K063035 · Transnoma Medical, Inc. · MXC · Oct 1, 2007 · Cardiovascular
Device Facts
Record ID
K063035
Device Name
SLEUTH IMPLANTABLE ECG MONITORING SYSTEM
Applicant
Transnoma Medical, Inc.
Product Code
MXC · Cardiovascular
Decision Date
Oct 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Indications for Use
The Transoma Medical Sleuth 2010 System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for: - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or . - patients who experience transient symptoms that may suggest a cardiac arrhythmia.
Device Story
Sleuth Implantable ECG Monitoring System consists of Model 2010 Implantable Monitoring Device (IMD), Model 4000 Activator, and Model 5000 Base Station. IMD is a titanium structure implanted subcutaneously in left pectoral region. Device continuously monitors R-R intervals; automatically stores ECG segments when intervals exceed pre-defined limits. Patients use hand-held Activator to manually trigger ECG storage during symptomatic events. Data transfers via telemetry from IMD to Activator, then via Base Station to a service center for physician review. System enables remote monitoring of arrhythmic events; assists physicians in diagnosing unexplained syncope or pre-syncope. Benefits include continuous 24/7 monitoring and capture of both symptomatic and asymptomatic events.
Clinical Evidence
Clinical study conducted at 4 medical centers with 28 patients presenting with unexplained syncope/pre-syncope or arrhythmia risk. Patients monitored 24/7 for 7-10 days, 1, 3, and 6 months. Study demonstrated diagnostic viability; 13 patients were diagnosed with arrhythmias using Sleuth records. Two adverse events reported over 4,700 patient days, deemed not serious or definitively linked to the device.
Technological Characteristics
Titanium IMD structure; subcutaneous implantation. Sensing principle: R-R interval monitoring. Connectivity: Telemetry (IMD to Activator), Bluetooth (Base Station to service center). System includes IMD, Activator, and Base Station. Software-based event detection and storage.
Indications for Use
Indicated for patients at increased risk of cardiac arrhythmias or those experiencing transient symptoms suggestive of cardiac arrhythmia.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
Medtronic, Inc., Reveal® Plus Insertable Loop Recorder (ILR) System (K003667, K994331 and K972242)
Brentwood Medical Technology Corporation - IOmark™ Digital Holter, (K031466)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
K063035
page 1/3
#### General Provisions
Submitter's Name and Address
Transoma Medical, Inc. 4358 West Round Lake Road Arden Hills, Minnesota 55112 Phone: 651-481-7410 Fax: 651-481-7415
Contact Person
Angela Raun Phone: 651-481-7460 Fax: 651-481-7415 Fmail: araun(@TransomaMedical.com
OCT 1 * 2007
Date of Summary
September 18, 2007
| Proprietary Name of Device | Sleuth Implantable ECG Monitoring System |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name | Implantable ECG Monitoring System |
| Classification Name | Cardiac Implantable Event Recorder<br>Product Code - MXC<br>Regulation Number 21 CFR Part 870.2800<br>Device Class II |
| Legally Marketed Device(s) to<br>Which Substantial Equivalence<br>is Claimed | Medtronic, Inc., Reveal® Plus Insertable<br>Loop Recorder (ILR) System (K003667,<br>K994331 and K972242) |
| | Brentwood Medical Technology Corporation<br>- IOmark™ Digital Holter, (K031466) |
#### Device Description
The Transoma Medical Sleuth Implantable ECG Monitoring System is designed to measure and store the patient s electrocardiogram (ECG) during symptomatic and asymptomatic events. The Sleuth System consists of the Model 2010 Implantable Monitoring Device (IMD), the Model 4000 Activator, and the Model 5000 Base Station (Bluetooth Modem). The IMD is an anatomically-shaped Litanium structure which is implanted subcutancously, typically in the left pectoral region. The Activator is a hand-held device carried by the patient.
The IMD continuously measures the R-R interval and automatically stores ECG segments when this interval falls outside of pre-defined limits. Using the Activator, the patient can also request storage of ECG segments when symptoms are experienced. The IMD stores a limited quantity of events, which are · ansferred to the Activator via telemetry throughout the day. On a daily basis, the Activator transfers events through the Base Station to a service center for review and analysis.
{1}------------------------------------------------
#### Intended Use
The Transoma Medical Sleuth 2010 System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for:
KOG: 3035
page 2/3
- patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or .
- patients who experience transient symptoms that may suggest a cardiac arrhythmia. 133035 .
## Summary of Technological Characteristics
The Transoma Medical Sleuth 2010 System incorporates substantially equivalent technology. comparable features, labeling, and intended use, and is similar to the predicate devices currently available on the market.
### Non-clinical Test Summary
The substantial equivalence of the Transoma Medical Sleuth 2010 System has been demonstrated via bench and animal testing. Testing included:
In vitro/ bench testing of the IMD, the Activator, the Base Station and the system In vivo canine testing Biocompatibility testing Software testing Electromagnetic compatibility testing Electrical safety testing Sterilization validation Shelf-life testing
### Clinical Study Summary
A clinical study was conducted at 4 major medical centers in Panama. The primary objectives were to evaluate the diagnostic viability of the ECG signals and the performance of the system. Twenty-eight (28) patients were enrolled in the study. The patients had unexplained syncope/ pre-syncope or were at risk of arrhythmias. Patients were monitored continuously (24/7) by the Sleuth device for arrhythmic events. Clinic follow-ups were 7-10 days. 1, 3, and 6 months after implantation.
The cardiac monitoring center has been able to analyze and categorize the ECG waveform data. The Sleuth system has successfully captured events mitiated manually, as well as asymptomatic events which are detected and recorded automatically by the system. As designed, the Sleuth system has also successfully stored and transferred trending data.
The Sleuth ECG system has performed well following implantation in the initial 28 patients. The system has diagnostic viability, since thirtcen of the patients were discovered during monitoring to have arrhythmus, and the Sleuth records were important in helping the physicians arrive at a firm diagnosis. The occurrence of two adverse events within > 4,700 collective patient days of use is within expectation, and these events were neither serious, nor definitively shown to derive from the Sleuth system.
# Performance Standards
Performance standards have not been established by the Food and Drug Administration for these devices under Section 514.
# Conclusion (Statement of Equivalence)
Extensive bench and animal testing have demonstrated the Sleuth Model 2010 System functions in accordance with product specifications. Additionally, testing demonstrated the Sleuth System functions equivalently to comparative systems, i.e., the Brentwood IQmark Digital Holter (K031466) and the Medtronic Reveal Plus Insertable Loop Recorder (K003667, K994331 and K972242). These data
{2}------------------------------------------------
support a determination of substantial equivalence and subsequently market clearance of the Transoma Medical Sleuth System, comprised of the Model 2010 IMD, Model 4000 Activator, and Model 5000 Base Station.
Kax 3035
page 3/3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wing segments, creating a sense of movement or flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Transoma Medical, Inc. c/o Ms. Angie Raun Vice President Regulatory Affairs and Quality Assurance 4358 West Round Lake Road Arden Hills, MN 55112
2007
Re: K063035
Trade/Device Name: Sleuth Implantable ECG Monitoring System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC Dated: June 29, 2007 Received: July 2, 2007
Dear Ms. Raun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfimmimorfor
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
#### Section 4: Indications for Use Statement
510(k) Number_ K 063035
Device Name: Transoma Medical Sleuth Implantable ECG Monitoring System
#### Indications for Use:
The Transoma Medical Sleuth Implantable ECG Monitoring System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for:
- Patients with clinical syndromes or situations at increased risk of cardiac . arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac . arrhythmia
Prescription Use X (21 CFR 801 Subpart D)
OR
Over The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
.. .--------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blymlamor
ardiovascular Devic
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.