiment with the use of only Neatstitch device in 6 access ports). The scope of this study was to evaluation the feasibility, casety, efficacy of the NeatClose Suturing Device with compared to hand suturing, and additional three currently marketed devices for suturing of the laparoscopic access site (none of them was one of the predicate devices of this notification). Two pigs were used for this comparison, with 8 access sites in each. The measurements were for ease of use, duration, safety and efficacy.
It was concluded therefore that the NeatClose Suturing Device is a very effective, completely safe, reliable, fast and user-friendly device for the closure of port access in laparoscopic procedures. The suture from the sterile and packed NeatClose Sutruing Device was
tested for Sterility (USP general chapter 71), Tensile strength (USP
{2}------------------------------------------------
<063462 Page 3
NeatClose Suturing Device 11 Nov 2006
general chapter 881), Sutures-Diameter (USP general chapter 861) and Suture-Needle Attachment (USP general chapter 871). The sterilization was validated per international standards. Packaging and shelf-life validation are in progress and will be completed before commercial marketing.
# Section 5. 510(k) Summary
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NeatStitch Ltd % Yossi Muncher, Ph.D. Vice President, Clinical and Regulatory Affairs Hahadas 1 ST. Or Akiva, Israel 30600
JAN 1 2 2007
Re: K063462
Trade/Device Name: NeatClose Suturing Device Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM, HCF Dated: November 11, 2006 Received: November 22, 2006
Dear Dr. Muncher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 Yossi Muncher, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkersen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
NeatClose Suturing Device 11 Nov 2006
510(k) number (if known): ____________________________________________________________________________________________________________________________________________________
Device name: NeatStitch NeatClose suturing Device
Indication for use:
The NeatStitch NeatClose suturing device is intended for the approximation and/or ligation of soft tissues in laparoscopic procedures.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------------------------------|---------|
| Division of General, Restorative, and Neurological Devices | |
| 510(k) Number | 1663462 |
| Page 1 of | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.