Retrospective clinical specimens from patients with signs/symptoms of Hepatitis; Retrospective clinical specimens from high-risk populations (e.g., IV drug users, prison history, MSM)
Retrospective clinical specimens were used to evaluate the clinical sensitivity and specificity of the MONOLISA™ Anti-HAV EIA compared to a predicate device.
Retrospective study; Clinical performance; Hepatitis A; Diagnostic aid
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-center prospective and retrospective study; Multi-center prospective and retrospective observational study
Individuals with signs or symptoms of Hepatitis and individuals at high risk for Hepatitis infection (e.g., IV drug users, prison history, MSM); Sample Size: 1,248 total specimens (404 US, 844 Europe); Number of Sites: 3 geographical areas (US, France, Italy)
Comparative anti-HAV assay
Clinical sensitivity and specificity (positive/negative percent agreement)
Indications for Use
The MONOLISA™ anti-HAV EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of total antibodies (anti-HAV IgG and IgM) to Hepatitis A virus (anti-HAV) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD). This kit can be used as an aid in the diagnosis of acute or past Hepatitis A Virus (HAV) infection or as an aid in the identification of HAV-susceptible individuals for vaccination. However, any diagnosis should take into consideration the patient’s clinical history and symptoms, as well as serological data. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, and core blood or neonatal specimens. WARNING: This assay is not intended for screening blood or solid or soft tissue donors.
Device Story
MONOLISA™ Anti-HAV EIA is a competitive enzyme immunoassay (EIA) for qualitative detection of total anti-HAV antibodies in human serum or plasma. Procedure involves incubating patient specimens, calibrator, and controls with HAV antigen in microwells coated with mouse monoclonal anti-HAV antibodies. Conjugate (HRP-labeled mouse monoclonal anti-HAV) is added; it binds to HAV antigen in wells where patient anti-HAV is absent. TMB chromogen/substrate solution is added; HRP catalyzes color change (blue to yellow after stop solution). Color intensity measured spectrophotometrically at 450 nm and 615/630 nm. Absorbance compared to cutoff derived from calibrator mean. Used in clinical laboratories by trained personnel. Results aid diagnosis of acute/past HAV infection or identification of susceptible individuals for vaccination. Clinical decision-making requires integration with patient history, symptoms, and other serological data.
Clinical Evidence
Multi-center prospective/retrospective study (N=1,248 total specimens: 404 US, 844 Europe). US population included symptomatic and high-risk subjects; European population included symptomatic, high-risk, hospitalized, and recovered subjects. Positive percent agreement 99.5% (931/936); negative percent agreement 96.2% (378/393). Performance evaluated in pediatric subjects (N=60) and vaccinated individuals (VAQTA, HAVRIX, TWINRIX). Seroconversion panel testing showed sensitivity equivalent or superior to comparative assay. Cross-reactivity testing (N=255) included various viral/autoimmune conditions.
Indicated for use as an aid in the diagnosis of acute or past Hepatitis A Virus (HAV) infection or as an aid in the identification of HAV-susceptible individuals for vaccination in adult and pediatric populations. Not established for immunocompromised/immunosuppressed patients, cord blood, or neonatal specimens.
Regulatory Classification
Identification
HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.
Special Controls
*Classification.* Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K063318
B. Purpose for Submission:
New device clearance
C. Measurand:
Total Antibody to Hepatitis A virus
D. Type of Test:
Qualitative, ELISA
E. Applicant:
Bio-Rad.
F. Proprietary and Established Names:
MONOLISA™ Anti-HAV EIA
G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.3310, Hepatitis A virus (HAV) Serological Reagents
2. Classification:
Class II
3. Product Code:
LOL
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The MONOLISA™ anti-HAV EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of total antibodies (anti-HAV IgG and IgM) to Hepatitis A virus (anti-HAV) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD). This kit can be used as an aid in the diagnosis of acute or past Hepatitis A Virus (HAV) infection or as an aid in the identification of HAV-susceptible individuals for vaccination. However, any diagnosis should take into consideration the patient’s clinical history and symptoms, as well as serological data.
Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, and core blood or neonatal specimens.
WARNING: This assay is not intended for screening blood or solid or soft tissue donors.
2. Special condition for use statement(s):
{1}
The new device is not intended to be sold over the counter and is for prescription use only. Warnings on applicant labeling state "Under United States federal law restricts this device to sale by or on the order of a licensed practitioner or physician".
3. Special instrument Requirements:
Bio-Rad microwell plate or strip reader or equivalent. The spectrophotometer should have the following specifications at wavelength $450\mathrm{nm}$ :
Bandwidth: $10\mathrm{nm}$ HBW (Half Band Width) or equivalent
Absorbance Range: 0 to 3.0 Repeatability: $\pm (0.5\% + 0.005)$ Linearity or
Accuracy: $1\%$ from 0 to 3.0
The instrument should contain a reference filter for reading at 615 to $630\mathrm{nm}$ . An instrument without a reference filter can be used; however, areas in the bottoms of the wells that are opaque, scratched or irregular may cause absorbance readings that are falsely elevated.
I. Device Description:
Enzyme immunoassay (competitive assay format) for the detection of total antibodies to Hepatitis A virus.
J. Substantial Equivalence Information:
1. Predicate device name(s):
DiaSorin ETI-AB-HAVK PLUS
2. Predicate K number(s):
P890019
3. Comparison with predicate:
Table 1: Similarities between kit components and materials
| Similarities in Components / Materials | MONOLISA™ Anti-HAV EIA | ETI-AB-HAVK PLUS |
| --- | --- | --- |
| Solid Phase | Microplate wells coated with mouse Monoclonal anti-HAV antibodies. | Microplate wells coated with mouse Monoclonal anti-HAV antibodies. |
| Conjugate | Peroxidase-labeled mouse monoclonal antibody to HAV. | Peroxidase-labeled mouse monoclonal antibody to HAV. |
| Negative Control | Human serum, negative for total anti-HAV antibodies. | Human serum/plasma, negative for total anti-HAV antibodies. |
| Calibrator | Human serum, positive for anti-HAV antibodies, diluted in human serum pool negative for anti-HAV antibodies. | Human serum/plasma, containing anti-HAV antibodies. |
{2}
Table 2: Differences between kit components and materials
| Differences in Components / Materials | MONOLISA™ Anti-HAV EIA Catalog# 72496 | ETI-AB-HAVK PLUS Catalog# P001926 |
| --- | --- | --- |
| Conjugate | Ready-to-use. | To be diluted. |
| Incubation buffer | NA | Buffer, containing protein stabilizers and an inert blue dye. |
| Viral Antigen / Neutralizing Solution | Tris-buffer, containing inactivated HAV-virus, proteins and sample indicator dye. | Buffer, containing HAV, human serum/plasma and protein stabilizers. |
| Stopping Solution | 1N H2SO4. | 0.4N H2SO4. |
# K. Standard/Guidance Document Referenced (if applicable):
Class II Special Control Guidance Document, Hepatitis A Virus Serological Assays, issued February 9, 2006.
# L. Test Principle:
The MONOLISA™ Anti-HAV EIA is an enzyme immunoassay (competitive assay format) for the detection of total antibodies to Hepatitis A virus. In the assay procedure, patient specimens, a calibrator and controls are incubated with HAV antigen in microwells that have been coated with mouse monoclonal anti-Hepatitis A antibodies. Antibodies to HAV present in a specimen or control will complex with the HAV antigen reagent and with antibodies coated on the microwells. Excess sample and HAV Viral antigen reagent are removed by a wash step. The conjugate (containing horseradish peroxidase-labeled mouse monoclonal antibody to HAV) is subsequently added to the microwells and incubated. The conjugate binds to the HAV antigen bound to the microwell in the absence of antibodies to HAV from the specimen. Excess conjugate is removed by a wash step, and a TMB Chromogen / Substrate solution is added to the microwells and allowed to incubate. If a sample does not contain anti-HAV antibodies, the bound enzyme (HRP) causes the colorless
{3}
Page 4 of 18
tetramethylbenzidine (TMB) in the Chromogen solution to change to blue. The blue color turns yellow after the addition of a Stopping Solution. If a sample contains anti-HAV antibodies, the Chromogen / Substrate Solution in the well remains colorless during the substrate incubation, and after the addition of the Stopping Solution. The color intensity is measured spectrophotometrically.
Absorbance value readings for patient specimens are compared to the Cutoff value determined by the mean of the Calibrator absorbance values.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-Laboratory Precision Study:
A 21-member panel was tested: serum samples with the 6 corresponding plasma samples (EDTA K2, EDTA K3, Sodium Citrate, Sodium Heparin, Lithium heparin, ACD) at 3 different levels (1 negative, 1 negative near the cutoff, 1 low positive near the cutoff) were tested on 1 lot, in duplicate, in 2 different runs per day (am and pm), by the same operator for a period of 20 days. The data were analyzed following the CLSI guidance EP5A2. The mean ratio, the Standard Deviation (SD) and percent coefficient of variation (%CV) were calculated for each panel member. The data summary is shown in the following tables.
MONOLISA™ Anti-HAV EIA Precision Results by Panel Member Cutoff to Signal (CO/S)
| Panel Member | N | Mean CO/S | Within run¹ | | Between Run² | | Between Day³ | | Total⁴ | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| Negative Control C0 | 40 | 0.282 | NA | NA | 0.02 | 6.4% | 0.01 | 5.0% | 0.02 | 8.1% |
| Positive Control C1 | 40 | 4.093 | NA | NA | 0.54 | 13.5% | 0.00 | 0.0% | 0.49 | 12.3% |
| Serum 1 | 80 | 0.395 | 0.02 | 4.1% | 0.03 | 6.3% | 0.01 | 1.9% | 0.03 | 7.7% |
| EDTA K2 1 | 80 | 0.379 | 0.02 | 5.0% | 0.03 | 6.8% | 0.01 | 1.4% | 0.03 | 8.6% |
| EDTA K3 1 | 80 | 0.376 | 0.01 | 3.4% | 0.04 | 9.3% | 0.00 | 0.0% | 0.04 | 9.9% |
| Sodium Citrate 1 | 80 | 0.387 | 0.04 | 10.1% | 0.01 | 3.3% | 0.02 | 5.1% | 0.05 | 11.8% |
| Sodium Heparin 1 | 80 | 0.363 | 0.01 | 3.4% | 0.03 | 7.6% | 0.00 | 0.0% | 0.03 | 8.3% |
| Lithium Heparin 1 | 80 | 0.364 | 0.01 | 3.4% | 0.03 | 7.0% | 0.01 | 3.5% | 0.03 | 8.5% |
| ACD 1 | 80 | 0.402 | 0.02 | 4.3% | 0.04 | 10.4% | 0.00 | 0.0% | 0.05 | 11.3% |
| Serum 2 | 80 | 0.691 | 0.03 | 4.9% | 0.06 | 9.6% | 0.01 | 2.3% | 0.06 | 11.0% |
| EDTA K2 2 | 80 | 0.657 | 0.02 | 3.2% | 0.05 | 6.6% | 0.01 | 1.8% | 0.05 | 7.5% |
| EDTA K3 2 | 80 | 0.686 | 0.03 | 4.9% | 0.06 | 8.2% | 0.00 | 0.0% | 0.07 | 9.5% |
| Sodium Citrate 2 | 80 | 0.636 | 0.03 | 3.7% | 0.05 | 7.0% | 0.03 | 4.8% | 0.06 | 9.2% |
| Sodium Heparin 2 | 80 | 0.628 | 0.02 | 3.1% | 0.04 | 6.0% | 0.03 | 4.2% | 0.06 | 7.9% |
| Lithium Heparin 2 | 80 | 0.685 | 0.05 | 6.6% | 0.06 | 8.6% | 0.00 | 0.0% | 0.08 | 10.8% |
| ACD 2 | 80 | 0.746 | 0.04 | 5.6% | 0.05 | 6.4% | 0.03 | 4.7% | 0.07 | 9.7% |
| Serum 3 | 80 | 1.506 | 0.06 | 4.2% | 0.14 | 9.4% | 0.00 | 0.4% | 0.15 | 10.3% |
| EDTA K2 3 | 80 | 1.261 | 0.07 | 4.7% | 0.11 | 7.0% | 0.07 | 4.8% | 0.15 | 9.7% |
| EDTA K3 3 | 80 | 1.257 | 0.04 | 2.4% | 0.09 | 6.0% | 0.05 | 3.6% | 0.11 | 7.4% |
| Sodium Citrate 3 | 80 | 1.462 | 0.08 | 5.1% | 0.13 | 8.7% | 0.07 | 4.9% | 0.17 | 11.2% |
| Sodium Heparin 3 | 80 | 1.380 | 0.11 | 7.5% | 0.12 | 8.0% | 0.05 | 3.4% | 0.17 | 11.4% |
| Lithium Heparin 3 | 80 | 1.346 | 0.08 | 5.6% | 0.11 | 7.1% | 0.03 | 1.8% | 0.14 | 9.2% |
| ACD 3 | 80 | 1.344 | 0.05 | 3.4% | 0.08 | 5.6% | 0.09 | 6.0% | 0.13 | 8.9% |
{4}
Page 5 of 18
NA: Not Applicable
1. Within Run: variability of the assay performance from replicate to replicate
2. Between Run: variability of the assay performance from Run to Run
3. Between Day: variability of the assay performance from Day to Day
4. Total: total variability of the assay performance includes within run, between run and between day.
## Reproducibility Study:
A 6 member panel consisting of diluted plasma specimens (negative and different levels of positive) was tested in triplicate, once a day for 3 days on 3 lots* of MONOLISA™ Anti-HAV EIA at 3 separate clinical trial sites. Each panel was coded with a different number on each day tested in order to blind the operator to the expected value of the sample.
*:3 different lots were used at the Bio-Rad site and 2 lots were used on each of the external sites.
The data from all reagent lots and sites were combined to obtain Standard Deviation (SD) and percent coefficient of variation (CV) for within run, between day, between lot, between site and total variance. The data were analyzed according to the principles described in the Clinical Laboratory Standards Institute guidance EP5-A2, revised November 2004 and ISO/TR 22971:2005. The PROC GLM procedure in SAS® was used to estimate the variance components of the model. The model was $y = \text{site} + \text{lot}(\text{site}) + \text{day}(\text{lot site}) + \text{error}$.
The summaries are shown in the following tables.
MONOLISA™ Anti-HAV EIA Reproducibility Results by Panel Member Cutoff to Signal (CO/S)
| Test site | Panel Member | N | Mean CO/S | Within Run¹ | | Between Day² | | Between Lot³ | | Total⁴ | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Site #1 | P1 | 18 | 0.33 | 0.01 | 3.03 | 0.09 | 28.1 | 0⁵ | 0 | 0.09 | 28.3 |
| | P2 | 18 | 0.68 | 0.03 | 4.9 | 0.07 | 10.0 | 0⁵ | 0 | 0.07 | 11.1 |
| | P3 | 18 | 1.02 | 0.05 | 5.0 | 0.05 | 5.2 | 0⁵ | 4.1 | 0.08 | 8.3 |
| | P4 | 18 | 1.98 | 0.09 | 4.7 | 0.32 | 16.0 | 0⁵ | 0 | 0.33 | 16.7 |
| | P5 | 18 | 2.48 | 0.18 | 7.4 | 0.36 | 14.8 | 0⁵ | 0 | 0.41 | 16.5 |
| | P6 | 18 | 3.66 | 0.23 | 6.2 | 0.20 | 5.5 | 0⁵ | 0 | 0.3 | 8.3 |
| Site#2 | P1 | 18 | 0.35 | 0.01 | 1.6 | 0.02 | 4.7 | 0.01 | 1.2 | 0.02 | 5.1 |
| | P2 | 18 | 0.92 | 0.03 | 3.0 | 0.07 | 8.0 | 0⁵ | 0 | 0.08 | 8.5 |
| | P3 | 18 | 1.28 | 0.04 | 3.5 | 0.00 | 0.0 | 0.01 | 0.62 | 0.05 | 3.6 |
| | P4 | 18 | 2.32 | 0.08 | 3.6 | 0.20 | 8.5 | 0.02 | 0.95 | 0.21 | 9.3 |
| | P5 | 18 | 3.10 | 0.13 | 4.1 | 0.20 | 6.4 | 0⁵ | 0 | 0.23 | 7.5 |
| | P6 | 18 | 4.16 | 0.13 | 3.2 | 0.36 | 8.7 | 0⁵ | 0 | 0.39 | 9.3 |
| Site #3 | P1 | 27 | 0.36 | 0.01 | 4.0 | 0.02 | 5.4 | 0.02 | 6.6 | 0.03 | 9.4 |
| | P2 | 27 | 0.81 | 0.03 | 3.4 | 0.03 | 3.9 | 0.06 | 7.2 | 0.07 | 8.8 |
| | P3 | 27 | 1.27 | 0.08 | 6.6 | 0.04 | 3.6 | 0.14 | 11.2 | 0.17 | 13.5 |
| | P4 | 27 | 2.16 | 0.11 | 5.0 | 0.05 | 2.2 | 0.34 | 15.7 | 0.36 | 16.6 |
| | P5 | 27 | 3.09 | 0.11 | 3.7 | 0.15 | 4.9 | 0.49 | 15.7 | 0.52 | 16.9 |
| | P6 | 27 | 4.47 | 0.11 | 2.5 | 0.33 | 7.3 | 1.01 | 22.5 | 1.06 | 23.8 |
¹ Within Run: variability of the assay performance from replicate to replicate
² Between Day: variability of the assay performance from Day to Day
³ Between Lot: variability of the assay performance from Lot to Lot
⁴ Total: total variability of the assay performance includes within run, between day and between lot.
⁵ Negative variances were rounded to zero, per statistical convention.
{5}
b. Linearity/assay reportable range: NA
c. Traceability, Stability, Expected values (controls, calibrators, or method): Controls and calibrators are provided specifically designed for use in conjunction with the performance of the assay.
d. Detection limit: See assay cut-off below
e. Analytical specificity: The potential for cross reactivity to other disease states, or viruses was evaluated for the MONOLISA™ Anti-HAV EIA Assay and the comparative assay. In addition, samples containing rheumatoid factors, auto-antibodies, anti-mouse antibodies were tested. In total, 255 specimens (including both serum and plasma) from 16 groups of potential cross-reactivity were tested. FDA approved methods were used to confirm the disease state of each specimen. The results are summarized in the following table.
Potential cross reactivity study:
| Clinical Condition | Comparative assay Positive | | | Comparative assay BRD | | | Comparative assay Negative | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | Total |
| | R | BRD | NR | R | BRD | NR | R | BRD | NR | |
| Hepatitis C (HCV) | 7 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 7 | 15 |
| Hepatitis B (HBV) HBs Ag | 9 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 6 | 15 |
| Hepatitis B (HBV) anti HBc | 10 | 0 | 0 | 0 | 0 | 0 | 3 | 1 | 1 | 15 |
| Human Immunodeficiency Virus (HIV) | 6 | 0 | 1 | 1 | 0 | 0 | 0 | 0 | 7 | 15 |
| Epstein Barr Virus (EBV) IgG | 1 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 13 | 15 |
| Epstein Barr Virus (EBV) IgM | 15 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 15 |
| Cytomegalovirus (CMV) IgG | 6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 9 | 15 |
| Cytomegalovirus (CMV) IgM | 7 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 8 | 15 |
| Rubella IgG | 5 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 9 | 15 |
| Rubella IgM | 10 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 10 | 15 |
{6}
| Toxoplasmosis IgG | 10 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 5 | 15 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Toxoplasmosis IgM | 8 | 2 | 0 | 0 | 0 | 1 | 0 | 0 | 4 | 15 |
| Mumps IgG | 3 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 11 | 15 |
| Varicella Zoster Virus(VZV) IgG | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 14 | 15 |
| Varicella Zoster Virus(VZV) IgM | 6 | 0 | 0 | 1 | 0 | 1 | 0 | 0 | 7 | 15 |
| Anti Nuclear Antibody (ANA) | 7 | 0 | 0 | 0 | 0 | 1 | 0 | 0 | 7 | 15 |
| Human Anti Mouse Antibody (HAMA) | 2 | 0 | 0 | 0 | 0 | 1 | 0 | 0 | 12 | 15 |
| Rheumatoid Arthritis | 12 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 3 | 15 |
| Total | 115 | 2 | 3 | 3 | 0 | 4 | 4 | 1 | 123 | 255 |
7 samples were discrepant: 4 reactive on MONOLISA™ Anti-HAV EIA, nonreactive on comparative assay and 3 were nonreactive on MONOLISA™ Anti-HAV EIA and reactive on comparative assay.
Warning: Assay interference for anti EBV IgM, anti HBc Ab, HBs Ag and rheumatoid factor, has not been evaluated sufficiently. The user is responsible for establishing cross-reactivity performance with these cross-reactants.
f. Assay cut-off:
The assay calibrator is equivalent to $20\mathrm{mIU / mL}$ standardized to the WHO $2^{\mathrm{nd}}$ Reference Standard for Anti-Hepatitis Immunoglobulin. However, assay results cannot be considered quantitative and no clinical claims for immunity can be determined from the cutoff.
2. Comparison studies:
a. Method comparison with predicate device: MONOLISA™ Anti-HAV EIA were compared to the DiaSorin ETI-AB-HAVK PLUS assays.
b. Matrix comparison N/A
3. Clinical studies:
a. Clinical sensitivity: See Performance Characteristics below
b. Clinical specificity:
{7}
Page 8 of 18
See Performance Characteristics below
c. Other clinical supportive data (when a and b are not applicable):
## Clinical Studies
A multi-center prospective and retrospective study was conducted to evaluate the clinical performance of the MONOLISA™ Anti-HAV EIA assay among individuals with signs or symptoms and those at high risk of Hepatitis infection. Specimens were collected in 3 different geographical areas: 404 specimens were collected in the US and 928 were collected in Europe (France and Italy).
The US population consisted of 174 subjects with signs and symptoms of Hepatitis.
Of these, 60% were male and 40% were female, and they ranged in age from 17 to 72 years (mean age of 38). The group was Caucasian (13.2%), Black or African American (4.6%), Hispanic or Latino (2.9%), and Asian (41.9%), with 1.1% represented by multiple ethnic groups. The remaining 36.8% were unknown. Among these 174 subjects, 23 (13.2%) were pediatric samples.
The 230 subjects from the high-risk group for Hepatitis A include intravenous drug users (N=55), homosexual males (N=15), sex workers (N=39), prison history (N=92), high-risk sex partners (N=25), and high-risk occupation/health care workers (N=4). Many had more than 1 high-risk behavior or risk factor. The group was Caucasian (7.4%), Black or African American (74.3%), Hispanic or Latino (15.2%), Asian (0.4%), Native Hawaiian or other Pacific Islander (0.4%), and American Indian or Alaska native (0.9%), with the remaining (1.3%) represented by multiple ethnic groups. Of these, 81% were male and 19% were female, and they ranged in age from 18 to 70 years (mean age of 45). Among these 230 subjects, 2 (0.9%) were pediatric samples.
The European population consisted of 252 specimens collected from patients with signs and symptoms of Hepatitis. Of these, 51% were male and 49% were female, and they ranged in age from 1 to 105 years (mean age of 53). Sixty-two (62) specimens were collected from a population at high risk for hepatitis composed of intravenous drug users (30), subjects who had clotting factor disorders (7) and MSM patients (25). The group was 87% male and 13% female, and ranged in age from 21 to 75 years (mean age of 40).
Three hundred and forty five (345) specimens were from an asymptomatic hospitalized population. Of these, 51% were male and 49% were female, and they ranged in age from 18 to 87 years (mean age of 59). Thirty four specimens were from healthcare workers (for HAV pre-vaccination screening). One hundred and fifty one (151) patients had recovered HAV
{8}
infection. Among these 844 european samples, 35 (4.1%) were from pediatric subjects.
## Vaccinated subjects:
Sixty-two (62) pre- and post-vaccination samples from 38 individuals were tested. Fourteen (14) individuals were enrolled in a vaccination program. They received the TWINRIX® vaccine, a combined Hepatitis A and Hepatitis B vaccine from GlaxoSmithKline. A pre-vaccination sample was collected the day of the first vaccination dose. A second sample was collected before the second vaccination dose was injected (one month after the first dose). A third dose of vaccine was scheduled 6 months after the first injection. The sample after the third vaccination dose was not available.
Twenty samples were collected from 10 subjects, aged 24 to 45 years, who had received the HAVRIX® vaccine. These subjects received HAVRIX® 1440, an inactivated Hepatitis A vaccine from GlaxoSmithKline, in a two-dose schedule (at 0 and 6 to 12 months). For each subject, a pre-and a post-vaccination specimen was obtained. All post-vaccination samples were obtained 4 weeks after vaccination. Fourteen purchased post-vaccination samples were tested; 8 were from individuals vaccinated with HAVRIX® and 6 were from individuals vaccinated with VAQTA® from Merck &Co.
MONOLISA™ Anti-HAV EIA versus the comparative assay Results in the US Population (N=404)
| Subject category | Comparative assay: Positive | | | Comparative assay Borderline | | | Comparative assay: Negative | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | |
| | R | BRD | NR | R | BRD | NR | R | BRD | NR | |
| Subjects with signs and symptoms | 123 | 0 | 2 | 1 | 0 | 0 | 1 | 3 | 44 | 174 |
| Subjects with high risk for Hepatitis | 114 | 0 | 0 | 2 | 1 | 1 | 4 | 1 | 107 | 230 |
| Total | 237 | 0 | 2 | 3^{b} | 1^{d} | 1^{c} | 5 | 4^{a} | 151 | 404 |
| | Positive percent agreement | 95% Exact Confidence interval | Negative percent agreement | 95% Exact Confidence interval |
| --- | --- | --- | --- | --- |
| Total | 98.8% (237/240) | 96.4 – 99.7 | 92.6% (151/163) | 87.5 – 96.1 |
R: Reactive, NR: Nonreactive, BRD: Borderline
{9}
Page 10 of 18
a the Borderline results with MONOLISA™ Anti-HAV EIA were considered as false positives.
b the specimens that were Borderline with the comparative assay and reactive with MONOLISA™ Anti-HAV EIA were considered as false positives with MONOLISA™ Anti-HAV EIA.
c the specimens that were Borderline with the comparative assay and nonreactive with MONOLISA™ Anti-HAV EIA were considered as false negative with MONOLISA™ Anti-HAV EIA.
d the results that were borderline with both the MONOLISA™ Anti-HAV EIA and with the comparative assay were not included in the negative agreement or the positive agreement calculations.
Comparison of Results for MONOLISA™ Anti-HAV EIA versus the comparative assay in the European Population (N= 844)
| Subject category | Comparative assay: Positive | | | Comparative assay: Borderline | | | Comparative assay: Negative | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | |
| | R | BRD | NR | R | BRD | NR | R | BRD | NR | |
| General hospitalized population | 236 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 108 | 345 |
| Sign / Symptoms of Hepatitis | 190 | 0 | 0 | 0 | 2 | 1 | 1 | 1 | 57 | 252 |
| Subjects with high risk for Hepatitis | 28 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 34 | 62 |
| Healthcare workers | 6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 28 | 34 |
| Infected/ recovered HAV | 150 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 151 |
| Total | 610 | 0 | 1 | 1^{b} | 2^{d} | 1^{c} | 1 | 1^{a} | 227 | 844 |
| | Positive percent agreement | 95% Exact Confidence interval | Negative percent agreement | 95% Exact Confidence interval |
| --- | --- | --- | --- | --- |
| Total | 99.7% (610/612) | 98.8 – 99.9 | 98.7% (227/230) | 96.2 – 99.7 |
R: Reactive, NR: Nonreactive, BRD: Borderline
a the Borderline result with MONOLISA™ Anti-HAV EIA was considered as false positive
b the specimen that was Borderline with the comparative assay and reactive with MONOLISA™ Anti-HAV EIA was considered as false positive with MONOLISA™ Anti-HAV EIA.
c the specimen that was Borderline with the comparative assay and nonreactive with MONOLISA™ Anti-HAV EIA was considered as false negative with MONOLISA™ Anti-HAV EIA.
d the 2 borderline results with both MONOLISA™ Anti-HAV EIA and with the comparative assay were not included in the calculation of the negative agreement or the positive agreement.
## Acute HAV Infection:
Among the retrospective samples, 84 were from subjects with a medical history and laboratory results indicative of acute Hepatitis A. The subjects
{10}
included 56% male, 37% female; the gender was not available for 7%. The mean age was 21, and subjects ranged from 1 to 55 years. Among them 39 were pediatric subjects.
The results are presented in the following table:
Comparison of Results for MONOLISA™ Anti-HAV EIA versus the comparative assay on Acute HAV infection in the adult and pediatric European Population (N= 84):
| | Comparative assay: Positive | | | Comparative assay: Borderline | | | Comparative assay: Negative | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | Total |
| | R | BRD | NR | R | BRD | NR | R | BRD | NR | |
| Adults | 45 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 45 |
| Pediatrics | 39 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 39 |
| Total | 84 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 84 |
R: Reactive, NR: Nonreactive, BRD: Borderline
The positive agreement was 100% (84/84) with a 95% exact confidence interval of 96.5% to 100%.
## Performance of MONOLISA™ Anti-HAV EIA in pediatric subjects:
Sixty (60) pediatric samples were tested during the US and European clinical studies in addition to the 39 pediatric samples from acute HAV infection.
Among the US population, 23 had signs and symptoms of hepatitis and 2 were from the high risk group. In the European population, 3 belonged to the general hospitalized population, 22 had signs and symptoms of hepatitis, 2 were from the high risk group, 3 were healthcare workers, 5 had recovered from Hepatitis A infection. The results from these pediatric samples are summarized in the following table.
Comparison of Results for MONOLISA™ Anti-HAV EIA versus the comparative assay in the Pediatric European and US Population (N= 60)
| Subject category | Comparative assay: Positive | | | Comparative assay: Borderline | | | Comparative assay: Negative | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | |
| | R | BRD | NR | R | BRD | NR | R | BRD | NR | |
| European Pediatrics | 16 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 19 | 35 |
{11}
Including the combined US and European Sites, the positive percent agreement of the MONOLISA™ Anti-HAV EIA with the comparative anti-HAV assay was $99.5\%$ (931/936) with a $95\%$ exact confidence interval of $98.8\%$ to $99.8\%$ . The negative percent agreement of the MONOLISA™ Anti-HAV EIA with the comparative anti-HAV assay was $96.2\%$ (378/393) with a $95\%$ exact confidence interval of $93.8\%$ to $97.9\%$ .
# STUDY ON VACCINATED SUBJECTS:
The HAV antibody response to vaccination was evaluated with 3 different vaccines that are currently licensed in the US: VAQTA® from Merck & Co, HAVRIX® 1440 from Glaxo SmithKline and TWINRIX® from Glaxo SmithKline.
For VAQTA® vaccine, 6 post-vaccination samples from US subjects were available.
For HAVRIX® vaccine, 10 matched sets of pre- and post-vaccination samples from European subjects and 8 post-vaccination samples from US subjects were available.
For TWINRIX® vaccine, 14 matched sets of pre-vaccination and post first dose samples from European individuals were available.
The following results were obtained:
MONOLISA™ Anti-HAV EIA Results on Vaccinated Subjects versus the comparative assay - All testing sites
| | Comparative assay: Positive | | | Comparative assay: Borderline | | | Comparative assay: Negative | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | MONOLISA™ Anti-HAV EIA | | | |
| | | R | BRD | NR | R | BRD | NR | R | BRD | | NR |
| VAQTA | Post-vaccination | 6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 6 |
| | Post-injection | 6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 6 |
| TWINRIX® | Vaccination | 6 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 6 |
{12}
Page 13 of 18
| HAVRIX | Pre-vaccination | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 10 | 10 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Post-vaccination | 18 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 18 |
| TWINRIX | Pre-vaccination | 1 | 0 | 0 | 0 | 0 | 0 | 1* | 0 | 12 | 14 |
| | Post 1^{st} injection | 9 | 1 | 0 | 0 | 0 | 2 | 0 | 0 | 2 | 14 |
R: reactive, NR: Nonreactive, BRD: Borderline
* Result close to the cutoff value (CO/S=1.2)
In pre-vaccination samples, MONOLISATM Anti-HAV EIA was in overall agreement with the comparative assay for 21/22 (95.5%) of samples tested. For TWINRIX® vaccine on post first dose vaccination, MONOLISATM Anti-HAV EIA demonstrated reactivity in 9/14 (64.3%) samples. The reference method demonstrated reactivity in 10/14 (71.4%) samples. For HAVRIX® post-vaccination samples, MONOLISATM Anti-HAV EIA demonstrated reactivity in 18/18 (100%) samples. The reference method demonstrated reactivity in 18/18 (100%) samples. For VAQTA® post-vaccination samples, MONOLISATM Anti-HAV EIA demonstrated reactivity in 6/6 (100%) samples. The reference method demonstrated reactivity in 6/6 (100%) samples.
4. Clinical cut-off:
See assay cut-off previously described in this document
5. Expected values/Reference range:
The expected results of the MONOLISA™ Anti-HAV EIA assay were determined in presumably healthy individuals from the Mid-west US (St Louis, Missouri), the Western US (California and Washington) and from Europe (Parma, Italy). In the Mid-west, the population was 55% female and 45% male, with ages ranging from 1 to 96 years. 48% (134) were pediatric specimens.
The majority of the subjects were White/Caucasian (64%), and 32% were black or African American; for 4% data were not available. In this study, 41% were found reactive for Anti-HAV total antibodies, and 57% were found nonreactive.
In the Western US, 73% were from California, 27% were from Washington. The population was 56% female and 44% male, and their ages ranged from 15 to 90 years.
In this population, 38% were found reactive for Anti-HAV total antibodies, and 62% were found nonreactive. In Europe, the population was 50% female and 50% male, with ages ranging from 18 to 87 years. In this group, 69% were found reactive for Anti-HAV total antibodies and 31% were found nonreactive. The expected results for the US and for presumably healthy individuals living in Europe are presented below (Tables 5, 6 and 7).
{13}
Expected Results for MONOLISATM Anti-HAV EIA in subjects from the Mid-west US (N= 280)
| MONOLISA™ Anti-HAV EIA | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Age Range | Gender | Reactive | | Borderline | | Nonreactive | | Total |
| | | N | % | N | % | N | % | |
| < 10 | Female | 10 | 28.6% | 0 | N/A | 25 | 71.4% | 35 |
| | Male | 7 | 18.4% | 2 | 5.3% | 29 | 76.3% | 38 |
| 10 -19 | Female | 14 | 36.8% | 2 | 5.3% | 22 | 57.9% | 38 |
| | Male | 9 | 39.1% | 0 | N/A | 14 | 60.9% | 23 |
| 20- 29 | Female | 3 | 60.0% | 0 | N/A | 2 | 40.0% | 5 |
| | Male | 3 | 100.0% | 0 | N/A | 0 | N/A | 3 |
| 30 -39 | Female | 5 | 50.0% | 0 | N/A | 5 | 50.0% | 10 |
| | Male | 3 | 33.3% | 0 | N/A | 6 | 66.7% | 9 |
| 40 -49 | Female | 3 | 23.1% | 0 | N/A | 10 | 76.9% | 13 |
| | Male | 4 | 50.0% | 0 | N/A | 4 | 50.0% | 8 |
| 50 -59 | Female | 10 | 55.6% | 0 | N/A | 8 | 44.4% | 18 |
| | Male | 8 | 47.1% | 0 | N/A | 9 | 52.9% | 17 |
| 60 -69 | Female | 8 | 57.1% | 0 | N/A | 6 | 42.9% | 14 |
| | Male | 4 | 30.8% | 0 | N/A | 9 | 69.2% | 13 |
| 70-79 | Female | 6 | 66.7% | 1 | 11.1% | 2 | 22.2% | 9 |
| | Male | 5 | 83.3% | 0 | N/A | 1 | 16.7% | 6 |
| 80-89 | Female | 9 | 69.2% | 0 | N/A | 4 | 30.8% | 13 |
| | Male | 3 | 50.0% | 0 | N/A | 3 | 50.0% | 6 |
| >=90 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 1 | 50.0% | 0 | N/A | 1 | 50.0% | 2 |
| Total | | 115 | 41.1% | 5 | 1.8% | 160 | 57.1% | 280 |
Expected Results for MONOLISA™ Anti-HAV EIA in subjects from the Western US (N= 245)
| MONOLISA™ Anti-HAV EIA | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Age Range | Gender | Reactive | | Borderline | | Nonreactive | | Total |
| | | N | % | N | % | N | % | |
| <19 | Female | 3 | 60.0% | 0 | N/A | 2 | 40.0% | 5 |
| | Male | 1 | 20.0% | 0 | N/A | 4 | 80.0% | 5 |
| 20- 29 | Female | 11 | 42.3% | 0 | N/A | 15 | 57.7% | 26 |
| | Male | 5 | 20.8% | 0 | N/A | 19 | 79.2% | 24 |
| 30 -39 | Female | 10 | 50.0% | 0 | N/A | 10 | 50.0% | 20 |
| | Male | 5 | 27.8% | 0 | N/A | 13 | 72.2% | 18 |
| 40 -49 | Female | 6 | 33.3% | 0 | N/A | 12 | 66.7% | 18 |
| | Male | 10 | 45.5% | 0 | N/A | 12 | 54.5% | 22 |
| 50 -59 | Female | 15 | 38.5% | 1 | 2.6% | 23 | 59.0% | 39 |
| | Male | 5 | 23.8% | 0 | N/A | 16 | 76.2% | 21 |
| 60 -69 | Female | 6 | 50.0% | 0 | N/A | 6 | 50.0% | 12 |
| | Male | 4 | 33.3% | 0 | N/A | 8 | 66.7% | 12 |
| 70-79 | Female | 1 | 11.1% | 0 | N/A | 8 | 88.9% | 9 |
| | Male | 1 | 50.0% | 0 | N/A | 1 | 50.0% | 2 |
| 80-89 | Female | 6 | 100% | 0 | N/A | 0 | 1. N/A | 6 |
| | Male | 3 | 75.0% | 0 | N/A | 1 | 25.0% | 4 |
| >=90 | Female | 0 | N/A | 0 | N/A | 1 | 100.0% | 1 |
| | Male | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| Unknown | Female | 0 | N/A | 0 | N/A | 1 | 100.0% | 1 |
| Total | | 92 | 37.6% | 1 | 0.4% | 152 | 62.0% | 245 |
{14}
Page 15 of 18
Expected Results for MONOLISA™ Anti-HAV EIA in subjects from Italy, Europe (N=285)
| MONOLISA™ Anti-HAV EIA | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Age Range | Gender | Reactive | | Borderline | | Nonreactive | | Total |
| | | N | % | N | % | N | % | |
| < 19 | Female | 0 | N/A | 0 | N/A | 1 | 100.0% | 1 |
| | Male | 0 | N/A | 0 | N/A | 1 | 100.0% | 1 |
| 20-29 | Female | 1 | 33.3% | 0 | N/A | 2 | 66.7% | 3 |
| | Male | 0 | N/A | 0 | N/A | 2 | 100.0% | 2 |
| 30-39 | Female | 1 | 14.3% | 0 | N/A | 6 | 85.7% | 7 |
| | Male | 2 | 28.6% | 0 | N/A | 5 | 71.4% | 7 |
| 40-49 | Female | 7 | 33.3% | 0 | N/A | 14 | 66.7% | 21 |
| | Male | 3 | 15.8% | 0 | N/A | 16 | 84.2% | 19 |
| 50-59 | Female | 10 | 45.5% | 0 | N/A | 12 | 54.5% | 22 |
| | Male | 14 | 51.9% | 0 | N/A | 13 | 48.1% | 27 |
| 60-69 | Female | 37 | 86.0% | 0 | N/A | 6 | 14.0% | 43 |
| | Male | 23 | 85.2% | 0 | N/A | 4 | 14.8% | 27 |
| 70-79 | Female | 31 | 96.9% | 0 | N/A | 1 | 3.1% | 32 |
| | Male | 32 | 86.5% | 0 | N/A | 5 | 13.5% | 37 |
| 80-89 | Female | 13 | 100.0% | 0 | N/A | 0 | N/A | 13 |
| | Male | 23 | 100.0% | 0 | N/A | 0 | N/A | 23 |
| Total | | 197 | 69.1% | 0 | N/A | 88 | 30.9% | 285 |
## Adult Subjects At High Risk For Viral Hepatitis:
Expected results of asymptomatic prospective high-risk subjects, determined from a multi-center study in the US and in Europe, are reported in the following tables. A total of 230 US subjects were at high risk for viral hepatitis including intravenous drug users (N=55), homosexual males (N=15), sex workers (N=39), prison history (N=92), high-risk sex partners (N=25), and high-risk occupation/health care workers (N=4). Many had more than 1 high-risk behavior or risk factor. Subjects were from Los Angeles, CA, (86.5%), Santa Ana, CA (4.3%), or Miami, FL (9.1%). The group was Caucasian (7.4%), Black or African American (74.3%), Hispanic or Latino (15.2%), Asian (0.4%), Native Hawaiian or other Pacific Islander (0.4%), and American Indian or Alaska native (0.9%), with the remaining (1.3%) represented by multiple ethnic groups. Of these subjects, 81% were male and 19% were female, and they ranged in age from 18 to 70 years (mean age of 45). The data are reported in Table below. The percent of Anti-HAV reactive results with MONOLISA™ Anti-HAV EIA in this high-risk asymptomatic population was 53%.
The European group (N=62) was 87% male and 13% female, and ranged in age from 21 to 75 years (mean age of 40). It consisted of intravenous drug users (30), subjects who had clotting factor disorders (7) and MSM patients (25). The data are reported in Table 9.
The percent of Anti-HAV reactive results with MONOLISA™ Anti-HAV EIA in this high-risk asymptomatic population was 45%.
{15}
Expected results for MONOLISATM Anti-HAV EIA in the US High Risk Group for Viral Hepatitis A (N=230)
| MONOLISA™ Anti-HAV EIA | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Age Range | Gender | Reactive | | Borderline | | Nonreactive | | Total |
| | | N | % | N | % | N | % | |
| < 19 | Female | 1 | 100.0% | 0 | N/A | 0 | N/A | 1 |
| | Male | 1 | 100.0% | 0 | N/A | 0 | N/A | 1 |
| 20-29 | Female | 1 | 33.3% | 0 | N/A | 2 | 66.7% | 3 |
| | Male | 1 | 50.0% | 0 | N/A | 1 | 50.0% | 2 |
| 30-39 | Female | 4 | 57.1% | 0 | N/A | 3 | 42.9% | 7 |
| | Male | 12 | 33.3% | 0 | N/A | 24 | 66.7% | 36 |
| 40-49 | Female | 15 | 62.5% | 0 | N/A | 9 | 37.5% | 24 |
| | Male | 37 | 43.5% | 1 | 1.2% | 47 | 55.3% | 85 |
| 50-59 | Female | 6 | 85.7% | 0 | N/A | 1 | 14.3% | 7 |
| | Male | 31 | 60.8% | 1 | 2.0% | 19 | 37.3% | 51 |
| 60-69 | Female | 1 | 100.0% | 0 | N/A | 0 | N/A | 1 |
| | Male | 9 | 90.0% | 0 | N/A | 1 | 10.0% | 10 |
| 70-79 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 2 | 100.0% | 0 | N/A | 0 | N/A | 2 |
| Total | | 121 | 52.6% | 2 | 0.9% | 107 | 46.5% | 230 |
Expected results for MONOLISATM Anti-HAV EIA in the European High Risk Group for Viral Hepatitis A (N=62)
| MONOLISA™ Anti-HAV EIA | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Age Range | Gender | Reactive | | Borderline | | Nonreactive | | Total |
| | | N | % | N | % | N | % | |
| < 19 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| 20-29 | Female | 0 | N/A | 0 | N/A | 5 | 100.0% | 5 |
| | Male | 0 | N/A | 0 | N/A | 11 | 100.0% | 11 |
| 30-39 | Female | 1 | 50.0% | 0 | N/A | 1 | 50.0% | 2 |
| | Male | 5 | 35.7% | 0 | N/A | 9 | 64.3% | 14 |
| 40-49 | Female | 1 | 100.0% | 0 | N/A | 0 | N/A | 1 |
| | Male | 9 | 64.3% | 0 | N/A | 5 | 35.7% | 14 |
| 50-59 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 9 | 81.8% | 0 | N/A | 2 | 18.2% | 11 |
| 60-69 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 1 | 50.0% | 0 | N/A | 1 | 50.0% | 2 |
| 70-79 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 2 | 100.0% | 0 | N/A | 0 | N/A | 2 |
| >80 | Female | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| | Male | 0 | N/A | 0 | N/A | 0 | N/A | 0 |
| Total | | 28 | 45.2% | 0 | N/A | 34 | 54.8% | 62 |
{16}
Page 17 of 18
N. Proposed labeling:
The labeling is sufficient and it satisfies the requirement of 21 CFR Part 809.10.
WARNINGS and PRECAUTIONS:
For in vitro diagnostic use only
1. The MONOLISA™ Anti-HAV EIA contains human source material used in the preparation of Negative Control (C0), Positive Control (C1) and Calibrator (C2) that has been tested with either FDA or CE approved methods and found non-reactive for Hepatitis B surface antigen (HBsAg), antibodies to Hepatitis C Virus (HCV) and antibodies to Human Immunodeficiency Viruses (HIV-1 and HIV-2).
2. The HAV Viral Antigen reagent (R6) has been treated with formalin to inactivate the virus.
3. No known test method can offer complete assurance that infectious agents are absent. Therefore, all human blood derivatives, reagents and human specimens should be handled as if capable of transmitting infectious disease. It is recommended that reagents and human specimens be handled in accordance with the OSHA Standard on Bloodborne Pathogens. Biosafety Level 2 or other appropriate biosafety practices should be used for materials that contain or are suspected of containing infectious agents.
4. The following is a list of potential chemical hazards contained in some kit components (See section 4: REAGENTS):
4.1 ProClin™ 300 (0.1% and 0.25%) are biocidal preservatives that may cause sensitization by skin contact; prolonged or repeated exposure may cause allergic reaction in certain sensitive individuals.
4.2 The 1N Sulfuric Acid (H₂SO₄) Stopping Solution is irritating to skin and severely irritating or corrosive to eyes, depending on the amount and length of exposure; greater exposures can cause eye damage, including permanent impairment of vision. In case of contact with eyes, rinse immediately with plenty of water and seek medical advice. Keep away from strong bases, reducing agents and metals; do not pour water into this component. Waste from this material is considered hazardous acidic waste. However, if permitted by local, regional, and national regulations, it can be neutralized to pH 6-9 for non-hazardous disposal if operators are trained and equipped to do so.
4.3 Sodium azide (< 0.1%), a biocidal preservative, may be detrimental if enough is ingested. Sodium azide may react with certain metals, including lead or copper often found in plumbing, to form highly explosive metal azides. Flush with copious amounts of water when pouring dilute solutions down the drain to prevent explosive build-up.
5. Biological spills: Human source material spills should be treated as potentially infectious. Spills not containing acid should be immediately decontaminated, including the spill area, materials, and any contaminated surfaces or equipment, with an appropriate chemical disinfectant that is effective for the potential
{17}
Page 18 of 18
biohazards relative to the samples involved (commonly a 1:10 dilution of bleach, 70-80% Ethanol or Isopropanol, an iodophor [such as 0.5% Wescodyne™ Plus], or a phenolic, etc.) and wiped dry.¹⁸⁻²⁰
6. Spills containing acid should be appropriately absorbed or neutralized, and wiped dry. The area should be decontaminated with an appropriate agent. Materials used to absorb the spill should be disposed of as biohazardous waste.
NOTE: DO NOT PLACE SOLUTIONS CONTAINING BLEACH INTO THE AUTOCLAVE.
7. Dispose of all specimens and material used to perform the test as though they contain an infectious agent. Laboratory chemical or biohazardous wastes must be handled and discarded in accordance with all local, regional and national regulations.
Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.