K062881 · Collagen Matrix, Inc. · NPL · Oct 11, 2006 · Dental
Device Facts
Record ID
K062881
Device Name
COLLAGEN DENTAL MENBRANE - CONFORMABLE II
Applicant
Collagen Matrix, Inc.
Product Code
NPL · Dental
Decision Date
Oct 11, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagen Dental Membrane-Conformable II is intended for use in dental surgery procedures as a material for placement in the area of dental implant, bone defect or ridge reconstruction to aid in wound healing post dental surgery.
Device Story
Collagen Dental Membrane-Conformable II is a white, nonfriable, resorbable membrane matrix derived from highly purified bovine Achilles tendon type I collagen. Designed for dental surgery, the membrane is flexible and conforms to the contours of defect sites. It is supplied sterile, non-pyrogenic, and for single-use. The device acts as a barrier or scaffold to aid in wound healing during dental implant, bone defect, or ridge reconstruction procedures. It is intended for use by dental surgeons in a clinical setting.
Clinical Evidence
Bench testing only. In vitro product characterization studies and biocompatibility testing per ISO 10993-1 and FDA Blue Book Memorandum G95-1 were performed to demonstrate safety and substantial equivalence.
Technological Characteristics
Resorbable membrane matrix; material: highly purified type I collagen from bovine Achilles tendon; form: flexible, nonfriable, conformable sheet; sterilization: sterile; single-use; non-pyrogenic.
Indications for Use
Indicated for use in oral surgical procedures as a resorbable material for placement in the area of dental implant, bone defect, or ridge augmentation to aid in wound healing.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K062881
# 510(k) Summary of Safety and Effectiveness
# OCT 1 1 2006
| Applicant Name and Address: Collagen Matrix, Inc. | 509 Commerce Street<br>Franklin Lakes, New Jersey 07417 |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Peggy Hansen, RAC<br>Sr. Director, Clinical, Regulatory, and Quality Assurance<br>Tel: (201) 405-1477<br>Fax: (201) 405-1355 |
| Date of Summary: | September 25, 2006 |
| Device Common Name: | Collagen Dental Membrane |
| Device Trade Name: | Collagen Dental Membrane - Conformable II |
| Device Classification Name: | Bone Grafting Material<br>Class II<br>872.3930<br>LYC |
| Predicate Device(s): | Collagen Dental Membrane, K011695 |
#### Description of the Device
Collagen Dental Membrane-Conformable II is a white, nonfriable, conformable, resorbable, membrane matrix engineered from highly purified type I collagen derived from bovine Achilles tendon. It is flexible and conforms to the contours of the defect site. Collagen Dental Membrane-Conformable II is supplied sterile, non-pyrogenic, in various sizes, and for single use only.
#### Intended Use
Collagen Dental Membrane-Conformable II is intended for use in dental surgery procedures as a material for placement in the area of dental implant, bone defect or ridge reconstruction to aid in wound healing post dental surgery.
## Summary/Comparison of Technical Characteristics
Collagen Dental Membrane-Conformable II and its predicate have similar technological characteristics. In particular, the Collagen Dental Membrane-Conformable II and its predicate are similar with respect to intended use, material, form, sizes, thickness, physical integrity, permeability and conformability.
{1}------------------------------------------------
### Safety
A Collagen Dental Membrane-Conformable II equivalent has been evaluated by a I Contagen Domar Messa its safety/biocompatibility. The device passed all applicable FDA Blue Book Memorandum G95-1and ISO 10993-1 testing for the biological evaluation of medical devices.
### Conclusion
The results of the in vitro product characterization studies show that the device Firstitution of the Collagen Dental Membrane-Conformable II is safe and substantially equivalent to the original device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville, Maryland 20850
OCT 1 0 2007
Ms. Peggy Hansen Senior Director, Clinical, Regulatory and Quality Assurance Collagen Matrix, Incorporated 509 Commerce Street Franklin Lakes, New Jersey 07417
Re: K062881
Trade Name: Collagen Dental Membrane-Conformable II Regulation Number: 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: 2 Product Code: NPL Dated: September 25, 2006 Received: September 27, 2006
Dear Ms. Hansen:
This letter corrects our substantially equivalent letter of October 11, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the require apple to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address 1 http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Syrite Y. Michieoms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Image /page/3/Picture/7 description: The image is a circular logo or emblem. The text "1906 - 2006" is at the top of the circle. The letters "FDA" are in the center of the circle in a bold, stylized font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are below the word "Centennial".
Protecting and Promoting Public Health
{4}------------------------------------------------
# Indications for Use
K062881
510(k) Number (if known):
Collagen Dental Membrane - Conformable II Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
Collagen Dental Membrane – Conformable 11 is intended for use in oral surgical Collegon Donal resorbable material for placement in the area of dental implant, bone defect, or ridge augmentation to aid in wound healing.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Byrne
on Sign-Off) of Anesthesiology, General Hospital, . Non Control, Dental Devis
Number:
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.