ACON UTI URINARY TRACT INFECTION TEST STRIPS

K063295 · ACON Laboratories, Inc. · LJX · Oct 12, 2007 · Hematology

Device Facts

Record IDK063295
Device NameACON UTI URINARY TRACT INFECTION TEST STRIPS
ApplicantACON Laboratories, Inc.
Product CodeLJX · Hematology
Decision DateOct 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7675
Device ClassClass 1

Indications for Use

The ACON UTI Urinary Tract Infection Test Strips is for the qualitative detection of Leukocyte and Nitrite in urine as an aid in the screening of urinary tract infection (UTI). It is intended for over-the-counter home use only.

Device Story

Disposable plastic test strips with reagent pads for Nitrite and Leukocyte detection; used by lay persons at home. Principle: Nitrite test converts nitrate to nitrite via Gram-negative bacteria; nitrite reacts with p-arsanilic acid and 1 N-(1-naphthyl)-ethylenediamine to form pink color. Leukocyte test detects granulocyte esterases; esterases cleave derivatized pyrazole amino acid ester to liberate hydroxy pyrazole, which reacts with diazonium salt to produce beige-pink to purple color. User compares strip color to bottle label blocks. No instrumentation required. Results aid in UTI screening; positive nitrite indicates presence of nitrate-reducing bacteria; positive leukocyte indicates granulocyte esterase activity. Benefits include rapid, non-invasive home screening for potential infection.

Clinical Evidence

Clinical studies conducted at Beta sites evaluated accuracy of lay users compared to professionals using the predicate device. Results indicated lay users obtained comparable testing data to professionals. Performance verified via sensitivity, reproducibility, interference, temperature flex, read time flex, and stability studies.

Technological Characteristics

Plastic test strips with nitrite and leukocyte reagent pads. Nitrite detection: p-arsanilic acid/1-N-(1-naphthyl)-ethylenediamine coupling. Leukocyte detection: derivatized pyrazole amino acid ester/diazonium salt reaction. Qualitative, visual read. Standalone, non-electronic. Shelf life: 24 months (accelerated).

Indications for Use

Indicated for symptomatic individuals (e.g., pain/burning during urination, frequency, fever, back/groin pain, cloudy/foul-smelling urine) for qualitative screening of urinary tract infections via detection of Nitrite and Leukocytes in urine.

Regulatory Classification

Identification

A leukocyte peroxidase test is a device used to distinguish certain myeloid cells derived from the bone marrow, i.e., neutrophils, eosinophils, and monocytes, from lymphoid cells of the lymphatic system and erythroid cells of the red blood cell series on the basis of their peroxidase activity as evidenced by staining. The results of this test are used in the differential diagnosis of the leukemias.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k063295 B. Purpose for Submission: New device C. Measurand: Urinary Nitrite and Leukocytes D. Type of Test: Qualitative E. Applicant: ACON Laboratories, Inc. F. Proprietary and Established Names: ACON UTI Urinary Tract Infection Test Strips G. Regulatory Information: 1. Regulation section: 21 CFR §862.1510, Nitrite (nonquantitative) test system 21 CFR §864.7675, Leukocyte peroxidase test 2. Classification: Class I (meets the limitations of exemptions in 21 CFR 862.9 (c)(9) 3. Product code: LJX JMT 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See Indications for use below. 2. Indication(s) for use: The ACON UTI Urinary Tract Infection Test Strips are for the qualitative detection of Nitrite and Leukocytes in urine as an aid in the screening of urinary tract infection (UTI). It is intended for over-the-counter home use only. {1} 3. Special conditions for use statement(s): For over-the-counter use. In the limitations section of the labeling, there are statements such as, “There is the possibility that this test may produce false results”, “You may get a negative result if you have a UTI caused by bacteria that does not change nitrate to nitrite, when urine has not been held in the bladder for more than 4 hours, when taking antibiotics, or when your diet does not include nitrates” and “Consult your physician before making any medical decisions.” 4. Special instrument requirements: None required; this is a self-contained visually read device. I. Device Description: ACON UTI Urinary Tract Infection Test Strips are plastic strips with a nitrite and a leukocyte reagent pad attached. They are packaged in a plastic bottle that contains desiccant and a separately packaged color chart. The reagent pads react with urine and change color. Test results are obtained by comparing the test strip to the color blocks printed on the color chart. The presence of nitrites in urine will turn the nitrite test pad pink to red. The presence of leukocytes will turn the leukocyte pad a shade between beige-pink and dark purple. J. Substantial Equivalence Information: 1. Predicate device name(s): Bayer Multistix 10 SG Reagent Strips for Urinalysis 2. Predicate 510(k) number(s): k905396 3. Comparison with predicate: The predicate and the proposed device have the following similarities: assay format, test methodology, and assay matrix. The assays differ in their intended use setting (professional use in point-of care setting versus over-the-counter) and their intended use (the predicate device also detects additional analytes that may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance in addition to information about urinary tract infections). K. Standard/Guidance Document Referenced (if applicable): None referenced. L. Test Principle: Nitrite: Gram negative bacteria in urine convert nitrate to nitrite. In an acidic medium, nitrite reacts with p-arsanilic acid to form a diazonium compound. The diazonium compound in turn couples with 1-N-(1-naphthyl)-ethylenediamine to produce a pink color. Nitrite is not detectable in normal urine therefore a consistent pink to red color evidences nitrite in urine. {2} Leukocytes: This test reveals the presence of granulocyte esterases in urine. If esterases are present, they cleave a derivatized pyrazole amino acid ester to liberate derivatized hydroxy pyrazole. This pyrazole reacts with a diazonium salt to produce a beige-pink to purple color. Normal urine is usually negative for leukocyte esterases. Therefore, a color change to +, ++, or +++ may have clinical significance; a color change to 'trace' (±) is considered clinically negative for leukocytes but may have clinical significance if the user gets repeated trace results. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: **a. Precision/Reproducibility:** Reproducibility of ACON UTI Test Strips was tested by professional users using a negative control solution and a moderately positive control solution (positive for nitrites, 2+ for leukocytes). Three lots of strips were used. Each control was tested in duplicate by five operators per day for ten days (total strips tested = 600). Results were read by comparing color reaction to the color blocks on the canister label according to the package insert instructions. Test results agreed with the expected results each day for each lot tested at both levels tested. The ability of lay users to read the test results is demonstrated in Section 3.c. "Other clinical supportive data" below. **b. Linearity/assay reportable range:** Not applicable. These tests are intended for qualitative use. **c. Traceability, Stability, Expected values (controls, calibrators, or methods):** No traceability was provided. Stability testing protocol and acceptance criteria were provided. The sponsor's accelerated stability testing suggests that the strips will have a 24 month shelf life when stored unopened at room temperature. Real time stability testing is currently underway. **d. Detection limit:** See Assay Cutoff section below. **e. Analytical specificity:** Potentially interfering endogenous and exogenous substances at the concentrations in the table below were added to normal urine, urine spiked with leukocyte positive urine (125 Leu/uL), and nitrite positive urine (1.0 mg/dL). All tests with normal (negative) urine samples returned negative results except that sodium nitrite produced a positive result for the nitrite test. All tests with positive urine samples returned positive results except albumin, ascorbic acid, glucose, sodium bicarbonate and riboflavin. These compounds were tested further as described below. 3 {3} ACON UTI Test Strips: Compounds Tested for Potential Interference | Potential Interferent | Tested Concentration (mg/dL) | | Potential Interferent | Tested Concentration (mg/dL) | | --- | --- | --- | --- | --- | | Albumin | 1000 | | Lactose | 1 | | Ammonium Chloride | 200 | | Lithium Acetoacetate | 250 | | Ascorbic Acid | 200 | | Oxalic Acid | 70 | | Bilirubin | 4 | | Phenolphthalein | 1200 | | Calcium Chloride | 80 | | Potassium Chloride | 1200 | | Citric Acid | 65 | | Riboflavin | 10 | | Creatinine | 600 | | Sodium Chloride | 1800 | | Fructose | 1.2 | | Sodium Nitrate | 0.3 | | Galactose | 0.5 | | Sodium Nitrite | 0.1 | | Glucose | 4100 | | Sodium Phosphate | 500 | | Glycine | 450 | | Theophylline | 100 | | Hemoglobin | 5 | | Urea | 4000 | A series of experiments were performed to determine if common urine components, including the compounds that caused positive test results in the study above, would affect the performance of the ACON UTI test strips. The components tested with negative urine, samples spiked with nitrite at 0.05, 0.1, and $1.0\mathrm{mg / dL}$, and samples spiked with leukocytes at 15, 70, 125, and 500 Leu/uL. The concentrations of the interferents that caused positive test results above are generally much higher than would be found in urine of typical users. The labeling includes cautions that ascorbic acid and riboflavin may interfere with the test. The results of these studies are summarized in the following table: Interferents: ACON UTI Test Strips | Interferent | Effect on Nitrite Test | | Interferent | Effect on Leukocyte Test | | --- | --- | --- | --- | --- | | Specific Gravity ≥ 1.030 | 2/5 positive at 0.05 mg/dL | | Specific Gravity ≥ 1.025 | negative results at 15 cells/uL | | Ascorbic Acid ≥ 30 mg/dL | 0/5 positive at 0.05 mg/dL | | Ascorbic Acid | no effect at tested concentrations | {4} | Interferent | Effect on Nitrite Test | | Interferent | Effect on Leukocyte Test | | --- | --- | --- | --- | --- | | pH | no effect at tested concentrations | | pH ≥ 8.0 | 1 of 5 tests read positive at 15 cells/uL | | Riboflavin ≥ 10 mg/dL | 0/5 positive at 0.05 mg/dL | | Riboflavin ≥ 10 mg/dL | 1/5 negative at 15 cells/uL | | Sodium Bicarbonate ≥ 1000 mg/dL | 0/5 positive at 0.05 mg/dL | | Sodium Bicarbonate | no effect at tested concentrations | | Glucose | no effect at tested concentrations | | Glucose ≥ 2000 mg/dL | lower than expected readings at all tested concentrations | | Albumin | no effect at tested concentrations | | Albumin ≥ 1000 mg/dL | 0/5 positive at 0.05 mg/dL | The effect of the temperature of the test solution, the length of time the strip was in contact with the urine sample ('dip time') and the test read time after dipping were also evaluated. Calibrator solutions (negative, low positive and strongly positive) were used in the tests. Storage of the calibrator solutions at temperatures ranging from $4^{\circ}\mathrm{C}$ to $45^{\circ}\mathrm{C}$ for 30 minutes before using them to test the strips did not affect the strip's detection of negative or positive samples. Dip times of the recommended 2 seconds up to 10 seconds did not alter the expected result of the test at any read time tested (15 seconds to 30 minutes). Dip times of less than two seconds only affected the test result when the leukocyte test was read at 15 or 30 seconds (recommended read time is two minutes). Read times greater than five minutes caused the negative nitrite and leukocyte tests to read 'positive' and the positive leukocytes to read one color block higher than expected. The labeling instructs users to read the nitrite test at 1 minute and the leukocyte test at 2 minutes but not to read either test beyond 3 min. # f. Assay cut-off: The sensitivity of each test was assessed by testing a range of urine samples spiked to known concentrations. The sponsor defined the minimum sensitivity level as the lowest level at which over $67\%$ of the test results show a color change equivalent to the first color block on the color chart, i.e a uniform pink for the nitrite test and the beige-pink of the trace block for the leukocyte test. Three lots of strips were tested. For three consecutive days each sample was tested 21 times by three operators (7 strips per operator per day). A total of 189 strips were tested at each concentration (3 operators x 3 days x 7 strips x 3 lots). Results were read by comparing color reactions on the strip to the color chart according to the package instructions. Results are shown in the tables below: {5} Nitrite Sensitivity: ACON UTI test strips | Nitrite Conc. (mg/dL) | Negative (yellow) | Positive (pink) | % Positive | | --- | --- | --- | --- | | 0.025 | 189 | 0 | 0 | | 0.05 | 3 | 186 | 98.4 % | | 0.06 | 0 | 189 | 100 % | | 0.08 | 0 | 189 | 100 % | | 0.1 | 0 | 189 | 100 % | Leukocyte Sensitivity: ACON UTI test strips | Leukocyte Conc. (Cells/uL) | 0 Cells/uL | Trace * (15Cells/uL) | + (70Cells/uL) | ++ (125Cells/uL) | % Agreement | | --- | --- | --- | --- | --- | --- | | 3.75 | 189 | 0 | 0 | 0 | 0 | | 7.5 | 86 | 103 | 0 | 0 | 54 | | 9 | 8 | 181 | 0 | 0 | 96 | | 12 | 0 | 189 | 0 | 0 | 100 | | 15 | 0 | 189 | 0 | 0 | 100 | * "Trace" in the ACON UTI strip is claimed as equivalent to 9 – 15 cells/uL; "trace" in the Bayer predicate is claimed as equivalent to 5- 15 cells/uL. In clinical interpretation of both tests a reading of "Trace" is negative for leukocytes. The labeling instructs users to read the test as positive when the leukocyte result is $\geq +$ . The sponsor claims a minimum analytical detection limit of $0.05\mathrm{mg / dL}$ for the nitrite test and 9 - 15 cells/uL for the leukocyte test. 2. Comparison studies: a. Method comparison with predicate device: See clinical studies section below. b. Matrix comparison: Not applicable; this device is for use with urine samples only. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): One hundred symptomatic* inexperienced lay users tested their own freshly collected urine specimens with the ACON UTI Urinary Tract Infection Test Strips. Their results were compared to results obtained by laboratory {6} personnel using the ACON UTI Urinary Tract Infection Test Strips and the predicate tests. Results are shown below: Leukocyte Results: ACON/Lay User vs. ACON/Professional | | Professional | | | | | | --- | --- | --- | --- | --- | --- | | Lay User | Neg | Trace (±) | + | ++ | +++ | | Neg | 70 | | | | | | Trace | | 4 | | | | | + | | 2 | 8 | 2 | | | ++ | | | | 12 | | | +++ | | | | | 2 | | % Exact Agreement | 96% (96/100) | | | | | | % Agreement(±One Color Block) | 100% (100/100) | | | | | Nitrite Results: ACON/Lay User vs. ACON/Professional | | Professional | | | --- | --- | --- | | Lay User | Nitrite (-) | Nitrite (+) | | Nitrite (-) | 96 | 0 | | Nitrite (+) | 0 | 4 | | % Exact Agreement | 100% (100/100) | | | % Agreement (± One Color Block) | 100% (100/100) | | Leukocyte Results: ACON/Professional vs. Bayer/Professional | | Bayer | | | | | | --- | --- | --- | --- | --- | --- | | ACON | Neg | Trace (±) | + | ++ | +++ | | Neg | 69 | 1 | | | | | Trace | | 6 | | | | | + | | | 0 | 8 | | | ++ | | | 1 | 10 | 3 | | +++ | | | | 1 | 1 | | % Exact Agreement | 86% (86/100) | | | | | | % Agreement(±One Color Block) | 100% (100/100) | | | | | Nitrite Results: ACON/ Professional vs. Bayer/Professional | | Bayer | | | --- | --- | --- | | ACON | Nitrite (-) | Nitrite (+) | | Nitrite (-) | 96 | 0 | | Nitrite (+) | 0 | 4 | | % Exact Agreement | 100% (100/100) | | {7} * UTI symptoms included at least one of the following: burning or pain during urination, frequent urination, fever, back or groin pain, or cloudy, dark or bloody urine, or bad smelling urine. There were no follow up culture results. Because there were so few nitrite positive samples in the clinical study, the sponsor recruited 40 lay-users to participate in a readability study with spiked urine specimens at different concentrations of leukocyte and nitrite: | Spiked Sample # | Spiked Concentrations | | | --- | --- | --- | | | Leukocyte | Nitrite | | 1 | Negative | Negative | | 2 | Trace | Negative | | 3 | + | 1^{st} positive color block | | 4 | +++ | 2^{nd} positive color block | Each participant tested a panel of 4 coded and blinded spiked urine samples with the ACON test using the package insert. The same panel of spiked urine samples was also tested at the same time by a professional with the ACON test and the predicate. The test results are summarized in the following tables. | ACON Test: Lay User vs. Expected | Spiked Sample #1 | | Spiked Sample #2 | | Spiked Sample #3 | | Spiked Sample #4 | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Leu (-) | Nit (-) | Leu (trace) | Nit (-) | Leu (+) | Nit (+) | Leu (3+) | Nit (+) | | % Agreement (Exactly Match) | 100% | 100% | 83% (33 “trace”, 4 “+” & 3 “-”) | 100% | 85%* (34 “+” & 6 “++”) | 100% | 98%* (39 “+++” & 1 “++”) | 100% | | % Agreement (± One Color Block) | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | * These disagreements of the readings do not result in a change of the interpretation of the test result. {8} 9 | ACON Test: Prof. vs. Expected | Spiked Sample #1 | | Spiked Sample #2 | | Spiked Sample #3 | | Spiked Sample #4 | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Leu (-) | Nit (-) | Leu (trace) | Nit (-) | Leu (+) | Nit (+) | Leu (3+) | Nit (+) | | % Agreement (Exactly Match) | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | | Predicate Test: Prof. vs. Expected | Spiked Sample #1 | | Spiked Sample #2 | | Spiked Sample #3 | | Spiked Sample #4 | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Leu (-) | Nit (-) | Leu (trace) | Nit (-) | Leu (+) | Nit (+) | Leu (3+) | Nit (+) | | % Agreement (Exactly Match) | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | | ACON Test: Lay User vs. Prof. | Spiked Sample #1 | | Spiked Sample #2 | | Spiked Sample #3 | | Spiked Sample #4 | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Leu (-) | Nit (-) | Leu (trace) | Nit (-) | Leu (+) | Nit (+) | Leu (3+) | Nit (+) | | % Agreement (Exactly Match) | 100% | 100% | 83% (33 "trace", 4 "+" & 3 "-") | 100% | 85%* (34 "+" & 6 "+") | 100% | 98%* (39 "+" & 1 "+") | 100% | | % Agreement (± One Color Block) | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | * These disagreements of the readings do not result in a change of the interpretation of the test result. | ACON vs. Bayer by Professional | Spiked Sample #1 | | Spiked Sample #2 | | Spiked Sample #3 | | Spiked Sample #4 | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Leu (-) | Nit (-) | Leu (trace) | Nit (-) | Leu (+) | Nit (+) | Leu (3+) | Nit (+) | | % Agreement (Exactly Match) | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 4. Clinical cut-off: Not applicable. {9} 5. Expected values/Reference range: Normal urine is usually negative for leukocytes and nitrites. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 10
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...