ARCHITECT INTACT PTH REAGENTS, CALIBRATORS (A-F) AND CONTROLS (LOW, MEDIUM, HIGH), MODELS 8K25-20, 8K25-25, 8K25-01

K063232 · Biokit, S.A. · CEW · Jun 19, 2007 · Clinical Chemistry

Device Facts

Record IDK063232
Device NameARCHITECT INTACT PTH REAGENTS, CALIBRATORS (A-F) AND CONTROLS (LOW, MEDIUM, HIGH), MODELS 8K25-20, 8K25-25, 8K25-01
ApplicantBiokit, S.A.
Product CodeCEW · Clinical Chemistry
Decision DateJun 19, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1545
Device ClassClass 2

Indications for Use

The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System. The ARCHITECT Intact PTH assay is intended to be used as an aid in the differential diagnosis of hypercalcemia, hypocalcemia and parathyroid disorders. The ARCHITECT Intact PTH Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of intact PTH in human serum and plasma. The ARCHITECT Intact PTH Controls are for the use in quality control to monitor the accuracy and precision of the ARCHITECT Intact PTH assay on the ARCHITECT i System for human serum and plasma.

Device Story

ARCHITECT Intact PTH assay is a two-step sandwich chemiluminescent microparticle immunoassay (CMIA) performed on the ARCHITECT i System. Input: human serum or plasma sample. Process: sample, assay diluent, and anti-PTH coated paramagnetic microparticles combined; intact PTH binds to microparticles; wash cycle; anti-PTH acridinium-labeled conjugate added; second wash cycle; pre-trigger and trigger solutions added. Output: chemiluminescent signal measured as relative light units (RLUs). System optics detect RLUs, which are directly proportional to intact PTH concentration. Results compared against calibration curve to determine PTH concentration. Used in clinical laboratory settings by trained personnel. Output assists clinicians in differential diagnosis of calcium and parathyroid disorders.

Clinical Evidence

Bench testing only. Performance demonstrated via correlation studies comparing the ARCHITECT Intact PTH assay to the Roche Elecsys PTH predicate. Results showed a correlation coefficient of 0.99 for both routine and STAT protocols.

Technological Characteristics

Chemiluminescent microparticle immunoassay (CMIA) technology. Utilizes anti-PTH coated paramagnetic microparticles and acridinium-labeled conjugate. Detection via chemiluminescent reaction measured as relative light units (RLUs). Designed for use on the ARCHITECT i System. In vitro diagnostic use.

Indications for Use

Indicated for use as an aid in the differential diagnosis of hypercalcemia, hypocalcemia, and parathyroid disorders in patients requiring quantitative determination of intact PTH in human serum and plasma.

Regulatory Classification

Identification

A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Predicate Devices

Submission Summary (Full Text)

{0} 1 of 13 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k063232 B. Purpose for Submission: New device for intact parathyroid hormone (PTH), calibrator and control materials C. Measurand: PTH D. Type of Test: Quantitative chemiluminescent microparticle immunoassay E. Applicant: BIOKIT S.A. F. Proprietary and Established Names: ARCHITECT Intact PTH reagents, calibrators (A-F) and controls (Low, Medium, High), models 8K25-20, 8K25-25, 8K25-01 G. Regulatory Information: 1. Regulation section: 21CFR Sec. 862.1545-Parathyroid hormone test system 21CFR Sec. 862.1150 Calibrator 21CFR Sec. 862.1660 Quality control material (assayed and unassayed) 2. Classification: Class 2, 2, 1 (reserved), respectively 3. Product code: CEW - radioimmunoassay, parathyroid hormone JIT - calibrator, secondary JJX - single (specified) analyte controls (assayed and unassayed) 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See indication(s) for use below 2. Indication(s) for use: {1} 2 of 13 Reagents The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System. The ARCHITECT Intact PTH assay is intended to be used as an aid in the differential diagnosis of hypercalcemia, hypocalcemia and parathyroid disorders. Calibrators The ARCHITECT Intact PTH Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of intact PTH in human serum and plasma. Controls The ARCHITECT Intact PTH Controls are for the use in quality control to monitor the accuracy and precision of the ARCHITECT Intact PTH assay on the ARCHITECT i System for human serum and plasma. For in vitro diagnostic use 3. Special conditions for use statement(s): Prescription use only The ARCHITECT Intact PTH assay has not been clinically validated for intra-operative use. 4. Special instrument requirements: ARCHITECT i System I. Device Description: ARCHITECT Intact PTH Reagent Kit - MICROPARTICLES 1 or 4 Bottle(s) (6.6 mL each) - Anti-PTH (goat, polyclonal) coated microparticles in TRIS buffer Preservative: sodium azide - CONJUGATE 1 or 4 Bottle(s) (5.9 mL each) Anti -PTH (goat, polyclonal) acridinium -labeled conjugate in MES buffer with protein (bovine, goat) stabilizer. Preservative: sodium azide - ASSAY DILUENT 1 or 4 Bottle(s) (10.0 mL each) Intact PTH Assay Diluent containing phosphate buffer with protein (bovine, goat) stabilizer Preservative: sodium azide ARCHITECT® Intact PTH Calibrators contents: 6 bottles (4.0 mL each) of ARCHITECT Intact PTH Calibrators Calibrator A contains Bis Tris Propane buffer with protein (bovine) stabilizer Calibrators B-F contain PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin® 300 {2} The calibrators yield the following concentrations: | Calibrator | Target concentration (pg/mL) | | --- | --- | | A. | 0.0 | | B. | 4.8 | | C. | 24.0 | | D. | 120.0 | | E. | 600.0 | | F. | 3000.0 | ARCHITECT® Intact PTH control contents: 3 x 1 Bottle (8.0 mL each) of ARCHITECT Intact PTH controls contain PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin® 300. The following concentration ranges may be used for individual replicate control specifications on the ARCHITECT i System when used with the Routine protocol: | Control | Target Concentration (pg/mL) | Range (pg/mL) | | --- | --- | --- | | Low | 10.0 | 6.0 – 14.0 | | Med | 65.0 | 42.3 – 87.8 | | High | 250.0 | 162.5 – 337.5 | The following concentration ranges may be used for individual replicate control specifications on the ARCHITECT i System when used with the STAT protocol: | Control | Target Concentration (pg/mL) | Range (pg/mL) | | --- | --- | --- | | Low | 8.3 | 5.0 – 11.7 | | Med | 54.2 | 35.2 – 73.1 | | High | 208.3 | 135.4 – 281.2 | Although control materials produced different target values and ranges, human serum and plasma specimens tested with the STAT or Routine protocol will produce equivalent results. The different ranges for the Controls in the STAT and Routine protocol are associated with the matrix used for the Intact PTH Controls. J. Substantial Equivalence Information: 1. Predicate device name(s): Roche Elecsys Parathyroid Hormone Test System 2. Predicate 510(k) number(s): k992680 3. Comparison with predicate: {3} Reagents: ARCHITECT vs. ROCHE Elecsys Similarities: | CHARACTERISTICS | DEVICE | | PREDICATE | | --- | --- | --- | --- | | | STAT | ROUTINE | Intact PTH | | Product Type | Immunoassay | Immunoassay | Immunoassay | | Methodology | Chemiluminescent Microparticle Immunoassay (CMIA) | Chemiluminescent Microparticle Immunoassay (CMIA) | Electrochemiluminescence Immunoassay (ECLIA) | | Intended Use | The ARCHITECT® Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System. | The ARCHITECT® Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System. | The Roche Elecsys Immunoassay for the in vitro quantitative determination of intact parathyroid hormone in human serum and plasma for the differential diagnosis of hypercalcemia and hypocalcemia. The Electrochemiluminescence Immunoassay (ECLIA) is intended for use on the Roche Elecsys 1010/2010 and modular analytics E170 (Elecsys module) immunoassay analyzers. | | Assay Protocol | Direct sandwich immunoassay | Direct sandwich immunoassay | Sandwich principle | | Interpretation of Results | Standard Curve | Standard Curve | Standard Curve | Differences: | Characteristics | Device | | Predicate | | --- | --- | --- | --- | | | STAT | ROUTINE | | | Components | Microparticles - 1or 4 Bottle(s) (6.6 mL each) Anti-PTH (goat, polyclonal) coated microparticles in TRIS buffer. Preservative: sodium azide Conjugate -1or 4 Bottle(s) (5.9 mL each) Anti-iPTH (goat, polyclonal) acridinium-labeled conjugate in MES buffer with protein | Microparticles - 1or 4 Bottle(s) (6.6 mL each) Anti-PTH (goat, polyclonal) coated microparticles in TRIS buffer. Preservative: sodium azide Conjugate - 1or 4 Bottle(s) (5.9 mL each) Anti-iPTH (goat, polyclonal) acridinium-labeled conjugate in MES buffer | Microparticles - 1 bottle 6.5mL: Streptavidin-coated microparticles, 0.72 mg/mL; binding capacity: 470 ng biotin/mg microparticles; preservative R1 - Anti -PTH-Ab-biotin -1 bottle (10.0 mL) Biotinylated monoclonal anti-PTH antibody (mouse) 2.3 mg/L; phosphate buffer | {4} Calibrators: | Characteristics | Device | Prodince | | --- | --- | --- | | Intermediate Use | STAT | Prodince | | Intermediate Use | Deviator Components | Prodince | Differences: | Characteristics | Device | Prodince | | --- | --- | --- | | Calibrator Components | Synthetic peptide PTH in a Bis Tris Propane buffer | Synthetic peptide in a human serum matrix | | Calibrator Concentrations | 6 levels | 2 levels | {5} Controls: Similarities: | Characteristics | Device | Predicate | | --- | --- | --- | | Intended Use | The ARCHITECT® Intact PTH Controls are for the use in quality control to monitor the accuracy and precision of the ARCHITECT Intact PTH assay on the ARCHITECT i System for human serum and plasma. | Elecsys® PreciControl Bone is used for the quality control of PTH on Elecsys immunoassay systems. | | Control Concentrations | 3 levels | 3 levels | Differences: | Characteristics | Device | Predicate | | --- | --- | --- | | Control Components | PTH synthetic peptide in Bis Tris Propane buffer | Lyophilized control serum based equine serum | # K. Standard/Guidance Document Referenced (if applicable): CLSI - Interference Testing in Clinical Chemistry - EP07-A2 CLSI - Evaluation of Precision Performance of Clinical Chemistry Devices - EP05-A2 FDA - Points to Consider Guidance Document on Assayed and Unassayed Quality Control Material; Draft # L. Test Principle: The ARCHITECT Intact PTH assay is a two-step sandwich immunoassay for the quantitative determination of intact PTH in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology. In the first step, sample, assay diluent, and anti -PTH coated paramagnetic microparticles are combined. Intact PTH present in the sample binds to the anti-PTH coated microparticles. After washing, anti-PTH acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pretrigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). The ARCHITECT Intact PTH can be used with both STAT and Routine Protocols. The STAT protocol has a shorter incubation time in comparison to the Routine protocol. Routine and STAT protocols require separate calibrations due to different incubation time but require only one reagent kit. A direct relationship exists between the amount of intact PTH in the sample and the RLUs detected by the ARCHITECT i System optics. {6} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: A study was performed for the ARCHITECT Intact PTH assay, based on guidance from the Clinical Laboratory Standards Institute (CLSI) Protocol EP5-A. Multiple ARCHITECT Intact PTH control lots were assayed using two lots of reagents in replicates of two at two separate times per day for 20 days at one site on two instruments using the STAT protocol and at a second site on two instruments using the Routine protocol. In addition, a second precision study was performed, where two lots of reagents were assayed in four replicates per run at two separate times per day over 10 days on one instrument at a third site using the STAT protocol and on one instrument at a fourth site using the Routine protocol. Each reagent lot used a single calibration curve throughout the study. Data from both studies are summarized in the following tables. | ARCHITECT Intact PTH Precision Using STAT Protocol | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Sample | n | Mean Conc. (pg/mL) | Within Run | | Total | | | | | | SD | %CV | SD | %CV | | Low Control | 878 | 8.5 | 0.74 | 8.7 | 0.74 | 8.7 | | Medium Control | 880 | 56.5 | 2.34 | 4.1 | 2.35 | 4.2 | | High Control | 880 | 208.9 | 8.62 | 4.1 | 8.62 | 4.1 | | ARCHITECT Intact PTH Precision Using Routine Protocol | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Sample | n | Mean Conc. (pg/mL) | Within Run | | Total | | | | | | SD | %CV | SD | %CV | | Low Control | 880 | 10.7 | 0.65 | 6.1 | 0.69 | 6.4 | | Medium Control | 880 | 69.6 | 2.28 | 3.3 | 2.31 | 3.3 | | High Control | 880 | 255.8 | 7.40 | 2.9 | 7.56 | 3.0 | A third study was performed with high concentrated samples with one lot of reagents in replicates of two at two separate times per day for 20 days at one site on one instrument using the STAT protocol and the Routine protocol. Each reagent lot used a single calibration curve throughout the study. Data from both studies are summarized in the following tables. 7 of 13 {7} 8 of 13 | ARCHITECT Intact PTH Precision Using STAT Protocol | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Mean Conc. | | Within Run | | Total | | | Sample | n | (pg/mL) | SD | %CV | SD | %CV | | High Concentration | 80 | 828.6 | 25.9 | 3.2 | 31.8 | 4.0 | | High Concentration | 80 | 2039.9 | 49.5 | 2.5 | 67.4 | 3.4 | | ARCHITECT Intact PTH Precision Using Routine Protocol | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Mean Conc. | | Within Run | | Total | | | Sample | n | (pg/mL) | SD | %CV | SD | %CV | | High Concentration | 80 | 988.7 | 36.8 | 4.6 | 41.7 | 5.2 | | High Concentration | 80 | 2392.4 | 78.5 | 3.9 | 88.8 | 4.4 | b. Linearity/assay reportable range: MEASUREMENT RANGE The measurement for the ARCHITECT Intact PTH assay is STAT: 4.0 pg/mL to 2227 pg/mL Routine: 3.0 pg/mL to 2257 pg/mL The range of measurement was established with dilution recovery studies and detection limit (functional sensitivity) below. The high and low calibrators have been diluted to yield the below expected values dilutions. The obtained dilutions have been measured with the ARCHITECT Instrument using both STAT and Routine protocols. The percent recovery has been calculated. | | Routine expected observed | | | | | --- | --- | --- | --- | --- | | | | | % recovery | % CV | | dil 1 | 3000 | 2994 | 99 .8 | 0 .1 | | dil 2 | 1502.4 | 1551 | 103 .3 | 0 .5 | | dil 3 | 2251.2 | 2312 | 102 .7 | 0 .1 | | dil 4 | 1876.8 | 1841 | 98 .1 | 6 .2 | | dil 5 | 1689.6 | 1680 | 99 .5 | 0 .2 | | dil 6 | 1596.1 | 1647 | 103 .2 | 1 .4 | | dil 7 | 753.6 | 762 | 101 .2 | 0 .2 | | dil 8 | 379.2 | 362 | 95 .6 | 3 .8 | | dil 9 | 192 | 181 | 94 .4 | 1 .1 | | dil 10 | 98.4 | 92 | 93 .6 | 0 .5 | | dil 11 | 4.8 | 5 | 109 .4 | 6 .7 | {8} | | STAT expected observed | | | | | --- | --- | --- | --- | --- | | | | | % recovery | % CV | | dil 1 | 2500 | 2480 | 99 .2 | 2 .1 | | dil 2 | 1252 | 1263 | 100 .8 | 0 .4 | | dil 3 | 1876 | 1796 | 95 .7 | 1 .2 | | dil 4 | 1564 | 1447 | 92 .5 | 0 .1 | | dil 5 | 1408 | 1349 | 95 .8 | 0 .5 | | dil 6 | 1330 | 1285 | 96 .6 | 1 .6 | | dil 7 | 628 | 584 | 93 .0 | 15 .8 | | dil 8 | 316 | 306 | 97 .0 | 4 .3 | | dil 9 | 160 | 154 | 96 .1 | 0 .6 | | dil 10 | 82 | 76 | 93 .2 | 1 .2 | | dil 11 | 4 | 4 | 98 .8 | 1 .8 | c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrators and controls are manufactured using Human Parathyroid Hormone and are traceable to WHO International reference preparation of Parathyroid Hormone, Human, for Immunoassay code: 79/500. Value assignments and verification processes for ARCHITECT Intact PTH calibrators and controls are done using an internal protocol. A study was conducted to determine the shelf-life integrity of the ARCHITECT Intact PTH Reagent Kit, Calibrator, and Controls when stored at the recommended storage condition of 2 to $8^{\circ}\mathrm{C}$ . A real-time stability study supports a shelf life of 13 months for the ARCHITECT Intact PTH Reagent, Calibrator, and Control kit at the recommended storage condition of 2 to $8^{\circ}\mathrm{C}$ . The expiration date of 12 months is being assigned for the ARCHITECT Intact PTH Reagent, Calibrator and Control kit. d. Detection limit: A study was performed using human samples with concentrations targeted at $3\mathrm{pg / mL}$ , $4\mathrm{pg / mL}$ and $5\mathrm{pg / mL}$ of intact PTH. Those samples were tested in replicates of two over 10 days using two reagent lots on two instruments using the STAT protocol and on one instrument using the Routine protocol. Functional sensitivity (lowest concretion with a $\mathrm{CV} \leq 20\%$ ) was determined to be $= 3 \, \mathrm{pg/mL}$ for the Routine protocol and $= 4 \, \mathrm{pg/mL}$ for the STAT protocol. e. Analytical specificity: Specificity The specificity of the ARCHITECT Intact PTH assay is designed to have a {9} cross-reactivity = 0.01% when tested with structurally similar compounds listed in the table below. A study was performed with the ARCHITECT Intact PTH assay based on guidance from the for Clinical Laboratory Standards Institute (CLSI) Protocol EP7-A. Aliquots of ARCHITECT Intact PTH Calibrator A were supplemented with potential cross-reactants at the concentrations listed and tested for intact PTH. Data from this study are summarized in the following table. | PTH fragment | Concentrations | % Cross-Reactivity | | --- | --- | --- | | 1-34 | 100000 pg/mL | 0.00 | | 39-68 | 100000 pg/mL | 0.00 | | 53-84 | 100000 pg/mL | 0.00 | | 44-68 | 100000 pg/mL | 0.00 | | 39-84 | 100000 pg/mL | 0.00 | $$ \% \text{ Cross-Reactivity} = \frac{\text{Mean Value spiked} - \text{Mean Value non spiked (pg/mL)}}{\text{Concentration of Cross-Reactant (pg/mL)}} \times 100 $$ ## Interference studies Potential interference in the ARCHITECT Intact PTH assay from hemoglobin, bilirubin, triglycerides, and protein at the levels indicated below. Interference was demonstrated by a study based on guidance from the Clinical Laboratory Standards Institute (CLSI) Protocol EP7-A. There was no significant interference observed. Data from this study are summarized in the following table. Potentially Interfering | Substance | Concentration | % Mean Recovery | | --- | --- | --- | | Hemoglobin | 500 mg/dL | 102 | | Bilirubin | 20 mg/dL | 98 | | Triglycerides | 5000 mg/dL | 105 | | Protein low | 4 g/dL | 106 | | Protein high | 9.5 g/dL | 93 | | Protein high (Routine protocol) | 10.5 g/dL | 94* | When using the STAT protocol, interference with high levels of protein may be observed. ## Limitations to the procedure The applicant states Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or to animal serum products can be prone to this interference and anomalous values may be observed. Additional information may be required for diagnosis. The ARCHITECT Intact PTH assay has not been clinically validated for intra-operative use. 10 of 13 {10} f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: The comparison studies were comprised of Serum and EDTA plasma samples. A study was performed comparing the ARCHITECT Intact PTH assay to the predicate assay. In the study 467 specimens were analyzed with the STAT Protocol and 442 with the Routine Protocol, In this evaluation, specimen concentrations ranged from ≤4 to 2027.2 pg/mL using the STAT Protocol, from ≤3 to 2326.9 pg/mL using the Routine Protocol and from 1.2 to 1813.0 pg/mL with the predicate. | ARCHITECT Intact PTH vs. Predicate Assay Using STAT Protocol | | | | | | --- | --- | --- | --- | --- | | Regression Method | n | Slope | Intercept | Correlation Coefficient | | Passing-Bablok | 467 | 1.28 | 0.45 | 0.99 | | Least Squares | 467 | 1.20 | 11.00 | | | ARCHITECT Intact PTH vs. Predicate Assay Using Routine Protocol | | | | | | --- | --- | --- | --- | --- | | Regression Method | n | Slope | Intercept | Correlation Coefficient | | Passing-Bablok | 442 | 1.29 | -0.66 | 0.99 | | Least Squares | 442 | 1.19 | 11.40 | | 200 EDTA plasma specimens were tested in singlicate using both, the STAT and the ROUTINE protocols and the predicate device. The results ranged from 5.5 pg/mL to 1252.5 pg/mL and are presented in the table below: | | | Intercept | Slope | r | | --- | --- | --- | --- | --- | | STAT | Passing Bablok | 3.003 | 1.192 | 0.9946 | | STAT | Linear | 16.898 | 1.096 | 0.9946 | | Routine | Passing Bablok | 3.978 | 1.263 | 0.9956 | | Routine | Linear | 19.235 | 1.152 | 0.9956 | b. Matrix comparison: To evaluate the use of different anticoagulants, a study was conducted using 20 lithium heparin samples, 20 sodium heparin samples, and 20 serum samples collected in serum separator tubes (SST) and analyzed in both the stat and routine mode. The results were compared to serum samples collected in {11} plain tube with no additives. Ten of the twenty samples were spiked with PTH to cover the assay range. The results are presented below: For the STAT protocol: | Test Condition | N | Range of % differences | | | --- | --- | --- | --- | | Normal range (24.9 pg/mL to 97.1 pg/mL) | | | | | Serum no additive | 10 | Control Condition | | | SST | 10 | -15.9 % to -6.3 % | | | Li-Heparin | 10 | -3.6 % to 6.4 % | | | Sodium-Heparin | 10 | -8.4 % to 9.7 % | | | Upper range (121.7 pg/mL to 2681.1 pg/mL) | | | | | Serum no additive | 10 | Control Condition | | | SST | 10 | -10.4 % to 6.4 % | | | Li-Heparin | 10 | -8.4 % to 3.3 % | | | Sodium-Heparin | 10 | -8.3 % to 6.2 % | | For the ROUTINE protocol: | Test Condition | N | Range of % differences | | | --- | --- | --- | --- | | Normal range (22.7 pg/mL to 92.6 pg/mL) | | | | | Serum no additive | 10 | Control Condition | | | SST | 10 | -20.5 % to -5.6 % | | | Li-Heparin | 10 | -11.1 % to 9.0 % | | | Sodium Heparin | 10 | -17.8 % to 5.3 % | | | Upper range ( 120.4 pg/mL to 2641.1 pg/mL) | | | | | Serum no additive | 10 | Control Condition | | | SST | 10 | -10.1 % to 4.9 % | | | Li-Heparin | 10 | -7.5 % to 6.6 % | | | Sodium-Heparin | 10 | -7.8 % to 6.0 % | | 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable {12} 13 of 13 5. Expected values/Reference range: A reference range study was conducted with USA population samples, testing a total of 168 samples from apparently healthy adults. These samples gave the values summarized in the following table. | | N | Median | Intact PTH (pg/mL) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | 2.5^{th} percentile | | 97.5^{th} percentile | | | | | STAT | Routine | STAT | Routine | STAT | Routine | | Healthy Adults | 168 | 27.2 | 28.9 | 8.5 | 8.7 | 72.5 | 77.1 | A reference range study was conducted with European population for the ARCHITECT Intact PTH assay to establish the reference range from apparently healthy adults. Data from this study are summarized in the following table. | | N | Median | Intact PTH (pg/mL) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | 2.5^{th} percentile | | 97.5^{th} percentile | | | | | STAT | Routine | STAT | Routine | STAT | Routine | | Healthy Adults | 143 | 37.25 | 35.70 | 15.64 | 14.04 | 69.95 | 67.84 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...