SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH LINEZOLID
Device Facts
| Record ID | K062705 |
|---|---|
| Device Name | SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH LINEZOLID |
| Applicant | Dade Behring, Inc. |
| Product Code | LON · Microbiology |
| Decision Date | Oct 20, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1645 |
| Device Class | Class 2 |
Indications for Use
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive enterococci and staphylococci. After inoculation, panels are incubated for 4.5 - 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert. This particular submission is for the addition of the antimicrobial linezolid, at concentrations of 0.5 to 16 ug/ml Long Dilution Sequence and 0.5 - 4 ug/ml 4-Dilution Breakpoint Sequence, for enterococci and staphylococci, to the test panel. The Gram-positive organisms which may be used for linezolid susceptibility testing in this panel are: Enterococcus faecium (vancomycin-resistant) Staphylococcus aureus (methicillin-susceptible and methicillin-resistant)
Device Story
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels determine antimicrobial susceptibility of Gram-positive cocci. Input: isolated colonies from solid media, prepared via turbidity method in saline/Pluronic® broth. Operation: panels are rehydrated/inoculated, then incubated in WalkAway®-SI system. System uses colorimetric optics (color wheel/lamp/photosensor) to detect growth via light detection. Output: quantitative MIC values and qualitative susceptibility interpretations. Used in clinical labs; operated by technicians. Rapid read (<16 hours) or overnight (16-24 hours) methods available. Output informs antibiotic therapy decisions; benefits include rapid susceptibility determination for targeted treatment.
Clinical Evidence
Bench testing only. Validation conducted using fresh and stock efficacy isolates and challenge strains. Performance compared against frozen reference panels. Results demonstrated acceptable essential agreement and categorical agreement for linezolid. Reproducibility testing confirmed acceptable performance using turbidity inoculum preparation and WalkAway® SI System. Quality control testing met acceptance criteria.
Technological Characteristics
Miniaturized broth dilution susceptibility test; dehydrated antimicrobial agents in microtiter plates. Rehydrated with Synergies plus™ Pos Broth. Incubation at 35°C ± 1°C. Automated reading via WalkAway® SI System or visual reading. Connectivity: Standalone or integrated with laboratory instrumentation.
Indications for Use
Indicated for testing Linezolid (0.5–16 µg/mL) on Gram-positive MIC/Combo panels for Enterococcus faecium (vancomycin-resistant) and Staphylococcus aureus (methicillin-susceptible and methicillin-resistant). For prescription use only. Resistance detection for Linezolid in Gram-positive isolates is unknown due to limited resistant strains; non-susceptible results require retesting and reference lab confirmation.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
- MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels
- MicroScan® Dried Gram-Positive Panels