LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067
Device Facts
| Record ID | K063409 |
|---|---|
| Device Name | LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067 |
| Applicant | Agamatrix |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Jan 12, 2007 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Indications for Use
AgaMatrix Liberty™ Blood Glucose Monitoring System: AgaMatrix Liberty™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip, palm, or forearm. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Blood Glucose Meter: AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Blood Glucose Test Strips: AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Control Solutions (Normal level and High level): AgaMatrix Liberty™ Control Solutions are intended for use with the AgaMatrix Liberty™ Meter and AgaMatrix Liberty™ Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.
Device Story
Portable, battery-operated blood glucose meter; uses test strips to measure glucose in fresh capillary whole blood samples (fingertip, palm, or forearm). Operated by patients at home or healthcare professionals in clinical settings. Device provides quantitative glucose concentration readout to assist in diabetes management. System includes meter, test strips, lancing device, and control solutions. No complex AI/ML processing described; standard electrochemical glucose monitoring principle.
Clinical Evidence
Clinical study of 49 subjects (Type II diabetes majority) compared forearm vs. fingertip glucose readings. Linear regression analysis showed slope 0.98 (lay user) and 1.01 (HCP), with 94-96% of results within ±20% of reference. Median glucose 152 mg/dL (range 84-320 mg/dL).
Technological Characteristics
Portable, battery-operated electrochemical glucose meter. Materials and design identical to predicate. Standards: ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, IEC 61000-4-3.
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood from fingertip, palm, or forearm. Intended for home (OTC) use by persons with diabetes or clinical use by healthcare professionals to monitor diabetes control effectiveness.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Liberty™ Blood Glucose Monitoring System (k052762)