SEJOY DIGITAL THERMOMETER, MT SERIES (MT-201, MT-301, MT-402, MT-403 AND MT-405)
K062784 · Sejoy Electronics & Instruments Co., Ltd. · FLL · May 4, 2007 · General Hospital
Device Facts
Record ID
K062784
Device Name
SEJOY DIGITAL THERMOMETER, MT SERIES (MT-201, MT-301, MT-402, MT-403 AND MT-405)
Applicant
Sejoy Electronics & Instruments Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
May 4, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The devices Model MT series (MT-201, MT-402, and MT-403) are intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the devices are reusable for clinical or home use on people of all ages. The device model MT-405 is intended to measure temperature orally, and the device is reusable for clinical or home use for children less than 4 years old.
Device Story
Digital thermometers (MT series) measure human body temperature; input via thermistor resistance change; principle of operation involves converting resistance change to frequency changes in an R-C oscillator circuit; temperature derived from pulse number measurement over time; output displayed to user; used in clinical or home settings by patients or clinicians; provides rapid, reliable temperature readings to assist in clinical decision-making.
Clinical Evidence
Controlled human clinical studies conducted on MT series. Data evaluated clinical bias, clinical uncertainty, and clinical repeatability per Sejoy Clinical Test Protocol. Results demonstrated performance consistent with predicate devices.
Technological Characteristics
Thermistor-based sensing; R-C oscillator circuit for signal conversion; digital display; battery-powered; reusable; models include standard, flexible, and pacifier form factors. Compliance with ASTM E1112, IEC 60601-1, and IEC 60601-1-2.
Indications for Use
Indicated for human body temperature measurement. MT-201, MT-301, MT-402, MT-403: oral, rectal, or axillary use for all ages. MT-405: oral use for children < 4 years. Reusable for clinical or home settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Digital Pacifier Thermometer Model NT-01 (K041694)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY - 4 2007
18/300/1800
្រះ
K062784
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92.
The assigned 510(k) number is: ___________________________
#### l. Submitter's Identification:
Name: SEJOY ELECTRONICS & INSTRUMENTS CO., LTD Address: 4th Floor, B2 Building, Feng-Tan-Lu Industrial Zone, West Lake District, Hangzhou, P. R. China
#### 2. Name of the Device:
| Digital thermometer Model MT Series | |
|-------------------------------------|--------|
| Including: Digital Thermometer | MT-201 |
| Basal Digital Thermometer | MT-301 |
| Flexible Digital Thermometer | MT-402 |
| Instant Flexible Thermometer | MT-403 |
| Pacifier Thermometer | MT-405 |
#### 3. Predicate Device Information:
For submitted devices MT-201 and MT-301: * BD DIGITAL THERMOMETER 510(k) number: K945427 Applicant: BD BECTON DICKINSON VACUTAINER SYSTEMS PREANALYIC Owner: BECTON, DICKINSON & CO.
For submitted devices MT-402 and MT-403:
* BD FLEXIBLE DIGITAL THERMOMETER 510(k) number: K902624 Applicant: BD BECTON DICKINSON VACUTAINER SYSTEMS PREANALYIC Owner: BECTON, DICKINSON & CO.
For submitted device MT-405:
* Digital Pacifier Thermometer Model NT-01
510(k) number: K041694
Applicant: Hangzhou Hua'an Medical & Health Instrument Company Limited Owner: Hangzhou Hua'an Medical & Health Instrument Company Limited
{1}------------------------------------------------
#### 4. Device Description:
Digital Thermometer Model MT series (MT-201, MT-301, MT-402, MT-403 and MT-405) enable fast and reliable measurements. These thermometers provide very high clinical accuracy, and have been designed to provide maximum userfriendliness.
The basic principle of these thermometers is that a change of thermistor resistance, caused by changes of temperature, is converted to changes of frequency of R-C oscillator circuit. Therefore, temperature can be given by measuring the frequency of the oscillator.
For a given time period by applying to R-C oscillator circuit, changes of temperature will correspond to changes of pulse number.
#### 5. Intended Use:
The devices Model MT series (MT-201, MT-402, and MT-403) are intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the devices are reusable for clinical or home use on people of all ages. The device model MT-405 is intended to measure temperature orally, and the device is reusable for clinical or home use for children less than 4 years old.
## 6. Comparison to Predicate Devices:
The device models MT-201, MT-301 are similar in design and intended use to the BD DIGITAL THERMOMETER, BECTON, DICKINSON & CO., differing only in physical dimensions, display resolution, battery replacement method.
The device models MT-402, MT-403 are similar in design and intended use to the BD FLEXIBLE DIGITAL THERMOMETER, BECTON, DICKINSON & CO., differing only in physical dimensions, display resolution, battery replacement method.
The device model MT-405 Pacifier Thermometer is similar in design and intended use to the Digital Pacifier Thermometer Model NT-01, Hangzhou Hua'an Medical & Health Instrument Company Limited, differing only in physical dimensions, battery replacement method.
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ASTM E I 112, as well as IEC60601-1 and IEC60601-1-2 requirements. Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(K) Submissions for Clinical Electronic Thermometers.
{2}------------------------------------------------
## 8. Discussion of Clinical Tests Performed:
Controlled human clinical studies were conducted using the digital thermometer model MI series. Clinical data was presented which evaluated clinical bias, clinical uncertainty and clinical repeatability per the Sejoy Clinical Test Protocol outline.
#### 9. Conclusions:
The digital thermometer model MT series, Model MT-201 and MT-301 have the same intended use and similar technological characteristics as the BD DIGITAL THERMOMETER.
The Model MT-402 and MT-403 have the same intended use and similar technological characteristics as the BD FLEXIBLE DIGITAL THERMOMETER.
Pacifier Thermometer MT-405 has the same intended use and similar technological characteristics as the Hua'an Medical & Health Instrument Co. LTD's Digital Pacifier Thermometer Model NT-01.
Moreover, information contained in this submission demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Digital Thermometer Model MT Series Thermometers are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The caduceus is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sejoy Electronics & Instruments Company Limited C/O Mr. Paul Ware President PW Resources, Incorporated 34 McNamara Street Stoughton, Massachusetts 02072
- 4 2007 MAY
Re: K062784
Trade/Device Name: Digital Thermometer Model MT Series Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Dated: May 4, 2007 Received: May 4, 2007
Dear Mr. Ware:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Ware
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Runne
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Digital Thermometer Model MT series Indications For Use:
The model MT series digital thermometers (MT-201, 301, 402 and 403) are intended to measure the human body temperature in regular mode orally, rectally or under the arm. The devices are reusable for clinical or home use on people of all ages.
The device model MT-405 is intended to measure oral temperature only. It is reusable for clinical or home use for children under the age of four.
Prescription Use AND/OR Over-The-Counter Use V (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Susa Bunne
esthesiology, General Hospital, trol, Dental Devices
Page 8 of 45
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.