K062755 · Ikonisys, Inc. · JOY · Jan 4, 2007 · Hematology
Device Facts
Record ID
K062755
Device Name
IKONISCOPE ONCOFISH BLADDER TEST SYSTEM
Applicant
Ikonisys, Inc.
Product Code
JOY · Hematology
Decision Date
Jan 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Bladder cancer cell detection
Automated cell locating/counting using pattern recognition algorithms
—
99% concordance with manual FISH analysis
—
—
Clinical trial of 100 patient samples compared with standard manual FISH analysis.
>1 (cytotechnologists)
Indications for Use
The Ikoniscope® oncoFISH™ Bladder Test System is an automated scanning microscope coupled with image analysis, acquisition and display functions. It is intended for in-vitro diagnosis as an aide to the technologist or pathologist in the detection, classification and enumeration of cells of interest based on particular characteristics such as intensity, size, shape or fluorescence. The Ikoniscope® oncoFISH™ Bladder Test System is intended to detect cells, derived from urine samples, stained by FISH using commercially available direct labeled DNA probes or chromosomes 3, 7, 17 and loss of 9p21 locus. Following identification, a summary report is generated which may provide the basis for a diagnostic determination by the genetics professional and a gallery of the images scanned is presented for review to permit the professional to confirm or deny the diagnosis.
Device Story
Automated scanning microscope system; processes urine samples stained with FISH probes for chromosomes 3, 7, 17, and 9p21 locus. System performs automated slide loading/handling; executes low and high magnification scanning; uses pattern recognition algorithms to identify and capture images of cells meeting predetermined signal characteristics. Output: digital image gallery of scorable cell nuclei and summary report. Used in clinical laboratory settings; operated by cytotechnologists, cytogeneticists, or pathologists. Healthcare providers review the image gallery to confirm or deny diagnostic findings. Device increases efficiency by reducing manual slide scanning time; supports diagnostic determination by genetics professionals.
Clinical Evidence
Clinical trial of 100 patient samples compared automated Ikoniscope results to manual FISH analysis, showing 99% diagnostic concordance. Reproducibility study of 50 slides across two systems showed 98% concordance. Bench testing compared automated dot counting to expert cytotechnologist review of 1,200 nuclei, demonstrating >79% agreement.
Technological Characteristics
Automated scanning microscope with integrated slide loader/handler. Features digital image acquisition and workstation for image analysis. Employs pattern recognition algorithms for cell locating and signal classification. Standalone workstation configuration. No specific material standards or software architecture details provided.
Indications for Use
Indicated for use as an aid to technologists or pathologists in the detection, classification, and enumeration of cells in urine samples stained with FISH probes for chromosomes 3, 7, 17, and 9p21 locus loss.
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
Predicate Devices
Ikoniscope® fastFISH™ Auto/Amniocyte Test System (K052577)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
K062755
B. Purpose for Submission:
Original 510(k)
C. Manufacturer and Instrument Name:
Ikonisys, Inc. Ikoniscope® oncoFISH™ Bladder Test System
D. Type of Test or Tests Performed:
The Ikoniscope oncoFISH Bladder Test System detects cells of interest from urine samples, stained by FISH using commercially available direct labeled DNA probes or chromosomes 3, 7, 17, and loss of 9p21 locus.
E. System Descriptions:
1. Device Description:
The Ikoniscope onco FISH Bladder Test System is an automated scanning microscope system incorporating automated slide loading and handling, low and high magnification scanning to identify targets of interest and digital image acquisition, coupled with an image analysis workstation. Microscope slides, prepared according to the DNA probe manufacturers' specifications, are placed into a multiple slide cassette, and loaded into the Ikoniscope oncoFISH Bladder Test system. The system unloads each slide, scans each one, and returns it to the cassette automatically. During scanning, images of cells exhibiting the predetermined characteristics for FISH signals are digitally photographed and stored. After all the slides are scanned, the workstation provides an image gallery for each slide that displays the image of each cell meeting predetermined characteristics and quantity and places scanned nuclei into scorable categories, established according to the specifications in the DNA probes FDA cleared labeling. The operator/reader can then evaluate the cell nuclei, and make the diagnostic determination accordingly.
{1}
2
2. Principles of Operation:
The Ikoniscope oncoFISH Bladder Test System combines elements of existing technologies to perform its function. Fluorescence In-Situ Hybridization (FISH) uses commercially available, FDA cleared, DNA probes (not supplied with the test system) for labeling chromosomes 3,7,17, and the loss of 9p21 locus (UroVysion® Multicolor DNA Probe Kit, Vysis, Downer’s Grove, IL). Automated cell locating/counting using pattern recognition algorithms to identify the signal characteristics of interest. The software incorporated in the system automatically captures an image of each cell containing FISH signals and stores its location on the slide. These images are then presented to the operator, using a computer workstation, for analysis. Images are displayed in scorable categories according to the specifications of the probe developer.
3. Modes of Operation:
N/A
4. Specimen Identification:
Barcode
5. Specimen Sampling and Handling:
Samples should be obtained and handled according to the laboratory’s standard operating procedures and following the protocol described in the package insert for the UroVysion probe kit.
6. Calibration:
Calibration of the Ikoniscope is done at the time of installation by Ikonysis.
7. Quality Control:
ProbeChek® quality control slides by Abbott should be used with the UroVysion probes as recommended by the probe manufacturer.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☑ or No ☐ Comprehensive software documentation at a Moderate Level of Concern was provided.
{2}
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5260, Automated cell-locating device
2. Classification:
Class II
3. Product code:
JOY
4. Panel:
Hematology (81)
G. Intended Use:
1. Indication(s) for Use:
The Ikoniscope® oncoFISH™ Bladder Test System is an automated scanning microscope coupled with image analysis, acquisition and display functions. It is intended for in-vitro diagnosis as an aide to the technologist or pathologist in the detection, classification and enumeration of cells of interest based on particular characteristics such as intensity, size, shape or fluorescence. The Ikoniscope® oncoFISH™ Bladder Test System is intended to detect cells, derived from urine samples, stained by FISH using commercially available direct labeled DNA probes or chromosomes 3, 7, 17 and loss of 9p21 locus. Following identification, a summary report is generated which may provide the basis for a diagnostic determination by the genetics professional and a gallery of the images scanned is presented for review to permit the professional to confirm or deny the diagnosis.
2. Special Conditions for Use Statement(s):
N/A
{3}
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
BioView, Inc., Duet™ System (K050840)
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Illumination | Halogen Lamp | Same |
| Basic Components | Automated slide loading
Automated microscope
Camera, PC, Keyboard
and control panel, Color monitor, Color printer. | Same
Same |
| Cells Targeted | Urine | Same |
| Clinical Trial Comparison | Test device compared with standard manual FISH analysis. | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Method of Operation | Automated epifluorescent microscopy with monochrome digital image capture of wavelength specific fluorescent signals. | Automated microscopy in bright-field and fluorescent illumination with color digital image capture of color specific fluorescent signals. |
| Microscope Objectives | 10X, 100X (50X actual resolution). | 10X, 60X |
| Camera | Monochrome, Digital | Color, Digital |
| Image Presentation | Pseudo-color image | Color image |
| Clinical Trial Size | 100 slides for 100 patients | 135 slides for 135 patients |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | The Ikoniscope®
oncoFISH™ Bladder Test System is an automated scanning microscope coupled with image analysis, acquisition and display functions. It is intended for in-vitro diagnosis as an aide to the technologist or pathologist in the detection, classification and enumeration of cells of interest based on particular characteristics such as intensity, size, shape or fluorescence.
The Ikoniscope®
oncoFISH™ Bladder Test System is intended to detect cells, derived from urine samples, stained by FISH using commercially available direct labeled DNA probes or chromosomes 3, 7, 17 and loss of 9p21 locus. Following identification, a summary report is generated which may provide the basis for a diagnostic determination by the genetics professional and a gallery of the images scanned is presented for review to permit the professional to confirm or deny the diagnosis. | The Duet™ System is an automated scanning microscope and image analysis system. It is intended for in-vitro diagnostic use an aiding tool to the pathologist in the detection, classification and counting of cells of interest based on particular color, intensity, size, shape and pattern.
The Duet™ System is intended to detect the following cell types:
1. Hematopoietic cells stained by Giemsa stain, immunohistochemistry or ISH (with bright field or fluorescent) prepared from cell suspension.
2. Amniotic cells stained by FISH (using direct-labeled DNA probed for chromosomes X, Y, 13, 18 and 21. Cells in urine specimens, stained by FISH using the Vysis UroVysion™ Bladder Cancer Recurrence Kit) for chromosomes 3, 7, 17 and 9p21 locus) from subjects with transitional cell carcinoma of the bladder. |
5
{5}
I. Special Control/Guidance Document Referenced (if applicable):
N/A
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
The Ikoniscope® oncoFISH™ Bladder Test System was evaluated in a clinical trial to determine the accuracy of the system compared with FISH analysis of the same samples using standard manual evaluation. In this trial, results of FISH analysis of 100 patient samples evaluated using the Ikoniscope® oncoFISH™ Bladder Test System were compared with the results of the evaluation of the same samples using standard manual analysis to perform the FISH analysis. In this trial there was a 99% concordance between the two methods in terms of diagnostic result.
Results: One hundred consecutive samples were analyzed as the study samples. 96 samples were identical as compared to the manual method. Of the four samples that did not concur, three were read as inconclusive and the manual method identified it as negative. This would indicate a requirement for a manual review of the sample. The oncoFISH Bladder Test system demonstrated a 99% concordance rate.
b. Precision/Reproducibility:
A trial, designated the reproducibility trial, tested between system agreement. A total of 50 individual patient slides processed as part of the accuracy study were evaluated on each of two test systems, with the order of the evaluation determined randomly.
Results: Of the 50 samples analyzed 49 of the samples were concordant. The one slide that was not concordant was a negative and an inconclusive. The most probable reason for this is sample degradation. There was 98% concordance of diagnostic outcome between the two evaluations for all of the slides
c. Linearity:
N/A
{6}
d. Carryover:
N/A
e. Interfering Substances:
N/A
2. Other Supportive Instrument Performance Data Not Covered Above:
a. Comparison of manual dot count with automated dot count.
The purpose of this study was to demonstrate that oncoFISH dot counting is equivalent to the fastFISH auto/amniocyte dot counting. Two independent-expert cytotechnologists evaluated images of nuclei obtained using the oncoFISH system and compared those results to the results of automated scanning of the same images. The sample consisted of scan data for five slides of which a minimum of 1200 nuclei were to be selected and reviewed.
Results: The study provides evidence that automated scanning, signal recognition and enumeration software produces FISH signal counts that are in agreement with those determined by experienced cytotechnologists that have examined the same images. The overall agreement of the human evaluators and the automated evaluations was 79% and higher for each chromosome signal analyzed.
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{7}
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.