NOBELPERFECT CONICAL CONNECTION
Device Facts
| Record ID | K062936 |
|---|---|
| Device Name | NOBELPERFECT CONICAL CONNECTION |
| Applicant | Nobel Biocare AB |
| Product Code | DZE · Dental |
| Decision Date | Dec 19, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Nobel Biocare's NobelPerfect Conical Connection implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The NobelPerfect Conical Connection implants are indicated for single or multiple unit restorations. The NobelPerfect Conical Connection implants can be used in splinted or non-splinted applications. The NobelPerfect Tapered Conical Connection implant may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
Device Story
Threaded, root-form dental implants; machined from commercially pure titanium (grade 4); feature tapered body and TiUnite surface treatment. Surgically placed in jaw bone by dental professionals to support prosthetic devices (artificial teeth). Restores esthetics and chewing function. Combines design features of NobelPerfect Groovy implants with internal conical connection of SFB implants. Supports immediate loading protocols contingent on stability requirements.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and technological similarities to legally marketed predicate devices.
Technological Characteristics
Material: Commercially pure titanium, grade 4. Surface: TiUnite. Design: Threaded, root-form, tapered body. Connection: Internal conical connection. Energy source: None (mechanical device).
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple unit dental restorations in upper or lower jaw arches. Suitable for splinted or non-splinted applications. Immediate placement and function permitted if initial stability criteria are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.