K063305 · Medtronic Xomed, Inc. · ETN · Nov 29, 2006 · Ear, Nose, Throat
Device Facts
Record ID
K063305
Device Name
STIMULATING BUR GUARD
Applicant
Medtronic Xomed, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Nov 29, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
When used with both the Medtronic Nerve Integrity Monitor (NIM) and XPS Drill Systems, the Stimulating Bur Guard is intended to stimulate cranial and peripheral motor nerves, including spinal nerve roots, with a standard bur / blade for location and identification during the incision and removal of soft and hard tissue or bone. The Stimulating Bur Guard is indicated for nerve monitoring during the incision and removal of soft and hard tissue or bone with a standard bur / blade during otology, neurotology, sinus, laryngeal, nasopharyngeal, head and neck, general and plastic, and orthopedic surgical procedures, including spinal applications.
Device Story
Stimulating Bur Guard is a sterile, single-use accessory for XPS 3000 Drill System and Medtronic Nerve Integrity Monitor (NIM). Device locks onto drill handpiece; features anterior opening to accept standard surgical bur/blade. Principle of operation: delivers electrical stimulating current from NIM console through the guard to the bur/blade tip to locate/identify cranial, peripheral, and spinal motor nerves during tissue/bone removal. Used in OR by surgeons. Output is electrical stimulation of nerves; allows surgeon to identify nerve proximity during drilling, preventing accidental nerve damage. Benefits patient by reducing risk of iatrogenic nerve injury during complex surgical procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included reliability, useful life, bur/blade compatibility, current delivery, noise, temperature, current drop out, and electrical isolation. Biocompatibility testing performed per ISO 10993-1 and FDA G95-1. EMC and electrical safety testing conducted per EN 60601-1-2 and EN/IEC 60601-1.
Technological Characteristics
Sterile, single-use, monopolar nerve stimulator accessory. Materials biocompatible for limited duration (<24 hrs) contact. Connects to NIM console for energy delivery. Dimensions/form factor designed to lock onto XPS 3000 handpiece. Sterilization via ETO. No software or firmware components.
Indications for Use
Indicated for nerve monitoring during incision/removal of soft/hard tissue or bone using a standard bur/blade. Patient population includes those undergoing otology, neurotology, sinus, laryngeal, nasopharyngeal, head and neck, general, plastic, or orthopedic surgical procedures, including spinal applications.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
K063305
## 510(k) Summary
NOV 2 9 2006
| 510(k) Owner | Medtronic Xomed, Inc<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216-0980 USA<br>904-296-9600<br>904-296-2386 (FAX) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name | Antoine Kouchakjy<br>Senior Regulatory Affairs Specialist<br>Medtronic Xomed, Inc |
| Date Summary Prepared | October 31, 2006 |
| Proprietary Name | Stimulating Bur Guard (Final name TBD) |
| Common Name | Stimulator, Nerve |
| Classification Name | Primary · Surgical nerve stimulator / locator<br>(21 CFR 874.1820, Product Code ETN, Class II)<br>Secondary · Drill, Surgical, ENT (electric or pneumatic) including<br>handpiece<br>(21 CFR 874.4250, Product Code ERL, Class II)<br>Secondary · Surgical instrument motors & accessories / attachments<br>(21 CFR 878.4820, Product Code HWE, Class I) |
### Marketed device claiming equivalence to:
The Stimulating Bur Guard is equivalent to the Medtronic Xomed Monopolar Stimulator Probe, K992869. The guard is also an accessory to the XPS 3000 Drill System covered under K041523.
### Device Description
The Stimulating Bur Guard is a sterile, single use device used in conjunction with the XPS 3000 BF Powered Drill System and the Medtronic Xomed Nerve Monitoring System. It locks onto a handpiece and has an opening on the anterior end to accept a standard bur / blade. The primary purpose of the Stimulating Bur Guard is to deliver the stimulating current from the Nerve Monitoring console to the standard bur / blade on the drill handpiece. The second purpose of the Stimulating Bur Guard is to provide support for the extended length burs.
#### Intended Use
When used with both the Medtronic Nerve Integrity Monitor (NIM) and XPS Drill Systems, the Stimulating Bur Guard is intended to stimulate cranial and peripheral motor nerves, including spinal nerve roots, with a standard bur / blade for location and identification during the incision and removal of soft and hard tissue or bone.
# Page 9 of 298
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten alphanumeric string. The string is "K063305". The characters are written in a cursive style, with some connections between the numbers.
### Indications for Use
The Stimulating Bur Guard is indicated for nerve monitoring during the incision and removal of soft and hard tissue or bone with a standard bur / blade during otology, neurotology, sinus, laryngeal, nasopharyngeal, head and neck, general and plastic, and orthopedic surgical procedures, including spinal applications.
| Characteristic | New Device:<br>Medtronic Xomed<br>Stimulating Guard | Predicate:<br>Medtronic Xomed Ball-Tip<br>Monopolar Stimulating<br>Probe [K992869] |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | When used with both the Medtronic<br>Nerve Integrity Monitor (NIM) and<br>XPS Drill Systems, the Stimulating<br>Guard is intended to stimulate cranial<br>and peripheral motor nerves,<br>including spinal nerve roots, with a<br>standard bur / blade for location and<br>identification during the incision and<br>removal of soft and hard tissue or<br>bone. | To stimulate cranial and<br>peripheral motor nerves for<br>location and identification<br>during surgery, including<br>spinal nerve roots |
| Insulated | Yes | Yes |
| Stimulator connector | Yes | Yes |
| Biocompatible | Yes | Yes |
| Sterile | Yes | Yes |
| Sterilization Method | ETO | ETO |
| Shelf Life | 3 months (initially) | 8-year |
| Single Use | Yes | Yes |
| Monopolar | Yes | Yes |
### Summary of Technological Characteristics
### Biocompatibility
The Stimulating Guard is considered biocompatible for use as an external communicating device with tissue / bone contact of limited duration (< 24 hrs). The appropriate tests were performed on all patient contacting materials according to:
- ISO 10993-1:2003, Biological evaluation of medical devices Part 1 Evaluation . and Testing
- FDA G95-1, Required Biocompatibility Training and Toxicology Profiles for . Evaluation of Medical Devices, May 1, 1995
### Shelf Life Aging
Accelerated aging studies for the Stimulating Guard will ensure three (3) months shelf life prior to a limited market release and a minimum shelf life of one (1) year at the time of full market release. Shelf life testing will be ongoing post-launch to ultimately achieve a four (4) year shelf life.
# Page 10 of 298
{2}------------------------------------------------
# KOG 3305
### Operating Life
The device is provided sterile and intended for single use. No reprocessing instructions are given. The device met the acceptance criteria for worst case use during a surgical procedure.
### Bench Testing
The device met the acceptance criteria for: Reliability and Useful life, Bur / blade Compatibility, Current delivery, Current delivery compared to 2.3mm ball tip probe, Noise, Temperature, Current drop out, Electrical Isolation
### Electromagnetic Compatibility and Electrical Safety Testing
The Stimulating Guard is not a source of electromagnetic interference in and of itself. However, the devices are evaluated for EMC as a system with the nerve monitoring and drill system consoles in accordance with EN 60601-1-2. Electrical safety testing is conducted in accordance with EN / IEC 60601-1 and electrical safety is achieved through levels of protection built within the BF Rated Consoles. Both the XPS 3000 and NIM systems are BF rated.
### Software
The device does not contain any software nor does it require any software changes to be made to the XPS Drill System or the Nerve Monitoring System (NIM)
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medtronic Xomed, Inc. c/o Antoine Kouchakiy Senior Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville, Florida 32216-0980
NOV 2 9 2006
Re: K063305
Trade/Device Name: Stimulating Bur Guard Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: October 31, 2006 Received: November 1, 2006
Dear Mr. Kouchakjy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2 - Antoine Kouchakjy
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eychlemin IWD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
Kolo 3305 510(k) Number (if known):
Stimulating Bur Guard (Final name to be determined) Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
When used with both the Medtronic Nerve Integrity Monitor (NIM) and XPS Drill Systems, the When used with bour the modified North late cranial and peripheral motor nerves, including Stimulating Dur Guard is meenated bur / blade for location and identification during the incision and removal of soft and hard tissue or bone.
The Stimulating Bur Guard is indicated for nerve monitoring the incision and removal of The Stimulating Dur Guard is marcated bur / blade during otology, neurotology, sinus, Son and hard though of oche wad neck, general and plastic, and orthopedic surgical procedures, including spinal applications.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kenneth Baker
510(k) Number K063305
Prescription Use
(Per 21 CFR 801.109) X
Page 8 of 298
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.