K062736 · Astron Clinica Limited · PSN · Sep 14, 2007 · General, Plastic Surgery
Device Facts
Record ID
K062736
Device Name
SIASCOPE V
Applicant
Astron Clinica Limited
Product Code
PSN · General, Plastic Surgery
Decision Date
Sep 14, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment.
Device Story
SIAscope V is a non-invasive skin analysis system used by clinical practitioners to image skin conditions. The device utilizes Spectrophotometric Intracutaneous Analysis (SIAscopy) to investigate the skin's interior structure. It captures input via a 3.2-megapixel color digital camera while illuminating the skin with visible and near-infrared LEDs. The device transforms remitted light intensity measurements into synthesized graphical outputs called 'SIAscans'. These scans display the relative distribution of blood, collagen, and pigment. The device connects to a PC via USB 2.0 for processing and display. By providing these visual maps, the device assists clinicians in evaluating skin conditions, potentially aiding in clinical decision-making compared to traditional visual inspection or biopsy. The system is designed for ease of use with an updated user interface compared to its predecessor.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison and design specifications.
Technological Characteristics
Non-invasive skin analysis system. Materials: Glass, ABS, and TPE plastics. Illumination: Visible and near-infrared LEDs. Imager: 3.2-megapixel CMOS sensor. Connectivity: USB 2.0 to PC. Power: USB-powered. Form factor: Handheld, 250g. Software: DERMETRICS™ platform with unchanged SIAscopy algorithms.
Indications for Use
Indicated for non-invasive skin analysis to visualize the relative location of blood, collagen, and pigment in the skin of patients.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with two snakes entwined around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2017
Astron Clinica Limited c/o Smith Associates Mr. E.J. Smith 1676 Village Green, Suite A Crofton. Maryland 21114
Re: K062736
Trade/Device Name: SIAscope V Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: II Product Code: PSN Dated: June 15, 2007 Received: June 18, 2007
Dear Mr. Smith:
This letter corrects our substantially equivalent letter of September 14, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical
{1}------------------------------------------------
Page 2 - Mr. E. J. Smith
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known): K062736
Device Name: SIAscope V
Indications for Use:
The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment'.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sig: Off) | Page 1 of 1 |
|------------------------------------------------------------|-------------|
| Division of General, Restorative, and Neurological Devices | |
| 10(k) Number | L062736 |
3
{3}------------------------------------------------
62736
SEP 1 4 2007
# 510(k) Summary
#### September 6, 2007
Company Name & Address: Astron Clinica Ltd The Mount Toft Cambridge, United Kingdon CB3 7RL
Contact Name:
E.J. Smith 1676 Village Green, Suite A Crofton, MD 21114 Phone: 888-729-9674 Fax: 410-793-0448 Email: ESmith9746@aol.com
Date Summary Prepared:
August 31, 2006
Device Identification:
| Classification Name: | Light, Surgical, Floor Standing |
|------------------------------------|---------------------------------|
| Common/Usual Name: | Skin Analysis System |
| Proprietary Name: | SIAscope |
| Establishment Registration Number: | 3004519907 |
| Regulation Number: | 21 CFR 878.4580 |
| Product Code: | FSS |
| Classification Panel: | General & Plastic Surgery |
| Device Classification: | Class II |
#### Intended Use
The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment.
CAUTION : This device should only be used to scan the patient's skin.
{4}------------------------------------------------
# KO62736
#### Device Overview
The SIAscope is a quick, non-invasive device for imaging skin conditions. It uses a technique known as 'Spectrophotometric Intracutaneous Analysis' (SIAscopy) to identify and display graphically the separate components of the skin. SIAscopy uses a digital camera and light (both visible and near-infrared) to investigate the skin's interior structure.
The SlAscope operates by illuminating the skin with LEDs and measuring the intensity of remitted light. The intensity of the illumination is similar to the emission from a typical hand held torch or remote control unit.
#### Description of SIAscans
Unlike biopsy, SIAscopy provides an array of readings of the skin that are displayed in graphical form creating a synthesized image called a SIAscan.
#### Predicate Product Comparison
The SIAscope V operates in a very similar fashion to the predicate device. Where alterations have been make this have been design to improve the imaging performance, reliability and safety of the device over and above the predicate device.
Although the design of the user interface has changed, these changes have been made to make the device easier and more intuitive to use for the clinical practitioner. However, the software algorithms and theory of SIAscopy is the same as found in the predicate device.
We assert that the SIAscope V substantially equivalent to the SIAscope II (K023729)
| Parameter | SIAscope V | SIAscope II (K023729) |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FSS | FSS |
| Intended Use | SIAscope is a non-invasive skin<br>analysis system, which provides<br>color bitmaps called 'SIAscans' that<br>show the relative location of blood,<br>collagen and pigment. | SIAscope is a non-invasive skin<br>analysis system, which provides a<br>synthesized 'image' showing the<br>relative location of blood, collagen<br>and pigment. |
| Illumination For: | Skin | Skin |
| Illumination: | LED | LED |
| Imager Chip: | CMOS | CCD |
### Summary comparison of features and specification SIAscope V vs. SIAscope II
{5}------------------------------------------------
# 长662736
| Power Source: | USB via PC | AC |
|------------------------------------|------------------------------------------------------------------------|-----------------------------|
| External Materials:<br>(Caseworks) | Glass, ABS & TPE Plastics | Glass, Aluminium |
| PC Connection: | USB 2.0 | Proprietary Umbilical Cable |
| Weight: | 250grams | 1Kg |
| Calibration Method: | Factory Calibrated | Factory Calibrated |
| Viewfinder: | Colour (3.2 MegaPixel) | Monochrome (0.7 MegaPixel) |
| Software: | DERMETRICS™<br>- New User Interface<br>- SIAscopy algorithms unchanged | SIAscope II |
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.