An ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. Ossicular replacement prostheses are indicated for the functional restoration of the ossicular chain when a conductive hearing loss is present. Indications for use include: (a) Chronic middle ear disease, (b) Otosclerosis, (c) Congenital fixation of the stapes, (d) Secondary surgical intervention to correct for a significant and persistent conductive hearing loss from prior otologic surgery, and (e) Surgically correctible injury to the middle ear from trauma.
Device Story
Family of Nitinol-based ossicular replacement prostheses (ORPs) designed for middle ear reconstruction. Devices include Incudo-Stapedial Joint, Angular Piston, Piston, Malleus Piston, PORP, and Footplate Shoe. Implants facilitate sound conduction from tympanic membrane to inner ear. Used in surgical settings by otolaryngologists. Attachment via manual or heat-activated crimping. Nitinol provides shape-memory properties for secure fixation to ossicles. Benefits include restoration of hearing in patients with conductive loss. No software or electronic components.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Nitinol (ASTM F2063-05), Unalloyed Titanium (ASTM F67), Titanium alloy (ASTM F136), Hydroxylapatite (ASTM F1185), Fluoroplastic (ASTM D1710, ASTM F754), Medical grade silicone. Mechanical fixation via manual or heat-activated crimping. Sterile, single-use implants. No software or energy source.
Indications for Use
Indicated for patients with conductive hearing loss due to chronic middle ear disease, otosclerosis, congenital stapes fixation, prior failed otologic surgery, or middle ear trauma requiring functional reconstruction of the ossicular chain.
Regulatory Classification
Identification
A partial ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. The device is made of materials such as stainless steel, tantalum, polytetrafluoroethylene, polyethylene, polytetrafluoroethylene with carbon fibers composite, absorbable gelatin material, porous polyethylene, or from a combination of these materials.
{0}------------------------------------------------
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS Grace Medical Partial Ossicular Replacement Prostheses
GEC - 6 2006
| Trade Name: | The family of Grace Medical Nitinol Partial Ossicular Replacement<br>Prostheses consists of:<br>Nitinol Locking Incudo-Stapedial Joint Nitinol Locking Angular Piston Nitinol Locking Piston Nitinol Locking Malleus Piston Nitinol Locking PORP Footplate Shoe |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Partial Ossicular Replacement Prostheses |
| Classification Name: | Partial Ossicular Replacement Prostheses (CFR 21 §<br>874.3450) |
| Official Contact: | Jeff Cobb<br>Vice President of Regulatory Affairs & Quality<br>Grace Medical, Inc.<br>8500 Wolf Lake Drive, Suite 110<br>Memphis, TN 38133 |
| Telephone: | (901) 380-7000 |
| Telefax: | (901) 380-7001 |
| Date Prepared: | November 8, 2006 |
### Predicate Devices
- 1) SMart ISJ Prosthesis manufactured by Gyrus ENT LLC and cleared via 510(k) No. K033554.
- 2) SMart Piston manufactured by Gyrus ENT LLC and cleared via 510(k) No. K003214.
- 3) Angular Prosthesis (Plester) manufactured by Heinz Kurz GmbH Medizintechnik and cleared via 510(k) No. K972492.
- 4) CliP Piston MVP manufactured by Heinz Kurz GmbH Medizintechnik and cleared via 510(k) No. K042503.
- 5) CliP Piston aWengen manufactured by Heinz Kurz GmbH Medizintechnik and cleared via 510(k) No. K021479.
- 6) Adjustable and Fixed-Length PORPs manufactured by Grace Medical, Inc. and cleared via 510(k) No. K061853.
- 7) Partial Ossicular Replacement Prosthesis manufactured by Grace Medical, Inc. and cleared via 510(k) K972815.
Intended Use - The Grace Medical Partial Ossicular Replacement Prostheses have the same primary intended use as the predicate devices.
An ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner car. Ossicular replacement prostheses are indicated for the functional restoration of the ossicular chain when a conductive hearing loss is present. Indications for use include:
{1}------------------------------------------------
- (a) Chronic middle ear disease,
- (b) Otosclerosis,
- (c) Congenital fixation of the stapes,
- (d) Secondary surgical intervention to correct for a significant and persistent conductive hearing loss from prior otologic surgery, and
- (e) Surgically correctible injury to the middle ear from trauma.
Material(s) - The Grace Medical Partial Ossicular Replacement Prostheses are manufactured from the same materials as the predicate devices:
- (i) Nitinol (ASTM F2063-05)
- (ii) Unalloyed titanium (ASTM F67)
- (iii)Titanium alloy (ASTM F136)
- (iv) Hydroxylapatite (ASTM F1185)
- (v) Fluoroplastic (USP Class VI for plastics, ASTM D1710 and ASTM F754)
- (vi) Medical grade silicone
### Comparison Charts
.
### NITINOL LOCKING INCUDO-STAPEDIAL JOINT
| | Nitinol Locking Incudo-Stapedial Joint | SMart<br>ISJ<br>(Gyrus) | SMart<br>Piston<br>(Gyrus) | Angular<br>Prosthesis<br>(Kurz) | Similarities or Differences |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----------------------------|---------------------------------|------------------------------------------------------------------------------------------------|
| Intended Use | An ossicular replacement prosthesis is a device<br>intended to be implanted for the functional<br>reconstruction of segments of the ossicular chain and<br>facilitates the conduction of sound wave from the<br>tympanic membrane to the inner ear. | Same | Same | Same | Same |
| Material(s) | Nitinol | Nitinol | Nitinol<br>Fluoroplastic | Titanium | Substantially Equivalent |
| Method of<br>Attachment | Manual or Heat Crimp-Assist | Same | Same | Manual | Same as SMart ISJ and<br>SMart Piston. Different<br>than but SE to Kurz<br>Angular Prosthesis. |
| How<br>Supplied | Sterile | Sterile | Sterile | Sterile | Same |
### NITINOL LOCKING ANGULAR PISTON
| | Nitinol Locking Angular Piston | SMart<br>ISJ<br>(Gyrus) | SMart Piston<br>(Gyrus) | Angular<br>Piston<br>(Kurz) | Similarities or<br>Differences |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------|-----------------------------|--------------------------------|
| Intended<br>Use | An ossicular replacement prosthesis is a device intended to be<br>implanted for the functional reconstruction of segments of the<br>ossicular chain and facilitates the conduction of sound wave from<br>the tympanic membrane to the inner car. | Same | Same | Same | Same |
| Material(s) | Nitinol<br>Titanium | Nitinol | Nitinol<br>Fluoroplastic | Titanium | Substantially<br>Equivalent |
| Lengths | 3.0mm to 9.0mm | n/a | 3.0mm to<br>6.0mm | 4.25mm to<br>4.75mm | Substantially<br>Equivalent |
| How<br>Supplied | Sterile | Sterile | Sterile | Sterile | Same |
{2}------------------------------------------------
# NITINOL LOCKING PISTON
| | Nitinol Locking Piston | SMart<br>ISJ<br>(Gyrus) | SMart Piston<br>(Gyrus) | Angular<br>Piston<br>(Kurz) | Similarities or<br>Differences |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------|-----------------------------|--------------------------------|
| Intended<br>Use | An ossicular replacement prosthesis is a device intended to<br>be implanted for the functional reconstruction of segments<br>of the ossicular chain and facilitates the conduction of<br>sound wave from the tympanic membrane to the inner ear. | Same | Same | Same | Same |
| Material(s) | Nitinol<br>Titanium<br>Fluoroplastic | Nitinol | Nitinol<br>Fluoroplastic | Titanium | Substantially<br>Equivalent |
| Lengths | 3.0mm to 9.0mm | n/a | 3.0mm to 6.0mm | 4.25mm to<br>4.75mm | Substantially<br>Equivalent |
| How<br>Supplied | Sterile | Sterile | Sterile | Sterile | Same |
# NITINOI. LOCKING MALLEUS PISTON
| | Nitinol Locking Malleus Piston | CliP Piston<br>MVP<br>(Kurz) | CliP Piston<br>aWengen<br>(Kurz) | Similarities or<br>Differences |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------|--------------------------------|
| Intended<br>Use | An ossicular replacement prosthesis is a device intended to be implanted for<br>the functional reconstruction of segments of the ossicular chain and facilitates<br>the conduction of sound wave from the tympanic membrane to the inner ear. | Same | Same | Same |
| Material(s) | Nitinol<br>Titanium<br>Fluoroplastic | Titanium | Titanium | Substantially<br>Equivalent |
| Lengths | 3.0mm - 9.0mm | 5.00mm -<br>6.25mm | 4.0mm -<br>5.5mm | Substantially<br>Equivalent |
| How<br>Supplied | Sterile | Sterile | Sterile | Same |
### Nitinol Locking PORP
| | Nitinol Locking PORP | SMart<br>ISJ<br>(Gyrus) | SMart Piston<br>(Gyrus) | Grace Adjustable and<br>Fixed-length PORPs | Similarities or<br>Differences |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------|-----------------------------------------------------------------------------------------|--------------------------------|
| Intended<br>Use | An ossicular replacement prosthesis is<br>a device intended to be implanted for<br>the functional reconstruction of<br>segments of the ossicular chain and<br>facilitates the conduction of sound<br>wave from the tympanic membrane to<br>the inner ear. | Same | Same | Same | Same |
| Shaft<br>Material(s) | Nitinol<br>Titanium<br>Silicone (sleeve on adjustable models) | Nitinol | Nitinol<br>Fluoroplastic | Titanium<br>Silicone (sleeve on adjustable<br>models) | Substantially<br>Equivalent |
| Head<br>Material(s) | Titanium<br>Hydroxylapatite | No<br>head | No head | Titanium<br>Hydroxylapatite | Same |
| Lengths | Fixed Sizes: 0.5mm to 9.0mm<br>Adjustable model - Trimmed<br>intraoperatively to length | n/a | 3.0mm to 6.0mm | Fixed Sizes: 0.5mm to 9.0mm<br>Adjustable model - Trimmed<br>intraoperatively to length | Substantially<br>Equivalent |
| How<br>Supplied | Sterile | Sterile | Sterile | Sterile | Same |
{3}------------------------------------------------
#### FOOTPLATE SHOE
| | Footplate Shoe | Grace HA Footplate Shoe<br>K972815 | Similarities or<br>Differences |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|--------------------------------|
| Intended Use | An ossicular replacement prosthesis is a device intended to<br>be implanted for the functional reconstruction of segments<br>of the ossicular chain and facilitates the conduction of sound<br>wave from the tympanic membrane to the inner ear. | Same | Same |
| Material(s) | Nitinol<br>Titanium<br>Hydroxylapatite | Hydroxylpaptite | Substantially<br>Equivalent |
| Accommodates<br>Shafts with<br>Diameters: | 0.3mm to 1.07mm | 1.07mm | Substantially<br>Equivalent |
| How Supplied | Sterile | Sterile | Same |
Differences between the Grace Medical Modified Partial Ossicular Replacement Prostheses and the predicate devices should not affect the safety or effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Grace Medical, Inc. c/o Jeff Cobb Regulatory Affairs & Quality 8300 Wolf Lake Drive Suite 110 Memphis, TN
DEC = 6 2006
Re: K063374
Trade/Device Name: Partial Ossicular Replacement Prostheses, Nitinol Incudo-Stapedial Joint, Nitinol Angular Piston, Nitinol Piston, Nitinol Malleus Piston, Nitinol PORP and Footplate Shoc Regulation Number: 21 CFR 874.3450 Regulation Name: Partial ossicular replacement prosthesis Regulatory Class: Class II Product Code: ETB Dated: November 8, 2006 Received: November 8, 2006
Dear Mr. Cobb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premated nevice results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eychler S.M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number: ¥063374
Device Name: Partial Ossicular Replacement Prostheses including: Nitinol Incudo-Stapedial Joint, Nitinol Angular Piston, Nitinol Piston, Nitinol Malleus Piston, Nitinol PORP, and Footplate Shoe
## Indications for Use:
An ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. Ossicular replacement prostheses are indicated for the functional restoration of the ossicular chain when a conductive hearing loss is present. Indications for use include:
- (a) Chronic middle ear disease,
- (b) Otosclerosis.
- (c) Congenital fixation of the stapes,
- (d) Secondary surgical intervention to correct for a significant and persistent conductive hearing loss from prior otologic surgery, and
- (e) Surgically correctible injury to the middle ear from trauma.
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2/96)
Karus boha
X
e and Throat Devis
510(k) Number K063374
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.