CARDIAC MARKERS ON ABX PENTRA 400 CLINICAL CHEMISTRY ANALYZER
Applicant
Horiba Abx
Product Code
CGS · Clinical Chemistry
Decision Date
Jun 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
Cardiac Markers reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure cardiac marker analytes. ABX PENTRA CK-NAC CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of the total creatine kinase in human serum and plasma based on an optimized UV test. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. The ABX PENTRA CK Control is for use in quality control by monitoring accuracy and precision for the quantitative ABX PENTRA CK-NAC method. ABX PENTRA Myoglobin CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of myoglobin (an oxygen storage protein found in muscle) in human serum and plasma based on a latex-enhanced immunoturbidimetric assay. Measurements of myoglobin aids in the rapid diagnosis of heart or renal disease. The ABX PENTRA Myoglobin Cal is a calibrator for use in the calibration of quantitative Horiba ABX PENTRA Myoglobin CP method on Horiba ABX clinical chemistry analyzers. The ABX PENTRA Immuno II Control L/H is for use in quality control by monitoring accuracy and precision. The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers. The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision. The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.
Device Story
System consists of reagents, calibrators, and controls for use on ABX PENTRA 400, a discrete photometric benchtop clinical chemistry analyzer. CK-NAC CP assay uses optimized UV test; Myoglobin CP assay uses latex-enhanced immunoturbidimetric test. Input: human serum or plasma samples. Analyzer performs photometric measurements; processes data to provide quantitative analyte concentrations. Used in clinical laboratory settings by trained personnel. Output: numerical concentration values for CK and myoglobin. Healthcare providers use results to assist in diagnosis/management of myocardial infarction, muscle diseases, and heart/renal conditions. Benefits include rapid, automated diagnostic testing.
Clinical Evidence
Bench testing only. Performance data includes: CK-NAC CP (n=350, r^2=0.9930, CV < 4.65%, range 8-1500 U/l) and Myoglobin CP (n=180, r^2=0.9756, CV < 5.24%, range 20.7-500 µg/l). Testing confirmed accuracy, precision, linearity, and stability against predicate methods.
Indicated for quantitative in vitro diagnostic determination of total creatine kinase and myoglobin in human serum and plasma. Used for diagnosis and treatment of myocardial infarction, muscle diseases (e.g., Duchenne-type muscular dystrophy), and heart or renal disease. For prescription use only on the ABX PENTRA 400 analyzer.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062737
B. Purpose for Submission:
Notification of intent to manufacture and market the devices: a group of reagents and their associated controls and calibrators for use on the ABX PENTRA 400 - ABX PENTRA CK NAC CP, ABX PENTRA Myoglobin CP, ABX PENTRA CK Control, ABX PENTRA Immuno II Control L/H, ABX PENTRA Myoglobin Cal, ABX PENTRA N Control, ABX PENTRA P Control, and ABX PENTRA Multical.
C. Measurand:
Total creatine kinase and myoglobin
D. Type of Test:
Quantitative
E. Applicant:
Horiba ABX
F. Proprietary and Established Names:
ABX PENTRA CK NAC CP
ABX PENTRA Myoglobin CP
ABX PENTRA CK Control
ABX PENTRA Immuno II Control L/H
ABX PENTRA Myoglobin Cal
ABX PENTRA N Control
ABX PENTRA P Control
ABX PENTRA Multical
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1215 : Creatine phosphokinase/creatine kinase or isoenzymes test system
21 CFR 866.5680 : Myoglobin immunological test system
21 CFR 862.1150 : Calibrator
21 CFR 862.1660 : Quality control materials (assayed)
2. Classification:
Class II (Assays)
Class II (Calibrator)
Class I (Control materials), reserved
{1}
3. Product code:
CGS, NAD reduction/NADH oxidation, Cpk or isoenzymes
DDR, Myoglobin, antigen, antiserum, control
JJY, Quality control material (assayed and unassayed)
JIX, Calibrator, multi-analyte mixture
JIT, Secondary calibrator
4. Panel:
Clinical Chemistry (75)
Immunology (82)
H. Intended Use:
1. Intended use(s):
Cardiac Markers reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure cardiac marker analytes.
ABX PENTRA CK-NAC CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of the total creatine kinase in human serum and plasma based on an optimized UV test. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
The ABX PENTRA CK Control is for use in quality control by monitoring accuracy and precision for the quantitative ABX PENTRA CK-NAC method.
ABX PENTRA Myoglobin CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of myoglobin (an oxygen storage protein found in muscle) in human serum and plasma based on a latex-enhanced immunoturbidimetric assay. Measurements of myoglobin aids in the rapid diagnosis of heart or renal disease.
The ABX PENTRA Myoglobin Cal is a calibrator for use in the calibration of quantitative Horiba ABX PENTRA Myoglobin CP method on Horiba ABX clinical chemistry analyzers.
The ABX PENTRA Immuno II Control L/H is for use in quality control by monitoring accuracy and precision.
The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers.
The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.
2
{2}
The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.
2. Indication(s) for use:
See Intended Use above.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
ABX PENTRA 400
I. Device Description:
All the reagents, controls and calibrators included in this submission are for use on the ABX PENTRA 400 (k052007), which is a discrete photometric benchtop clinical chemistry analyzer.
The ABX PENTRA CK NAC CP is an in vitro diagnostic assay for the quantitative determination of total creatine kinase in human serum and plasma based on an optimized UV test. The assay is composed of a bi-reagent cassette, with 26 ml and 6.5 ml compartments. Reagents are chemical solutions with additives.
The ABX PENTRA Myoglobin CP is an in vitro diagnostic assay for the quantitative determination of myoglobin in human serum and plasma based on a latex-enhanced immunoturbidimetric test. The assay is composed of a bi-reagent cassette, with 15 ml and 9.5 ml compartments. Reagents are chemical solutions with chemical additives and substances of animal origin.
The ABX PENTRA Myoglobin Cal is a liquid calibrator prepared from a dilution of purified myoglobin positive human sera. It is used for the calibration of the myoglobin assay. The assigned values are given on the vials. This calibrator is provided in five vials of 1 mL.
The ABX PENTRA CK Control is a lyophilized assayed control prepared from a bovine serum albumin with chemical additives and material of biological origin. It has to be used for the quality control of the creatine kinase assay. The assigned values are given in the enclosed annex. This calibrator is provided in 4 vials of 3 mL.
The ABX PENTRA Immuno II Control L/H is a lyophilized assayed control prepared from a stabilized pool of human sera. It has 2 levels (Low and High) to be used for the quality control of the myoglobin assay. The assigned values are given in the enclosed annex. Each level of this control is provided in one vial of 3 mL.
The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator
3
{3}
components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. This calibrator is provided in ten vials of 3 ml. All products derived from blood are prepared exclusively from the blood of donors tested individually and shown by FDA-approved methods to be free from HBsAg and antibodies to HCV and HIV.
The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml. All products derived from blood are prepared exclusively from the blood of donors tested individually and shown by FDA-approved methods to be free from HBsAg and antibodies to HCV and HIV.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche CK NAC Reagent Set, Beckman Coulter ACCESS Myoglobin Reagent, Pointe Scientific CK-MB Control Set Level I/Level II, Bio-Rad Liquicheck Cardiac Markers Control Level 2, Beckman Coulter ACCESS Myoglobin Calibrators, ABX Pentra Multical and ABX Pentra N Control.
2. Predicate 510(k) number(s):
k834502, k021229, k954074, k961828, k021229, k060854
3. Comparison with predicate:
| | Predicate device (k834502): | Device : |
| --- | --- | --- |
| Device Name | CK NAC | ABX Pentra CK NAC CP |
| Manufactured by | Roche, USA | HORIBA ABX, France |
| Instrument | COBAS MIRA | ABX PENTRA 400 |
| Analytes | Creatine Kinase | Total Creatine Kinase |
| Method : | Optimized UV test | Optimized UV test |
| Specimen : | Serum | Serum and plasma |
| Number of tests | - | 125 tests |
| Sample volume | 4.4 μL/test | 8 μL/test |
| Detection limit | - | 8 U/L |
| Precision | CV Total < 1.6% | CV Total < 4.65% |
| Measuring range | - | 8 U/L – 1500 U/L |
| Upper linearity limit | 2000 U/L | 1500 U/L, and with automatic post-dilution: 4500 U/L |
| Calibration stability | - | 8 days |
| Closed reagent stability | Until the expiration date when stored at 2-8°C after reconstitution: | 18 months at 2-8°C |
{4}
| | Predicate device (k021229): | Device : |
| --- | --- | --- |
| Device Name | CK NAC | ABX Penta CK NAC CP |
| Open reagent stability | 6 days at 2-8C 24 hours at 15-25°C | on-board stability (refrigerated area): 64 days |
| | Predicate device (k021229): | Device : |
| --- | --- | --- |
| Device Name | ACCESS Myoglobin Reagent | ABX Penta Myoglobin CP |
| Manufactured by | BECKMAN COULTER Inc. | HORIBA ABX, France |
| Instrument | ACCESS Immunoassay systems (Access II analyzer) | ABX PENTRA 400 |
| Analytes | Myoglobin | Myoglobin |
| Method : | Paramagnetic particle and chemiluminescent immunoassay. | Latex-enhanced immunoturbidimetric assay. |
| Specimen : | Serum and plasma | Serum and plasma |
| Number of tests | 100 tests | 100 tests |
| Sample volume | 20 μL/test | 14 μL/test |
| Detection limit | < 1 μg/L (ng/mL) | 6.7 μg/L (ng/mL) |
| Precision | CV Total < 4.54% | CV Total < 5.24% |
| Measuring range | 13.8 μg/L – 3650.9 μg/L | 20.4 μg/l – 500 μg/L |
| Upper linearity limit | 4000 μg/L | 500 μg/l, and with automatic post-dilution: 2500 μg/L |
| Calibration stability | 56 days | 21 days |
| Closed reagent stability | Until the expiration date when stored at 2-10°C | 12 months at 2-8°C |
| Open reagent stability | Open pack: 56 days at 2-10°C | On-board stability (refrigerated): 35 days |
| | Predicate device (k954074): | Device : |
| --- | --- | --- |
| Device Name | CK-MB Control Set Level I, Level II | ABX Penta CK Control |
| Manufactured by | POINTE SCIENTIFIC, INC. AALTO SCIENTIFIC, INC, | HORIBA ABX, France |
| Instrument | COBAS MIRA Plus | ABX PENTRA 400 |
| Method : | Single parameter control by monitoring the performances of CK-MB determination with Creatine Kinase-MB Reagent Set | Bi-parameter control by monitoring accuracy and precision for the quantitative Methods as specified in the enclosed value sheet. Concentrations and activities are mostly in the normal or at the normal/pathological threshold. |
| | | |
{5}
| | Predicate device (k954074): | Device : |
| --- | --- | --- |
| Device Name | CK-MB Control Set Level I , Level II | ABX Pentra CK Control |
| Biological additives | Human derived isoenzymes | CK-MM (Human origin) CK-MB (porcine brain origin) |
| Controlled molecules | CK-MB: the exact control values are given on the vial label. | Total Creatine Kinase (CK-NAC): the exact control values are given in the enclosed annex. |
| Theoretical values and confidence intervals | - The assay value and range are established using the manufacturer's CK-MB reagent. | - The assigned values are determined by calibration with a reference material in accordance with established protocols. |
| | - The assigned values and range are indicated on the vial label. | - The assigned values and precise confidence interval are indicated in the annex enclosed in the kit. |
| | - The values are lot-specific. | - The assigned values are lot-specific. |
| Closed stability | Until the expiration date when stored at 2-8°C | 18 months at 2-8°C |
| Components stability after reconstitution of the calibrator | At 2 - 8°C : 7 days | At 15 - 25°C : 24 hours At 2 - 8°C : 3 days |
| | Predicate device (k961828): | Device : |
| --- | --- | --- |
| Device Name | Liquicheck Cardiac Markers Control, level 2 | ABX Pentra Immuno II Control L/H |
| Manufactured by | Bio-Rad | HORIBA ABX, France |
| Instrument | ACCESS Immunoassay systems (Access II analyzer) | ABX PENTRA 400 |
| Method : | Quality control by monitoring the precision of laboratory testing procedures listing in the package insert | Quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed value sheet. |
| Controlled molecules | Myoglobin, Troponin I, Troponin T, CK Total, LD-1 isoenzyme, CK-MB isoenzyme. The exact control values are given in the notice. | Myoglobin, Ferritin. The exact control values are given in the enclosed annex. |
| Assigned values | - The assigned values are determined by calculating the mean value obtained from replicate analyses. - The assay values are listed in the notice enclosed in the kit. | - The assigned values specified are determined by calculating the mean value obtained from multiple determinations. - The assigned values are indicated in the notice enclosed in the kit. |
{6}
| | Predicate device (k961828): | Device : |
| --- | --- | --- |
| Device Name | Liquicheck Cardiac Markers Control, level 2 | ABX Penta Immuno II Control L/H |
| Closed stability | - The values are lot specific | - The assigned values for both Low and High controls are lot-specific |
| Open stability | 2 years at -20°C to -10°C | 24 months at 2-10°C |
| | Stability of Myoglobin, Troponin I, Troponin T: At 2 to 8°C : 10 days | 2 weeks at 2-10°C |
| | Stability of Total CK, CK-MB, LD-1: At 2 to 8°C : 20 days | 3 months at -20°C |
| | Predicate device (k021229): | Device : |
| --- | --- | --- |
| Device Name | ACCESS Myoglobin Calibrators | ABX Penta Myoglobin Cal |
| Manufactured by | BECKMAN COULTER Inc. | HORIBA ABX, France |
| Instrument | ACCESS Immunoassay systems (Access II analyzer) | ABX PENTRA 400 |
| Method : | Calibration of ACCESS Myoglobin assay | Calibration of HORIBA ABX Myoglobin methods |
| Calibrated molecules | Myoglobin | Myoglobin |
| Format | Liquid | Liquid |
| Calibration value | - Assigned by procedure traceable to the manufacturer's working calibrators and based on prEN ISO 17511. | - Determined using primary calibration with in-house calibrator adjusted to nephelometric method. |
| | - The exact assigned values are indicated on each vial or on calibration card (5 levels: approximately 50, 200, 800, 2000 and 4000 ng/mL) | - The assigned values are indicated on each vial (5 levels: 0, 62.5, 125, 250 and 500 ng/mL) |
| Closed stability | Up to the expiration date at -20°C | 12 months at 2-10°C |
| Open stability | 60 days at 2-10°C | 7 weeks at 2-10° |
ABX Penta Multical and ABX Penta N Control: Addition of creatine kinase.
# K. Standard/Guidance Document Referenced (if applicable):
1) Valtec protocol for evaluation of device performance: Vassault et al., Ann. Biol. Clin., 1986, (44), 686-745).
{7}
2) CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods.
3) CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach.
4) CLSI EP9-A2, Method Comparison and Bias Estimation Using Patient Samples.
5) CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation.
# L. Test Principle:
For total creatine kinase, the reagents supplied cause a chemical reaction which produces an amount of gluconate-6-phosphate $+\mathrm{NADPH} + \mathrm{H}^{+}$ which is proportional to the amount of creatine kinase in the sample.

For myoglobin, the method used in this assay is a turbidimetric immunoassay. When an antigen-antibody reaction occurs between Mb in a sample and anti-Mb antibody which has been sensitized to latex particles, agglutination results. This agglutination is detected as an absorbance change, with the magnitude of the change being proportional to the quantity of Mb in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
# Within Run:
For the CK NAC (total CK) reagent, within-run precision was determined using 2 controls (ABX PENTRA N and P Controls) and 3 serum specimens of low, medium and high concentrations were tested 20 times in a single run for each sample (in accordance with the Valtec guideline (Vassault et al., Ann. Biol. Clin., 1986, (44), 686-745). Results are described below.
| In U/L | N Control | P Control | Sample 1 | Sample 2 | Sample 3 |
| --- | --- | --- | --- | --- | --- |
| Mean | 166 | 474 | 46 | 115 | 347 |
| SD | 2 | 4 | 1 | 1 | 3 |
| %CV | 1.2 | 0.92 | 2.54 | 1.14 | 0.79 |
For the Myoglobin reagent, within-run precision was determined using 2 controls (Immuno II Control L/H) and 3 serum specimens of low, medium and high concentrations tested 20 times in a single run for each sample (in accordance with the Valtec guideline (Vassault et al., Ann. Biol. Clin., 1986,
{8}
(44), 686-745). Results are described below.
| In ng/mL | L Control | H Control | Sample 1 | Sample 2 | Sample 3 |
| --- | --- | --- | --- | --- | --- |
| Mean | 66.9 | 207.4 | 27.9 | 46.7 | 330.3 |
| SD | 1.0 | 2.4 | 1.2 | 1.7 | 1.8 |
| %CV | 1.53 | 1.18 | 4.26 | 3.56 | 0.54 |
# Between Run:
For the CK NAC reagent, between-run precision NCCLS (CLSI) EP-5A was followed using two serum specimens of low & high levels and 2 controls tested in duplicate for 20 days, two series per day (for a total of 80). Results are described below.
| In U/L | | N Control | P Control | Sample 1 | Sample 2 |
| --- | --- | --- | --- | --- | --- |
| | Mean | 162 | 489 | 81 | 311 |
| Between Run | SD | 3.7 | 8.34 | 2.26 | 3.62 |
| | %CV | 2.28 | 1.78 | 2.76 | 1.13 |
| Total | SD | 4.14 | 11.71 | 3.78 | 8.12 |
| | %CV | 2.56 | 2.50 | 4.65 | 2.61 |
For the Myoglobin reagent, between-run precision NCCLS (CLSI) EP-5A was followed using two serum specimens of low & high levels and 2 controls tested in duplicate for 20 days, two series per day (for a total of 80). Results are described below.
| In ng/mL | | L Control | H Control | Sample 1 | Sample 2 |
| --- | --- | --- | --- | --- | --- |
| | Mean | 67.5 | 201.2 | 58.8 | 389.3 |
| Between Run | SD | 1.66 | 5.02 | 1.65 | 10.00 |
| | %CV | 2.47 | 2.49 | 2.81 | 2.57 |
| Total | SD | 2.99 | 10.55 | 2.46 | 17.08 |
| | %CV | 4.43 | 5.24 | 4.18 | 4.39 |
# b. Linearity/assay reportable range:
Linearity for CK NAC was assessed for non-post dilution samples in accordance with CLSI EP6-A. Samples used for this study were in house preparations of CK aqueous solution. Two different studies with ten different levels of total CK (each tested 4 times except for lowest dilution of the low level study which was sampled twice) were prepared based on the dilution of the highest concentration solution level with a solution near the low limit of detection resulting in a range of samples from 8.48 to $147\mathrm{U / L}$ and 81.38 to $1510~\mathrm{U / L}$ . The linearity of the CK NAC assay was evaluated by comparing observed versus expected values across the expected range. A linear regression analysis was performed on the data and plotted. The sponsor's acceptable bias for recovery of the samples is $\pm 8\%$ . The observed linearity across the 8.48 to $147\mathrm{U / L}$ range has a slope of 0.993, and an intercept of 0.5558. The observed linearity across the 81.38 to $1510~\mathrm{U / L}$ range has a slope
{9}
of 0.9995, and an intercept of 0.3936. The assay range was demonstrated to be linear from 8.48 to 1510 U/L. The sponsor will claim a range of 8 (the detection limit) to 1500 U/L. Recovery of post dilution serum samples was also assessed comparing manual versus automatic dilution by the PENTRA 400. Five serum samples were each tested 4 times and ranged from 1610 to 4263 U/L. The sponsor's acceptable bias for recovery of the samples is ±10%. The data supports the sponsor's claim of a post dilution limit of 4500 U/L.
Linearity for Myoglobin was assessed for non-post dilution samples in accordance with CLSI EP6-A. Samples used for this study were human serum. Eleven different levels of myoglobin (each tested 4 times) were prepared based on the dilution of the highest concentration solution level with a solution near the low limit of detection resulting in a range of samples from 48 to 482.6 ng/mL. The linearity of the assay was evaluated by comparing observed versus expected values across the expected range. A linear regression analysis was performed on the data and plotted. The sponsor's acceptable bias for recovery of the samples is ±10%. The observed linearity across the reportable range has a slope of 0.9998, and an intercept of 0.0667. The assay range was demonstrated to be linear from 48 to 482.6 ng/mL. The sponsor will claim a range of 20.4 (the lowest value of the comparison study) to 500 ng/mL (the value of the highest calibrator). The lower values of 20.4 to 48 where found to be acceptable for the claimed range as the predicate device included this range in its claim and the sponsor showed equivalence at these levels. Recovery of post dilution serum samples was also assessed comparing manual versus automatic dilution by the PENTRA 400. Seven samples were each tested 4 times and ranged from 513.9 to 2301.1 ng/mL. The sponsor's acceptable bias for recovery of the samples is ±10%. The data supports the sponsor's claim of a post dilution limit of 2500 ng/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Real time stability at 8°C was investigated for Myoglobin calibrator. Recovery of the values for the calibration points in Myoglobin was assayed for 3 lots periodically over a period of 18 months. The closed stability study was performed over a period providing data to support the sponsor's claim of closed stability between 2 - 10°C for 12 months. Real time stability studies after opening were performed. Recovery of the values for the calibration points in Myoglobin was assayed for 1 lot periodically over a period of 7 weeks, providing data supporting a stability period of 7 weeks at 8°C.
Real time stability at 8°C was investigated for Immuno II Control L/H. Recovery of the values in Myoglobin and Ferritin for the Low and High controls was assayed for 3 lots during 27 to 29 months. The closed stability study was performed over a period providing data to support the sponsor's claim of closed stability of 27 months at 8°C. Stability after reconstitution and storage frozen at -20°C was investigated for Immuno II Control L/H.
10
{10}
Stability studies after reconstitution were performed providing data supporting a stability of 3 months at -20°C for ABX Pentra Immuno II Control L/H after reconstitution and a closed stability of 24 months between 2 - 10°C.
For ABX PENTRA N Control & ABX PENTRA P Control traceability, the values of the ABX PENTRA Controls are assigned from the ABX PENTRA calibrator, reagents and analyzers. The target value is determined by the median of results from 150 measurements/parameters. Confidence range is determined as the calculated range in percent which is based on the experimental results from the previous target value trials. The range declared in the target value sheet is equal to the assigned value +/- 3 standard deviations (3 SD).
For ABX Pentra Multical traceability, the ABX Pentra Multical is prepared from reference materials. Commercial calibrators are standardized by means of a master lot which is stored at -80°C. Two controls are used to ensure that the calibration values of the master lot, as well as the entire measurement system (calibrator, reagent, and analyzer), remain stable during the storage period. The target value is determined by the median of results from 150 measurements/parameters.
For the ABX PENTRA N Control, ABX PENTRA P Control, and ABX PENTRA Multical, protocols and acceptance criteria for open and closed stability of the controls and calibrators were described and found to be acceptable.
d. Detection limit:
The detection limit for the CK NAC was determined by measuring 30 measurements of saline water (NaCl 0.9 g/L) + 4.65 SD. The limit of detection for the CK NAC assay was determined to be 8 U/L.
Because the measurement of saline water is not possible with the normal application of the Myoglobin assay (no results available when sample concentration is lower than the calibration curve), usual measurement of MDL (30 measurements of saline water) is not possible. For the specific proteins the Minimum Interpretation Limit (MIL) is used. Like MDL, MIL is the lower concentration that could be distinguished from 0, concentration calculated using the rate absorbency measurement of known concentration's samples and saline water. The limit of detection for the Myoglobin assay was determined to be 6.7 ng/mL.
e. Analytical specificity:
For CK NAC: In accordance with the Valtec guideline, the tested interferants were added to base serum at two different CK concentrations (normal and high). The base serum with each substance was then serially diluted with the same base serum and saline to adjust CK concentration. Hemoglobin up to 55
11
{11}
$\mu \mathrm{mol} / \mathrm{l}$ (97 mg/dL), total bilirubin up to 125 $\mu \mathrm{mol} / \mathrm{L}$ (7.3 mg/dL), direct bilirubin up to 100 $\mu \mathrm{mol} / \mathrm{l}$ (5.9 mg/dL) and triglycerides (as Intralipid®, representative of lipemia) up to 7 mmol/L (612.5 mg/dL) do not interfere with CK determination by this test.
For Myoglobin: Substances were added to the base serum at two different myoglobin concentrations (normal and high). The base serum with each substance was then serially diluted with the same base urine serum and saline to adjust myoglobin concentration. Hemoglobin up to 195 $\mu \mathrm{mol} / \mathrm{l}$ (336 mg/dL), total bilirubin up to 500 $\mu \mathrm{mol} / \mathrm{L}$ (29.3 mg/dL), direct bilirubin up to 500 $\mu \mathrm{mol}$ (29.3 mg/dL) and triglycerides (as Intralipid®, representative of lipemia) up to 7 mmol/L (612.5 mg/dL) do not interfere with myoglobin determination by this test. The sponsor also determined a prozone effect for the Myoglobin assay to occur at 1374 ng/mL with dilution being effective in overcoming the effect.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 175 serum and plasma samples were run using the CK NAC reagents on the PENTRA 400 and a reference method. Samples ranged from 8.0 to 1408 U/L. Diluted or spiked samples (59) were used to cover the range of the assay. Linear regression analysis gave the following relationship: Device = 1.056(Predicate) - 2.253; r = 0.9967.
A total of 90 serum and plasma samples were run using the Myoglobin reagents on the PENTRA 400 and a reference method. Clinical non-spiked samples ranged from 20.4 to 484 ng/mL. Linear regression analysis gave the following relationship: Device = 0.920(Predicate) + 22.251; r = 0.9876.
b. Matrix comparison:
The sponsor demonstrated equivalence of CK NAC results in serum and Lithium Heparin plasma samples by using 42 clinical samples spanning a range of 29 to 1197 U/L. Linear regression showed a slope of 0.96 and an intercept of 1.04, and a correlation coefficient of 0.999.
The sponsor demonstrated equivalence of Myoglobin results in serum and Lithium Heparin plasma samples by using 69 clinical samples spanning a range of 19 to 236 ng/mL. Linear regression showed a slope of 1.01 and an intercept of 2.12, and a correlation coefficient of 0.9959.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
12
{12}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The reference range for CK NAC is indicated to be:
Adults:
Women < 145 U/L
Men < 171 U/L
Reference:
IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37°C; Part 2; Clin Chem Lab Med 2002; 40(6): 635-642.
The reference range for Myoglobin is indicated to be:
70 – 110 ng/mL (μg/L)
Reference:
Thomas L. Clinical Laboratory Diagnostics. 1st ed. Frankfurt: TH-Books Verlagsgesellschaft; 1998. p. 106.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.