TOP NEUROPOLE NEEDLES, MODELS ST, X, RC, XE AND TL

K062946 · Top Corp. · GXI · Feb 15, 2007 · Neurology

Device Facts

Record IDK062946
Device NameTOP NEUROPOLE NEEDLES, MODELS ST, X, RC, XE AND TL
ApplicantTop Corp.
Product CodeGXI · Neurology
Decision DateFeb 15, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Top Neuropole Needles are used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning.

Device Story

Top Neuropole Needles are Teflon-insulated injection needles designed for percutaneous nerve blocks and radiofrequency (RF) lesioning. Available in five models (ST, X, RC, XE, TL) with gauges 20-24 and lengths 30-200 mm. Procedure involves inserting needle to target nerve; localization achieved via electrostimulation through needle or contrast medium injection under radiography. Once localized, clinician performs nerve block by injecting local anesthetic or creates RF lesion. Used in clinical settings for pain management or local anesthesia. Single-use, sterile devices.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Teflon-insulated injection needles; 20-24 gauge; 30-200 mm length. Five models (ST, X, RC, XE, TL). Single-use, sterile. Functions as probe for radiofrequency lesioning and conduit for local anesthetic or contrast media. No software or electronic components.

Indications for Use

Indicated for percutaneous nerve blocks with local anesthetic solution or radiofrequency lesioning. Used for nerve localization via electrostimulation or contrast medium injection with radiography. Patient population includes individuals requiring regional anesthesia or pain management procedures.

Regulatory Classification

Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ L062946 #### 510(k) SUMMARY -EB 1 5 2007 ### This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c) | Submitter's Name And Address: | Top Corporation<br>19-10 Senjunakai-cho<br>Adachi-ku<br>Tokyo, Japan 120-0035 | |-------------------------------|-------------------------------------------------------------------------------| | Contact Name: | Toshimitsu Suzuki | | Tel. | +81 3-3882-3101 | | Submission Date: | April 24, 2006 | **Device** | Trade or (Proprietary) Name: | Top Neuropole Needles | |-----------------------------------------------------------------|--------------------------------------------------------------------------| | Common or usual name: | Pole Needles | | Classification Name: | Probe, radiofrequency lesion<br>Class II devices. (21 C.F.R. § 882.4725) | | Product Code: | GXI | | Legally Marketed Device To Which Claim Substantial Equivalence: | Radionics (K870028) | #### DEVICE DESCRIPTION Top Neuropole Needles are teflon insulated injection needles for safe thermolesion procedures and regional anesthesia. There are 5 models offered, the ST, X, RC, XE and TL. The Top Neuropole Needle models TL, XE and RC are for pain management, while models ST and X are for local anesthesia. The needles are available in multiple sizes. The needle gauge ranges from 20 to 24 and the length from 30 to 200 mm. They are for single use only and supplied sterile. They may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. The nerve is localized either through the needle or by injecting contrast medium through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic or a radiofrequency lesion may be made. #### INTENDED USE The Top Neuropole Needles are used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. {1}------------------------------------------------ # TECHNOLOGICAL CHARACTERISTICS Top Neuropole Needles have the same device characteristics, materials, dimensions and intended use as the predicate device. Top Neuropole Needles have been tested to ensure the devices comply with applicable industry standards and US regulations ## CONCLUSIONS The intended use and performance characteristics of the Top Neuropole Needles are the same as the predicate and raise no new questions of safety and effectiveness. The Top Neuropole Needles are substantially equivalent to the predicate device. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Top Corporation % Cambria Regulatory Consulting, Inc. Ms. Cathryn N. Cambria 5536 Trowbridge Drive Dunwoody, Georgia 30338 FEB 1 5 2007 Re: K062946 Trade/Device Name: Top Neuropole Needles Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency lesion probe Regulatory Class: II Product Code: GXI Dated: February 2, 2007 Received: February 5, 2007 Dear Ms. Cambria: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Cathryn N. Cambria This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Page 1 of 1 #### 510(k) Number (if known): K062946 Device Name: Top Neuropole Needles # Indications for Use: The Top Neuropole Needles are injection needles used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made. # (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | X | OR | Over-The-Counter | |----------------------|---|----|------------------| | Use | | | | | (Per 21 CFR 801.109) | | | | (Division Sign-O) Division of Gene al, Restorative, and Neurological Devices | 510(k) Number | 1062946 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...