System reagent for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on OLYMPUS analyzers Measurements of Uric Acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
Device Story
Olympus Uric Acid Reagent is an in vitro diagnostic reagent used on Olympus clinical chemistry analyzers. It enables quantitative measurement of uric acid levels in human serum, heparinized plasma, and urine samples. The device is intended for use by clinical laboratory professionals to assist in the diagnosis and management of renal and metabolic conditions, including gout and renal failure, and to monitor patients undergoing cytotoxic therapy. The reagent facilitates automated biochemical analysis within a clinical laboratory setting, providing quantitative data that informs clinical decision-making regarding patient metabolic status and treatment efficacy.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
In vitro diagnostic reagent for quantitative uric acid determination. Designed for use on Olympus automated clinical chemistry analyzers. Principle of operation involves enzymatic or chemical reaction for uric acid quantification in serum, plasma, or urine.
Indications for Use
Indicated for quantitative determination of uric acid in human serum, heparinized plasma, and urine to aid in diagnosis/treatment of renal and metabolic disorders (e.g., renal failure, gout, leukemia, psoriasis, wasting conditions) and monitoring patients on cytotoxic drugs.
Regulatory Classification
Identification
A uric acid test system is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
Submission Summary (Full Text)
{0}
510(K) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062862
B. Purpose for Submission
New device
C. Measurand:
Uric acid
D. Type of Test:
Quantitative, enzymatic, colorimetric assay
E. Applicant:
Olympus America Inc.
F. Proprietary and Established Names:
Olympus Uric Acid Reagent
G. Regulatory Information:
1. Regulation Section
21 CFR §862.1775, uric acid test system
2. Classification:
Class I, Reserved
3. Product Code:
KNK
4. Panel
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s)
See Indications for use below.
2. Indication(s) for use:
Olympus Uric Acid Reagent is for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on OLYMPUS analyzers.
Measurements of Uric Acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
3. Special conditions for use statement(s):
For Prescription Use only
4. Special instrument requirements:
AU400/400°, AU600/640/640°, AU2700/5400 system
1
{1}
2
I. Device Description:
This *in vitro* diagnostic reagent is packaged as two reagents, R1 and R2. The final concentrations of the reactive ingredients are presented in the table below:
Phosphate buffer (pH 7.5) 42 mmol/L
Peroxidase ≥ 5.9 kU/L
MADB 0.15 mmol/L
4-Aminophenazone 0.30 mmol/L
EDTA 0.44 mmol/L
Uricase ≥ 250 U/L
Preservatives
J. SUBSTANTIAL EQUIVALENCE INFORMATION
1. Predicate device name(s):
Olympus Uric Acid Reagents
2. Predicate 510(k) number(s):
k961274
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Method Principle | This Olympus Uric Acid procedure is a modification of the Fossati method. Uric acid is converted by uricase to allantoin and hydrogen peroxide. Hydrogen peroxide reacts with 4-aminoantipyrine (4-AAP) in the presence of N,N-bis(4-sulfobutyl)-3,5-dimethylaniline, disodium salt (MADB) to produce a chromophore which is read bichromatically at 660/800 nm. The amount of dye formed is proportional to the uric acid concentration in the sample. | This Olympus Uric Acid procedure is a modification of the Fossati method. Uric acid is converted by uricase to allantoin and hydrogen peroxide. Hydrogen peroxide reacts with 4-aminoantipyrine (4-AAP) in the presence of N,N-bis(4-sulfobutyl)-3,5-dimethylaniline, disodium salt (MADB) to produce a chromophore which is read bichromatically at 660/800 nm. The amount of dye formed is proportional to the uric acid concentration in the sample. |
| Calibrator Traceability | National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 909b | National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 909b |
| On-board stability | 30 days | 30 days |
{2}
3
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | System reagent is for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on OLYMPUS analyzers. | System reagent is for the quantitative determination of Uric Acid in human serum, and urine on OLYMPUS analyzers. |
| Specimen | Serum, plasma, & urine | Serum, & urine |
# K. STANDARD/GUIDANCE DOCUMENT REFERENCED (IF APPLICABLE)
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline.
# L. Test Principle
This Olympus Uric Acid procedure is a modification of the Fossati method. Uric acid is converted by uricase to allantoin and hydrogen peroxide. Hydrogen peroxide reacts with 4-aminoantipyrine (4-AAP) in the presence of N,N-bis(4-sulfobutyl)-3,5-dimethylaniline, disodium salt (MADB) to produce a chromophore which is read bichromatically at 660/800 nm. The amount of dye formed is proportional to the uric acid concentration in the sample.
# M. Performance Characteristics (if/when applicable):
1. Analytical Performance:
a. Precision /Reproducibility:
Precision was performed by testing four sample pools two-fold twice a day for twenty days. The within-run and total CVs were calculated according to CLSI EP5-A protocol. The within run precision for serum samples is less than 2% and total precision is less than 3% on the AU400/400e, AU600/640/640e, and AU2700/5400 instruments. The within run precision for urine samples is less than 3% and total precision is less than 5% on the AU400/400e, AU600/640/640e, and AU2700/5400 instruments. The results are shown in the table below:
AU400/400ᵉ Serum
| N = 80 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 3.9 | 0.02 | 0.53 | 0.05 | 1.31 |
| 7.3 | 0.05 | 0.64 | 0.07 | 0.98 |
| 9.8 | 0.06 | 0.59 | 0.12 | 1.17 |
AU600/640/640ᵉ Serum
| N = 80 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 3.8 | 0.01 | 0.38 | 0.03 | 0.80 |
| 7.2 | 0.03 | 0.43 | 0.06 | 0.89 |
| 9.7 | 0.05 | 0.53 | 0.08 | 0.85 |
{3}
AU2700/5400 Serum
| N = 80 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 3.8 | 0.02 | 0.48 | 0.06 | 1.49 |
| 7.3 | 0.02 | 0.34 | 0.11 | 1.52 |
| 9.8 | 0.05 | 0.50 | 0.16 | 1.60 |
$\mathrm{AU400 / 400^{e}}$ Urine
| N = 80 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 21.5 | 0.23 | 1.09 | 0.36 | 1.66 |
| 58.0 | 1.06 | 1.84 | 1.37 | 2.36 |
| 88.4 | 1.80 | 2.03 | 1.91 | 2.16 |
AU600/640/640e Urine
| N = 100 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 21.8 | 0.24 | 1.12 | 0.35 | 1.61 |
| 58.4 | 0.95 | 1.63 | 1.33 | 2.28 |
| 89.3 | 1.78 | 2.00 | 2.01 | 2.25 |
AU2700/5400 Urine
| N = 80 | Within run | | Total | |
| --- | --- | --- | --- | --- |
| Mean, mg/dL | SD | CV% | SD | CV% |
| 21.5 | 0.25 | 1.17 | 0.37 | 1.73 |
| 58.2 | 0.76 | 1.31 | 1.27 | 2.19 |
| 88.8 | 1.71 | 1.93 | 1.98 | 2.23 |
b. Linearity/ Assay reportable range.
A linearity study was conducted on the Olympus Uric Acid reagent using prepared standard solutions. Samples were prepared by diluting a high pool containing 33.49 mg/dL with a 0 mg/dL sample and assayed. The linearity was established for serum as 1.5 to 30 mg/dL. For urine linearity, samples were prepared by diluting a high pool containing 123 mg/dL with a 0 mg/dL sample. The linearity was established for urine as 0.7 to 100 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators or methods)
The frequency of calibration is every 30 days. Calibration of the Olympus Uric Acid reagent is accomplished by use of the Olympus Chemistry Calibrator (Cat # DR0070 or DR0071), which is traceable to the National Institute of Standard and Technology (NIST) Standard Reference Material (SRM) 909b for serum and plasma specimens. For urine specimens the calibration is established using the Olympus Urine Calibrator (Cat # DR0090).
A stability study was conducted on the Olympus Uric Acid reagent whereby both fresh reagent and reagent at the end of its on-board stability (30 days) were assayed for uric acid recovery on prepared control material and to confirm that the linearity of the reagent was maintained after 30 days on-board,
{4}
The manufacturer recommends in the labeling that during operation of the Olympus analyzer at least two levels of an appropriate quality control material should be tested a minimum of once a day. In addition, these controls should be performed after calibration, with each new lot of reagent, and after specific maintenance or troubleshooting steps described in the appropriate Olympus User's Guide. Quality control testing should be performed in accordance with regulatory requirements and each laboratory's standard procedure.
## d. Detection Limit
The lowest detectable level represents the lowest measurable level of uric acid that can be distinguished from zero. It is calculated as the absolute mean plus three standard deviations of 20 replicates of an analyte free sample.
The lowest detectable level using serum settings was estimated as follows:
| Analyzer | Lowest Detectable Level (mg/dL) |
| --- | --- |
| AU600/640/640^{e} | 0.02 |
| AU400/400^{e} | 0.02 |
| AU2700/5400 | 0.02 |
## Functional Sensitivity:
### Serum:
Precision results (40-fold determination) for a level <0.8 mg/dL are shown below with a CV of <20% for each application.
| Analyzer | Mean Concentration (mg/dL) | SD | CV% |
| --- | --- | --- | --- |
| AU600/640/640^{e} | 0.24 | 0.013 | 5.5 |
| AU2700/5400 | 0.24 | 0.015 | 6.2 |
| AU400/400^{e} | 0.23 | 0.012 | 5.4 |
### Urine:
Precision results (40-fold determination) for a level <3 mg/dL are shown below with a CV of <20% for each application.
| Analyzer | Mean Concentration (mg/dL) | SD | CV% |
| --- | --- | --- | --- |
| AU600/640/640^{e} | 0.64 | 0.058 | 9.0 |
| AU2700/5400 | 0.64 | 0.038 | 5.9 |
| AU400/400^{e} | 0.63 | 0.053 | 8.5 |
## e. Analytical Specificity
A study of common interfering substances was conducted. The results of the study were evaluated following the recommendations of CLSI EP7-A on the AU400/400e, AU600/640/640e, and AU2700/5400. The following substances interfere with this uric acid method:
### AU400/400e
Ascorbic Acid: Interference less than 5% up to 20 mg/dL Ascorbate
Bilirubin: Interference less than 5% up to 40 mg/dL Bilirubin
Hemolysis: Interference less than 5% up to 500 mg/dL Hemolysate
Lipemia: Interference less than 5% up to 1000 mg/dL Intralipid
{5}
AU600/640/640
Ascorbic Acid: Interference less than 5% up to 20 mg/dL Ascorbate
Bilirubin: Interference less than 5% up to 40 mg/dL Bilirubin
Hemolysis: Interference less than 5% up to 500 mg/dL Hemolysate
Lipemia: Interference less than 5% up to 1000 mg/dL Intralipid
AU2700/5400
Ascorbic Acid: Interference less than 5% up to 20 mg/dL Ascorbate
Bilirubin: Interference less than 5% up to 40 mg/dL Bilirubin
Hemolysis: Interference less than 5% up to 500 mg/dL Hemolysate
Lipemia: Interference less than 5% up to 1000 mg/dL Intralipid
f. Assay Cut-off
Not applicable
2. Comparison Studies:
a. Method comparison with predicate device:
Serum
Patient samples were used to compare the Olympus Uric Acid on the AU600 to another Olympus analyzer Uric Acid method (Method 2). Further studies were conducted as outlined below:
| Y Method | AU600/640/640^{e} | AU400/400^{e} | AU2700/5400 |
| --- | --- | --- | --- |
| X Method | Method 2 | AU600/640/640^{e} | AU600/640/640^{e} |
| Slope | 1.014 | 0.973 | 0.980 |
| Intercept | -0.226 | 0.250 | 0.253 |
| Correlation Coeff. (r) | 1.000 | 1.000 | 1.000 |
| No. of Samples (n) | 122 | 122 | 122 |
| Range (mg/dL) | 1.9 – 29.4 | 1.7 - 29.4 | 1.7-29.4 |
Urine
Urine samples were used to compare this Olympus Uric Acid on the AU640 to another Olympus analyzer Uric Acid method (Method 2). Further studies were conducted as outlined below:
| Y Method | AU600/640/640^{e} | AU400/400^{e} | AU2700/5400 |
| --- | --- | --- | --- |
| X Method | Method 2 | AU600/640/640^{e} | AU600/640/640^{e} |
| Slope | 0.981 | 1.034 | 1.014 |
| Intercept | 0.496 | -0.199 | -0.195 |
| Correlation Coeff. (r) | 0.998 | 0.999 | 0.999 |
| No. of Samples (n) | 159 | 159 | 159 |
| Range (mg/dL) | 0.9 – 98.6 | 0.8 - 96.7 | 0.8 – 96.7 |
b. Matrix comparison
Matched serum and heparinized plasma samples were assayed on the AU640 Olympus analyzer to demonstrate that no matrix effect exists for this Olympus Uric acid reagent, between serum and heparinized plasma samples.
{6}
| Slope | 1.018 |
| --- | --- |
| Intercept | -0.241 |
| Correlation Coeff. (r) | 1.000 |
| No. of Samples (n) | 47 |
| Range (mg/dL) | 1.51 – 29.1 |
3. Clinical Studies:
a. Clinical Sensitivity: Not applicable
b. Clinical Specificity: Not applicable
c. Other Clinical Supportive Data (when a. and b. are not applicable) Not applicable
4. Clinical cut-off: Not applicable
5. Expected Values/ Reference Ranges: In the labeling the expected values are provided from Teitz, Guide to Laboratory Tests, 4th Edition, Saunders, 2006 as follows:
Serum: Adults:
Female: 2.3 - 6.6 mg/dL
Male: 4.4 - 7.6 mg/dL
Urine:
Female: 250 – 750 mg/24 hours
Male: 250 – 800 mg/24 hours
Excretion may decrease by 20 to 25% on a purine-free diet to less than 400 mg/24 hours. Expected values may vary with age, sex, diet and geographical location. Each laboratory should determine its own expected values as dictated by good laboratory practice.
N. Proposed Labeling:
The labeling is sufficient to and satisfies the requirements of 21 CFR Part 809.10
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.