The Pajunk EpiLong Sets Pediatric are intended for delivery of continuous conduction of epidural anesthesia. The catheter has to be removed or replaced after 72 hours. They are to be used with pediatric patient population. Prescription Use (Per 21 CFR 801.109)
Device Story
Pajunk EpiLong Sets Pediatric are single-use, sterile, non-pyrogenic kits for continuous epidural anesthesia in pediatric patients. Kits include an epidural needle (e.g., Tuohy, optionally NanoLine/lacquer coated), a polyamide indwelling catheter, LOR syringe, flat filter, and Tuohy Borst adaptor. Catheters feature depth graduations, closed or open tips, and optional integrated spirals or stylets. Used by clinicians in clinical settings to access the epidural space and deliver anesthesia. The needle is used for access and then withdrawn; the catheter remains in the epidural space for up to 72 hours. Depth markings allow for precise positioning. The device facilitates longer anesthetic effects compared to single-shot techniques, benefiting pediatric patients requiring extended pain management.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics, design, and literature review.
Technological Characteristics
Kits include epidural needles (Tuohy, optional NanoLine/lacquer coating), polyamide catheters with depth graduations, LOR syringes, flat filters, and Tuohy Borst adaptors. Catheters may include integrated spirals or stylets. Sterilization via EtO. Components are standard anesthesia conduction kit parts.
Indications for Use
Indicated for pediatric patients requiring continuous epidural anesthesia. Contraindicated for use beyond 72 hours without removal or replacement.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{0}------------------------------------------------
K062902
Premarket Notification Submission
Image /page/0/Picture/3 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a dark background. Below that, in smaller letters, is the word "MEDIZINTECHNOLOGIE".
JUN - 7 2007
# 510(k) Premarket Notification Submission:
## Summary of Safety and Effectiveness
Date of Preparation: May 2nd , 2007
| Submitter Information/ production site: | USA Contact: |
|----------------------------------------------|--------------------------------|
| Pajunk GmbH Medizintechnologie | Pajunk Medical Systems |
| Karl-Hall-Strasse 01 | German American Trade Center |
| 78187 Geisingen | 5126 South Royal Atlanta Drive |
| Germany | 30084 Tucker, Georgia |
| Fon: +49(0)7704-9291-0 | USA |
| Fax: +49(0)7704-9291-610 | |
| Establishment Registration Number: 9611612 | |
| Contact: | Contact |
| Christian Quass, Director Regulatory Affairs | Stefan Dayagi |
| Fon: +49(0)7704-9291-586 | Fon: +01(0)770-493-9305 |
| Fax: +49(0)7704-9291-605 | |
E-Mail: stefan.dayagi@pajunk-usa.com
Contract Sterilizer: Sterigenics Germany GmbH FDA-ERN: 3002807090
## Device Information:
E-Mail: christian.quass@pajunk.com
Trade Names: Pajunks EpiLong Set Pediatric, EpiLong Soft Pediatric, EpiLong Pediatric Kit acc. Marhover Common Name: Pediatric Epidural Anesthesia Sets Pediatric
- Classification Name: Anesthesia Conduction Kit
Classification 21 CFR § 868.5140, April 1, 2005,
Reference:
Poposed Regulatory Class II Classification:
Proposed Product CAZ, Kit, Conduction Anesthesia Classification Code:
### Panel: Anesthesiology
Predicate Devices:
- 1. Pajunks EpiLong Set K060311 2. Portex Pediatric Epidural and Peripheral Block Anesthesia Catheter K033080
- 3. MEDDEVICE's BIO-MATE pediatric epidural catheter K901129
- 4. BBrauns PERIFIX pediatric epidural kit K962696
-008-
{1}------------------------------------------------
# Premarket Notification Submission
Image /page/1/Picture/2 description: The image shows the word "PAJUNK" in large, bold, white letters. Below that, in smaller letters, is the word "MEDIZINTECHNOLOGIE". The background of the image is a textured pattern of black and white.
#### Device Description
Pajunk's EpiLong Anesthesia Sets Pediatric are single use, sterile, non-pyrogenic and latex free medical device kits for use with pediatric patients.
They are intended for continuous epidural anesthesia deliver y using the Polyamide indwelling catheter. The catheter has to be removed or replaced after 72 hours.
The set is equipped with an epidural needle, for example the Tuohy. The Tuohy needle is a proven and established needle variant used in epidural anaesthesia.
PAJUNK® offers the catheters for pediatric use in different lenghts and diameters. The catheter itself is equipped with ascending de pth graduations. With these depth gradua tions the position of the catheter can be exactly determined at any time. A marker at the end of the catheter indicates how deep the catheter can be introduced. Optionally the catheter can be employed with a stylet, an integrated spiral (called Epilong Soft Pediatric) or open tip.
#### Predicate Devices
The devices Pajunk claims substantial equivalence with is Pajunks Epilong Set cleared under K060311, BBrauns Perifix Pediatric epidural kit K962696, Portex´s Pediatric Epidural and Peripheral Block Anesthesia Catheter K033080, MEDDEVICE 's Bio-mate pediatric epidural catheter K901129.
The indication for use of Pajunks EpiLong Sets already cleared for market for an unspecified patient population is enhanced for pediatric patients.
The detailed discussion of substantial equivalence can be fo und in Section 12 of this submission. Because there is no change in technology or material the focus is set on the clinical literature review in section 10 of this submission.
#### Sterilization
#### Sterilization method: EtO
The contract sterilizer and the sterilizing process are identical to the ones used for all of Pajunks devices provided sterile, especially the anesthesia conduction devices cleared for the US market in several Premarket Notification submissions.
By annual validation and quarterly verification as well as by shelflife testing the sterilization procedure is claimed to be safe and effective for several years now.
#### Technology Characteristics:
The Pediatric epidural Anesthesia Set provides a standard epidural needle. Variant: The EpiLong Marhofer Kit. Within this Kit the needle is (NanoLine or lacqueur) coated.
The catheter comes with a catheter container introductory aid for better handling and shape security, a LOR syringe, a flat filter and a Tuohy Borst adaptor. The catheter is closed at tip and equipped with three lateral holes, optional with open tip, an integrated spiral (for enhanced stability) and a stylet. There is no change in components compared to the EpiLong sets already cleared for market.
All components are available seperately.
#### Conclusion:
The comparison between the predicate devices and the proposed devices in section 12.0 of this submission demonstrates that the proposed devices are at least as safe and effective as, and substantially equivalent to the predicate devices.
```
AI-K062902
```
Al-III-mod.doc
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN -7 2007
Mr. Christian Quass Director Regulatory Affairs, Ass. Safety Official PAJUNK® GmbH Medizintechnologie Karl-Hall-Strasse 01 78187 Geisingen, GERMANY
Re: K062902
Trade/Device Name: Pajunks EpiLong Sets Pediatric, Pajunk EpiLong Sets Pediatirc acc. Marhover, Pediatric Epidural Catheter, Pajunks Pediatric Epidural Needles
Regulation Number: 21 CFR 868.5140
Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ
Dated: May 31, 2007
Received: June 04, 2007
Dear Mr. Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Quass
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
CARS
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
# Premarket Notification Submission
Image /page/4/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a dark background. Below this, the word "MEDIZINTECHNOLOGIE" is written in smaller, white letters. The logo appears to be for a medical technology company.
## Indications for use
| 510(k) Number: | |
|----------------|--|
| Device Name: | |
K062902 Pajunks EpiLong Sets Pediatric
#### Indications for Use:
The Pajunk EpiLong Sets Pediatric are intended for delivery of continuous conduction of epidural anesthesia. The catheter has to be removed or replaced after 72 hours.
They are to be used with pediatric patient population.
Prescription Use (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
sion Sign-Off) l Vision of Anesthesiology, General Hospital, icrection Control, Dental Devices
510(k) Number. K062902
Page 1 of 1
A1-K062902
-003-
{5}------------------------------------------------
## Premarket Notification Submission
Image /page/5/Picture/2 description: The image shows the logo for PAJUNK MEDIZINTECHNOLOGIE. The word "PAJUNK" is in large, bold, white letters against a dark background. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, black letters.
## Indications for use
| 510(k) Number: | |
|----------------|--|
| Device Name: | |
K062902 Pajunks EpiLong Sets Pediatric acc. Marhover
Indications for Use:
The Pajunk EpiLong Sets Pediatric acc. Marhover are intended for delivery of continuous conduction of epidural anesthesia. The Marhover-Set contains a NanoLine coated cannula. The catheter has to be removed or replaced after 72 hours.
They are to be used with ped iatric patient population.
Prescription Use (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
ാന Sign-Off) sion of Anesthesiology, General Hospital, Intection Control, Dental Devices
510(K) Number K062603
Page 1 of 1
Al-III-mod.doc
{6}------------------------------------------------
## Premarket Notification Submission
Image /page/6/Picture/2 description: The image shows the word "PAJUNK" in large, bold, white letters against a dark background. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, dark letters. The image appears to be a logo or brand name.
#### Indications for use
| 510(k) Number: | K062902 |
|----------------|-----------------------------|
| Device Name: | Pediatric epidural catheter |
#### Indications for Use:
The Pajunk pediatric epidural catheter is placed in the epidural space to facilitate a longer anesthetic effect.
It is intended for use with pediatric patients.
After the anesthesia conduction needle has be en withdrawn from the patient, the catheter tip can remain in the epidural space for as long as determined by the professional anesthetist and the instructions for use. It has to be removed or replaced after 72 hours.
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ion Sign-Off) sion of Anesthesiology, General Hospital, control. Dental Devices
Number K062902
-005-
Al-K062902
Al-III-mod.doc
{7}------------------------------------------------
## Premarket Notification Submission
Image /page/7/Picture/2 description: The image contains the word "PAJUNK" in large, bold, white letters against a dark background. Below this, the word "MEDIZINTECHNOLOGIE" is written in smaller, black letters. The text appears to be a logo or heading, possibly related to medical technology.
#### Indications for use
#### 510(k) Number: Device Name:
Pajunks Pediatric Epidural needles
#### Indications for Use:
Pajunks pediatric epidural needles are anesthesia conduction needles e mployed to identify and to access epidural space.
It is intended for use with pediatric patients.
After epidural space is accessed and the catheter is inserted the anesthesia conduction needle has to be withdrawn from the patient.
Prescription Use *X*
(Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qld
்பு Sign-Off) of Anesthesiology, General Hospital. Stion Control, Dental Devices
3(k) Number ...
#### Page 1 of 1
-006-
Al-III-mod.doc
Al-K062902
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.