K063375 · Philips Medical Systems · MKJ · Jan 11, 2007 · Cardiovascular
Device Facts
Record ID
K063375
Device Name
HEARTSTART MRX, MODELS M3535A OR M3536A
Applicant
Philips Medical Systems
Product Code
MKJ · Cardiovascular
Decision Date
Jan 11, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. AED Therapy To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless. Manual Defibrillation Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation. Non-invasive External Pacing Therapy The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early. Pulse Oximetry The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level. Non-invasive Blood Pressure Monitoring The NBP option is intended for non-invasive measurement of a patient's arterial blood pressure. End-tidal CO2 The EtCO2 option is intended for non-invasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate. 12-Lead ECG The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements. Q-CPR The Q-CPR™ option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time). The Q-CPR option is contraindicated as follows: The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less that 25 kg). The Q-CPR option is not for use when CPR is contraindicated. Invasive Pressure The Invasive Pressure option is indicated for measuring arterial, venous, intracranial and other physiological pressures on patients. Temperature The Temperature option is indicated for measuring temperature in patients.
Device Story
HeartStart MRx Monitor/Defibrillator is a multi-parameter patient monitor and defibrillator; used by qualified medical personnel in clinical settings. Device inputs include ECG signals, SpO2, NBP, EtCO2, invasive pressure, temperature, and chest compression/ventilation data via Q-CPR sensors. Device processes physiological signals to provide real-time monitoring, diagnostic 12-lead ECG reports, and resuscitation feedback. Output is displayed on-screen for clinician review to guide clinical decision-making during cardiac arrest, bradycardia, or routine monitoring. Software release 7.01 adds Bluetooth wireless technology for data transfer as an alternative to RS232 cables. Benefits include improved resuscitation performance via real-time feedback and enhanced data connectivity for patient management.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, integration tests, environmental tests, and safety testing based on hazard analysis. Pass/Fail criteria were based on predicate device specifications; results demonstrated substantial equivalence and adherence to performance claims.
Technological Characteristics
Multi-parameter monitor/defibrillator. Features Bluetooth wireless connectivity for data transfer. Includes modules for ECG, NBP, SpO2, EtCO2, invasive pressure, temperature, and Q-CPR feedback. Software-controlled operation. No changes to core hardware sensing/actuation principles compared to predicate.
Indications for Use
Indicated for patients requiring cardiac monitoring, defibrillation, pacing, or physiological parameter measurement. AED therapy for patients >= 8 years old, unresponsive, pulseless, and apneic. Q-CPR contraindicated for neonates/pediatrics < 8 years or < 25 kg.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
K063375
#### 8.0 510(k) Summary
JAN 1 1 2007
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. 8807.92.
- 1. The submitter of this premarket notification is Philips Medical Systems
This summary was prepared on 24 October 2006
- 2. The name of this device is the HeartStart MRx Monitor/Defibrillator Software Release 7.01. Classification names are as follows:
| Classification | ProCode | Description |
|----------------|---------|--------------------------------------------------------------|
| 870.2340, II | 74 DPS | Electrocardiograph device |
| 870.1130, II | 74 DXN | Non-invasive blood pressure |
| 870.2700, II | 74 DQA | Pulse oximeter |
| 870.2300, II | 74 MWI | Monitor, Physiological, Patient |
| 870.2300, II | 74 MSX | System, Network and Communication, Physiological<br>Monitors |
| 868.1400, II | 74 CCK | End-tidal Carbon Dioxide |
| 870.5550, II | 74 DRO | External Trascutaneous Pacemaker (Non-invasive) |
| 870.5300, II | 74 LDD | Low-energy defibrillator |
| 870.1025, III | 74 MKJ | Defibrillators, Automatic, External |
| 870.5200, III | 74 LIX | Cardiopulmonary Resuscitation Aid |
| 870.1110, II | 74 DSK | Blood Pressure Computer |
| 880.2910, II | 80 FLL | Clinical Electronic Thermometer |
- 3. The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software cleared under K031187 and K051134, as well as K061707).
- 4. The modification is a change that provides an optional Bluetooth™ wireless technology option to transfer data from the MRx to a Bluetooth enabled device. This wireless connection is an alternative to the current RS232 cable connection capability.
- 5. The new device has the same Indications for Use as the legally marketed predicate device.
- 6. The new device has the same technological characteristics as the legally marketed predicate device.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved system level tests, integration tests, environmental tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the device meets all performance claims.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 1 2007
Philips Medical Systems Ms. Denise Halev Sr. Quality and Regulatory Engineer 3000 Minuteman Road Andover, MA 01810
Re: K063375
Trade Name: HeartStart MRx Models M3535A/M3536A Regulation Number: 21 CFR 870.5310 Regulation Name: Automatic External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: December 12, 2006 Received: December 12, 2006
Dear Ms. Haley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Denise Haley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
. A. Hillelenne
for
E
Γ
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### 2.0 Indications for Use
K063375 510(k) Number (if known):
Device Name: HeartStart MRx Monitor/Defibrillator
### Indications for Use:
The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
## AED Therapy
To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.
## Manual Defibrillation
Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation.
### Non-invasive External Pacing Therapy
The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.
### Pulse Oximetry
The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level.
### Non-invasive Blood Pressure Monitoring
The NBP option is intended for non-invasive measurement of a patient's arterial blood pressure.
### End-tidal CO2
The EtCO2 option is intended for non-invasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.
### 12-Lead ECG
The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements.
[continued....]
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|------------------------------------|--------------------------------------------------------|
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | Page 1 of 2 |
| 510(k) Number | K063375 |
PHILIPS MEDICAL SYSTEMS
{4}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
HeartStart MRx Monitor/Defibrillator Device Name:
Indications for Use: [continued ... ]
# Q-CPR
The Q-CPR™ option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).
The Q-CPR option is contraindicated as follows:
- The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less that 25 kg).
- The Q-CPR option is not for use when CPR is contraindicated. -
## Invasive Pressure
The Invasive Pressure option is indicated for measuring arterial, venous, intracranial and other physiological pressures on patients.
### Temperature
The Temperature option is indicated for measuring temperature in patients.
Prescription Use Yes Over-The-Counter Use No AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2
MeleMay
(Division Sign-Off)
510(k) Number K063375
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.