K063294 · Radiological Specialists, Inc. · LLZ · Jan 24, 2007 · Radiology
Device Facts
Record ID
K063294
Device Name
TOPACS
Applicant
Radiological Specialists, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 24, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Thoth ToPACS is an image management system whose intended use is to provide (scaleable ) DICOM compatible PACS solutions for hospital and related institutions/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR and other devices) and information systems. Lossy Compressed Mammographic Images and Digitized Film Screen Images must not be reviewed for primary Image interpretations. Mammographic Images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other specifications reviewed and accepted by FDA
Device Story
ToPACS is a standalone software-based Picture Archiving and Communication System (PACS) for hospitals and clinical institutions. It ingests medical images and data from various modalities (CR, CT, DR, MR) via DICOM protocols. The software provides tools for image processing, measurement, communication, and storage on general-purpose computing hardware. It is operated by healthcare professionals; the system does not contact the patient or control life-sustaining equipment. Physicians use the displayed output to interpret clinical information and make diagnostic decisions. The device facilitates clinical workflow by enabling the retrieval and distribution of diagnostic images across hospital sites. It benefits patients by providing timely access to medical imaging data for clinical review.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Standalone software package for general-purpose computing hardware. DICOM compatible. Functions include image processing, measurement, communication, and storage. Software-based; no specific hardware requirements other than general-purpose compatibility. Level of concern: minor.
Indications for Use
Indicated for use by licensed professionals in hospital and clinical settings to archive, distribute, retrieve, and display medical images and data from various modalities (CR, CT, DR, MR). Not for primary interpretation of lossy compressed mammographic images or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
### SECTION 9
### SUMMARY OF SAFETY AND EFFECTIVENESS
# 510(k) Summary
JAN 2 4 2007
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 ad 21CFR 807.92.
Lossy Compressed Mammographic Images and Digitized Film Screen Images must not be reviewed for primary Image Interpretations. Mammographic Images may only be interpreted using an FDA approved Monitor that offers at least 5 Mpixel resolution and meets other specifications reviewed and accepted by FDA
## I. General Information
### Establishment:
| | Radiological Specialists, Inc. | |
|----------|--------------------------------------------------|--|
| Address: | 5920 Noble Avenue, Van Nuys,<br>CA 91411, U.S.A. | |
| Phone: | (818) 908-9000 | |
| Fax: | (818) 908-9947 | |
### Registration Number:
Contact Person: Mr. Frank McMurray President, Phone: +1 (818) 908-9000 Fax: +1 (818) 908-9947 E Mail: radspecinc@aol.com
Date of Summary Preparation: July 27, 2006 Device Name:
> ●Trade Name: ToPACS
●Common Name: Picture Archiving Communications System
### ●Classification Number:
CFR 892.2050, System, Image Processing ●Classification: Class II
### ●Performance Standards:
None established under Section 514 the Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
## II. Safety and Effectiveness Information.
#### Device Description: ●
· Device Description:
ToPACS is a stand-alone software package that is used on general purpose computing I of ACO is a bains aronivimum hardware requirements are met, the user or system integrator is free to choose his/her own hardware platform micegrator is free to enouse minage processing, measurement, communication and The soltware and is alguing to the specifications that are documented in a Storage. 1011 System Test Plan. Testing is an integral part of Thoth software Thours Process. The software does not contact the patient, nor does it control any life-sustaining devices. A physician has ample opportunity for competent human intervention while interpreting images and clinical information.
#### Intended Use: ●
The Thoth ToPACS is an image management system whose intended use is to provide (scaleable ) DICOM compatible PACS solutions for hospital and related (Sections/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR and other devices) and information systems.
- Technological Characteristics (comparison with predicate device): .
- Predicated Device: . k030781: KODAK DirectView 5 k023460: PACSPLUS k031562: Ramsoft PACS
#### Statement of Substantial Equivalence: .
Thoth ToPACS is substantially equivalent to RamSoft PACS, PACSPLUS, and Kodak DirectView PACS. The determination of substantial equivalence is not based on an assessment of performance tests. It is our conclusion that there is no software component in ToPACS product or hardware component which would be used in conjunction with ToPACS product that we know of whose failure or latent design flaw would be expected to result in death or injury to a patient. Thus, the "Level of Concern" of ToPACS product is "minor."
#### General Safety and Effectiveness Concerns: ●
ToPACS does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo or emblem that appears to be associated with a health organization or government entity. The logo features a stylized graphic of three curved lines, possibly representing waves or abstract shapes, arranged vertically. To the left of the graphic, there is text that reads "MINISTRY OF HEALTH", arranged vertically along the curve of the graphic. The text is in a simple, sans-serif font and is oriented to be read from bottom to top.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Frank McMurray President Radiological Specialists, Inc. 5920 Noble Avenue VAN NUYS CA 91411
JAN 2 4 2007
Re: K063294
Trade/Device Name: ToPACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 1, 2006 Received: November 8, 2006
Dear Mr. McMurray: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendount microads commitered have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mail. therefore, market the device, subject to the general controls provisions of the Act. The energly controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top and "Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. There are three stars at the very bottom of the logo.
*Protecting and Promoting Public Health*
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2 -
{4}------------------------------------------------
#### SECTION 6
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Korzzay
Device Name: ToPACS
Indications for Use Statement:
The Thoth ToPACS is an image management system whose intended use is to provide (scaleable ) DICOM compatible PACS solutions for hospital and related institutions/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR and other devices) and information systems.
Lossy Compressed Mammographic Images and Digitized Film Screen Images must not be reviewed for primary Image interpretations. Mammographic Images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other specifications reviewed and accepted by FDA
CAUTION: Federal (US) law restricts the use of this device to licensed professionals.
Description Use
David G. Syzon
ision of Reprodu
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.