K063291 · Choice Smart Health Care Co. , Ltd. · FLL · Nov 21, 2006 · General Hospital
Device Facts
Record ID
K063291
Device Name
INFRARED EAR THERMOMETER, MODEL IRT-21 AND IRT-22
Applicant
Choice Smart Health Care Co. , Ltd.
Product Code
FLL · General Hospital
Decision Date
Nov 21, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Infrared Ear Thermometer (IRT-21 and IRT-22), designed with infrared technology, is used to measure body temperature from ear canal. Press the ON/OFF button to turn on the power, insert the probe into the ear canal then press the SCAN button to measure the temperature, when you hear a beep sound, the measurement is complete. Infrared Ear Thermometer (IRT-21 and IRT-22) is intended for household and hospital use on people of all ages.
Device Story
Infrared Ear Thermometer (IRT-21 and IRT-22) measures natural thermal radiation from tympanic membrane and adjacent ear canal surfaces; probe cover used for hygiene. User powers device via ON/OFF button; inserts probe into ear canal; presses SCAN button; device emits beep upon measurement completion. Output displayed as temperature reading (0.1°F/°C resolution). Used in household and hospital settings by patients or clinicians. Provides rapid temperature assessment to support clinical decision-making.
Clinical Evidence
Clinical study enrolled 100 patients to compare IR thermometer against a reference thermometer. Results demonstrated clinical repeatability and bias/uncertainty within acceptable limits for diagnostic purposes, conforming to ASTM E1965-98 (Re-approved 2003).
Technological Characteristics
Infrared sensing technology; measures thermal radiation from tympanic membrane. Operating range 32°C–42.9°C (89.6°F–109.2°F). Ambient operating range 10°C–40°C (50°F–104°F). Conforms to ASTM E1965-1998, ASTM E1104-1998, IEC 60601-1, IEC 60601-1-2, and ISO 10993-1:2003. Includes memory capacity for readings.
Indications for Use
Indicated for body temperature measurement via the ear canal in patients of all ages in household and hospital settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
## 510(K) Summary
NOV 2 1 2006
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92, and the relevant 510(K) submission guidance.
The assigned 510(K) number is:
1. Submitter's Identifications: Mr. Mok Chi Wing Choice Smart Health Care Company Limited. RM 1901, CC WU BUILDING, 302 HENNESSY ROAD, WANCHAI Hong Kong
Date Summary Prepared: April 25, 2006
#### 2. Name of the device:
Trade name: Infrared Ear Thermometer (IRT-21 and IRT-22) Common Name: Clinical Electronic Thermometer Classification Name: Thermometer, Electronic, Clinical
- 3. Information of the 510(K) Cleared Device(Predicate Device): Infrared Ear Thermometer (GT-302) (K010383).
#### 4. Device Description:
Infrared Ear Thermometer (IRT-21 and IRT-22) is used to measure body temperature from ear canal. Operation is based on measuring the natural thermal radiation emitted from the tympanic membrane and adjacent surfaces. Probe cover is used for measuring to ensure accurate temperature measurement. For hygienic purpose, it is advised to discard the probe cover after each use.
Press the ON/OFF button to turn on the power, the symbol 1888 appears immediately. Insert the probe into the ear canal then press the SCAN button to measure the temperature, when you hear a beep sound, the measurement is The temperature is graduated on 0.1 Tor 0.1 ℃, reading a range of complete. 89.6°F -- 109.2°F (32°C -- 42.9°C ). The ambient temperature environment is intended for used is 50°F -- 104°F (10°C-40°C).
- 5. Intended Use:
Infrared Ear Thermometer (IRT-21 and IRT-22), designed with infrared technology, is used to measure body temperature from ear canal. Press the ON/OFF button to turn on the power, the symbol #88 appears immediately.
{1}------------------------------------------------
Insert the probe into the ear canal then press the SCAN button to measure the temperature, when you hear a beep sound, the measurement is complete. Infrared Ear Thermometer (IRT-21 and IRT-22) is intended for household and hospital use on people of all ages.
#### 6. Comparison to the 510(K) Cleared Device(Predicate Device):
Infrared Ear Thermometer (IRT-21 and IRT-22) has the same intended use and technological characteristics as the cleared device of Infrared Ear Thermometer (GT-302) (K010383). Although Infrared Ear Thermometer (IRT-21 and IRT-22) has slightly broader ambient operating temperature, slightly broader display temperature range and more memorization capacity compared with legally marketed one, these differences do not affect the safety, performance of the subject device. So the new device is substantial equivalence to the selected predicate device.
## 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Various testing was performed as per ASTM E1965-1998 (Re-approved 2003) and ASTM E1104-1998 (Re-approved 2003). Accuracy testing was done at 35°C/95°F, 37°C/98.6°F, 41°C/105.8°F according to the requirements of ASTM E1965-98 (Re-approved 2003) and results show that no individual reading was in error by more than the acceptable maximum error. Declaration of conformity letter of conformed standards sees Appendix VIII.
Electrical safety conforms to applicable voluntary standards IEC 60601-1, IEC 60601-1-2. And Biocompatibility conforms to applicable voluntary standards ISO 10993-1:2003 biocompatibility testing on skin irritation, in vitro cytotoxicity and sensitivity.
For predicate device (K010383), it is complied with voluntary standards includes ASTM 1965-98, ASTM 1104, IEC 60601-1 and IEC 60601-1-2 based on its 510K summary.
So the new device is substantial equivalence to the selected predicate device.
#### 8. Discussion of Clinical Study:
A comparison study and clinical repeatability testing was performed between Infrarcd thermometer and reference thermometer. 100 patients were enrofled for this trial. The repeatability result can be considered reasonably small and will not pose a problem for diagnostic purposes. The results of clinical bias and
{2}------------------------------------------------
uncertainty indicate that this IR thermometer can be used reliably. Clinical result conform ASTM E1965-98 (Re approved 2003).
- 9. Conclusions:
Infrared Ear Thermometer (IRT-21 and IRT-22) has the same intended use and technological characteristics as the cleared device of Infrared Ear Thermometer Various performance testing data which conducted (GT-302) (K010383). according to ASTM E1965 and ASTM E1104 standards, such as accuracy test, storage test, shock test demonstrate the same safety and effectiveness as that of cleared device. In the other words, the Infrared Ear Thermometer (IRT-21 and IRT-22) is substantial equivalence to predicted device.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Choice Smart Health Care Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated 2600 Northwest Lake Road Camas, Washington 98607-9526
NOV 2 1 2006
Re: K063291
Trade/Device Name: Infrared Ear Thermometer (IRT-21 and IRT-22) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 11, 2006 Received: November 1, 2006
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Cuy Duleom for
Shin-Hwa Phat
Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K063299
1 of 7
# Indication for Use
### 510(k) Number (if known):
Device name: Infrared Ear Thermometer (IRT-21 and IRT-22)
#### Indications for Use:
Infrared Ear Thermometer (IRT-21 and IRT-22), designed with infrared technology, is used to measure body temperature from ear canal. Press the ON/OFF button to turn on the power, insert the probe into the ear canal then press the SCAN button to measure the temperature, when you hear a beep sound, the measurement is complete. Infrared Ear Thermometer (IRT-21 and IRT-22) is intended for household and hospital use on people of all ages.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Md C. A. f. Amw 11/21/06
I of Anesthesiology. General Hospital ontrol. Dent
K063291
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.