ULTRASONIC BIOMETER/PACHYMETER FOR OPTHALMOLOGY ODM 2100 AND ODM 2200
Applicant
Meda , Ltd.
Product Code
IYO · Radiology
Decision Date
Jan 29, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The intended use of the ODM-2100 & ODM-2200 includes the location and visualization of ophthalmic disorders and measurement of ocular distances.
Device Story
The ODM-2100/2200 is an ophthalmic diagnostic ultrasound system combining B-mode imaging and A-mode biometry. The device utilizes a 10MHz transducer to emit ultrasound pulses into the eye and orbit. In B-mode, reflected echoes are processed to create real-time cross-sectional images of ocular tissues displayed on a monitor. In A-mode, the system measures time-of-flight for echoes reflected by the anterior chamber, lens, and vitreum to calculate axial lengths. The ODM-2100 features a portable housing with a built-in 10-inch monitor, while the ODM-2200 utilizes an external SVGA LCD monitor; both share identical internal hardware and software. Operated by clinicians, the device provides visual and quantitative data to assist in the localization of ophthalmic disorders and ocular measurements, supporting clinical decision-making regarding eye health and surgical planning.
Clinical Evidence
No clinical data required. Substantial equivalence is supported by non-clinical bench testing, including safety (IEC 60601-1), EMC (IEC 60601-1-2), acoustic output measurements per FDA guidance, and biological safety (ISO 10993).
Technological Characteristics
System includes 10MHz A-scan and B-scan transducers. B-mode provides cross-sectional imaging; A-mode provides axial biometric measurements. Connectivity includes SVGA output (ODM-2200). Complies with IEC 60601-1, IEC 60601-1-2, and ISO 10993. Operates via electric pulse excitation of transducers.
Indications for Use
Indicated for ultrasound imaging of the eye and axial biometric parameter measurement of the eye in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K063433
MEDA. CO., LTD
510(k) Premarket Notifica Model ODM-2100/2200 Ultrasonic A/B Scanner For Ophthalmology
## Attachment 8 510(K) Summary
## 510(k) SUMMARY
JAN 2 9 2007
#### 1) Submitter Infortation
Name: MEDA. Co., Ltd. Address: Room D, F3, Building C2, Xinmao Science Skill Park, Huayuan Industry Development Area, Tianjin, China
Phone: 86-22-83713808 Fax: 86-22-83713880
Contact person: Edward A. Kroll Spectre Solutions, Inc. 5905 Fawn Lane Cleveland, Ohio 44141 Phone: (440) 546-9810 Fax: (440) 546-9124
Data Prepared: July 10, 2006
#### 2) Name of Device
Trade Name: ODM-2100 Ultrasonic A/B Scanner for Ophthalmology ODM-2200 Ultrasonic A/B Scanner for Ophthalmology Common Name: Ultrasonic A and B Scan System for Ophthalmology Classification Name: Ultrasonic Pulsed Echo Imaging System Regulation Number: 892.1560 Product Code: 90-IYO
#### 3) Predicate Devices
Sonomed Medical E-Z Scan™ AB5500*, K040668
#### 4) Intended Use
The intended use of the ODM-2100 & ODM-2200 includes the location and visualization of ophthalmic disorders and measurement of ocular distances.
{1}------------------------------------------------
#### 5) Device Description
The ODM-2100/2200 combines a B-scanner used for the visualization by ultrasound of the eye and orbit and an A-scan used for intraocular measurements. The intended used of this system includes the localization and visualization of ophthalmic disorders and measurement of the eye and orbit.
The ODM-2200 uses the same software and hardware (Switch power supply, P.C. board, user's interface, transducer etc.) as ODM-2100 does. The only difference is that:
ODM-2100 uses a portable housing and built-in 10-inch video monitor; ODM-2200 uses standard SVGA. LCD monitor.
#### 6) Technological characteristics
#### a) Devices Description
The ODM-2100/2200 includes B-mode ultrasonic cross-section imaging and A-mode axial biometric parameter measurement. Users can switch between these two modes through keyboard.
In B-mode, the 10MHz transducer emits ultrasound under electric pulse and then receives and stores its echo reflected by interfaces of different tissues. Through the sector scanning of transducer, several echo lines create real time image of eye tissue and the image is displayed on the monitor.
In A-mode, the 10MHz transducer transmits ultrasound into eye tissue and receives its echo reflected back by anterior chamber, lens and vitreum. Each length and their sum (AL) are calculated by measuring the time spent in different parts.
#### b) Performance Data
- 1) Non-clinical test
Compliance Tests for ODM-2100 are as follows: Safety Test (IEC 60601-1, Tested by TÜV) EMC Test Report (IEC 60601-1-2, Tested by TÜV) Acoustic output Test (FDA Guidance Document "Information for Manufacture Seeking Marketing Clearance of Diagnostic Ultrasound System and Transducers") Biological Safety Test (ISO 10993, Tested by TÜV)
2) Clinical test: Not required
#### 3) Conclusions
The ODM-2100/2200 Ultrasonic A/B Scanner for Ophthalmology is equivalent in safety and efficacy to the legally marketed predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines extending from its lower body, resembling a flag or banner.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Meda. Ltd. % Mr. Edward A. Kroll President Spectre Solutions, Inc. 5905 Fawn Lane CLEVELAND OH 44141
JAN 2 9 2007
Re: K063433
Trade Name: Model ODM 2100 and 2200 Ultrasonic A/B Scanner for Ophthalmology Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYO and ITX Dated: November 8, 2006 Received: December 5, 2006
Dear Mr. Kroll:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Model ODM 2100 and 2200 Ultrasonic A/B Scanner for Ophthalmology, as described in your premarket notification:
Transducer Model Number
A-Scan (10MHz) B-Scan (10MHz)
Image /page/2/Picture/13 description: The image is a circular seal with the letters "FDA" in a stylized font in the center. Below the letters, the word "Centennial" is written in a cursive font. Above the letters, the numbers "12X - XX" are written. The seal is surrounded by a circular border with text that is difficult to read due to the image quality.
ling Public . 1
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{4}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (240) 276-3666.
V
Sincerely yours,
David W. Logeman
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
# Indications for Use
100 K063433 510(k) Number (if known):
Device Name: Model ODM 2100 and 2200 Ultrasonic A/B Scanner for Ophthalmology
Indications for Use:
- Ultrasound Imaging of the eye .
- Axial biometric parameter measurement of the eye .
Prescription Use X (Part 21 CFR 801 Subpart D)
4. 14. 12
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
1 - 1 - 1 - 1 -
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Geluidh. Segner
Division Sign Off
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Pelvic Devices
K063433
{6}------------------------------------------------
Attachment 1
## Ko63433
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
System: ODM-2100/2200 Ultrasonic A/B Scanner for Ophthalmology Transducer:
#### Intended Use: Diagnostic ultrasound imaging
| Clinical Application | Mode Of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other*<br>(Specify) |
| Ophthalmic | | N | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neuonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardicac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Contents:
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Gavin A. Segrem
Diff. V
Reproductive, Abdominal,
Various
Page l
{7}------------------------------------------------
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM Kob3433
ODM-2100/2200 Ultrasonic A/B Scanner for Ophthalmology System: Transducer: __ A-Scan (10MHz)
#### Intended Use: Diagnostic ultrasound imaging
| Clinical Application | Mode Of Operation | | | | | | | | | Combined<br>(specify) | Other*<br>(Specify) |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|--|-----------------------|---------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | | | |
| Ophthalmic | N | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neuonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Contents:
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
David G. Ingram
(Division Sign-Off) Division of Reproductive, Abdomi and Radinlogical Devices 50 g. Sumber _________________________________________________________________________________________________________________________________________________________________
Page 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.