← Product Code [JJY](/productcode/JJY) · K062916

# CLINIQA LIQUID QC CARDIAC MARKER CONTROL SP, LEVELS 1,2, &3 (K062916)

_Cliniqa Corporation · JJY · Oct 23, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062916

## Device Facts

- **Applicant:** Cliniqa Corporation
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Oct 23, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, & 3 are intended for use as assayed quality control material for CK-MB, hs-CRP, Digoxin, Myoglobin, cTnl, Troponin T, and NT ProBNP analysis.

## Device Story

CLINIQA Liquid QC Cardiac Marker Control SP is a liquid, ready-to-use quality control material provided in three levels. It is used by clinical laboratory professionals to monitor the precision and accuracy of analytical methods for cardiac markers, including CK-MB, hs-CRP, Digoxin, Myoglobin, cTnl, Troponin T, and NT ProBNP. The control is processed similarly to patient samples on clinical chemistry analyzers. By comparing the measured values of the control against established target ranges, laboratory personnel verify the performance of their diagnostic systems. This ensures the reliability of patient test results, aiding in the clinical assessment of cardiac conditions.

## Clinical Evidence

No clinical data; bench testing only. The device is a quality control material; performance is established through analytical verification of target values and stability testing.

## Technological Characteristics

Liquid, ready-to-use, multi-level (1, 2, & 3) control material. Contains human-serum based matrix with specific cardiac marker analytes (CK-MB, hs-CRP, Digoxin, Myoglobin, cTnl, Troponin T, NT ProBNP). Designed for use with clinical chemistry analyzers. No electronic or software components.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k062916

B. Purpose for Submission:
New Device

C. Measurand:
Cardiac Markers: CK-MB, hs-CRP, Digoxin, Myoglobin, cTnI, Troponin T, and NT ProBNP

D. Type of Test:
Quality control material

E. Applicant:
Cliniqa Corporation

F. Proprietary and Established Names:
CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1660 Quality control material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJY, multi-analyte controls, all kinds (assayed and unassayed)
4. Panel:
Clinical Chemistry (75)

H. Intended Use:
1. Intended use(s):
See Indication(s) for use section below.
2. Indication(s) for use:
CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3 are intended for use as assayed quality control material for CK-MB, hs-CRP, Digoxin, Myoglobin, cTnI, Troponin T, and NT ProBNP analysis.
3. Special conditions for use statement(s):
Prescription Use Only
4. Special instrument requirements:
The controls material is assayed for the following instruments:
CK MB – Beckman Access

{1}

hs-CRP - Roche Integra

Digoxin - Beckman Access

Myoglobin - Beckman Access

cTnI - OCD Vitros ECi

TnT - Roche Elecsys

NT ProBNP - OCD Vitros ECi

# I. Device Description:

CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3 are human serum based, containing constituents of human origin and purified biochemicals.

Preservatives, stabilizers and sodium azide have been added to maintain product integrity. CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3 are provided in liquid form, and are ready to use.

# J. Substantial Equivalence Information:

1. Predicate device name(s): MAS Cardioimmune XL, Microgenics Corporation.
2. Predicate 510(k) number(s): k061196
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Cardiac Marker Control | Cardiac Marker Control  |
|  Origin | Human serum | Human serum  |
|  Analyte | CK MB hs-CRP Myoglobin cTroponin I Troponin T NT ProBNP | CK MB hs-CRP Myoglobin Troponin I Troponin T NT ProBNP  |
|  Stability (opened) | 30 days at 2-8 °C for: hs-CRP Myoglobin cTroponin I | 30 days at 2-8 °C for: hs-CRP Myoglobin Troponin I  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Analyte | Digoxin | Digitoxin  |
|   |  Not Present | BNP  |
|   |  Not Present | homocysteine  |
|  Stability (opened) | 30 days at 2-8 °C for: | 15 days at 2-8 °C for:  |
|   |  CK MB | CK MB  |
|   |  Troponin T | Troponin T  |
|   |  NT ProBNP | NT ProBNP  |

{2}

K. Standard/Guidance Document Referenced (if applicable):
Not applicable.

L. Test Principle:
Not applicable.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability
The sponsor has not provided any information regarding the traceability of the values assigned to the product. Assays used to establish the assignment of values will be run by at least two laboratories. A minimum of 12 data points will be used to determine the mean (expected) value. Within and between assay standard deviations and coefficient of variations (CV) will be calculated for each set of data. The Dixon Method for removing outliers from data sets of 3 through 25 observations will be used to analyze data sets with CVs greater than 10%. No more than 10% of a set of data will be removed as a statistical outlier. The resulting data will be statistically averaged to obtain a representative expected value for each constituent. All values are assigned with the instrument manufacturer's reagents available at the time of assay.

Stability
Stability characteristics of the CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3 were determined using the Arrhenius model of accelerated elevated temperature studies to predict estimated storage stability at 2 - 8 °C. All samples were tested on the Roche Integra, Ortho Clinical Diagnostics Vitros ECi, Roche Elecsys, or Beckman Coulter Access. The data submitted supports an open vial stability of 30 days and an unopened vial stability of 2 years at 2-8 °C for the CLINIQA Liquid QC Cardiac Marker Control SP, Levels 1, 2, &amp; 3. Real time stability studies are ongoing.

Expected Values
The expected values are derived from replicate analyses of representative samples of the product and are specific to each lot. Consensus testing data used to establish the expected values were derived from multiple laboratories. All values are assigned with the instrument manufacturer's reagents available

{3}

at the time of assay.

d. Detection limit:
Not applicable.

e. Analytical specificity:
Not applicable.

f. Assay cut-off:
Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:
Not applicable.

b. Matrix comparison:
Not applicable.

3. Clinical studies:

a. Clinical Sensitivity:
Not applicable.

b. Clinical specificity:
Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

5. Clinical cut-off:
Not applicable.

6. Expected values/Reference range:
Not applicable.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K062916](https://fda-staging.innolitics.com/device/K062916)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
