MICROSCAN MICROSTREP PLUS PANEL-CLARITHROMYCIN

K063299 · Dade Behring, Inc. · LRG · Nov 27, 2006 · Microbiology

Device Facts

Record IDK063299
Device NameMICROSCAN MICROSTREP PLUS PANEL-CLARITHROMYCIN
ApplicantDade Behring, Inc.
Product CodeLRG · Microbiology
Decision DateNov 27, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

To determine bacterial antimicrobial agent susceptibility

Device Story

MicroScan® MICroSTREP plus® Panel is a 96-well plastic dish containing dehydrated microdilutions of antimicrobics. Panels are rehydrated with Mueller-Hinton broth supplemented with 2-5% lysed horse blood and inoculated with standardized organism suspension (approx. 5 × 10⁵ CFU/mL). After 20-24 hours incubation at 35°C, growth is assessed. The device adds automated reading capability via the MicroScan® WalkAway System, which uses optics light detection to measure growth in wells. The system determines the Minimum Inhibitory Concentration (MIC) by identifying the lowest concentration showing growth inhibition. Results are used by clinicians to determine antimicrobial susceptibility of aerobic streptococci, aiding in appropriate antibiotic selection for patients.

Clinical Evidence

Bench testing only. Reproducibility study (10 isolates, 3 sites, 3 days, triplicate) showed >95% reproducibility for both manual and automated methods. Comparison study of 70 challenge isolates (including 53 S. pneumoniae) demonstrated 98.6% Essential Agreement (EA) for the automated method compared to reference broth dilution, with no very major discrepancies and acceptable minor/major error rates. No clinical efficacy studies were required as performance was established via manual reading in the predicate submission.

Technological Characteristics

Miniaturized broth microdilution panel. Rehydrated with Mueller-Hinton broth, 2-5% lysed horse blood, and 50 mM HEPES. Incubation: 20-24 hours at 35°C +/- 1°C. Readout: Manual visual or automated via MicroScan® WalkAway instrument. Connectivity: Integrated with WalkAway system.

Indications for Use

Indicated for testing Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococci (Groups C,F, and G) and viridans group streptococci. Prescription use only; turbidity method of inoculum preparation required.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} Page 1 of 6 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K063299 B. Purpose for Submission: To add the option for automated reading of Clarithromycin at 0.06—2 µg/mL to the MICroSTREP plus® Panel on the MicroScan® WalkAway System C. Measurand: Clarithromycin at 0.06—2 µg/mL D. Type of Test: Quantitative and Qualitative growth based detection algorithm using optics light detection E. Applicant: Dade Behring Inc, MicroScan® F. Proprietary and Established Names: MicroScan® MICroSTREP plus® Panel – Clarithromycin at 0.06—2 µg/mL G. Regulatory Information: 1. Regulation section: 21 CFR 866.1640 – Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product Code: LRG – Instrument for Auto Reader &amp; Interpretation of Overnight Antimicrobial Susceptibility System LTT – Panels, Test, Susceptibility, Antimicrobial 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): Clarithromycin at 0.06—2 µg/mL is for use with MICroSTREP plus® Panels. MICroSTREP plus® Panels are designed for use in determining quantitative {1} Page 2 of 6 and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including *Streptococcus pneumoniae*. 2. **Indication(s) for use:** This submission is for adding the option for automated reading of the antibiotic Clarithromycin at concentrations of 0.06 — 2 µg/mL to MICroSTREP plus® Panels on the MicroScan® WalkAway System for testing *Streptococcus pneumoniae*, *Streptococcus pyogenes*, *Streptococcus agalactiae*, *Streptococci* (Groups C,F, and G) and viridans group streptococci. 3. **Special condition for use statement(s):** Prescription Use Only Turbidity method of inoculum preparation only 4. **Special instrument Requirements:** Not Applicable I. **Device Description:** The MicroScan® MICroSTREP plus® Panel is a 96-well plastic dish which contains microdilutions of each antimicrobic in various concentrations dried in aqueous solutions. The panel is rehydrated and inoculated at the same time with a Mueller-Hinton broth supplemented with lysed horse blood (2 – 5%). The target inoculum concentration for each well should be approximately 5 × 10⁵ colony forming units (CFU)/mL. Panels are incubated in a 35°C non-CO₂ incubator for 20-24 hours. After incubation, the panels are read manually for growth. Additionally, panels may be incubated in and read by a MicroScan® WalkAway instrument. Each panel contains a “growth” but it does not contain a “no growth” control well. J. **Substantial Equivalence Information:** 1. **Predicate device name(s):** MICroSTREP plus® 2. **Predicate K number(s):** K021181 3. **Comparison with predicate:** | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Determination of susceptibility to antimicrobics with aerobic streptococci including *Streptococcus pneumoniae* | Same | | Isolates | For use with aerobic streptococci including *Streptococcus pneumoniae* isolated colonies from culture | Same | {2} | Results | Quantitative with qualitative interpretations | | Same | | --- | --- | --- | --- | | Incubation | 20 – 24 hours | | Same | | Panels | Clarithromycin dried in aqueous solution | | Same | | Differences | | | | | Item | Device | | Predicate | | Technology | Growth based using algorithm with optics light detection | | Growth based | | Reading | Overnight method Manual or automated | | Overnight method Manual read only | | Instrument | MicroScan® WalkAway System or Microdilution viewer | | Microdilution viewer | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S16) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." # L. Test Principle: The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with $115~\mu \mathrm{L}$ Mueller-Hinton broth supplemented with $2 - 5\%$ lysed horse blood (LHB), after inoculation of the broth with a standardized suspension of the organism. The target inoculum concentration for each well should be approximately $5\times 10^{5}$ colony forming units (CFU)/mL. After incubation in a non- $\mathrm{CO}_{2}$ incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organisms is determined by observing the lowest antimicrobial concentration showing inhibition of growth. Panels can be read manually using indirect light or the panels can be read on the MicroScan® WalkAway instrument using optics light detection. # M. Performance Characteristics (if/when applicable): # This submission is for the AST Panel only. The ID System was not reviewed. The Reproducibility studies, QC performance data, and Challenge isolates evaluated by the manual and automated reading methods are required to demonstrate that there is no difference between manual reading and automated reading in the MicroScan® WalkAway System. The clinical efficacy performance was previously established using the manual read method and was therefore not required for this submission. {3} Page 4 of 6 1. Analytical performance: a. Precision/Reproducibility: Reproducibility was demonstrated using 10 isolates tested at 3 sites on 3 separate days in triplicate. The study included testing on the MicroScan® WalkAway System with automated reading at 20-24 hours, and manual readings at 20-24 hours incubation. Both reading methods demonstrated &gt;95% reproducibility, and no differences were observed. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or method): The recommended QC isolate S. pneumoniae ATCC 49619 was tested a sufficient number of times with acceptable results on &gt;95% of the testing days with the reference method. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results on all testing days. The following table provides the frequency of results in each concentration with the expected range stated. Both reading methods produced the same mode. | Organism | Concentration μg/mL | Reference results | MicroScan® WalkAway results | | | --- | --- | --- | --- | --- | | | | | Manual Overnight | Instrument Overnight | | S. pneumoniae ATCC 49619 Expected range 0.03 – 0.12 μg/mL | <= 0.015 | 3 | | | | | 0.03 | 60 | 1 | 2 | | | 0.06 | 2 | 43 | 41 | | | 0.12 | | 21 | 22 | | | 0.25 | | | | | | 0.5 | | | | | | 1 | | | | Inoculum density control: A turbidity meter, which was verified each day of testing, was used for the turbidity inoculation method. Colony counts were performed weekly, on the ATCC S. pneumoniae with all results in the expected range of approximately 5 x 10⁵. The mode for the test device was slightly more resistant than the mode for the reference method results, if only by one dilution. d. Detection limit: Not applicable {4} Page 5 of 6 e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Clinical efficacy testing with manual result reading was conducted in the previous submission (K021181). In this submission, Challenge isolates were evaluated by the manual and automated reading methods to demonstrate that there is no difference between manual reading and instrument reading on the MicroScan® WalkAway System. There were 70 challenge isolates including 53 Streptococcus pneumoniae CDC Challenge Set strains tested at one site and compared to the reference broth dilution result. A comparison was done with readings on the instrument after 20 hours incubation, and also read manually when incubated for 20-24 hours. Performance by the automated reading method was acceptable with no differences. Clarithromycin was tested in doubling dilutions ranging from 0.015 – 2 µg/ml. Data are presented and clearance is requested for 0.06 – 2 µg/ml (mapped dilutions) on the MICroStrep plus® panel. The recommended CLSI reference method was followed with the exception of the use of a small amount (0.1%) of Pluronic (a wetting agent) in the final inoculum. A validation of the use of Pluronic in the frozen reference panel was conducted. QC was also performed with no difference apparent in the results. Read method comparison of Streptococcus species and Clarithromycin – Mapped Dilution Sequence | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Challenge Manual | 70 | 68 | 97.1 | 12 | 10 | 83.3 | 30 | 3 | 1 | 0 | | Challenge Automated | 70 | 69 | 98.6 | 14 | 13 | 92.9 | 30 | 3 | 1 | 0 | EA-Essential Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies Essential agreement (EA) is when the Microscan® MICroSTREP plus® panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Evaluable (Eval) are results that are within the test range and on scale. {5} Page 6 of 6 Automated reading results were the same as the manual reading results with no differences. Because of the limited amount of on-scale results, trending could not be determined. There were no vmj, and there were no additional minor errors or maj errors generated by the automated reading method. The overall EA% of 98.6% and Eval EA% of 92.9% for the automated reading were both acceptable. The test device had a growth rate of &gt;95% for both the manual reading and the automated reading methods. The comparison of the reading methods demonstrates that the manual reading method and the automated reading on the MicroScan® WalkAway System are no different. The efficacy data performed with the manual reading method would therefore be expected to have no differences. The performance data currently documented in the package insert will not change. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable d. Clinical cut-off: Not applicable 4. Expected values/Reference range: Streptococcus species Interpretive criteria: &lt;= 0.25 (Susceptible), 0.5 (Intermediate), &gt;= 1 (Resistant) The expected value range, interpretive criteria and QC are included in the package insert. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...