← Product Code [JWY](/productcode/JWY) · K062783

# SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE (K062783)

_Trek Diagnostic Systems, Inc. · JWY · Oct 26, 2006 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062783

## Device Facts

- **Applicant:** Trek Diagnostic Systems, Inc.
- **Product Code:** [JWY](/productcode/JWY.md)
- **Decision Date:** Oct 26, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Sensititre® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of Haemophilus influenzae; Streptococcus pneumoniae and Streptococcus species.

## Device Story

Sensititre® HP MIC Susceptibility plates are multi-well plastic microtitre plates containing dried, stabilized antimicrobics for broth dilution susceptibility testing. The device accepts clinical isolates of Streptococcus spp. prepared in Mueller-Hinton broth with 2–5% lysed horse blood. The system detects bacterial growth via fluorescence; specific surface enzymes produced by the organism cleave a fluorogenic substrate, releasing a fluorophore. The device is used in clinical laboratories; readings are performed manually or via automated instrumentation (Sensititre® AutoReader or ARIS®). The resulting MIC (minimum inhibitory concentration) informs clinicians of the lowest antimicrobial concentration inhibiting bacterial growth, aiding in the selection of appropriate antibiotic therapy.

## Clinical Evidence

No clinical trials were performed. Evidence consists of in vitro bench testing comparing the Sensititre® MIC results to a reference method. Performance was evaluated by testing clinical isolates of Streptococcus species against Gemifloxacin, Linezolid, and Levofloxacin. Results demonstrated essential agreement and categorical agreement between the Sensititre® system and the reference method, supporting the addition of these organisms to the device's indications.

## Technological Characteristics

Multi-well plastic microtitre plates; dried, stabilized antimicrobics. Sensing principle: fluorescence detection of bacterial growth via enzymatic cleavage of fluorogenic substrates. Energy source: automated reader instrumentation. Connectivity: standalone or integrated with Sensititre® AutoReader/ARIS®. Software: automated growth detection algorithm.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k062783

B. Purpose for Submission:
To add additional organism groups to the antibiotics gemifloxacin, linezolid, and levofloxacin on the Sensititre® Haemophilus influenzae / Streptococcus pneumoniae (HP) MIC and Sensititre® Susceptibility Plates

C. Measurand:
|  Measurand: | Range μg/mL  |
| --- | --- |
|  Gemifloxacin | 0.03 – 0.5  |
|  Linezolid | 0.5 – 4  |
|  Levofloxacin | 0.5 – 16  |

D. Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST) growth based fluorescence

E. Applicant:
TREK Diagnostic Systems, Inc.

F. Proprietary and Established Names:
Sensititre® Haemophilus influenzae / Streptococcus pneumoniae (HP) MIC plates

G. Regulatory Information:
1. Regulation section:
866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
II
3. Product code:
JWY-manual readings of AST testing of &gt;16 hour incubation
LRG Automated readings of AST of &gt;16 hour incubation.
4. Panel:
83 Microbiology

H. Intended Use:
1. Intended use(s):
The Sensititre® Haemophilus influenzae / Streptococcus pneumoniae (HP) MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility

{1}

testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.

## 2. Indication(s) for use:

This application is for the addition of Streptococcus species (spp.) to gemifloxacin (0.03 – 0.5 ug/mL), linezolid (0.5 – 4 ug/mL), and levofloxacin (0.5 – 16 ug/mL) for use with the Sensititre® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility Plates.

## 3. Special conditions for use statement(s):

For prescription use only

## 4. Special instrument requirements:

Automated readings are performed on the Sensititre® AutoReader or Sensititre® ARIS®.

## I. Device Description:

Sensititre® MIC Susceptibility plate MIC panels are multi-well plastic microtitre plates, precision dosed with dried, stabilized antimicrobics. This is a microversion of the classic broth dilution methods and can provide both qualitative and quantitative susceptibility results. The medium required for testing S. pneumoniae is Sensititre® Mueller-Hinton (MH) broth with 2 – 5% lysed horse blood with a final organism density of 5 X 10⁵ colony units (CFU/mL). This is then incubated in a non CO₂ incubator for 20 – 24 hour and read manually for growth or using the Sensititre® autoread.

## J. Substantial Equivalence Information:

1. Predicate device name(s):

Pasco MIC and MIC/ID Panels

2. Predicate 510(k) number(s):

K033119

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | an in vitro diagnostic product for clinical susceptibility testing of gram negative and gram positive organisms. | Same  |
|  Inoculum | Prepared from colonies using the direct inoculation method | Same  |

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Growth medium | Mueller – Hinton (MH) broth with 2 – 5% lysed horse blood | Same  |
|  Inoculation method | Direct equated to a 0.5 McFarland | Same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Type panel | Dried antibiotics | 100 μl/well frozen  |
|  Incubation | 20 - 24 hours | 16-24 hours  |
|  Technology | Fluorescence detection of growth | Turbidity detection of growth  |
|  Reading method | Visual growth and Auto read by instrumentation | Turbidity detection of growth  |

# K. Standard/Guidance Document Referenced (if applicable):

"Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S16) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard."

# L. Test Principle:

The Sensititre® Autoread System utilizes fluorescence technology to read 24 hour Streptococcus pneumoniae plates. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a non-fluorescent (fluorogenic) substrate. The non-fluorescent substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond, which prevents fluorescence. The fluorophore, is then said to be quenched. The substrate is added to the inoculum broth and dispensed into the test plates at the same time as the test organism or the plates can be prepared with the substrate already added to the plate. Enzymatic action of the bacterial surface enzymes on the specific substrates, cleave this bond releasing the fluorophore, which is now capable of fluorescence. The amount of fluorescence detected is directly related to the activity of bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism or automatically for the detection of fluorescence.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

a. Precision/Reproducibility:

A reproducibility study was performed on 25 Streptococcus spp. These isolates were tested once for each antimicrobial at each of the three sites on

{3}

the automated and manual read methods demonstrating &gt;95% reproducibility for both read methods.

b. Linearity/assay reportable range:
Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The recommended QC isolate, S. pneumoniae 49619 was tested daily with acceptable results. Quality control was also performed at all sites using both manual and autoread methods. The Sensititre® results demonstrated that the system can produce QC results in the recommended range for both manual and automated read methods. The mode is the same between the reference method and the two read methods.

An additional QC organism S. pneumoniae #5 was tested to provide on-scale results. The reference method was in range for every day tested.

Quality Control Table

|  Antimicrobial | ORGANISM | Conc ug/mL | Sensititre® Autoread | Sensititre® manual | Reference  |
| --- | --- | --- | --- | --- | --- |
|  |   |   |   |   |   |
|  Gemifloxacin | S. pneumoniae ATCC 49619
Exp. Range: 0.008 – 0.003 ug/ml | ≤0.03 | 60 | 60 | 60  |
|   |  |   |   |   |   |
|   |  S. pneumoniae #5
Exp. Range: 0.015 – 0.06 ug/ml | ≤0.03 | 60 | 60 | 60  |
|  |   |   |   |   |   |
|  Linezolid | S. pneumoniae ATCC 49619
Exp. Range: 0.5 – 2 μg/mL | 1 | 57 | 58 | 52  |
|   |   |  2 | 3 | 2 | 8  |
|   |  |   |   |   |   |
|   |  S. pneumoniae #5
Exp. Range: 0.5 – 2 μg/mL | 0.5 | 11 | 5 | 7  |
|   |   |  1 | 49 | 55 | 53  |
|  |   |   |   |   |   |
|  Levofloxacin | S. pneumoniae ATCC 49619
Exp. Range: 0.5 – 2 μg/mL | ≤0.5 | 60 | 60 | 60  |
|   |   |  |   |   |   |
|   |  |   |   |   |   |
|   |  S. pneumoniae #5
Exp. Range: 0.5 – 2 μg/mL | ≤0.5 | 1 | 2 | 14  |
|   |   |  1 | 58 | 58 | 46  |
|   |   | 2 | 1 |  |   |
|  |   |   |   |   |   |

Nephelometer was used at each site to standardize the inoculum and it was calibrated each time it was switched on. Colony counts from QC ATCC and in-house source was performed using direct inoculum method and the mean result was within the minimum and maximum ranges.

d. Detection limit:
Not Applicable

e. Analytical specificity:

{4}

Not Applicable

f. Assay cut-off:
Not Applicable

2. Comparison studies:
a. Method comparison with predicate device:

The recommended broth dilution reference panel was prepared according to the CLSI recommendation. Clinical testing included both fresh and stock clinical isolates and a set of challenge organisms. The broth reference panel for Streptococcus spp. was set up on MH supplemented with 2% to 5% lysed horse blood as recommended by CLSI and incubated in a non CO2 incubator for 20 – 24 hours. The comparison resulted in the following performance evaluations as reflected below.

Summary Table for Streptococcus spp. Other than S. pneumoniae (Manual Read Method)

|   | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #NS/ R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  *Gemifloxacin | 349 | 349 | 100 | 35 | 35 | 100 | - | - | - | - | - | -  |
|  *Linezolid | 349 | 349 | 100 | 344 | 344 | 100 | 349 | 100 | 1 | 0 | 0 | 0  |
|  Levofloxacin | 349 | 349 | 100 | 153 | 153 | 100 | 349 | 100 | 3 | 0 | 0 | 0  |

EA-Essential Agreement
vmj – very major discrepancies
CA-Category Agreement
maj – major discrepancies
R-resistant isolates
min – minor discrepancies
*Only EA can be determined since there are no interpretative criteria.
*NS-Not susceptible, only susceptible category exists

Summary Table for Streptococcus spp. Other than S. pneumoniae (AutoRead Method)

|   | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #NS/ R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  *Gemifloxacin | 349 | 349 | 100 | 33 | 33 | 100 | - | - | - | - | - | -  |
|  *Linezolid | 349 | 347 | 99.4 | 343 | 343 | 100 | 347 | 99.4 | 3 | 0 | 0 | 0  |
|  Levofloxacin | 349 | 349 | 100 | 141 | 141 | 100 | 349 | 100 | 3 | 0 | 0 | 0  |

EA-Essential Agreement
vmj – very major discrepancies
CA-Category Agreement
maj – major discrepancies
R-resistant isolates
min – minor discrepancies
*Only EA can be determined since there are no interpretative criteria.
*NS-Not susceptible, only susceptible category exists

EA is when there is agreement between the reference method and the Sensititre® panel within plus or minus one serial two-fold dilution of antibiotic. Category agreement (CA) is when the Sensititre® panel interpretative results, Sensitive, Intermediate, and Resistant (SIR) agrees exactly with the reference panel result interpretation. Evaluable EA is when the MIC result is on scale for both the Sensititre® and the reference and have on-scale EA. The EA% is acceptable when

{5}

compared to the reference method as described in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA".

The charts above demonstrated that the performance for Streptococcus spp. other than S. pneumoniae was acceptable for both methods of reading.

The growth rate for Streptococcus spp. is greater than 90%.

b. Matrix comparison: Not Applicable

3. Clinical studies:

a. Clinical Sensitivity: Not Applicable
b. Clinical specificity: Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable

4. Clinical cut-off: Not Applicable

5. Expected values/Reference range:

|  Antibiotic | Organism | Interpretative Criteria  |
| --- | --- | --- |
|  Gemifloxacin | Streptococcus pneumoniae | ≤0.12 (S), 0.25 (I), ≥0.5 (R)  |
|  *Linezolid | Streptococcus spp. including S. pneumoniae | ≤2 (S)  |
|  Levofloxacin | Streptococcus spp. including S. pneumoniae | ≤2 (S), 4 (I), ≥8 (R)  |

* (CLSI Comment) For some organism/antimicrobial agent combinations, the absence or rare occurrence of resistant strains precludes defining any results categories other than "susceptible." For strains yielding results suggestive of a "nonsusceptible" category, organism identification and antimicrobial susceptibility test results should be confirmed. Subsequently, the isolates should be saved and submitted to a reference laboratory that will confirm results using a CLSI/NCCLS reference dilution method.

N. Proposed Labeling:

The expected value range, interpretive criteria and QC for gram positive panels are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

{6}

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K062783](https://fda-staging.innolitics.com/device/K062783)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
