The Gambro Medical lines QuickSet® Bloodlines are single use sterile medical devices intended to provide extracorporeal blood circuit for hemodialysis treatment. QuickSet® Bloodlines can be safely connected to hemodialyzers, vascular accesses and various perfusion lines, under the responsibility of the physician in charge.
Device Story
Gambro QuickSet® Bloodlines are single-use sterile tubing sets for hemodialysis; they transport patient blood from vascular access to a dialyzer and return purified blood to the patient. The system consists of arterial and venous lines, chambers, and connectors. Operated by clinical staff in a hemodialysis setting, the device facilitates the extracorporeal circuit necessary for blood purification. The device is connected to hemodialyzers and vascular access points. By providing a sterile, reliable pathway for blood flow, the device enables effective hemodialysis treatment, directly supporting the removal of waste products and fluid management in patients with renal failure.
Clinical Evidence
Bench testing only. Testing included biocompatibility, sterilization validation, sterility, flow rate, injection port validation, machine compatibility, and mechanical hemolysis testing.
Technological Characteristics
Single-use sterile tubing sets. Materials: Soft PVC (with or without DEHP), SEBS (latex-free injection plugs). Components: Arterial/venous chambers (injection or blow molded), pinch/slide clamps. Sterilization: Radiation or ETO. Connectivity: Standard hemodialysis machine interfaces. Storage: 0°C to +30°C.
Indications for Use
Indicated for patients requiring hemodialysis treatment. Single-use sterile device for extracorporeal blood circulation.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Gambro G Series Bloodline Sets for Hemodialysis
Cobe Hemaflo™ Blood Tubing Sets
Submission Summary (Full Text)
{0}------------------------------------------------
# 5.0 510(K) SUMMARY
| Submitter's Name | Gambro Renal Products | JAN 3 0 2007 |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Address | 10810 West Collins Avenue<br>Lakewood, CO 80215 | |
| Establishment<br>Number | 1713683 | |
| Date of Submission | October 31, 2006 | |
| Contact Person | Thomas B. Dowell, Regulatory Affairs Project Manager | |
| Telephone Number<br>Fax Number | (303) 231-4094<br>(303) 542-5138 | |
| Name of the Device | Gambro QuickSet® Bloodlines | |
| Catalogue Numbers | 018430501: Gambro QuickSet® Post-pump arterial chamber<br>018440501: Gambro QuickSet® No arterial pressure monitoring<br>009445601: Gambro QuickSet® No arterial pressure monitoring<br>009558601: Gambro QuickSet® Pre-pump arterial chamber<br>009559601: Gambro QuickSet® Post-pump arterial chamber<br>009566601: Gambro QuickSet® Pre-pump pillow with post arterial<br>chamber<br>009558714: Gambro QuickSet® Pre-pump arterial chamber<br>009566714: Gambro QuickSet® Pre-pump pillow with post arterial<br>chamber | |
| Common or Usual<br>Name | Extracorporeal blood circuit for hemodialysers | |
| Classification Name | Classification Name: Hemodialysis system and accessories<br>Device Class: II<br>Product Code: FJK<br>Regulation Number: 21 CFR 876.5820 | |
| Identification of the<br>Legally Marketed<br>Device<br>(Predicate Device) | Gambro G Series Bloodline Sets for Hemodialysis<br>Cobe Hemaflo <sup>TM</sup> Blood Tubing Sets | |
{1}------------------------------------------------
## 510(k) SUMMARY, continued
# Device Description
The Gambro medical lines QuickSet® are tubing sets (bloodlines) employed in the hemodialysis equipments extracorporeal circulation: they convey the patient's blood from the arterial-venous access fistula to the dialyzing filter (arterial line) and back after purification (venous line). The Gambro Medical lines QuickSet® Bloodlines are single use sterile medical devices intended to provide extracorporeal blood circuit for hemodialysis treatment. QuickSet® Bloodlines can be safely connected to hemodialyzers, vascular accesses and various perfusion lines, under the responsibility of the physician in charge.
# Comparison Table
| | PREDICATE<br>G Series Blood Line | PREDICATE<br>Hemaflo™ Blood Tubing Set | MODIFIED DEVICE<br>QuickSet® Bloodlines<br>[currently distributed] | MODIFIED DEVICE<br>QuickSet® Bloodlines<br>[new models] |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Gambro G series blood lines are intended for use during hemodialysis in conjunction with an artificial kidney (hemodialyzer). The type of blood line for use is dependent upon the type of dialysis delivery system employed. | Cobe Hemaflo™ Disposable Blood Tubing Set is intended for use in hemodialysis treatment. | The Gambro Medical lines QuickSet® Bloodlines are single use sterile medical devices intended to provide extracorporeal blood circuit for hemodialysis treatment. QuickSet® Bloodlines can be safely connected to hemodialyzers, vascular accesses and various perfusion lines, under the responsibility of the physician in charge. | The Gambro Medical lines QuickSet® Bloodlines are single use sterile medical devices intended to provide extracorporeal blood circuit for hemodialysis treatment. QuickSet® Bloodlines can be safely connected to hemodialyzers, vascular accesses and various perfusion lines, under the responsibility of the physician in charge. |
| Currently in Distribution | NO | NO | YES | NO |
{2}------------------------------------------------
Traditional 510(k) Gambro QuickSet® Bloodlines
| | PREDICATE | PREDICATE | MODIFIED | MODIFIED |
|----------------------|---------------------|-------------------|-------------------|-------------------|
| | G Series Blood Line | Hemaflo™ | DEVICE | DEVICE |
| | | Blood Tubing | QuickSet® | QuickSet® |
| | | Set | Bloodlines | Bloodlines |
| | | | [currently | [new models] |
| | | | distributed] | |
| | | | 009445601 | |
| | | | 009558601 | |
| Catalogue No. | Not applicable | Not applicable | 009559601 | 009558714 |
| | | | 018430501 | 009566714 |
| | | | 018440501 | |
| | | | 009566601 | |
| Arterial Chamber | Injection Molded | Injection Molded | Blow Molded | Injection Molded |
| Venous Chamber | Injection Molded | Injection Molded | Blow Molded | Injection Molded |
| | | | | |
| Clamps | Pinch & Slide | Slide | Pinch | Pinch |
| Blood tubing | Soft PVC material | Soft PVC material | Soft PVC material | Soft PVC material |
| material | with DEHP | with DEHP | with DEHP | with DEHP-free |
| | plasticizer | plasticizer | plasticizer | plasticizer |
| Injection plug | Latex | Latex | Latex-free | Latex-free |
| material | | | (SEBS) | (SEBS) |
| Packaging | Blister | Pouch | Pouch | Blister |
| | | | | |
| Quantity per box | 16 | 10 | 16 | 14 |
| Sterilization Method | ETO | Radiation | Radiation | Radiation |
| Expiration | 3 years | 3 years | 3 years | 3 years |
| Single Use | Yes | Yes | Yes | Yes |
| Storage | Between | Between | Between | Between |
| Temperature | 10°C (50°F) | 10°C (50°F) | 0°C (32°F) | 0°C (32°F) |
| | and | and | and | and |
| | +24°C (75°F) | +24°C (75°F) | +30°C (86°F) | +30°C (86°F) |
Description and Conclusion of Testing
### Nonclinical Testing:
The non-clinical testing consisted of performance testing (bench) that included biocompatibility testing, validation of the sterilization process, sterility testing, flow rate testing, validation of needle and needle-less system injection ports, compatibility testing with different hemodialysis machines, and testing for mechanical hemolysis.
### Conclusion:
The successful non-clinical testing demonstrates the safety and effectiveness of the Gambro QuickSet® Bloodlines when used for the defined indications for use and demonstrates that the device for which the 510(k) is submitted performs as well as or better than the legally marketed predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's emblem in the center. The emblem is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Thomas B. Dowell Regulatory Affairs Project Manager GAMBRO Renal Products 10810 West Collins Avenue LAKEWOOD CO 80215
JAN 3 0 2007
Re: K063290
Trade/Device Name: Gambro QuickSet® Bloodlines Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FJK Dated: October 31, 2006 Received: November 1, 2006
Dear Mr. Dowell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo commemorates the centennial anniversary from 1906-2006. The letters "FDA" are in large bold font in the center of the logo. The word "Centennial" is written in cursive below the letters "FDA".
Protecting and Promoting Public Health
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K063290
Device Name: Gambro QuickSet® Bloodlines
Indications for Use:
The Gambro Medical lines QuickSet® Bloodlines are single use sterile medical devices intended to provide extracorporeal blood circuit for hemodialysis treatment. QuickSet® Bloodlines can be safely connected to hemodialyzers, vascular accesses and various perfusion lines, under the responsibility of the physician in charge.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.